BPAP A40

advertisement
BiPAP A40 Ventilatory
Support System
BY
AHMAD YOUNES
PROFESSOR OF THORACIC MEDICINE
Mansoura Faculty of Medicine
BiPAP S/T
SPECIFICATIONS:
• Modes : CPAP, S, ST
• Timed inspiration 0.5 to 3.0 seconds
• IPAP 4 to 25 cm H2O
• EPAP 4 to 25 cm H2O
• CPAP 4 to 20 cm H2O
• Breath rate 0 to 30 BPM
• Weight 1.36 kg Dimensions 18 cm x 14 cm x 10
cm
• Alarms Patient Disconnect, Apnea, Low Minute
Ventilation
Pressure curve when BiPAP = 10/5 cm H2O; The pressure
is higher on inspiration than on expiration, but both
pressures are above ambient .In this example IPAP is set
as 10 cm H2O and EPAP is set as 5 cm H2O.
BiPAP AVAPS
• Clinicians no longer need to choose between
pressure or volume ventilation in order to
deliver optimum therapy to NIV patients.
• Now, they can have both. The BiPAP AVAPS
(Average Volume Assured Pressure Support)
home ventilator enables clinicians to achieve
the prescribed target volume pressure
without compromising patient comfort.
• Automatically adapts to patient’s changing
needs
• Designed for clinical ease of use and patient
comfort
• Powered by advanced technologies
SPECIFICATIONS:
•
•
•
•
•
•
•
•
•
•
•
•
•
Modes CPAP, S, ST, PC, T
AVAPS available on S, ST, PC, and T modes
Bi-Flex Available from S mode
Target Tidal Volume from 200 to 1500 ml
IPAPmin From EPAP to IPAPmax
IPAPmax From IPAPmin to 25 cm H2O
Timed inspiration 0.5 to 3.0 seconds
IPAP 4 to 25 cm H2O
EPAP 4 to 25 cm H2O
CPAP 4 to 20 cm H2O
Breath rate 0 to 30 BPM
Weight 1.36 kg Dimensions 18 cm x 14 cm x 10 cm
Alarms Patient Disconnect, Apnea, Low Minute Ventilation, Low
Tidal Volume (AVAPS modes only)
BiPAP AVAPS
BiPAP A30 Ventilator
• For 24 Hours Use
• BiPAP A30 ventilator has been designed to combine
ease of use with technology advancement to deliver
enhanced therapy that adapts to patient’s condition.
• Being at home can work wonders so It delivers a
smooth transition from hospital to home for chronic
respiratory patients.
• Compatible with PSG Systems
• Optional Oximetry Module
• Intuitive, User Friendly & Colored Interface
• Integrated Heated Humidifier
• Graphical & Statistical Data Management on EncorePro
& Direct View Softwares
SPECIFICATIONS:
•
•
•
•
•
•
•
•
•
•
Modes CPAP, S, ST, PC, T , AVAPS
AVAPS available on S, ST, PC, and T modes
AVAPS Rate 0.5 to 5 cmH2O/min
Target Tidal Volume from 200 to 1500 ml
IPAP 4 – 30 cmH2O
EPAP 4 – 25 cmH2O
Breath Rate 0 – 40 BPM
Inspiratory Time 0.5 – 3 sec
Rise Time 1 (100 ms) – 6 (600 ms)
Monitoring Pressure, Tidal Volume, Minute Ventilation, Respiratory
Rate, I/E Ratio
• Weight 2.1 kg (with Power Supply) Noise Level < 30 dBA at 10 cmH2O
• Humidification System one Humidification & Dry Box technology
• Alarms Patient Disconnect, Apnea, Low Minute Ventilation, Low Tidal
Volume, High Respiratory Rate
BiPAP A40 Ventilator
• BiPAP A40 comes with well-known and
clinically proven Philips Respironics
technology such as Auto-Trak, AVAPS and
a Dry Box humidifier design.
• The device is capable of non invasive and
invasive pressure ventilation, up to 40
cmH2O, providing treatment for your
chronic insufficiency respiratory patients.
• The device features AVAPS-AE, the first
fully automatic ventilation mode, designed
to help clinicians during titration process,
while maintaining comfort and therapy
optimization at the lowest pressures.
BiPAP A40 Ventilator
• This new innovative ventilation mode helps in providing
long term therapy compliance regardless of changes to
the body position, sleep stages and respiratory
mechanics.
• Detachable Battery with up to 5 Hours Backup
• AVAPS-AE Automatically Adjust Ventilation to the
Patients Need
• Compatible with PSG Systems
• Optional Oximetry Module
• Intuitive, User Friendly & Colored Interface
• Integrated Heated Humidifier
• Graphical & Statistical Data Management on EncorePro
& DirectView Softwares
SPECIFICATIONS:
• Ventilation modes : CPAP, S, S/T, PC, T, AVAPS-AE
• Hybrid ventilation AVAPS (Average Volume Assured Pressure
Support) AVAPS-AE
• IPAP 4 – 40 cm H2O
• EPAP 4 – 25 cm H2O
• Target tidal volume (when AVAPS enabled) 200 – 1500 ml
• Breath rate 0 – 40 bpm (4 – 40 bpm in T mode)
• Inspiratory time 0.5 – 3 sec.
• Triggering and cycling Auto-Trak ,Sensitive Auto-Trak, Flow
triggering.
• Rise time 1 (100 ms) – 6 (600 ms)
• Size 21.6 cm W x 19 cm L x 11.5 cm H
• Weight 2.1 Kgs (with power supply)
• Noise level < 30 dBA at 10 cm H2O
•
SPECIFICATIONS:
• Humidification System One humidity control and ‘Dry Box’
technology
• Alarms Patient disconnection Apnea ,Low minute ventilation ,Low
tidal volume (with AVAPS/AVAPS-AE only) ,High respiratory rate
• Monitoring Pressure, tidal volume, minute ventilation, respiratory
rate, leak, I/E ratio
• Battery back up Detachable battery module: 5 hours at IPAP 12 cm
H2O / EPAP 7 cm H2O and 12 bpm
• DC power source 12 VDC, 5.0 A (external battery),
24 VDC, 4.2 A (power supply)
• Data management Encore Pro 2 and Direct View software
• Compatible with oximetry module
• Advanced detection of residual respiratory events
(Apnea/Hypopnea Index, Obstructed Airway Apnea, Clear Airway
Apnea, Hypopnea, Periodic Breathing, RERA, Large Leak and
Snore)
Intended Use of BiPAP A40 Home Ventilator
• Bilevel ventilator providing invasive and non
invasive ventilation, up to 40 cmH2O
AVAPS-AE mode
• Auto EPAP
• Auto Pressure support
• Auto Backup rate
• Easy-to-use interface and compatible with a
detachable battery module to offer increased
patient freedom and support
• Unique integrated heated humidifier to minimize
water ingress and rainout for patient comfort
Device Description
• The device augments patient breathing by supplying
pressurized air through a patient circuit.
• It senses the patients breathing effort by monitoring
airflow in the patient circuit and adjusts its output to
assist in inhalation and exhalation.
• The higher pressure makes it easier for you to inhale,
and the lower pressure makes it easier for you to exhale.
• This device can also provide a single pressure level,
known as CPAP (Continuous Positive Airway Pressure).
• The BiPAP A40 Ventilator is compatible with the System
One Heated Humidifier.
Device Description
• The BiPAP A40 ventilator introduces a new therapy mode
called AVAPS-AE.
• This therapy mode combines an improved AVAPS
algorithm with an auto-back up to treat hypoventilation.
• An auto-EPAP algorithm runs simultaneously with the bilevel therapy to deliver the pressure support at the
optimal PEEP.
• Similar to the Trilogy 200, the BiPAP A40 can
provide pressure up to 40cm, provide invasive
support if prescribed, be used in mobile
applications such as wheelchairs, and provides
an optional detachable battery.
• Additionally the ventilator can be operated using AC
power, a detachable battery, or an external battery.
The new device, the BiPAP A40, was
announced on World COPD Day 2012.
• According to a press release by the company, the
BiPAP A40 is a new bi-level, multi-modality
ventilator with an automation ventilation system.
• The automated ventilation mode is designed to
assist patients during the transition from a
managed healthcare facility or hospital to a
private residence.
• The new ventilation mode is called the AVAPS-AE.
• Its automated mode can be used to guide
physicians during the first stages of therapy. In
addition, the system is designed to improve
efficacy and comfort for a patient
The new device, the BiPAP A40, was
announced on World COPD Day 2012.
• By maintaining airway patency and delivering an
accurate amount of air for a patient, the device can help
stabilize a patient’s blood oxygen levels quickly and
easily.
• In addition, the BiPAP A40 comes with a specialized
battery module for portability. In addition to portability,
the battery mode can be a valuable lifeline for a patient
during a power outage.
• “Philips is proud to be able to provide access to new
and innovative solutions that improve quality of care for
respiratory patients – both in hospital and at home.”
Download