List of equipment and specifications:
SNO
Name of the Equipment & Brief specifications
Quantity
HPLC - PDA
1

PUMP – QUARTERNARY GRADIENT:— Integrated vacuum degassing (with < 500uL internal volume
per chamber)
— Automated, continuous solvent compressibility compensation
— Flow precision of < 0.075% RSD (over 0.200 to 5.000 mL/min
flow range)
— Effective delay volume of < 650uL , independent of
backpressure
— 5,000 psi (345 bar) maximum operating pressure for both
Solvent and Sample Management functions
— Wide range of programmable gradient profiles including linear,
step, concave and convex.
— Integrated active, programmable and continuous plunger sealwash
— Tool-free, simple access to plungers, plunger seals, plunger
wash seals

INTEGRATED AUTOMSAPLER:— Programmable wash / purge option to minimize carryover.
— Sample capacity Up to 120 x 2mL vials in preferably in carousel
format.
— Sys Prep that routinely automates wet prime and refreshes
syringe.
— Tool-free, simple replacement of needle wash frit.


1)
2)
3)
4)
5)
PHOTODIODE ARRAY DETECTOR:01) Sensitivity : 10 µAU or better
02) Lamp Optimization facility.
03) Simultaneous 2D & 3D
04) Wavelength Range: 190- 800 nm or better
05) Optical Resolution: 1.2 nm
06) Photodiodes: 512 or better
07) Digital Resolution: 1.2nm/pixel
08) Wavelength Accuracy: ± 1.0 nm
09) Wavelength repeatability: ± 0.1 nm
10) Data Acquisition: Up to 80 Hz.
HPLC CHROMATOGRAPHY SOFTWARE:The Chromatography software should be embedded with data base
preferably Oracle.
The software should provide all the functionality that a needed by
analysts to efficiently work with chromatography data including but not
limited to features like robust peak integration capabilities with effective
detection of low level peaks on noisy or sloping baselines.
The software should have the capacity to control single quadrupole
Mass detectors in a routine lab environment
Should have custom field calculation option with capability to control UV
/ PDA for MS detectors in future without any upgradation.
Computer hardware provided shall possess all the features required for
this software
1
EMD
IN RS.
20000
15000
2
HPLC – RID
Pump
1.
2.
3.
4.
5.
6.
7.
8.
9.
Programmable Flow Rate Range : 0.000-10.00ml/min.,in 0.001ml/min.,
increments,
Upgradable flow rate up to >20 ml / min
pulse free flow,
Standard 100ul heads;
Flow Precision - 0.1% RSD or two seconds SD, six replicates, based on
retention time or volumetric measures at 1ml/min.,
Should have wide range of programmable gradient curves including Linear ,
step, concave and Convex .
Flow Accuracy +/-1.0% or 30ul/min.,
Maximum Operating Pressure : More than 5000 psi
Pressure Rep : <2.0%
AUTOMSAPLER
1)
2)
3)
4)
Programmable wash / purge option to minimize carryover.
Sample capacity Up to 120 x 2mL vials in preferably in carousel format.
SysPrep that routinely automates wet prime and refreshes syringe.
Tool-free, simple replacement of needle wash frit.
Refractive Index Detector
1.
2.
3.
4.
5.
6.
7.
8.
9.
Refractive Index range: 1 to 1.75 RIU
Measurement range: 5.0x 10 -4 RIU to 7.0x10 -9 RIU
Linear dynamic range: ≤5% over ±5.0x 10 -4
Noise: ±1.5x10 -9 RIU mode
Flow Rate: 0.1 to 10.0 ml/min
Temperature control: 30 to >50ºC
Flowcell: Fused quartz
Cell Volume: 10µl
Cell Pressure limit: ≤100psi
Chromatography software
1.
2.
3.
4.
The Chromatography software should be embedded with data base
preferably Oracle.
The software should provide all the functionality that a needed by analysts to
efficiently work with chromatography data including but not limited to features
like robust peak integration capabilities with effective detection of low level
peaks on noisy or sloping baselines.
The software should have the capacity to control single quadrupole Mass
detectors in a routine lab environment
Should have custom field calculation option with capability to control UV /
PDA for MS detectors in future without any upgradation.
1
3
15000
GC FID Detector
1. GCwith FIDdetector&S/SLInjector
2. SystemSoftware
3. Auto Sampler
4. BrandedPC &Printer
5. SuitableOnline UPS
6. Cylinders&Regulators
7. GasPurification Panel
GasChromatograph systemhavingcapabilityfor both
qualitativeandQuantitativeanalysiswithvalidated
software[OperatedthroughexternalPC].
Main Systemwith FIDdetector&S/SLInjector shall have features similar to or
better than the following:
1. Column Oven: Temperature range:ambient +10°C/-60°C*to 450°C
Temperature programramps/holds:7/8, Maximumtemperatureramprate:100°C
/minforall voltages
Cool down rate:450°Cto 50°C in 5.2minutes
Temperatureset-point resolution:1°C
2.Injector:Split/Split lessInjector[S/SL]
Pressure range:Pressure range:0-100psi Total flow:500
mL/minat 10 psi Maximumtemperature:450°C
Split range:1-10,000(columndependent) Suitedfor
columns: Wide bore(0.53mm) Narrowbore(0.05to0.32 mm)
3. Detectors: FID FlameIonization Detector: Maximum
temperature:450°C Detectivity: 2pgC/sec
Lineardynamic range:107 or better
Flametiptype: ceramicOperational quality: Flame-out
detection Auto re-ignition
ElectronicPneumaticControl: forinletsanddetectors and Pneumatics
mustbeprogrammable.
4. AutomaticLiquidAuto Sampler:Sample capacity:100 x 2mL vials
1
Large solvent washvial:2x120mL(optional), Dual and duplicate mode,Internal
standardaddition
4
REFRACTO METER
Software interpretation of shadow line intersect eliminates operator error; 589 nm
LEDs with 589 nm filter; 1.30000 to 1.70000 refractive index range; Reads out to
0.00001nD or 0.01% solids (Brix); Up to 14 custom scales can be programmed;
Operating temperature range from 10-80°Calibrations from 1 to 7 points using
distilled water and NIST certified oils; Automatic Temperature Compensation
between 10-40°C.
5
5000
1
KEJLDHALPROTEIN ANALYZER
Manual distillation unit, 200-240V, 50-6-Hz comprising; Combi-adapter for 100,
250, 400 &750 ml tubes and digestion tubes for 100-250 ml. Distillation time: 3.5
min at 30mg N; distillation capacity: 40ml/min; measuring range: 0.1-200 mg N;
reproducibility: 1%RSD; Recovery:>99.5% at nitrogen levels between 1-200 mg
N; tube draining: 200 ml in less than 10 ml; Delay: 0-1800 sec; safe: 0-15sec,
power 200-240 V 50-60Hz,2200 watts.
Measuring range: 0.1 to 200 mg Nitrogen; Nitrogen recovery:> 99.5%; Detected
nitrogen resolution: 0.0001; Reproducibility: Relative Standard deviation)< 1%;
Distillation speed: from 35 to 45 ml/minute; Steam generator water consumption:
2 Litre/Hr; Evaluation precision: 1.5%; Complete Nitrogen recovery from the
sample; Minimum operator intervention; No calibration required; Minimum
analysis time
10000
1
6
Volumetric Karl Fischer Titrator: Samples with a water content between 100 ppm
and 100% are analyzed in single measurement. Glass beaker 150 ML.
Repeatability 0.3% at > 10mg H2O; Coulometric Karl Fischer Titrator:
measurement range -2000 to +2000mV. Glass beaker 200ml. Resolution 0.1 mg.
Samples from any industry with a water content 1 ppm and 5%.
7
8000
KARL FISCHER TITRATOR
1
5000
MICROSCOPE
Trinocular microscope 450 inclined, 360 rotatable. Wide Field eyepiece; HKW 10x
paired. Quadruple Nosepiece (Ball bearing type); VA series DIN Semi Plan
Achromatic objectives 10x, 40x 40x(Sring loaded) and 100x (Sring loaded, Oil
immersion), Stage rectangular 142 x 122 mm with mechanical stage cross travel
75x50 mm with co-axial control; Movable Sub-stage ABBE condenser N.A. 1.25,
with aspheric lens; Iris diaphragm and special day light blue filter in swing out
filter holder; Extra Accessories: Microphotographic equipment consists of digital
camera with a capacity of 12 mega pixel and 10x optical zoom and adopter for
connecting to microscope. Micro scope image projection system with USB
Interface. Image analysis software
1
Terms and Conditions:















The sealed tenders, mentioning clearly “Food Testing Lab Equipment” on the cover
along with the Tender No., Name of equipment, should be addressed and delivered to:
The Registrar
JNTUK, Kakinada
Kakinada – 533 003
Andhra Pradesh
Separate tenders should be submitted for each equipment as per the specifications
mentioning the additional features, if any.
The completely filed in tender documents must reach the Registrar, JNTUK, Kakinada,
on or before 01.03.2013, 5.00 pm.
The tender will be opened on 06.03.2013 at 14:30 hrsby the undersigned/his nominee
in the presence of the committee members and available tenderers or their
representatives.
The JNTUK, Kakinada reserves the right either to select or reject certain terms (single or
multiple units)in the schedule without assigning any reasons whatsoever.
Incomplete tenders, amendments and additions to tenders after opening or late tenders are
liable to be ignored, and rejected
The tenderer shall be responsible for the delivery of the material, its successful
demonstration, and onsite training.
A Demand Draft for Rs.1,000/- (Rupees One Thousand Only) drawn in favour of the
“Registrar, JNTUK, Kakinada” towards application fee(nonrefundable) must
accompany the tender(s) from each tenderer.
A Demand Draft for an amount, as specified for each equipment, drawn in favour of the
“Registrar, JNTUK, Kakinada” towards refundable EMD must also accompany the
tender. The tender without EMD will be rejected. The EMD for the unsuccessful
tenderers will be refunded.
JNTUK, Kainada has been registered with DSIR, GOI and exempted from customs and
central excise.
The prices should be inclusive of packing, forwarding, insurance, delivery at destination
site and all taxes(FOR Kakinada).
The tenders should be valid for acceptance up to a period of 3 months. The tenderers
should be ready to extend the validity, if required.
Tenderer should enclose their offers with full details of equipment with full
documentation / descriptive literature / leaflets indicating special features of their
equipment or systems.
Tenders submitted after due date/time will not be considered and no intimation will be
sent in this regard.
The Registrar, JNTUK, Kakinada does not bind himself to accept the lowest bid on any
tender and reserves the right of accepting the whole or any part of the tender or portion
of the quantity offered and the tenderer shall supply the same at the rate quoted.




Acceptance of the tender shall be communicated by letter of acceptance or formal
acceptance of the tender to the successful tenderer.
Any specific queries, communications, and references should be made only to, The
Registrar, JNTUK, Kakinada.
The Tenderer should produce copy of service Tax Certificate and PAN Card.
All Tenderers should give a warranty declaration as detailed below:
1. We shall abide by all the specifications, terms and conditions listed in the tender
Notice and tender documents.
2. We warrant that everything to be supplied by us hereunder shall be free from all
defects and faults in material, workmanship and shall be of the highest quality and
material of the type ordered, shall be in full conformity with the specifications.
3. Any deviations in the material and/or the specifications from the accepted terms are
liable to be rejected and the tenders need to supply all the goods in the specified form
to the satisfaction/specifications specified in the order/contract and demonstrated at
their own cost.