Capillary Electrophoresis SOP *Turn on and let warm up for 30 minutes. (There is a button on the front of the instrument) Set up: Prepare buffer and samples in 1.5mL glass vials Using a P1000 pipette add about 1.4mL (pipette 700mL two times) of Run Buffer A, Solution A (0.1M NaOH), Distilled Water, and 1.0M HCL Place cap on vials and take note of where they are placed in the loading tray. o In another vial, add about 1.0mL of Text Mixture B and place cap on On the computer: click direct control panel and then load o Wait until the buffer trays are brought forward and then open the transparent cover. o Insert buffer vials to the front left inlet buffer tray (BI) o Place another vial with Run Buffer A in the outlet buffer tray (BO) o Place sample in sample inlet tray (SI) o Place empty waste vial in sample outlet tray (SO) o Record final position of vials in notebook Running the sample: On computer, build method for running the sample based on following conditions DO NOT CHANGE THE INITIAL CONDITION PDA detector initial condition: use channel 214nm Time program: o Rinse 1 min (20 psi, forward pressure) from 1.0M HCl to waste o Rinse 1 min (20 psi, forward pressure) from DI water to waste o Rinse 2 mins (20 psi, forward pressure) from 0.1M NaOH to waste o Rinse 1 min (20 psi, forward pressure) from DI water to waste o Rinse 2 mins (20 psi, forward pressure) from Run Buffer A to waste o Inject (0.5 psi pressure, 10 seconds) from sample vial (text mixture B) to waste o Separate at 0.0 min (Duration: 7 mins, Voltage: 25 KV, Ramp: 0.17 min) from Run Buffer A to Run Buffer A o Autozero at time 1.50 mins o Stop data at 7.0 mins. o Rinse with DI water for 1 min at time 10 mins. ( or any time after sample is done running) After Running the sample Save and run the method (see manual for detailed instructions) Save and print electropherogram if printer works Exit program and turn off instrument