Shelby Bleile UV/VIS & Fluorescence Introduction: The technique using the UV/VIS is primarily quantitative. It is used to test within the range of electromagnetic spectrum from ~200nm to ~800nm. It measures the absorbance through the bonding electrons of a molecule that is being tested. Fluorescence can be used to detect very low levels of analyte molecules. Not all molecules can utilize fluorescing so the instrument is very selective in that way. Purpose: The purpose of this lab is to become familiar with the UV/VIS and fluorescence. To also run standards and samples on both of the instruments Procedure: UV-Vis 1. Investigate the absorbance properties of SA by preparing stock solutions of SA in hexane. Dilute the stock to an appropriate range. 2. Create an unknown to prove the curve Fluorescence: 1. Create standards 2. Run standards 3. Run unknown samples based on the calibration range Results: Table 1. Results from Fluorescence Conc. of SA Excitation Emission Wavelength Intensity 10 ppm 327 nm 379 nm 373 nm 355.612 15 ppm 281 nm 376 nm 370 nm 410.410 20 ppm 279 nm 375 nm 371 nm 427.215 25 ppm 277 nm 373 nm 372 nm 458.092 30 ppm 277 nm 371 nm 372 nm 444.247 Fig. 1 Salicylic acid at 10 ppm Fig. 2 – Salicylic acid at 15 ppm Fig. 3 – Salicylic acid at 20 ppm Fig. 4 – Salicylic acid at 25 ppm Fig. 5 – Aspirin sample Fig. 6 – Graph showing all absorbance spectra for samples and standards Discussion: This lab went pretty well. We became familiar with the UV/VIS and the fluorescence. We were able to successfully run standards and samples on both of the instruments. We were able to get absorbance and transmittance values on different concentrations of salicylic acid in hexane on the UV/VIS instrument. We were also able to obtain results from the fluorescence on samples. We obtained optimal wavelengths, intensities, excitation, and emission values for the different samples of salicylic acid.