SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV- Cabinet Prepared By Concerned Dept 1. Checked By Concerned Dept Reviewed By QA Approved By QA-Head SOP. No. : SPL/ SOP/QC/029 Page No. : 1 of 3 Supersedes : New Revision No. : 00 Effective Date Review Date Objective To define the procedure for Operating and Calibration procedure of UV Cabinet. 2. Scope This procedure is applicable to Quality Control department at Solitaire Pharmacia Pvt. Ltd., Baddi. 3. Responsibility Executive Quality Control 4. Accountability Head Quality Control 5. Procedure 5.1. Instrument Detail: Name UV-CABINET Make Model Sr. No. Sarthak STK-312 100303 Instrument ID. QCD/UVC/009 5.2. 5.2.1. Location Instrument Lab. Operation Procedure: Before operation check the UV cabinet and observation glass window for the cleanliness. If found unclean, clean it using dry cloth. 5.2.2. Connect the UV cabinet to the Mains and switch “ON” the mains. 5.2.3. Place the TLC plate / sample in the UV cabinet hood by opening the front door. Then close the door. Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV- Cabinet Prepared By Concerned Dept 5.2.4. Checked By Concerned Dept Reviewed By QA Approved By QA-Head SOP. No. : SPL/ SOP/QC/029 Page No. : 2 of 3 Supersedes : New Revision No. : 00 Effective Date Review Date Set the required wavelength (at which sample is to be observed) by pressing”254nm'' switch, which corresponds to UV short region or ``365 nm '' switch which corresponds to UV long region and allowed to stabilized for 10 minutes. 5.2.5. Observe the TLC plate / sample through glass window situated on the top of the instrument .Note the observations in the analytical work report. 5.2.6. Put off the wavelength switch. Remove the TLC plate/sample and Switch off the “MAIN”. 5.2.7. Take proper entry in the log book. 5.3. Calibration of UV - Cabinet: Weight 6. 7. Records/Formats: SPL/QC/046 - Log book for UV-Cabinet SPL/QC/047 - Calibration of UV-Cabinet Annexure: NA 8. Reference: Nil 9. Abbreviations: SPL - Solitaire Pharmacia Pvt. Ltd., Baddi QC - Quality Control SOP - Standard Operating Procedure Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV- Cabinet Prepared By Concerned Dept 10. Checked By Concerned Dept Reviewed By QA Approved By QA-Head SOP. No. : SPL/ SOP/QC/029 Page No. : 3 of 3 Supersedes : New Revision No. : 00 Effective Date Revision History: Revision No 00 Format No.SPL/QA/001 Effective date Reason for Revision New Document Review Date SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV- Cabinet SOP. No. Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA : SPL/ SOP/QC/029 Approved By QA-Head Log book for UV-Cabinet Format No : SPL/QC/046 Sr. No. Format No.SPL/QA/001 Date Name of Material A.R. No. Done By Checked By SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV- Cabinet SOP. No. Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA : SPL/ SOP/QC/029 Approved By QA-Head Calibration of UV-Cabinet Format No : SPL/QC/047 Date of Calibration : Next Due Date: Name of Instrument : UV-Cabinet Make : Sarthak User Department : Quality Control Instrument Code : AVL/QC/047 Calibration Schedule : Monthly Procedure: Weight of Sodium salicylate : mg Weight of Sodium salicylate : mg Apply to the plate 5 ul of a 0.04 % W/V solution of Sodium Salicylate in alcohol for lamp o maximum output at 254 nm and 5 ul of 0.2 % W/V Solution in alcohol for lamp of maximum output at 365 nm. Acceptance Criteria: The lamps should be capable of revealing without doubt the spots of Sodium Silicate. Remark: The instrument is working/ not working under satisfactory condition. Analyzed By: Checked By: Approved By: Date Date Date : Format No.SPL/QA/001 : :