IR, Raman, and IR Microscopy Lab Introduction: The Fourier Transform Infrared Spectroscopy probes the bond types of sample analyte. It also measures absorbance, emission, transmittance, and photoconductivity of a solvent or mixture of solvents. This is accomplished by infrared radiation passing through and a sample. As it passes through a sample, some radiation gets absorbed by the analyte depending on the chemical bonds that are specific to each functional group in a sample. The rest of the radiation is transmitted. The FTIR can be used to determine the components and concentration of a sample. The Raman Spectroscopy focuses the infrared beam on a specific area of a sample. This spectroscopy can locate and probe a small sample. It is a process of inelastic scattering on monochromatic laser light. This can analyze solids, liquids, and most gases. Purpose: The objective for this experiment is to become familiar with the FTIR and the Raman Spectroscopy. For the FTIR, we will run three organic solvents and identify components of the spectrum. We will then create a calibration curve by running different mixtures of the three solvents and observing their wavelength. We will then run an unknown mixture. For the Raman Spectroscopy, we will analyze fibers with the IR in Microscopy mode. Procedure: Procedure: FT-IR: 1. First always make sure the instrument is turned on, but it is usually left on at all times. 2. Check to see if black lever between IR and Ramen is pushed down. 3. Once this is done, open the “Varian Resolutions Pro” on the computer desktop; When opened click on ‘current scan’ and select ‘basic IR’; then at the top of the screen hit “collect” and select ‘rapid scan’ and follow the parameters. 4. Select optics tab and set parameters but not change them anytime during the experiment; click setup and should get a reading between 1 and 5, if this not happen adjust the measure on the right side until there’s a reading between 1 and 5. 5. Perform a background scan and make sure there is no sample in there, if using the trough make sure it is empty; when background is completed run your sample, go back to collect and rapid scan and select scan, a spectrum should appear and will be displayed as absorbance. To get it to % transmittance, select “transform” from the menu and then select %T. 6. To shut down close and save programs but do NOT shut off the IR. Raman: 1. Fill blue dewar with liquid nitrogen until it overflows, let it sit for 20 min. and then top it off with more liquid nitrogen. 2. When that is ready turn the laser power on and turn the key behind the instrument, it will display ”ready” when it has cooled down enough. Make sure the black lever between the IR and Raman is up and the one between Raman and microscope is down, the lever inside the Raman should be up as well. 3. Load the sample and make sure to center it on the red dot and then start the program. Click “Varian Resolutions Pro”, select current scan and switch to Raman Scan. Go to collect at the top of the menu and select Raman scan, go to the optics tab and check the parameters; you may have to change them from the IR parameters, but after that do not change them. 4. Select the laser tab and click “turn on diode”, press the shutter which on and set the Raman power to the highest of 3 ranges. 5. Adjust the power of “laser control current” until laser power reads 600-700mW. 6. Click setup and you should see a center burst, most aren’t really defined but are noticeable, if no center burst appears, move the X and Y nobs on the front of the Raman to center the sample more; you may have to open it again and re-center it by eye and then it should work. 7. Click ‘ok’ and hit scan and spectrum should appear; no background scan is required for this scan. 8. To shut down, turn off laser by clicking “turn off diode” in the laser tab, press the shutter button and remove sample, turn off laser power supply then the key and switch, do opposite of how it was turned on. Do NOT turn off computer or Raman/IR. Data: 2 butanone % % composition Transmittance 100 2 50 7 33 10 25 15 2 propanol % % Composition Transmittance 100 28 50 46 33 55 25 67 Toluene at 1500 cm-1 % % Composition Transmittance 100 11 50 30 33 41 25 48 Toluene at 1600 cm-1 % % Composition Transmittance 100 35 50 64 33 70 25 79 Results: 2-Butanone y = -0.151x + 16.351 R² = 0.8702 16 % Transmittance 14 12 10 8 6 4 2 0 0 20 40 60 80 100 120 % Composition 2-Propanol y = -0.4715x + 73.516 R² = 0.9324 80 % Transmittance 70 60 50 40 30 20 10 0 0 20 40 60 % Composition 80 100 120 Toluene at 1500 cm y = -0.4729x + 57.092 R² = 0.9731 60 % Transmittance 50 40 30 20 10 0 0 20 40 60 80 100 120 % Composition Toluene at 1600 cm y = -0.5624x + 91.244 R² = 0.9896 90 % Transmittance 80 70 60 50 40 30 20 10 0 0 20 40 60 % Composition Equations: 2- Butanone y = -0.151x + 16.351 2- Propanol y = -0.4715x + 73.516 Toluene at 1500 cm y = -0.4729x + 57.092 Toluene at 1600 cm y = -0.5624x + 91.244 80 100 120 Unknown C=O peak 9% = -0.151x + 16.351 x = 48 % 2 - Butanone OH peak 103% = -0.4715x + 73.516 x = error < 25 % Toluene 47% = -0.4729x + 57.092 x = 21 % Toluene Conclusion: For the 3100 Excalibur FT-IR, we ran three organic compounds and mixtures of 1:1:1, 2:1:1, 1:2:1, 1:1:2, and un unknown. Calibration curves were created. The Toluene peak has good results. However, the 2- Butanone and 2- Propanol curves did not have good R2 values. When calculating the percent composition yield of unknown, 21 % Toluene was determined. We calculated 48% 2- Butanone in the unknown but with the errors in the value, it is not likely. When calculating percentage of 2- propanol, an error was observed. There may have been many errors in this experiment. First the mixture concentrations could have not been precise and therefore altering the calibration curves. Also, the trough could not have been completely clean, therefore having contamination, and picking up incorrect peaks. While running the varian FT-raman, we had difficulty in successfully getting a scan. At first 2propanol was attempted and a centerburst did not appear. After asking for help and still not getting a centerburst, Acetaminophen was attempted. A centerburst appeared but was not very clear. A scan was run (in notebook) and peaks were observed but there was a lot of noise. After refilling the liquid nitrogen, acetaminophen was rerun and there was a little less noise. Methyl salicylate was then run and picked up a good scan, with little noise. Errors in this experiment could have been not centering the pink dot and not being able to find the centerburst. On the second day, we attempted to run the microscopy. We were not able to run this successfully. The voltage range should have been between 1.5 and 5 volts. Even after we had the highest sensitivity, the voltage was only around 1 volts. We attempted to scan fabrics although there was a low voltage. While attempting to run an IR, the results came out poor. We were able to use the microscope and get a clear image of the fabric. After capturing, we had difficulty finding the image and when printed, the ink ruined the image.