UV-VIS and Fluorescence Purpose: This lab is meant to familiarize us with using the UV/VIS and the Fluorescence. This was done by making different standards and dilutions and running them. UV/VIS background: The Ultraviolet/Visible Spectroscopy can be used on an electromagnetic spectrum ranging from 200nm to 800nm. This is typically a quantitative technique measuring absorbance and %T. From this you are able to find concentration of a solution. It is also possible to use the UV/VIS for qualitative data but it much more useful for quantitative data. The bonding of electrons in a molecule helps the energy absorb. This instrument is typically used with inorganic, organic, and biological species. The instrument has a source providing energy, a monochromator, a reference cell, a photomultiplier tube, an amplifier, and some type of read out device. The UV/VIS can either be used on its own or as a detector for other instruments. Fluorescence background: This is a quantitative technique typically used to determine large elements that form vapors and hydrides. Fluorescence is used to measure excitation and emission. Unlike the UV/VIS the Fluorescence has two monochromators and two photomultiplier tubes. This instrument tends to be more expensive and is not any less sensitive than any other options so it is less commonly used. Procedure: -Using the UV/VIS o o o o o o o o o o o o o o Turn on the instrument Open the “VisionPro” program on the computer Instrument will make a start up check Select the scan icon on the tool bar 4 viewing windows should appear Settings should be Wavelength: 190nm-400nm Scanning for: Absorbance Lamp: Tungsten Set the baseline correction for 100%T Place 1 blank in each sample holder Use a quartz cuvette filling it ¾ full with hexane Approve the blank request which will start the baseline scan If denied then press the baseline/zero icon Set the results dropdown menu to peak pick Prepare sample and standards and place it in the front holder Press the run icon Run each sample 3 times with out opening the instrument Click on sample name on left and right click on plot samples This puts all samples on one graph When finished close out of the program and turn off computer monitor (leave the computer on) -Using the Fluorescence o o o o o Turn on the instrument and computer Open up the PX-150X program Check the lower right corner to make sure the indication says that the instrument is on Then go to configure menu instrument and then make sure that on is selected Set to Quantitative analysis by going to acquire mode selecting spectrum o o o Fill a cuvette with 2ppm solution and run with the settings Spectrum type: excitation EM wavelength: 400nm EX wavelength: 350nm – wavelength containing peak of interest Sensitivity: low Scan Speed: fast Recording Range: -10 to amt. showing entire peak EX/EM Slit width: 10/10 Response time: 0.02 Repeat Scan/Auto file: no setting needed/disable Then click Search wavelength and click search Set as 350-600nm and 240-650nm Record optical wavelengths Run each standard 3 times recording wavelengths found for EX and EM Data: Refer to page 52 for the data from the Fluorescence. Results: We ran multiple samples using the UV/VIS but after two days and running different samples at different concentrations we were unable to come out with any usable results. We never actually received any absorbances therefore I was unable to create a calibration curve. Fluorescence Calibration Curve: Calibration Curve for Quinone 1050 Intensity (c/s) 950 850 y = 241.9x - 39.907 R² = 1 750 650 550 450 350 1 Concentration (ppm) 1.5 2 2.5 3 3.5 Concentration (ppm) 4 Intensity (c/s) 2 443.887 3 685.778 4 927.679 Tonic Water had an intensity of 463.827, which = y 463.827 = 241.9𝑥 − 39.907 𝑥 = 2.08𝑝𝑝𝑚 Therefore Tonic Water has a concentration of 2.08ppm of quinone 4.5 Conclusion: I found running the UV/VIS to be very complicated because no matter what we tried we were unable to get results and it took us a while to actually start using the instrument due to issues with finding a usable cuvette. We made all of our samples both lab periods and tried running them but never got results. The Fluorescence gave us odd results at first but after trying it multiple times we finally got the instrument to work. We ran less samples than we wanted but we were still able to make a calibration curve with great linearity. I found the Fluorescence to be an easy instrument to use for the most part the most difficult part was the age of the computer and the fact that it kept freezing. I would like to try using the UV/VIS again so that I can get usable results.