IR, Raman & IR Microscopy Introduction: Fourier Transform Infrared Spectroscopy (FTIR) detects the vibration characteristics of functional groups in a sample. Chemical bonds bend, contract, or stretch in such a manner that is unique to each functional group when they are exposed to the FTIR’s infrared light. This movement creates a spectrum of wavelength absorbances. This method is used for both qualitative and quantitative analysis of compounds. Raman Spectroscopy, on the other hand, is generally just used for qualitative analysis of compounds. It is based on the inelastic scattering of monochromatic light, which forms a spectrum that can be analyzed to give the functional groups in an analyte. Microscopy is used to focus an infrared beam on a very specific area of a sample in order to analyze the sample. This is commonly used to look at carpet fibers. Purpose: The purpose of this lab is to be able to run the FTIR and Raman instruments and interpret the data collected. Three organic solvents and mixtures will be analyzed with the IR. Different carpet fiber samples will be looked at using the microscope and then used to try to match an unknown. Procedure: 1. 2. 3. 4. 5. 6. 7. 8. Run three organic solvents through the IR Identify unique components of each spectrum Run a mixture of solvents through the IR Run different mixtures of the three solvents Run an unknown mixture and determine the concentration of each component Learn how to operate the IR in Microscopy mode Obtain spectra for known fibers Match fibers collected with the known fibers SOP FTIR: 1. Make sure than the instrument is on. 2. Make sure that the black lever between the FT-IR and the Raman is down 3. Double click on the Varian Resolutions Pro icon in the top right of the display. 4. In the Current scan menu on the bottom of the screen select IR scan. 5. From the Collect menu at the top of the screen select Rapid scan 6. The Electronics tab parameters should be: speed: 20 KHz Filter: 5 UDR: 2 7. The Optics tab parameters should by: IR Source: Mid-IR Beam: internal Detector: DTGS Beamsplitter: KBr Broadband Accessory: none Optical filter: none Aperture: 4 cm-1 at 2000 cm-1 8. Click setup. The voltage reading should be between 1-5V. If not, change the selectivity 9. Click background to run a background scan 10. Return to rapid scan from the collect menu and click scan The transformers menu allows you to convert to %T 11. To shut down, close Varian Resolutions Pro and LEAVE EVERYTHING ON Raman Spectroscopy: 1. Fill the blue dewar with liquid nitrogen and wait 20 minutes then top off. 2. Make sure the FTIR is on 3. Turn on the laser power supply switch and then turn the key to the on position 4. make sure the lever between the Raman and FTIR is up, the lever between the Raman and microscope is down, and the internal lever in the Raman should be up, meaning bypass closed 5. double click on the Varian Resolutions Pro icon to open software DON’T USE SOFTWARE UNTIL SAMPLE IS LOADED (SEE TROUBLESHOOTING) 6. From the current scan menu select Raman scan 7. The optics parameters should be: IR Source: off Beam: right Detector: Raman Ge Beamsplitter: quartz near IR Accessory: FT-Raman ATR Crystal: none Optical filter: holographic notch Aperture: open 8. Select the laser tab and click turn on diode 9. Press the shutter switch on the front of the Raman (to open) and turn the power to its highest, 3 10. Under the laser tab, adjust laser control current until the Raman display reads between 600-700 mW 11. Click setup. If filled with noise, click autoscale (the four arrows) 12. Adjust same with X, Y, and Z, knobs until the centerburst is at its maximum 13. Click scan 14. To shut down: collect menu> Raman scan> Laser tab and click turn diode off. Press the shutter button (closed), remove your sample, turn off the laser power supply and turn key to off position, and exit Varian Resolutions Pro. Leave everything else on! Data/Results: Day 1: Goal: Run three organic solvents through the FTIR. Then run different mixtures of solvents through the instrument. Run an unknown as well and compare to the different spectra to determine what the unknown is. Run samples through the Raman instrument. Today, we were able to successfully run both the FTIR and the Raman instruments. Our three organic solvents were 2-butanone, 2-propanol, and toluene. The spectra from the FTIR showed peaks that matched up to where peaks for each sample should be. After running the solvents, we ran mixtures of the solvents at different ratios. The four ratios were 1:1:1, 2:1:1, 1:2:1, and 1:1:2. An unknown mixture was also run. Next, we ran two samples on the Raman instrument. The spectrum collected from this instrument was not as readable as the spectrum from the FTIR. We first ran a solid acetaminophen sample. We ran this twice since the first time it was difficult to read the peaks due to the excess background noise. Afterwards, we ran a liquid methyl salicylate sample, to see if we could get a better spectrum with a liquid sample rather than a solid sample. This sample did produce a better spectrum, however, there was still a lot of background noise. Day 2: Goal: Look at fibers through the microscopy instrument Today, we attempted to use the FTIR microscope to look at fiber samples. First, we had trouble getting the desired voltage. The reading was supposed to be between 1.5 and 5. The highest value we could get was a little over 1 (about 1.030), however, Donna told us it should be fine to run as long as it stayed over 1. We ran a 50-poly, 50-cotton sample. We were able to take a picture of the sample using the microscope, however, it was difficult to see when we printed it out. The spectra of the data we collected did not turn out well either. There was a lot of noise and it was difficult to pick out distinguishable peaks. Conclusion: This week we learned how to use the FTIR, Raman, and FTIR microscope instruments. For the FTIR, we looked at three different organic solvents and then combined them at different ratios to compare the spectra. Then we ran an unknown mixture to compare to the known ratios and determine what it is. After comparing the spectra, it was determined that the unknown we ran was the 1:1:2 mixture. All of the spectra we collected were easy to read and comparable to one another. For the Raman instrument, we ran two different samples, one solid and one liquid. We had a little bit of trouble finding the center-burst and when we did find it, there was a lot of background noise on the spectra, which made it hard to analyze. The FTIR microscope was difficult to work with since no other group had used it yet. We were able to take a picture of a sample and run an analysis on it, however, the spectra was unreadable. There were a couple problems with the lab. At first, the student operating procedures were not online so we had to research how to use the instruments and look for the packets next to the instruments in lab. Another problem was with the Raman instrument. Our spectra, no matter if it was a solid or a liquid, had a lot of background noise which made it very difficult to analyze. Overall, the experiment was a success since we were able to successfully use all the instruments and collect data from each.