Tim Abell IR and IR Microscopy Introduction Fourier Transform Infrared Spectroscopy (FTIR) uses light in the infrared range to determine the functional groups of a sample. Each functional group emits a unique wavelength when exposed to infrared light. The IR microscope works in a similar way to the FTIR but is used to identify what fabrics are made up of. Purpose The purpose of this lab is to run both known and unknown samples on the FTIR and the IR Microscope. The spectra then were interpreted. Three organic solvents were run on the FTIR and fabric ssamples were looked at on the IR Microscope. Method IR Make sure than the instrument is on. Make sure that the black lever between the FT-IR and the Raman is down Double click on the Varian Resolutions Pro icon in the top right of the display. In the Current scan menu on the bottom of the screen select IR scan. From the Collect menu at the top of the screen select Rapid scan The Electronics tab parameters should be: speed: 20 KHz Filter: 5 UDR: 2 The Optics tab parameters should by: IR Source: Mid-IR Beam: internal Detector: DTGS Beamsplitter: KBr Broadband Accessory: none Optical filter: none Aperture: 4 cm-1 at 2000 cm-1 Click setup. The voltage reading should be between 1-5V. If not, change the selectivity Click background to run a background scan Return to rapid scan from the collect menu and click scan The transformers menu allows you to convert to %T To shut down, close Varian Resolutions Pro and LEAVE EVERYTHING ON IR-microscope Fill microscope with liquid nitrogen Instrument is full when liquid nitrogen comes out side of funnel’s black tube Let instrument stabilize for 15 minutes ATR crystal is located in Varian FTIR drawer Remove red cap from crystal Make sure notes are facing left, and slide crystal into microscope barrel, you should hear 2 clicks The black lever should be up between the IR and Raman, and between the Raman and IR microscope Open Raman lid, push the black lever in the back right corner down (set to Bypass Open) and close lid On the Raman instrument control panel (on front of instrument), set button to Bypass On microscope turn up brightness, using switch Open up Varian Resolution Pro program From drop down menu with Current Scan, select Microscope From Collect menu, select Rapid Scan, Under Electronics tab set parameters: o Speed: 20KHz o Filter: 5 o UDR: 2 o Save Range: 4000-700cm-1 Under Optics tab, set parameters o IR Source: Mid-IR o Beam: Right o Detector: UMA600MCT o Beamsplitter: KBr Broadband o Accessory: UMA-600 o ATR Crystal: Germanium o Optical Filter: None o Source: Open o Detector: Right o Optics Mode: Reflectance Click Setup Center burst should appear with voltage reading between 1.5-5 (if not adjust sensitivity), click OK Collecting Background, put nothing under scope From Collect Menu, select Rapid Scan > Setup A center burst should appear, click Background Load sample on gold plate From Collect menu, select Rapid Scan, and click Setup A center burst should appear In View menu (right hand side of screen) select Processed Carefully raise stage using Course adjustment knob until you feel minimal pressure on sample and slide IR spectrum should appear on screen, use fine adjustment knob to maximize signal Click scan to collect spectrum, graph should be in % Transmittance To shut down, turn down brightness on microscope, take out crystal and put red cap back on, and put in drawer. Put gold plate in drawer, put levers back and turn off bypass button. Close Varian Resolution Pro program The Raman was not working during the week of doing this lab, so we do not have a procedure or data for that instrument. Data All spectra are attached in my notebook p. 18 Results After analyzing known organic solvents on the FTIR we looked at two unknowns. The unknowns were just a mixture of the two knowns in different proportions. The first unknown was one part 2-propanol and two parts 2-butanone, a 1:2 ratio. Unknown 2 had a ratio of 1:1. We analyzed a few known samples on the IR Microscope but did not have time to analyze and unknown samples. Conclusion The instruments were pretty easy to use even the SOP was not very clear. All spectra that were obtained were easy to interpret and understand. There were a few problems with the lab. There was no student SOP online so we had to use then one in the lab. The other problem was that the Raman laser was not working so we did not get to analyze any samples using it.