FTIR, IR microscopy, and Raman Purpose: Become familiar with the FTIR and FTIR microscopy. For the FTIR, organic solvents will be examined and for the microscope portion, fibers will be analyzed. (Raman was down) Background: Fourir Transform infrared spectroscopy is generally used to probe the bond types of the sample analyte. This technique is generally used for qualitative analysis. Using FTIR microscopy allows us to focus the infrared bean on a very specific area of a sample, or locate and probe a small sample. A common use of this technique is to look at fibers. The Raman uses scattered radiation that corresponds to the vibrational modes of the molecule. Procedures: IR Open Varian Resolution Pro go to current scan at bottom > Basic IR Scan Click on the Center burst on towards bottom of the page Ensure the black flap on the right side of the FT-IR is pushed down Ensure you see the center burst, and a number in the left corner ranging from 1.2-5 Check the parameters on the right side View - Interferrogram Aperture - 2cm-1 at 2000cm-1 Sensitivity – 1 Hit OK button Collect > Electronics Speed – 20 KHz Filter – 5 UDR – 2 Resolution - 0.5 Sensitivity – 1 Scans to co add – 16 Save range – 4000-700cm-1 12. Check Parameters in Optics Tab IR-Source – Mid-IR Beam – internal Detector – DTGS-Det#1 Beamsplitter – KBR Broadband Accessory – None Source - 2cm-1 at 2000cm-1 Microscope Box – Detector: Left, Optics Mode: Reflectance 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 13. Insert blank 14. Go back to the graph in the resolutions pro program Select the polystyrene sample graph and change the absorbance to % transmittance by selecting “% transmittance” under the transforms menu at top 15. To Shut Down: close Varian resolutions Pro and leave everything on. IR Microscope 1. 2. 3. 4. 5. Fill the Microscope detector with liquid nitrogen Insert ATR crystal To set up software, open software by click ‘varian resolutons pro’ From ‘current scan’ drop down, select method file ‘microscope’ From collect menu, select rapid scan, check that parametres are set up same as above procedure states 6. Run a background scan and then run the samples Raman 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. Fill The RAMAN blue dewar with liquid nitrogen Make sure the FT-IR instrument is on Top off the dewar with (l) nitrogen after 10 min Turn on the power supply to the laser located on the floor The two levers on either side of the Raman must be switched to on. The lever inside the Raman should be up with the bypass closed. Prepare sample and put solid in capillary tube to a depth of 1 inch Center the sample on the red dot on the holographic filter, adjust X,Y, and Z knobs if needed Open Varian Resolution Pro on the desktop. Current Scan Menu IR source: off Beam: right Detector: Raman ge Beamsplitter: Quartz Near IR ATR Crystal: None Optical filter: Holographic Notch Aperture: Open Select laser tab click turn on diode Press the shutter switch in front of Raman Set the Raman power to the highest of 3 Return to software and the laser tab and adjust the value of the laser control current to 600-700 mW Click setup and center burst should appear. Adjust X,Y,Z knobs to find center burst 16. Click scan 17. Shutting down Turn of laser by selecting Raman scan, and laser tab click turn off diode Press the shutter button on the front of the Raman to close the shutter Remove the sample Turn off laser Do not turn off computer, FT-IR, or Raman accessory Data: FTIR: Toluene 2-Propanol 2-Butanone % %T % %T % %T 100 40.111 0 n/a 0 n/a 0 n/a 100 20.246 0 n/a 0 n/a 0 n/a 100 0.054294 25 74.956 25 68.76 50 4.1811 25 73.473 50 44.002 25 14.052 50 59.428 25 63.664 25 18.065 33 65.393 33 58.589 33 3.9065 55.80993568 59.525 7.072249589 75.143 42.44863388 8.8219 Samples: 1. 2. 3. 4. 5. 6. 7. 8. 2-Butanone 2-Propanol Toluene Mixture 1: Toluene, 2-Propanol, 2-Butanone (1:1:2) Mixture 2: Toluene, 2-Propanol, 2-Butanone (1:2:1) Mixture 3: Toluene, 2-Propanol, 2-Butanone (2:1:1) Mixture 4: Toluene, 2-Propanol, 2-Butanone (1:1:1) Unknown Calibration Curves: % Transmittance (%) Toluene 80 70 60 50 40 30 20 10 0 y = -0.4399x + 83.173 R² = 0.9698 0 20 40 60 80 100 120 Mass percent (%) 2-Propanol 80 y = -0.6095x + 79.454 R² = 0.9677 % Transmittance (%) 70 60 50 40 30 20 10 0 0 20 40 60 80 Mass percent (%) 100 120 2-Butanone %Transmittance (%) 20 15 y = -0.1833x + 16.593 R² = 0.5756 10 5 0 0 -5 20 40 60 Mass percent (%) 80 100 120 Conclusion: We were able to get results with the FTIR and with the IR microscope. The Raman was not working so we were not able to use it. We got good results without liquid samples on the FTIR but we ran into a problem at the end. The library on the computer was either not working or we were working it wrong although we followed the directions. For each sample, the library would give us over 1,000 hits instead of narrowing it down for us. If we had an actual unknown and not one with just unknown proportions, identifying the sample would be extremely difficult. Looking at the calibration curves, the one for the 2-Butanone was not very good. The R2 value was only 0.5756 which is really low. There could have been a contamination possibly between changing out samples which could result on sample not being pure.