Infrared Spectroscopy Chemistry 243 Infrared spectral regions cm -1 μm 10 1 4 Hz μm cm c cm /s What is IR measuring? Radiation not energetic enough to promote electronic transitions, but rather vibrational and rotational motions Absorption results in an increase in amplitude of the motion Selection rule: Molecule (mode) must undergo a net change in dipole moment to be IR active This determines whether a given mode absorbs in the IR or not Types of molecular vibrations Stretches and bends Total number of vibrational modes (not all active): 3N-6 3N-5 for linear molecules N = total number of atoms involved in motion Skoog, Fig. 16-2 Harmonic oscillator, Hook’s law and anharmonicity Harmonic oscillator model F ky E 1 2 ky 2 Anharmonic oscillator model in blue Skoog, Fig. 16-3 A real vibrational transition acts like an anharmonic oscillator due to coulombic repulsion at short distances, and dissociation at long distances; different limit conditions than a harmonic oscillator Vibrational frequencies: IR (and Raman – more later) For 2 atoms, vibration m 1 k 2 frequency, 1/ 2 (in Hz) k force constant m (wavenumbe -1 rs in cm ) c (N/m kg/sec reduced mass (kg) m 2 ) c speed of light (cm/sec) m1 m 2 m1 m 2 Vibrations are quantized—can only assume discrete energies If energy is absorbed in the IR or Raman region, that’s the E Vibrational energy levels are equally spaced and transitions can only occur if the vibrational quantum number changes by ± 1. “Forbidden” overtones are observed at higher frequencies as a result of anharmonicity (v = ± 2,3 observed at 2 and 3, respectively) Instrumentation for measuring IR absorption Dispersive spectrometers Fourier transform spectrometers Source deliberately has a broad emission spectrum Use a grating monochromator to select wavelength Wavelength scanned by rotating grating Multiplexed measurement that has fewer optical components No gratings, slits, etc. Most common; the Michelson interferometer for IR Nondispersive photometers Use filter (or absorbing gas as a filter) to select specific probe wavelengths for a given atmospheric gas Dispersive instruments for IR (Dispersive = intentional use of many wavelengths) Generally double-beam dispersive instruments Lessens demands on sources and detectors Double Beam in Time Double beam geometry compensates for drift issues Low intensity sources and low sensitivity detectors Compensates for environmental interferences (such as water and CO2 in the lab atmosphere) Sample located before monochromator Skoog, Fig. 16-11 % Transmittance Typical IR Spectrophotometer Output Skoog, Fig. 16-10 Why %T instead of Absorbance? Just tradition. IR sources—low intensity Inert solids heated between 1500-2200 oK Nernst glower Best for far-IR ( > 50 m) Tungsten filament lamp Resistively heated nichrome or rhodium wire—lower intensity but longer life High pressure Hg arc lamp Silicon carbide rod—better than Nernst for >5 m Incandescent wire Rare earth oxides that is resistively heated Globar Approximately blackbody, continuum peaked at ~2 m Convenient for near-IR—often comes with UV-vis spectrometers CO2 laser Tunable through ~100 discrete lines between 900-1100 cm-1 Very high power (~2 orders better than thermal sources) Valuable for light detection and ranging (LIDAR) IR transducers—low sensitivity Pyroelectric Material that has strong temperature dependent polarization IR absorption leads to heating that in turn causes a current change in the external circuit connected to the capacitor Fast response allows time-domain interferometric measurements Photoconducting Thin film of material that promotes valence electrons to conduction band upon absorption of IR PbS for near-IR @RT, MCT (HgCdTe) for mid- and far-IR @ 77oK Thermal—limited by surrounding thermal noise Thermocouples Junction of wires that generate a potential difference with temperature variation (30 ms response) Thermistor (Bolometer) Single material that has a resistance change as a function of temperature (few ms response) Result: Use FT spectrometers for IR absorption Typically a Michelson interferometer Light throughput (Jaquinot) advantage Reduced number of optical elements (no slits needed for time data that will be transformed into frequency data) leads to more transmitted light (greater S/N) Wavelength accuracy (Connes) advantage This is the approach that takes us from time domain measurements to frequency domain results Superior frequency data can be converted to Multiplex (Fellgett) advantage Entire spectrum taken simultaneously in ~ 1 second or less Many more spectra can be acquired in a given time frame and thus S/N increases as 𝑛 Advantages of FT spectrometers 2-3 orders of magnitude higher throughput Partially offset by lower sensitivity of fast-response detector No stray light High resolution (<0.1 cm-1) Highly accurate and reproducible frequency determination Rapid scanning—few seconds for good spectra Good S/N ratios Particularly in mid-IR (4000-200 cm-1) Radiant Power Measuring optical spectra in time rather than frequency space Skoog, Fig. 7-41 Measuring optical spectra in time rather than frequency space The time-domain signal contains the same information as the frequency domain signal Light throughput (Jaquinot) advantage Wavelength accuracy (Connes) advantage Multiplex (Fellgett) advantage You can convert one to another, in either direction P ( t ) k cos ( 2 1 t ) k cos ( 2 2 t ) k constant t time Measuring optical spectra in the time domain but converting to frequency domain Measuring optical spectra in the time domain but converting to frequency domain Here, w = frequency Problem!!! How can a signal fluctuating at 5.5 x 1013 Hz be measured in time? Would require a sub-picosecond response time photodetector! Michelson interferometer Converts high frequency optical signal (1012 - 1015 Hz) to observable frequency (otherwise too fast to observe directly) Radiation is split into two beam paths of varying path length. Recombination gives interference that is dependent upon difference in path length for a given optical frequency. Path length is dynamically changed to give a lower frequency read-out. could be called D, but it’s called the mirror retardation partially reflective; about 50/50 Fixed focusing mirror Skoog, Fig. 7-43 Interferograms – the output of a Michelson interferometer; typically for optical work Monochromatic Dual Many clumped ’s Observed pattern Detector signal versus d Detector signal FT into frequency spectrum Skoog, Fig. 7-44 Advantages of Fourier transform optical measurements Light throughput (Jaquinot) advantage Wavelength accuracy (Connes) advantage Reduced number of optical elements (no slits needed for time data that will be transformed into frequency data) leads to more transmitted light (greater S/N) Superior frequency data can be converted to Multiplex (Fellgett) advantage Entire spectrum taken simultaneously in ~ 1 second or less Many more spectra can be acquired in a given time frame and thus S/N increases as 𝑛 Advantage in IR spectroscopy, NMR, MS, MRI If data is squeezed in time, it is stretched out in Hz Gratings vs. Interferometers Frequency vs. Time Domain Measurements Gratings use frequency or Multiple wavelengths require scanning or array detectors Typically no real-time wavelength reference Easier conceptual operation Multiplex - disadvantage in IR - advantage in UV An interferometer uses the time domain Multiple wavelengths inherent; FFT yields a freq-sorted spectrum Real-time wavelength reference Higher light throughput at constant resolution Multiplex - advantage in IR - disadvantage in UV Shot and flicker noise, in contrast to detector noise, increase with radiant power Detector noise is prevalent in IR, so interferometry is better. Not so much for UV-Vis Single-beam FTIR spectrometer Skoog, Fig. 16-8 Double-beam FTIR spectrometer Why? Skoog, Fig. 16-9 General comments on analytical IR spectrometry More challenging than UV-vis because there aren’t many good solvents Water is a problem Useful for quantitative analysis Mid-IR Strong IR absorption and dissolves many window materials CO2 is a problem—gives background signal Near-IR Need high concentrations with low pathlength cells Mostly good for qualitative organic compound analysis Far-IR Useful for inorganic studies with low frequency vibrations Qualitative Mid-IR spectrometry Organic chemistry Skoog, Fig. 17-5 See Fig. 17-6 for more Reflective IR spectrometry Diffuse reflectance Very useful for powdered samples Peak positions (but not necessarily intensities) same as absorption spectra. Attenuated total reflection (ATR) Allows analysis of complex samples and is not sensitive to sample thickness Spectra similar to standard absorption with exceptions near strong absorption bands Skoog, Fig. 17-11 & 13 Raman scattering Raman scattering discovered by C.V. Raman in 1928— received 1931 Nobel Prize for physics Excitation by a high intensity source to a non-quantized “virtual state” Emission of a new photon at energies plus or minus the vibrational energy can be measured Emission at same wavelength is Rayleigh scattering Raman spectroscopy Stokes and anti-Stokes lines are observed. Skoog, Fig. 18-1 Ratio of Stokes and anti-Stokes is temperature dependent Why? Very weak—best case is 0.001% of incident power Similar information as in IR spectra, but different selection rules make techniques complementary Selection rule requires change in polarizability (induced dipole) For molecules with center of symmetry, same mode cannot be Raman and IR active Measurements can be made in the visible spectral region Emission Energies This is a good thing! For 2 atoms, vibration m 1 k 2 frequency, m same as m 1/ 2 (in Hz) k force constant (N/m kg/sec reduced mass (kg) m1 m 2 m1 m 2 2 ) Raman instrumentation Laser excitation required because high intensity is needed to get measurable scattering Spectra almost always collected at 90o to avoid intense excitation light Fiber optic grating and filterbased Raman spectrometer BP is a bandpass filter and BR is a band-rejection filter (usually a notch filter) used to remove Rayleigh scattering. FT-Raman Applications of Raman spectroscopy Inorganic species Organic species Can look at aqueous solutions Metal-ligand bonds at 100-700 cm-1 hard to study with IR absoprtion Valuable for qualitative analyses for IR silent modes Biological applications Quantitative applications Raman spectra are less crowded than IR. Resonance and surface-enhanced Raman spectroscopy (SERS) Resonance Raman Excitation near an electronic transition can enhance Raman signals from symmetric vibrations by 102-106 Surface-enhanced Raman scattering (SERS) When the scatterer is located proximal to metallic nanotructures that support large electromagnetic fields, Raman signals are enhanced up to 1012 This has allowed single molecule detection!!! Van Duyne, et.al.; Anal. Chem. 2005, 77, 338A-346A. Comparison of IR and Raman Complementary techniques used to probe vibrational energy levels of molecules For 2 atoms, vibration frequency, m Raman scattering is incredibly 1/ 2 weak, but has the advantage of 1 k (in Hz) m visible region operation 2 More applicable in aqueous 2 k force constant (N/m kg/sec ) (biological) environments More material (cell) flexibility m1 m 2 reduced mass (kg) More sensitive detectors m1 m 2 available Raman rapidly becoming more widespread as peripheral technologies improve