1 2 Automation of the cytokinesis-block micronucleus cytome assay by laser scanning cytometry and its potential application in radiation biodosimetry 3 Maxime François1, Kevin Hochstenbach1,2, Wayne Leifert1, and Michael Felix Fenech1 4 5 6 1 7 BioTechniques 57:309-312 (December 2014) doi 10.2144/000114239 8 Keywords: Micronucleus assay; automation; laser scanning cytometry CSIRO Food and Nutrition Flagship, Nutrigenomics and DNA damage, Adelaide, Australia and Section of Investigative Medicine, Department of Medicine, Imperial College London, London, United Kingdom 2 9 10 Supplementary Material and Methods 11 Blood collection, lymphocyte isolation and irradiation 12 Upon signing informed consent, 8 ml of fresh blood was drawn from 2 healthy male adult volunteers 13 into vacutainers (Greiner Bio-One, Frickenhausen, Germany) with lithium heparin anticoagulant. 14 Isolated lymphocytes were counted using a cell counter (Coulter Electronics model ZB1, Beckman 15 Coulter, Brea, CA, USA), gently resuspended into RPMI 1640 medium (Sigma-Aldrich, Castle Hill, 16 Australia) completed with 10% FBS, 2 mM L-glutamine and 1 mM sodium pyruvate solution to a 17 concentration of 1 X 106 per ml in a total of 750 μl medium. Cells were transferred as 750 μl aliquots 18 into four 24-well plates (Nunc, Roskilde, Denmark) and exposed to 0, 1, 2, or 4 Gy ionizing radiation 19 (0.57 Gy/min) using a Faxitron 650 X-ray machine (Faxitron, Tuscon, AZ, USA). 20 Cell culture and slide preparation 21 Lymphocytes were cultured at 37°C after mitotic stimulation using 30 µg/ml of Phytohaemagglutinin 22 (Thermo Fisher, Scoresby, Australia). 44 h after initiation of the lymphocyte cultures, cytochalasin B 23 (Sigma-Aldrich, Castle Hill, Australia) was added at a final concentration of 4.5 μg/ml. At 72 h, 24 lymphocytes were harvested, counted again and concentration was adjusted to 300,000 cells/ml 25 with complete RPMI 1640 medium, as pilot studies showed this to be the optimal concentration for 26 preparing slides for LSC analysis. Cells were cytocentrifuged on slides using a Shandon Cytospin 4 27 (Thermo Fisher, Scoresby, Australia) at 40 xg before being air-dried and fixed in methanol for 10 min 28 at -20°C. Cytoplasm was stained with 0.2% (w/v) Fast Green (Sigma-Aldrich, Castle Hill, Australia) in 29 order for the LSC to determine cytoplasm boundaries. Nuclei were counterstained with 4',6- 30 diamidino-2-phenylindole (DAPI; 0.4 μg/ml; Sigma-Aldrich, Castle Hill, Australia). Staining procedures 31 were performed in the dark at room temperature. Slides were stored at -20°C in a sealed 32 microscope slide box with desiccant until LSC and visual scorings were performed. 33 Laser Scanning Cytometer and visual scoring 34 The automated scoring of MN in lymphocytes was carried out with an iCyte imaging cytometer (iCys 35 7.0; Thorlabs, Sterling Virginia, VA, USA). DAPI and Fast Green were excited with 405 nm and 633 nm 36 lasers, respectively and their emitted fluorescence were collected with blue and long red photo 37 multiplier tubes, respectively. Predefined settings were used for all microscope slides analyzed. The 38 settings were empirically determined for the assay and can be modified in any ways by the operator. 39 However, once the settings were set for the experiment they were not modified. Therefore, data 40 generated for each donor were directly comparable. These settings included the selection of the 41 lasers used, voltage of the photomultiplier tubes, threshold values, color of the contours as well as 42 minimal area required for a nucleus or a micronucleus to be identified and contoured as such. Once 43 images were generated, the software applied a threshold preset manually to identify the signals of 44 interest (represented on the images by pixel values). The thresholds were set so that cytoplasm and 45 nuclear stains were identified and contoured appropriately. Data for endpoints of the assay (i.e. 46 number of cells, circularity values for each nucleus, number of micronuclei) were automatically 47 generated by the software and were then exported to excel. 48 49 For each volunteer and IR dose, duplicate cytospots on each slide were then blindly scored using a 50 Nikon Eclipse E-600 microscope (Nikon, Sydney, Australia) with 1000× magnification to determine 51 the presence or absence of MN in binucleated cells.