Capillary Electrophoresis and IC

advertisement
Ion Chromatography and Capillary Electrophoresis
Purpose: Become familiar with the IC and CE machines and understand how they work and
then run samples with them and record data.
Background:
The IC (ion Chromatography) is used to separate and determine ions on columns with
relatively low ion-exchange capacity. It can be used for almost any kind of charged particle
including large proteins, amino acids, and small nucleotides. It is often used in water analysis,
protein purification, or quality control. The mobile phase, or buffer solution, inside the IC carries
the sample to the column where it is then analyzed.
The CE or CZE (Capillary Electrophoresis, or Capillary Zone Electrophoresis) contains
different ionic components of mixture migrate according to its own mobility. The buffer
composition is constant throughout the separation. The CE is commonly used for the
identification and determination on proteins. A high voltage power supply is used to move the
sample through the instrument to the detector.
IC Method:



read manual and figure out how to use the IC, develop SOP
run sample
Run standard sample and water samples
Procedure-Ion Chromatography:
Standard Operating Procedure:
 Turn on helium in gas room and behind instrument
 Check helium tank pressure (80psi) and pressure on instrument (9psi)
 Ensure that regenerant and eluent bottles are more than half full
o If less than half full, they must be refilled
 Equilibrate the instrument
o After 20-30 min, test flow rate (1 mL / 1 min)
 Set up the software
 Run the Sequence
o Inject sample when prompted to do so
o Each run should take 16 min.
 Analyze Peaks
 Shut Down
o Select Stop in Batch program
o Leave Chromeleon running
o Be sure to turn off the helium knob behind the instrument
Lab Procedure:
 Run four different water samples
o Do not run salt water or flavored water as this will harm the instrument!
 Run two samples per day
Procedure-Capillary Electrophoresis:
Chemicals Required:
 Capillary Performance Test Mixture B (mixture of p-hydroxybenzoic acid and phydroxyphenylacetic acid)
 Capillary Performance Run Buffer A (pH 8.35)
 Capillary Electrophoresis System Solution A (0.1 M NaOH)
 Distilled water
 1.0 M HCl Solution
 All chemicals should be stored in the chemical refrigerator or under the instrument
Standard Operating Procedure:
 Turn on the instrument
 Let it warm up for 30 minutes
 Read Dr. Hu’s SOP while the instrument is warming up
 Prepare and label chemicals
 Load buffer and sample vials according to Dr. Hu’s SOP
 Build Method using Dr. Hu’s SOP
o Time Program:
 Rinse 1.0 min (20 psi, forward pressure) from 1.0 M HCl to Waste
 Rinse 1.0 min (20 psi, forward pressure) from distilled water to Waste
 Rinse 2.0 min (20 psi, forward pressure) from 0.1 M NaOH to Waste
 Rinse 1.0 min (20 psi, forward pressure) from distilled water to Waste
 Rinse 2.0 min (20 psi, forward pressure) from Buffer A to Waste
 Inject (0.5 psi pressure, 10 seconds) from sample vial (Test Mixture B) to
Waste
 Separate at 0.0 min (Duration: 7 min, Voltage: 25 KV, Ramp: 0.17 min)
from Run Buffer A (BI) to Run Buffer A (BO)
 Autozero at time 1.50 min
 Stop data at 7.0 min
 Rinse 1.0 min at time 8.0 min (20 psi, forward pressure) from distilled
water to Waste
 Save and Run Method
 Save and print electropherogram with report
 If you are the last group, exit the program and turn off the instrument
o Remove all vials, except the buffer vials
o Load the trays before turning off the instrument
Data Analysis:
 Use peak area to calculate the percentage of each compound
 Predict which is p-hydroxybenzoic acid and which is p-hydroxyphenylacetic acid
Data:
The results are in my lab notebook.
Conclusion:
The IC and the CE were relatively easy to use after becoming acquainted with their procedure.
For the IC we ran a seven anion standard and then the standards and a sample. After running all
of these, the graphs were then analyzed and compared to one another to find similar structures of
molecules. The samples were water taken from the different chemistry labs and the nearby water
fountain. They all appeared to high levels of chloride and phosphate. For the CE many standard
were run and a unknown called mixture B.
Download