Elemental, My Dear Watson Paul Middlestead University of Ottawa, G.G. Hatch Laboratory For 20th Continuous Flow Conference, Davis, 2014 Who/what is this talk for? This talk is for new users We will touch the basics on Elemental analysers A mix bag of tricks and advices Refer to manufacturer’s instruction manuals Only endorsed products are: Rickards Red and Rickards White ! Meet my family Large capacity EA TC/EA Combustion Pyrolysis NCHS HO S only TOC Stable isotopes 12000C 15000C Meet Grand Pa Leco is the most popular manufacturer of elemental analysers. Every department has one or two of those accumulating dust. Static combustion using oxygen: not suitable for IRMS. Let’s touch on •Elemental Analyser as an…instrument •Schematics •Autosamplers •Flash combustion •Chemicals & Configurations •Gas Chromatography •Thermal Conductivity Detector •Typical run •Common problems •EA-IRMS EA as an …. instrument Simultaneous determination of Nitrogen, Carbon, Hydrogen, Sulfur, Oxygen Measuring range: 100 ppm to 100% Sample size: 0.1 to 1000 mg / 0.1 to 25 ul Detection limit: 10 ppm Accuracy: 0.3% - 0.02% absolute Manufacturers, models blablabla CE instruments (Carlo Erba) models 1108, 1110, NA 1500, NA 2100, Flash 1112, Flash 2000 Costech Model ECS4010 Elementar cube family Eurovector models EuroEa3028-HT, Ea3024IRMS, pyrolysis model Sercon-Integra2 TC\EA Thermo … EA : Bulk analysis of NCHS/O Organic compounds Pharmaceuticals Organometallics Petrochemicals Gasoline & fuels Graphite Carbides & nitrides Metals & alloys Polymers Explosives Hydrocarbons Soils Coal & coke Liquids In short, you can combust your grandmother! Typical EA setup with IRMS Ref gas 1 Ref gas 2 Elemental Analyser (CNS) (OH) He 2m Interface Isotope Ratio Mass Spectrometer EA picture Elemental Analyser Schematic He O2 Autosampler Gas chromatography Column in oven Thermal Conductivity Detector Water Trap TCD N2 Combustion Chemicals 1020C Reduction Chemicals 650C CO2 H2O SO2 EA line out Found the problem with the previous slide? A beer if you do… (5 seconds) Elemental Analyser Schematic He O2 Autosampler Gas chromatography Column in oven Thermocouple Detector Water Trap TCD N2 Combustion Chemicals 1020C Reduction Chemicals 700C CO2 H2O SO2 EA line out A cold Rickard’s Red for me please. Typical Analysis Procedure for EA 1. 2. Sample/Std is weighed and wrapped in tin foil, placed in culture tray/autosampler Sample drops is flash-combusted in O2 in He stream, causing rapid cracking and oxidation of gases stoichiometrically equivalent to their elemental components, some gas oxides need to be reduced before obtaining final products (N2, CO2, H2O, SO2) i.e. oxidation and reducing reactions involved Typical Analysis Procedure for EA 3. Separation of gases by gas chromatography or chemical column traps 4. Detection by thermal conductivity detector (TCD) or IRMS. 5. Curve made from weighing certified stds (for isotope and/or quantitative measurements) Some torture tools… Well well, got culture? Super size me MAIS C’EST IMPOSSIBLE! C’EST FANTASTIQUE! Filters? Careful! Not flat please Autosamplers (AS200) Zero Blank Zero Blank modified Special tools required ECS 4010 Be good to yourself Take 20 minutes to study the gas schematics of your instrument Sample dropping Combustion starts Injected O2 Sample drops Flash should occur within 2-3 seconds of being dropped START 0 10 20 seconds 30 40 Why you Tin man? • • • • PLATINUM: Absorbs much of the heat of the reactor before passing it on to the sample, slowing down reaction ALUMINIUM: Supplies good oxidation flash and prime sample ignition but does not melt or mix with the sample thus does not promote inner oxidation SILVER: Melts at 9600C, does not take part in the combustion, retains trace of carbon when in molten state TIN: Inexpensive & takes active part in process. Melts at 2350C with very low enthalpy, intermixes with organic and inorganic substances, expedites the final oxidation reaction Let’s get cracking C N Combustion tube Reduction tube CO2 CO2 N2 + NxOx N2 H H2O H2O S SO2 + SO3 SO2 Chemicals and Configurations • System should be optimized for element(s) to be analyzed • One-tube system: both oxidant catalyst and reducing chemicals in one tube; usually used if S is to be analyzed • Two-tube system: one tube for oxidant, one tube for reducing chemicals; usually used for N, NC or NCH • Chemical traps: Mg Perchlorate or Anhydrone will trap H2O, and Carbosorb will trap CO2 (and SO2) Configuration CN Configuration S Small is beautiful More torture tools Using grinding tools and diluted Nitric acid, one can re-use combustion and reduction tubes Gas Chromatography Gas chromatographic columns will separate different components according to their polarity and molecular size. Factors influencing the quality of the chromatography: column length, size of packing, tube diameter, stationary phase type, flow rate, temperature. Packed column: packed polymer beads, different sizes available. High capacity , low resolution. Capillary column: small capillary with polymeric film on inner wall. High resolution, low capacity. Deactivated fused silica is free of adsorption problems encountered with most packings or capillaries columns. Gas Separation/ GC He + N2 , CO2 , H2O , SO2 + He He + SO2 + H2O + CO2 + N2 + He INCREASING POLARITY OF GASES What after we separate? He + SO2 + H2O + CO2 + N2 + He DETECTOR DETECTOR CO2 N2 SO2 H20 TIME Gas Separation / Purge & Trap TCD SO2 H2O CO2 Detectors Thermal conductivity detector (TCD) Heated filament from which heat is removed at a constant rate by He gas stream. Change in heat transfer is caused by presence of analyte molecules with different thermal conductivities than He. Relatively low sensitivity, excellent range and linearity. Non-destructive. Electron capture detector (ECD) Electrons are captured by organic species in ionized carrier; used for trace sulfur determination. Isotope Ratio Mass Spectrometer (IRMS) Typical output of EA, text book CO2 N2 H2O SO2 Balance Balance: Get the best A 0.01mg readability translates in an error of 0.25% on a 2 mg sample (on 100% carbon) IAEA (and NIST) standards Links available on Isogeochem STD: Making it on your own… Best : %N, 15N, %C, 13C Test different materials and pray Make your own: Caffeine L-glutamic acid Mixtures such as sucrose + potassium nitrate Mix in solution of natural + enriched/depleted material then dried, powdered, sieved STD: Not making it on your own… STD: Not making it on your own… Can also contact other university labs to buy or obtain their internal standards. ARNDT SCHIMMELMANN Indiana University http://mypage.iu.edu/~aschimme/hc.htm Calculation Methods 1. K-Factor (Single point calibration with blank substraction) Calculated % = K * (Area unk – Area blk) Weight unk Where K = Weight std * Theor std % Area std - Area blk Calculation Methods Wt * Th std % 2. Linear Fit (Least square linear regression) Calculated % = m * Area unk + b Weight unk Peak Area Calculation Methods Wt * Th std % 3. Quadratic Fit (Least square quadratic regression) Calculated % = a * Area unk2 + b * Area unk + c Weight unk Peak Area An update… Most elemental analysers manufacturers have considerably refined their calculation methods, with more complex algorithms, low-high ranges and statistical tools. Leak check (1) Pressure check EA •Cap exit port •Increase P of He to 1.3 bar, wait 3 minutes •Close regulator •Pressure gauge should not move for 5 minutes Flow check (requires electronic mass flow controller) EA •Cap exit port •Monitor flow for 3 minutes •Flow should drop to 0 ml/min Leak check (2) Leak check with IRMS Interface EA •EA has been leak checked •EA chemicals have outgassed (12 hours) •Tune to Ar (mass 40) •Should meet manufacturer’s requirements •Use Ar cylinder to hunt Ar MS Elemental Analyser Schematic for IRMS Autosampler Water trap Gas chromatography Column Thermocouple Detector TCD N2 Combustion tube Reduction tube CO2 To MS Interface Diluter & ref gas Considerations for EA-IRMS Water is removed via Mg Perchlorate/sicapent trap ---water and mass spectrometer do not mix---Configuration should be optimized for gas of interest Leak free (mass 28 & 40, use Argon as leak probe) Low and stable background (mass 28, 18, 40, 44) Dynamic range must be respected, do use target beam Best sequence is carefully planned ie known concentration of samples Garbage in……Garbage out Some useful info Lab No. Sample type Nitrogen Carbon Hydrogen Sulphur C/N Comments 10,20 50,97 na na 5,00 Whall JD, Trent U. 97-256-010 Adult mysids 97-220-003 Amphipods 6,81 33,20 5,66 0,00 4,88 Pastershank G., U of O/ Biology 97-256-004 Chironomids 8,98 37,23 na na 4,15 Whall JD, Trent U. 97-256-014 Clam 12,30 48,91 na na 3,98 Whall JD, Trent U. 97-220-004 Clam muscle 11,26 43,41 6,77 1,10 3,85 Pastershank G., U of O/ Biology 0,05 11,91 0,19 0,00 10,03 37,22 5,91 0,00 3,71 Pastershank G., U of O/ Biology 0,00 0,74 0,09 0,00 N.A. Bad trace 97-220-005 Clam shell 231,6 3 Pastershank G., U of O/ Biology 97-220-018 Cragon (shrimp) 97-220-020 Diatom 97-220-002 Egg white 11,85 45,83 6,93 1,14 3,87 Pastershank G., U of O/ Biology 97-220-001 Egg yolk 5,41 62,69 9,73 0,00 11,59 Pastershank G., U of O/ Biology 97-220-009 Fish bone 6,43 19,53 3,51 0,00 3,04 Pastershank G., U of O/ Biology 97-220-008 Fish eye 12,90 48,70 7,05 2,13 3,77 Pastershank G., U of O/ Biology 97-220-013 Fish fin 10,01 34,36 5,35 0,00 3,43 Pastershank G., U of O/ Biology 97-220-014 Fish gill 13,28 47,10 7,06 0,00 3,55 Pastershank G., U of O/ Biology 97-220-015 Fish gut content 9,56 44,88 7,07 1,13 4,69 Pastershank G., U of O/ Biology Filters, we can do that! Best are quartz filters; they are stable. However they are more expensive and offer less choice. Silver filters are great too. Mostly glass filters, cheap, huge choice. How much to use? Filters, we can do that! Area = P*R2 In this case, a punch is about 16mm2 and our whole filter is about 1490mm2 . Our punch hole is roughly 1% of the filter. Sediment contamination Inorganic Carbon Organic Carbon Average d13C: 0‰ Average d13C: -25‰ 0% 1% 5% 10% 50% + + + + + 100% 99% 95% 90% 50% = Total Carbon = = = = = -25 ‰ -24.75 ‰ -23.75 ‰ -22.50 ‰ -12.5 ‰ Detection Limit? Hmmmm… Every manufacturer has them, frankly they are all similar. In the end, you want an instrument that is linear with a decent sensitivity and a nice sample peak. Your source should be adjusted accordingly and you may have to sacrifice sensitivity in order to obtain stability and linearity. Hence, the detection limit may change according to the source parameters, the EA configuration and the condition of the chemicals. Be conservative. Optimal amount for me % Nitrogen 100 10 5 2 1 0.1 0.001 Wt required 0.1mg 1.0mg 5mg 7mg 10mg 100mg Yeah, right! Optimal amount of Nitrogen: 0.1mg Open split, magical stuff EA 100 ml/min He + CO2 + N2 + He 2mm or 1/16” SS line from EA Fused silica capillary 100 um, 2m, inserted about 30cm in EA line, 0.4 ml/min going into source of IRMS In real life…Open split Ref 3 He Ref 1 EA Ref 2 IRMS The joy of sniffing EA only CO2 TCD output from EA N2 Mass spectrometer output Mass 44,45,46 Mass 28, 29 Magnet Peakjump The joy of sniffing, total CO2 TCD output from EA N2 Mass spectrometer output Masses 28 & 29 Masses 44, 45 & 46 Mass 44,45,46 Mass 28, 29 Magnet Peakjump To dilute or Not to dilute? A useful suggestion… I strongly recommend that you create a basic document for your users explaining the limits and pitfalls of EA-IRMS analysis, this will save you (and your users) an enormous amount of time. Some unresolved issues with EA Removing inorganic carbon is not trivial. Alkalies (Sodium, Potassium, Calcium) are difficult to combust, catalist is definitively required to bind and help with oxydation. Oxygen contribution to d34S is still a problem, especially with organic samples. Last suggestion and only Official Endorsement The only product fully and officially endorsed by the author is: RickaRd’s White Thank you My thanks to: Wendy Abdi, Nik Binder, Fred Longstaffe, Scott Hughes, Gilles St-Jean, Peter Stow and Patricia Wickham for the use of material, brain power and time …