Brittany Hicks FTIR & Raman Lab- Dr. Foy Spring 2013 FTIR and Raman Introduction Fourier transform infrared spectroscopy is a fairly recent instrument that collects data from an inference pattern to a spectrum. Data is continuously collected and concerted from a spectrum with a wide range. FTIR uses a beam that contains multiple frequencies of light to measure how much of the light is absorbed from the beam by the sample. FTIR can identify unknown materials to determine the quality and components of a mixture within a desired sample. This procedure can reveal quantitative and qualitative measurements. Raman spectroscopy is a spectroscopic technique to look at frequency modes of vibrational and rotational systems. Vibrational information relates specifically to bonds and symmetry within molecules. The molecule can ultimately be defined through these vibrational modes. Solid, liquid and gaseous samples can all be analyzed quantitatively and qualitatively through Raman spectroscopy. The purpose of this experiment is to first familiarize ourselves with the instruments in ultimately learning how to analyze and operate these instruments. Data will be collected and analyzed with the help of a calibration curve. Procedure - FT-IR Need to install IR part first Open Varian Resolution Pro Then go to current scan at bottom > Basic IR Scan Click on the Center burst on towards bottom of the page Ensure the black flap on the right side of the FT-IR is pushed down Ensure you see the center burst, and a number in the left corner ranging from 1.2-5 Check the parameters on the right side View - Interferrogram Aperture - 2cm-1 at 2000cm-1 Sensitivity – 1 Hit OK button Collect > Electronics Speed – 20 KHz Filter – 5 UDR – 2 Resolution - 0.5 Sensitivity – 1 Scans to co add – 16 Save range – 4000-700cm-1 Check Parameters in Optics Tab IR-Source – Mid-IR Beam – internal Detector – DTGS-Det#1 Beamsplitter – KBR Broadband Accessory – None Source - 2cm-1 at 2000cm-1 Microscope Box – Detector: Left, Optics Mode: Reflectance Insert blank Go back to the graph in the resolutions pro program Select the polystyrene sample graph and change the absorbance to % transmittance by selecting “% transmittance” under the transforms menu at top FT-IR with Microscope Procedure *** Referred to SOP located within the lab drawer. Procedure - Varian FT-RAMAN Fill The RAMAN blue dewar with liquid nitrogen Make sure the FT-IR instrument is on Top off the dewar with (l) nitrogen after 10 min Turn on the power supply to the laser located on the floor The two levers on either side of the Raman must be switched to on. The lever inside the Raman should be up with the bypass closed. Prepare sample and put solid in capillary tube to a depth of 1 inch Center the sample on the red dot on the holographic filter, adjust X,Y, and Z knobs if needed Open Varian Resolution Pro on the desktop. Current Scan Menu IR source: off Beam: right Detector: Raman ge Beamsplitter: Quartz Near IR ATR Crystal: None Optical filter: Holographic Notch Aperture: Open Select laser tab click turn on diode Press the shutter switch in front of Raman Set the Raman power to the highest of 3 Return to software and the laser tab and adjust the value of the laser control current to 600-700 mW Click setup and center burst should appear. Adjust X,Y,Z knobs to find center burst Click scan Shutting down Turn of laser by selecting Raman scan, and laser tab click turn off diode Press the shutter button on the front of the Raman to close the shutter Remove the sample Turn off laser Do not turn off computer, FT-IR, or Raman accessory Data Collected Below are calibration curves for some of the known samples recorded. All other data recorded by the FT-IR is located in my notebook. All the spectra for the samples run are included in my notebook to show further analysis. Toluene 29.5 % Transmittance 27.5 25.5 23.5 21.5 19.5 20.5 25.5 30.5 35.5 40.5 Concentration (%) 45.5 50.5 2-Butanone 20 % Transmittance 18 16 14 12 10 8 20 25 30 35 Concentration (%) 40 45 50 40 45 50 2-Propanol 98 % Transmittance 93 88 83 78 73 68 20 25 30 35 Concentration (%) For the IR- Microscope, a total of 10 samples were analyzed plus an additional background scan to set as a baseline spectra for all other additional samples. Unknown fibers were compared to the known spectrum to look at similarities among peaks. *** All spectra are located within my lab notebook The Raman instrument was down and unfortunately we did not get the opportunity to use this instrument for analysis. Conclusion The FT-IR and IR-Microscope were very easy to use and actually quite interesting when analyzing data. For the FT-IR, we were able to obtain very good results for three known organic samples: Propanol, Butanone, and Toluene. The unknown samples ran had very similar peaks to these known samples which suggests they are comprised of similar components within their structures. The IR microscope was also very easy to use and produced great results. We analyzed 10 different fiber samples under the microscope. The unknown samples were compared to the spectra of the 8 known fibers ran. Unknown C had similar wavelength and transmittance to the fiber polyester sample. Unknown B looked similar to the PVC sample we recovered from the microscope. Overall this lab was fun and was more relatable to my interest as a biologist.