FTIR and Raman Fourier Transform Infrared Spectroscopy (FTIR) refers to a fairly recent development in the manner in which the data is collected and converted from an interference pattern to a spectrum. An FTIR spectrometer simultaneously collects spectral data in a wide spectral range. This technique shines a beam containing many different frequencies of light at once, and measures how much of that beam is absorbed by the sample. It can identify unknown materials, determine the quality or consistency of a sample and determine the amount of components in a mixture. Some of the major advantages of FTIR over the previous techniques include speed and sensitivity. It can be applied to the analysis of solids, liquids, and gasses. Raman spectroscopy is a spectroscopic technique used to study vibrational, rotational, and other low-frequency modes in a system. Raman spectroscopy is commonly used in chemistry, since vibrational information is specific to the chemical bonds and symmetry of molecules. Therefore, it provides a fingerprint by which the molecule can be identified. Raman spectroscopy can be used to study solid, liquid and gaseous samples. This instrument can be used for both qualitative and quantitative applications. Purpose: We will perform experiments using the FTIR and Raman that will help us learn how to operate these instruments. Data will be collected and a calibration curve will be made of the collected data. Method: FT-IR I. Open Resolution Pro Click on the Center burst on towards bottom of the page Ensure the black flap on the right side of the FT-IR is pushed down Ensure you see the center burst, and a number in the left corner ranging from 1.2-5 Check the parameters on the right side The scan dialog box should appear Check Parameters in Electronics Tab Scans to co add – 16 Save range – 4000-700cm-1 Check Parameters in Optics Tab IR-Source – Mid-IR Beam – internal Detector – DTGS-Det#1 Beamsplitter – KBR Broadband Accessory – None Source - 2cm-1 at 2000cm-1 Microscope Box – Detector: Left, Optics Mode: Reflectance o Go to Collect tab at the top Hit scan button Name the standard o Take a screen shot of your background & polystyrene Click the print screen key on the keyboard (located on the right side near top) Open up the Paint application in the start menu Hold ctrl+V to paste the picture into paint Save the image to the desktop and put it on a flashdrive to put into notebook Repeat the same steps for the polystyrene o You can now go back to the graph in the resolutions pro program Select the polystyrene sample graph and change the absorbance to % transmittance by selecting “% transmittance” under the transforms menu at top Varian FT-RAMAN Fill The RAMAN blue dewar with liquid nitrogen (wait 20 min) Make sure the FT-IR instrument is ON (located in back of instrument) Top off the dewer with (l) nitrogen after 10 min Turn on the power supply to the laser located on the floor (press switch and turn key to on) V. The two levers on either side of the raman must be switched to on. VI. The lever inside the raman should be up with the bypass closed. VII. Prepare sample and put solid in capillary tube to a depth of 1 inch VIII. Liquids are placed in a 4 inch nmr tube IX. There are diff. sample holders for the different tubes. X. Center the sample on the red dot on the holographic filter. XI. The red dot, is not actually red, it’s more of a pale pinkish. To find it, lower the lid about halfway and then adjust the X, Y, and Z knobs to center it. XII. Open Varian Resolution Pro on the desktop. XIII. From the current scan menu select raman scan 1. Ir source: off 2. Beam: right 3. Detector: raman ge 4. Beamsplitter: Quartz Near IR 5. ATR Crystal: None I. II. III. IV. 6. Otical filter: Holographic Notch 7. Aperture: Open XIV. Select laser tab click turn on diode XV. Press the shutter switch in front of raman XVI. Set the raman power to the highest of 3 XVII. Return to software and the laser tab and adjust the value of the laser control current to 600-700 mW XVIII. Click setup and center burst should appear. This is far less defined than the center burst seen with the ft-Ir. Click the autoscale icon at bottom of screen Using xyz knobs and adjust the position of the sample holder until the center burst peak is at a maximum. Adjusting the y position is most helpful. If you are having trouble finding the center burst, play around with the knobs for a few more minutes, but if you’re still unsuccessful, press the shutter button, open the lid, and re-center the sample on the red dot by eye. XIX. Click scan XX. Shutting down a. Turn of laser by selecting raman scan, and laser tab click turn off diode b. Press the shutter button on the front of the raman to close the shutter c. Remove the sample d. Turn off laser (turn of key, turn off switch) e. Do not turn off computer, ftir, or raman accessory Data: FTIR Microscope was currently not working so we could not perform an experiment on it. FTIR Sample One: Toluene Sample Two: 2-Butanone Sample Three: 2-Propanol Mixture One: (1:1:1) 33% Each Mixture Two: (1:1:2) 25% 2-Propanol/2-Butanone 50% Toluene Mixture Three: (1:2:1) 25% Toluene/2-Butanone 50% 2-Propanol Mixture Four: (2:1:1) 25% Toluene/2-Propanol 50% 2-Butanone Unknown (1 ppm) **NOTE: All spectral data is located in notebook Calibration Curves: 2-Propanol 98 % Transmittance 93 y = -0.7799x + 111.49 R² = 0.9829 88 83 78 73 68 20 25 30 35 40 Concentration (%) 50 2-Butanone 20 18 % Transmittance 45 y = -0.3426x + 26.702 R² = 0.936 16 14 12 10 8 20 25 30 35 40 Concentration (%) 45 50 Toluene 29.5 % Transmittance 27.5 25.5 y = 0.223x + 15.284 R² = 0.78 23.5 21.5 19.5 20.5 25.5 30.5 35.5 40.5 Concentration (%) 45.5 50.5 Results of Unknown: Contained all three compounds (Toluene, 2-butanone and 2-propanol) Toluene: 23% Transmittance 2-Butanone: 14% Transmittance 2-propanol: 96% Transmittance Raman: Sample One: Acetaminophen Sample Two: Toluene Sample Three: Methyl Salicylate **Data located in notebook** Conclusion: Unfortunately, we were only able to use two out of the three instruments because the microscope was not working. The FTIR and Raman were interesting to use. The most difficult part about the raman was trying to center the laser. That consumed the largest portion of our time. We realized that by aligning it with your eyes first before shutting it made a huge difference. The data for the raman was hard to be interpreted also. It was difficult to distinguish the correct peaks that uniquely identified that particular substance. We believe that a large amount of peaks were due to solvent effects. I am interested to see the variety of products that can be run through the FTIR because I could possibly use it in my independent study research.