SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer Prepared By Concerned Dept 1. Checked By Concerned Dept Reviewed By QA Approved By QA-Head SOP. No. : SPL/ SOP/QC/022 Page No. : 1 of 4 Supersedes : New Revision No. : 00 Effective Date Review Date Objective To define the procedure for Operating and Calibration of UV and Visible Spectrophotometer. 2. Scope This procedure is applicable to Quality Control department at Solitaire Pharmacia Pvt. Ltd., Baddi. 3. Responsibility Executive Quality Control 4. Accountability Head Quality Control 5. Procedure 5.1. Instrument Detail: Name UV- SPECTROPHOTOMETER Make SHIMADZU Model UV-1800 Sr.No. A11454703180 Instrument ID. QCD/UVS/002 Location Instrument Lab. 5.2. Operation Procedure: 5.2.1. Ensure that instrument is clean and free from dust. 5.2.2. Switch “ON” the main supply by pressing “ON/OFF” switch. 5.2.3. Now switch on the instrument by pressing “ON/OFF” switch just back of the instrument. Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA Approved By QA-Head 5.2.4. It takes some time form initialization. 5.2.5. After initialization main window will display. 5.2.6. Before start analysis by UV check date and time. SOP. No. : SPL/ SOP/QC/022 Page No. : 2 of 4 Supersedes : New Revision No. : 00 Effective Date Review Date 5.3. For Traspetancy (%T): 5.3.1. Click single wavelength > Click %T. 5.3.2. For Single Wavelength: 5.3.3. Click Photometer > Single wavelength. 5.3.4. Enter the required wavelength. 5.3.5. Click OK. 5.4 For Wavelength Scan: 5.4.1. Click Transpetance > wavelength Scan on main menu. 5.4.2. Enter the required range of wavelength. 5.4.3. Click OK. 5.5. Calibration Procedure: 5.5.1. Control of wavelength: Sample Preparation: Prepare a 4% w/v of Holium Oxide in 1.4M Perchloric Acid. Sr. No. 01 02 03 04 5.5.2. Wavelength 241.15 nm 287.50 nm 361.50 nm 536.30 nm Observed Wavelength Acceptance Criteria 240.15 to 242.15 nm 286.15 to 288.15 nm 360.15 to 362.50 nm 533.30 to 539.30 nm Control of Absorbance: Sample Preparation: 57.0 to 63.0mg of Potassium diChromate→1000 ml with 0.005M Sulphuric Acid Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA Sr. Wavelength No. 01 02 03 04 Observed Absorbance SOP. No. : SPL/ SOP/QC/022 Page No. : 3 of 4 Supersedes : New Revision No. : 00 Approved By QA-Head Effective Date Calculate Absorbance (1%,1cm) Standard Absorbance (1%,1cm) 124.50 144.00 48.00 106.60 235 nm 257 nm 313 nm 350 nm Calculation: Absorbance (1%, 1cm) Review Date Maximum Tolerance Limit 122.90 to 126.22 142.40 to 145.70 47.0 to 50.30 104.90 to 108.20 Absorbance x 100 x 100 Wt of Potassium dichromate Sample Preparation: 57.0 to 63.0 mg of Potassium dichromate →100 ml with 0.005M Sulphuric Acid Sr. Wavelength No. 01 Observed Absorbance = Calculate Absorbance (1%,1cm) 430 nm Calculation: Absorbance (1%, 1cm) = Standard Absorbance (1%,1cm) Maximum Tolerance Limit 15.90 15.70 to 16.10 Absorbance x 100 x 10 Wt of Potassium dichromate Limit of Stray Light: Sample Preparation: Dissolved 1.2 gm of Potassium Chloride in 100 ml with water Absorbance at 200 nm =……….. Limit = Greater than 2.0 Resolution Power: Sample Preparation: Dissolved 2.0 ml of Toluene with Hexane in 100 ml of volumetric flask→ further take 1.0 ml →100 ml with Hexane in 100 ml volumetric flask.. i. Maximum at 269 nm Absorbance = ….. ii Minimum at 266 nm Absorbance = ….. Ratio of Absorbance = …………….. (Limit: NLT 1.5) If instrument is working under satisfactory condition then affix Calibration Label. If it is not working under satisfactory condition then inform maintenance department. Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer Prepared By Concerned Dept 6. 7. Checked By Concerned Dept Reviewed By QA Approved By QA-Head SOP. No. : SPL/ SOP/QC/022 Page No. : 4 of 4 Supersedes : New Revision No. : 00 Effective Date Records/Formats: SPL/QC/032 - Log book for UV and Visible Spectrophotometer SPL/QC/033 - Calibration of UV and Visible Spectrophotometer Annexure: NA 8. Reference: Nil 9. 10. Abbreviations: SPL - Solitaire Pharmacia Pvt. Ltd., Baddi QC - Quality Control SOP - Standard Operating Procedure Revision History: Revision No 00 Format No.SPL/QA/001 Effective date Reason for Revision New Document Review Date SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer SOP. No. Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA : SPL/ SOP/QC/022 Approved By QA-Head Log book for UV and Visible Spectrophotometer Format No : SPL/QC/032 Operation Time Sr. No. Date Format No.SPL/QA/001 Product Name Batch No. On Time Off Time Operated By Check By Remarks SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer SOP. No. Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA : SPL/ SOP/QC/022 Approved By QA-Head Calibration of UV and Visible Spectrophotometer Format No : SPL/QC/033 Date of Calibration: Next Due Date: Name of Instrument : UV and Visible Spectrophotometer Make : Shimadzu Model No. : UV-1800 User Department : Quality Control Instrument Code : QCD/UVS/002 Calibration Schedule : Monthly Control of Wavelength: Sample Preparation: Prepare a 4% w/v of Holium Oxide in 1.4M Perchloric Acid Sr. No. Wavelength 01 241.15 nm 02 287.50 nm 03 361.50 nm 04 536.30 nm Control of Absorbance: Observed Wavelength Acceptance Criteria 240.15 to 242.15 nm 286.15 to 288.15 nm 360.15 to 362.50 nm 533.30 to 539.30 nm Sample Preparation: ……….. mg of Potassium diChromate→1000 ml with 0.005M Sulphuric Acid Sr. No. Wavelength 01 02 03 04 235 nm 257 nm 313 nm 350 nm Observed Absorbance Calculation: Absorbance (1%,1cm) = Calculate Absorbance (1%,1cm) Standard Absorbance (1%,1cm) 124.50 144.00 48.00 106.60 Maximum Tolerance Limit 122.90 to 126.22 142.40 to 145.70 47.0 to 50.30 104.90 to 108.20 Absorbance x 100 x 100 Wt of Potassium dichromate Sample Preparation: ……… mg of Potassium dichromate →100 ml with 0.005M Sulphuric Acid Format No.SPL/QA/001 SOLITAIRE PHARMACIA PVT. LTD. RESTRICTED CIRCULATION AUTHORISED PERSONS ONLY STANDARD OPERATING PROCEDURE Title: Operation & Calibration of UV & Visible Spectrophotometer SOP. No. Prepared By Concerned Dept Checked By Concerned Dept Reviewed By QA Calculation: Absorbance (1%,1cm) : SPL/ SOP/QC/022 Approved By QA-Head = Absorbance x 100 x 10 Wt of Potassium dichromate Limit of Stray Light: Sample Preparation: Dissolved …………. gm of Potassium Chloride in 100 ml with water. Absorbance at 200 nm = ……….. Limit = Greater than 2.0 Resolution Power: Sample Preparation: Dissolved 2.0 ml of Toluene with Hexane in 100 ml of volumetric flask→further take 1.0 ml →100 ml with Hexane in 100 ml volumetric flask.. i. Maximum at 269 nm Absorbance = ….. ii Minimum at 266 nm Absorbance = ….. Ratio of Absorbance = …………….. (Limit: NLT 1.5) Remark: The instrument is working/ not working under satisfactory condition. Analyzed By: Checked By: Approved By: Date Date Date : Format No.SPL/QA/001 : :