Capillary Electrophoresis SOP Instrument: P/ACE MDQ Capillary Electrophoresis System Instrument Prep: Turn on the instrument Let it warm up for 30 minutes Sample Prep: Prepare chemicals o Use five 1.5-mL vials o Add 1.4 mL each of Run Buffer A (x2), Solution A, distilled water, and 1.0 M HCl Do not overfill the vials o Cap these vials using the red caps from the drawer o Cap 2-3 empty glass vials to be used for waste o Use a plastic vial for your sample Fill with 1mL of Test Mixture B Cap using blue cap from the drawer Load buffer and sample vials according to Dr. Hu’s SOP o Direct Control/Load Wait until buffer trays are sent to the front, lift the cover, and insert vials as follows: BI:A1 – 0.1 M NaOH BI:B1 – 1 M HCl BI:C1 – distilled water BI: D1 – Buffer A BO:A1 – Waste BO:B1 – Waste BO:C1 – Buffer A SI:A1 – Test Mix B SO:A1 – Waste Record the position of each vial in your lab notebook Running the Samples: Build Method o Initial Condition: Leave all defaults o PDA Detector Initial Conditions: Use Channel 1 (check both boxes) at 214 nm, uncheck Acquisition Enabled (top left quadrant) o Time Program: Rinse 1.0 min (20 psi, forward pressure) from 1.0 M HCl to Waste Rinse 1.0 min (20 psi, forward pressure) from distilled water to Waste Rinse 2.0 min (20 psi, forward pressure) from 0.1 M NaOH to Waste Rinse 1.0 min (20 psi, forward pressure) from distilled water to Waste Rinse 2.0 min (20 psi, forward pressure) from Buffer A to Waste Inject (0.5 psi pressure, 10 seconds) from sample vial (Test Mixture B) to Waste Separate at 0.0 min (Duration: 7 min, Voltage: 25 KV, Ramp: 0.17 min) from Run Buffer A (BI) to Run Buffer A (BO) Autozero at time 1.50 min Stop data at 7.0 min Rinse 1.0 min at time 8.0 min (20 psi, forward pressure) from distilled water to Waste o Be sure to use the SO waste for the injection and be sure to use both of the BO waste vials throughout the time program Save and Run Method o Use Dr. Hu’s SOP for more details on how to do this Save and print electropherogram with report o If printer is not working, inform lab instructor or assistant Turning off the Instrument: If you are the last group, exit the program, remove your vials and turn off the instrument o Be sure to leave a buffer vial in the buffer inlet and buffer outlet trays o o o Choose to inject the tray Allow the tray to inject, and turn off the instrument This ensures that the capillary will not dry out CE Risk Assessment Form Definitions: H-Hazard P-Precautions needed R-Response to hazard Hydrochloric Acid H1.1 Inhalation P1.1 Work in fume hood R1.1 If inhaled, move person to fresh air H1.2 Corrosive-will burn skin P1.2 Wear gloves R1.2 Remove contaminated clothing, wash affected area with water for 15 minutes Sodium Hydroxide H2.1 Corrosive-will burn skin P2.1 Wear gloves R2.1 Remove contaminated clothing, wash affected area with water for 15 minutes