UV-VIS and Fluorescence Purpose: The purpose of this lab was to teach us how to use the UV-VIS and the Fluorescence. We ran various standards and dilutions on both instruments. Background: UV-VIS: The UV-Vis spectroscopy can be used on an electromagnetic spectrum ranging from 200nm to 800nm. This is a quantitative method that measures absorbance and %T. Using this information you can calculate the solution concentration. The UV-Vis can also be used for qualitative data, but is more used for quantitative data. The electron-electron bond helps with the absorption of energy in a molecule. The instrument generally uses inorganic, organic, and biological species as samples. The instrument has an energy source, a monochromator, a reference cell, a photomultiplier tube, an amplifier, and a read out device. The UV-Vis can either be used on its own or as a detector for other instruments. Fluorescence: Fluorescence is the emission of light by a substance that has absorbed light or other radiation. Fluorescence is a quantitative method used to measure both excitation and emission. The fluorescence has two monochromators and two photomultiplier tubes, unlike the UV-Vis. The instrument is less commonly used because it is more expensive and is not any less sensitive. Procedure: UV-VIS 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Turn on the instrument Open the “Vision Pro” program on the computer Select the scan icon on the tool bar Four viewing windows should appear Select parameters: a. Wavelength: 190nm-400nm b. Scanning for: Absorbance c. Lamp: Tungsten d. Set the baseline correction for 100%T Place a blank in the back holder a. Use a quartz cuvette and fill it ¾ full with hexane Approve the blank request which starts the baseline scan a. If denied, press the baseline/zero icon Set the results by clicking the “drop-down” menu to peak pick Prepare samples and place in the front holder Press the run icon Run each sample three times Click on sample name on left and right click on plot samples Close the program and turn off computer monitor (leave the computer on) Fluorescence 1. 2. 3. 4. 5. 6. Turn on the instrument and computer Open up the PX-150X program Check the lower right corner to make sure the instrument is on Go to configure menu instrument and then make sure that on is selected Set to Quantitative analysis: acquire mode select spectrum Fill a cuvette with 2ppm solution and run with the settings a. Spectrum type: excitation 7. 8. b. EM wavelength: 400nm c. EX wavelength: 350nm – wavelength containing peak of interest d. Sensitivity: low e. Scan Speed: fast f. Recording Range: -10 to amt. showing entire peak g. EX/EM Slit width: 10/10 h. Response time: 0.02 i. Repeat Scan/Auto file: no setting needed/disable Then click Search wavelength and click search a. Set as 350-600nm and 240-650nm b. Record optical wavelengths Run each standard three times recording wavelengths found for EX and EM Data: Refer to page 33 in my lab notebook for the data from the Fluorescence Results: Over the two days, we ran many samples with various concentrations on the UV-VIS. We were unable to get any reliable results and spectra. We never received any absorbance data therefore it was impossible to produce a calibration curve. Fluorescence Calibration Curve: Calibration Curve for Quinone 1050 Intensity (c/s) 950 850 y = 241.9x - 39.907 R² = 1 750 650 550 450 350 1 1.5 2 2.5 3 3.5 Concentration (ppm) Concentration (ppm) Intensity (c/s) 2 443.887 3 685.778 4 927.679 Tonic Water: intensity = 463.827 = y 463.827 = 241.9𝑥 − 39.907 𝑥 = 2.08ppm Tonic Water: concentration = 2.08ppm (quinine) 4 4.5 Conclusion: The UV-VIS was very complicated to work with because we were unable to get any results no matter what we tried. Also, the quartz cuvettes were hard to find which took us extra time before we could actually start using the instrument. We made samples during both days that week and still tried running them, but never got any results. At first, the Fluorescence kept giving us weird results but after multiple times we finally were able to get it to work. We weren’t able to run all of our samples but we were still able to create a calibration curve. I really didn’t work on the Fluorescence because I was trying to fix the issues with the UV-Vis. Our biggest issue was how old the computer was and due to the age of it, the computer kept freezing.