J. Marschinke
Production Leon, Leon plus, etc.
Central office Bad Ems
Anesthesia Workstation
for Pediatric and Adult care
•
•
•
•
Ventilator is working with an pneumatic drive
and hanging bellows
Inhalant anesthetic in a rebreathing system
Inhalant anesthetic in a semi-closed system
Low flow range
Minimal flow range
Non-rebreathing system
Bain
Magill
Jackson Rees
Kuhn
Ventilation Modes
• (S) IMV
(Synchronized)
Intermittend Mandatory Ventilation
• (S) PCV (Synchronized)
Pressure Controlled Ventilation
• PSV Pressure Supported
Ventilation
• Man. Manual Ventilation
• Spont. Spontaneous Ventilation
System Description
Patientmodul
• Fresh gas decoupled
• Tempered:
preventing condensation
warming respiratory gas
• APL decoupled during controlled
ventilation
Patientmodul
• Switch to semi-open system
• Inspiratory and expiratory flow
sensors
• Absorber can be exchanged during
operation
• Can be fully sterilized
• Compliance compensated
Features
•
•
•
•
•
•
•
15“ TFT color display
Operation via touchscreen and/or control knob
Electronic fresh gas blender
One-hour rechargeable battery buffering
Interface to control monitors of other
manufacturers
Ethernet interface
Gas monitor (choose between mainstream and
sidestream) with automatic detection of
anesthetic gas and anesthetic gas mixture
Features
• Receptacle for two anesthetic vaporizers
• Undercarriage with 4 antistatic rollers, all with
brakes
• 4 Drawers
• 1 cupboard reserve bottle connections for O² and
N²O
• Pull out writing surface, DIN A4
• Rails on right and left to fit support arms and
other accessories
O² Emergency Control
• O² Emergency control
• Settings between 0 –
15 l/min.
• Use for fail of fresh
gas blender
• Use for quick start
• Use for device is
turned off
Gas Management
• Sidestream Monitor
with real time
measurement and real
time graphs
• Paramagnetical O²
measurement
• Different water trap for
exactly measures in
patient category „child“
• Optional: Main stream
monitor
Configuration of the options panel
• Solution for half open breathing
system (Kuhn, Jackson Rees,
Magill, Bain)
Configuration of the options panel
• Solution „external O² outlet“
Pressure gauges and control
elements
O² flush
Vacuum
dosing
Turn
switch
for
vacuum
dosing
Vacuum
pressure
O² bottle
pressure
N² O
bottle
pressure
New Ventilation Mode
PSV Pressure Supported Ventilation
Psv provides pressure support during insufficient
spontaneous breathing.
If the Leon plus is not triggered by the patient after
the adjustable apnoe time, the Leon plus initiates
inspiration on ist own.
In addition, you can use a button to start a manual
breath, not activated by the patient.
Real time graphs and
complete Measurement
Fresh gas control
• Air or N²O as carrier gas
• Set percent oxygen to total
fresh gas flow
• Set total fresh gas flow
• Setting range
200 ml / min. – 18 l / min.
• Ensure O² concentration in O² /
N²O mixture of 25 %
• Option: 10 l Reserve gas bottle
with the pressures on the
display
Real time graphs
Values as real time graphs
•
•
•
•
•
•
•
•
•
•
•
Airway pressure
Flow
Volume
CO² (% or mmHg)
O²
N²O
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
Values displayed in
numerical form
• Real time
measurement:
Minute volume, tidal
volume, Peak /
Plateau / Mean
pressure, PEEP,
Frequency
• Calculated values
like MAC,
Compliance, C20/C,
Resistance
Numerical display on two pages
Loops
One Loop at a time can be displayed in full screen
Loops
Loop window
Three loops can be
displayed in the Loops
window
• Volume over pressure
• Flow over pressure
• Flow over volume
One Loop at a time can be
displayed in full screen
Alarm Limits
• Alarm limits for monitoring (adjustable)
• Function of „Autoset“
• Alarms with three different priorities
(high, medium, informal)
• Complete history of alarms in the Log
• Adjustable loudness of alarm tone
Different Solutions
Ceiling swingarm mount
Wall mount
More
Solutions