Medical Imaging In vivo microimaging in mice (EL582/BE620/GA4426) Ultrasound Imaging E10.5 embryo Reference Prince and Links, Medical Imaging Signals and Systems, Chapters 10 & 11 Acknowledgement Thanks to Professor Yao Wang for use of her course materials! Pulse-Echo Ultrasound Imaging Pulse Object Transducer RF Amplitude Excitation Pulse Scan Beam to build up image Pulse-echo Signal (Complex) We will represent the input signal as the Real part of a complex signal (Reflections) (Backscatter)1 2 Time = Depth Envelope Amplitude 1 2 1 line of image data Plane Wave Approximation Field Pattern and Pulse-Echo Equation General Pulse-Echo Equation Schematic: Ultrasound Imaging System Transducer field pattern TGC Plane wave assumption Functions of the transducer Single Element Transducer HUNT et al.: ULTRASOUND TRANSDUCERS Used both as Transmitter And Receiver Transducer compromises Transmission mode: converts an oscillating voltage into mechanical vibrations, which causes a series of pressure waves into the body Maximum frequency Enough penetration ? namic range f \ Receiving mode: converts backscattered pressure waves into electrical signals Maximize lateral resolution Attenuation (frequency shifts) Strong ~~~~focussing focussing.? Solve depth of field L LoseI sensitivity? limitations / Bandwidth Axial resolution f-number = f/2a FWHM = 1.41Af-number Fig. 1. The general components and beam properties of focused piezoelectric circular transducers. Fig. 3. The compromises for the sensitivity, tissue penetration, spatial resolution that are needed to design an efficient ultra device. DOF = Kλ (f-number)2 For example, if one desires to improve the lateral resol an Ultrasound Transducer by means of a low f-number aperture (i.e., f-number < 4) From: Hunt et al, IEEE Trans BME, of1983 field is reduced, considerably degrading the resol depth ObJectives_ in regions away from the focal plane. Thus, a means of mo - Region or organ - Penetr ntion the focal zone over the imaging depth is required. A des - Desired resolution tion of the depth-of-field problem and a detailed analys - Speed of dhta collection various transducers used to overcome this effect are prese HUNT et al.: ULTRASOUND TRANSDUCERS 455 in Section IV. Pulse-echo device Other imoging devices Different approoches Transducer Another compromises difficult decision is the selection of pulse frequ Genesis of te-cSingle element -Multiple element -Trsnsmission - Doppler and bandwidth. Three important factors affect the choic frequency: 1) tissue attenuation, 2) required depth of pen shifts) (frequency tion, and 3)Attenuation system dynamic range. There is no strict rul Compromises Strong thesesensitivity? equating variables and arriving at an optimum freque ~~~~focussing LoseI ^ ~~~~~~~~Sensitivity depth the highest frequency that provides adequate p Generally, Sptial rtiesolution field limitations / Dynomic ro nge tration is selected; this optimizes the lateral resolution [cf. f-number f/2a and Fig. 3. The compromises for the penetration, needed Thesensitivity, bandwidth pulse tissue to obtain good axial resol Transducer materiol FWHM 1.41Af-number spatial resolution that are needed to design an efficient ultrasound device. an altogether different problem. Here, the tenden poses Fig. 1. The general components and beam properties of focused piezofor sensitivity electric circular Ceramic-PZT transducers. to select the widest bandwidth (shortest pulse) because __ ~~~~PVDF for wide bond-pass and flexibility provides the best axial resolution in water. Unfortunat Others ? For example, if one desires to improve the lateral resolution wide bandwidth sometimes leads to reduced sensitivity and Genesis of Ultrasound Transducer by means of a low f-number aperture (i.e., f-number < 4), the of lateraldegrading resolution at depth in attenuating media [16]. the resolution depth of field is reduced, considerably Transducer fabrication Computer modelling ObJectives_ in regions away from the focal plane.effect means offrom movingthe selective attenuation of the hi Thus, aresults Region or organ latter -Electro-mechonical Bocking ntion X layers -1/4 Penetr characteristics the focal zone over the imaging depth is required. A descripDesired resolution frequency tion of the depth-of-field problem and a components detailed analysis ofof the pulse. The shift of i tow -Electrical -Beam distributions matching Speed of dhta collection various transducers used to lower overcomefrequencies this effect are presented is more marked for a wide than for a na in Section IV. Testing bandwidth pulse. From (1), it is easily seen that this woul Pulse-echo device Other imoging devices Different approoches Another difficult decision is the selection of pulse frequency -Sensitivity te-cSingle element -Trsnsmission Doppler manifested reduced as lateral resolution. In Section IV and bandwidth. Three factors affect the choice important of profile _ -Multiple element -Beam Therapy -Scattering -Electrical impedance frequency: 1) tissue attenuation, of penetra- shift problem in which an empir 2) required examine thedepth frequency -Bandwidth and pulse tion, and 3) system dynamic range. There is no strict rule for shape formula Compromises hasoptimum been frequency. devised for selecting the optimum bandwi these equating variables and at arriving an ^ ~~~~~~~~Sensitivity Generally, the highest frequency provides adequate Sptial rtiesolution Transducer 3) that Materials: peneHaving established the cri nge an ultrasound transducer designed Fig. 2. The genesis ofDynomic trationfor is selected; this optimizes the lateral resolution . pulse-echo [cf. (1)]the transducer parameters, piezoelectric material, the back The pulse bandwidth needed to obtain good axial resolution Transducer materiol imaging. or the quarter-wave matching materials must be selected an poses different altogether problem. the Here, is tendency Ceramic-PZT for sensitivity to select the widest bandwidth (shortest __ because this ~~~~PVDF for wide bond-pass Section II, pulse) we present a discussion of piezoelectric mater and flexibility provides the best axial resolution in water. Unfortunately, Others ? Overto the past 25 years, wide bandwidth sometimes leads reduced and loss ferroelectric ceramics have gaine sensitivity ofmodelling lateral resolution of image at depth in attenuating mediaacceptance most universal [16]. The as the active piezoelectric co etc.), theTransducer required penetration, theofspeed fabrication depth Computer latter effect results from thenents selectiveinattenuation of the higher -Electro-mechonical and the restrictions transducers. formation,-1/4Bocking The success of the ceramic materia resolution, physical Xthe layers desired characteristics frequency components of the pulse. The shift of i towards -Electrical matching distributions on transducer size and -Beam due to their largely extremely high piezoelectric and ele shape. lower frequencies is more marked for a wide than for a narrow Testing bandwidth pulse. have From (1),mechanical it is easily seen coupling that this would be 2) Transducer Compromises: Once these manifested coefficients. objectives Unfortunately, some of From: Hunt et al, IEEE -Sensitivity reduced lateral resolution. In Section IV, we -Beam profile a wide been, set,, -Electrical lesstheasconvenbenefit variety of pulse-echo and other of the large acoustic mismatch bet impedance examine frequency problemisin lost whichbecause an empirical Trans BME, shift 1983 -Bandwidth and pulse tional imaging shape techniques are available. After an appropriate formula has been devised forthe selecting the optimum ceramic the biological material. Special matc andbandwidth. Transducer Materials: arrangements 3)make Having establishedsuch the as critical been selected, it is for necessary quarter-wave layers or passive elect compro- the piezoelectric 2. The genesis has of an ultrasound transducer designed Fig.technique pulse-echo to transducer parameters, material, the backing, imaging. mises in the design of the transducer -to achieve circuits to improve the bandwidth and sensitivit adequatematching or the quarter-wave materialsare mustused be selected. In Section II, we present of piezoelectric materials. illustrates some a discussion spatial resolution and beam properties. Fig. 3Over transducers, but these complicate their construction. Recen the past 25 years, ferroelectric ceramics have gained alof the that confront thethelow-acoustic cyclic depth impedance paradoxes piezoelectric polymer polyvin designer. the required of penetration, the speed ofthe most universal etc.), active piezoelectric acceptance as imagetransducer compo-Scattering _ Therapy Maximum frequency Enough penetration ? namic range f \ Maximize lateral resolution L focussing.? Solve - of Bandwidth Transducer Design Concepts Axial resolution = Compromises in ultrasound imaging Resolution (axial and lateral) with frequency Penetration with frequency = an - - - - - Compromise between resolution and penetration - Lateral resolution with f-number Depth of field with f-number Compromise between focusing and DOF ro formation, the desired resolution, and the physical restrictions transducer size and shape. 2) Transducer Compromises: Once these objectives have been, set,, a wide variety of pulse-echo and other less conventional imaging techniques are available. After an appropriate technique has been selected, it is necessary to make compromises in the design of the transducer -to achieve adequate spatial resolution and beam properties. Fig. 3 illustrates some of the cyclic paradoxes that confront the transducer designer. on nents in transducers. The success of the ceramic materials is largely due to their extremely high piezoelectric and electromechanical coupling coefficients. Unfortunately, some of this benefit is lost because of the large acoustic mismatch between the ceramic and the biological material. Special matching arrangements such as quarter-wave layers or passive electrical circuits are used to improve the bandwidth and sensitivity of transducers, but these complicate their construction. Recently, the low-acoustic impedance piezoelectric polymer polyvinyli- Transducer Design Concepts ASOUND TRANSDUCERS Ultrasound Imaging Modes 455 Transducer compromises A-mode M-mode B-mode Maximum frequency Enough penetration ? namic range f \ Maximize lateral resolution Attenuation (frequency shifts) Strong ~~~~focussing focussing.? L LoseI sensitivity? Solve depth of field limitations / Bandwidth Axial resolution f/2a 1.41Af-number components and beam properties of focused piezo- Fig. 3. The compromises for the sensitivity, tissue penetration, and spatial resolution that are needed to design an efficient ultrasound device. electric circular transducers. Genesis of ce t nt an From: Hunt et al, IEEE Trans BME, 1983 Ultrasound Transducer ObJectives_ Region or organ Penetr ntion Desired resolution - Speed of dhta collection - - Other imoging devices -Trsnsmission -Scattering Different approoches _ Compromises ~~~~~~~~Sensitivity - Sptial rtiesolution - Dynomic ro nge Transducer materiol Ceramic-PZT for sensitivity ~~~~PVDF for wide bond-pass and flexibility Others ? - Doppler Therapy For example, if one desires to improve the lateral resolution by means of a low f-number aperture (i.e., f-number < 4), the depth of field is reduced, considerably degrading the resolution in regions away from the focal plane. Thus, a means of moving the focal zone over the imaging depth is required. A description of the depth-of-field problem and a detailed analysis of various transducers used to overcome this effect are presented in Section IV. Another difficult decision is the selection of pulse frequency and bandwidth. Three important factors affect the choice of A-Mode frequency: 1) tissueDisplay attenuation, 2) required depth of penetration, and 3) system dynamic range. There is no strict rule for these simplest Oldest, typearriving at an optimum frequency. equating variables and theof Generally, highest frequencyofthat provides adequate pene Display the envelope pulse-echoes vs. time, trationdepth is selected; d = ctthis /2 optimizes the lateral resolution [cf. (1)] . The pulse bandwidth to obtain good axial depth resolution – Measure theneeded reflectivity at different poses an below altogether problem. Here, the tendency is the different transducer position to select the widest bandwidth (shortest pulse) because this provides the best axial resolution in water. Unfortunately, wide bandwidth sometimes leads to reduced sensitivity and loss of lateral resolution at depth in attenuating media [16]. The cer fabrication Computer modelling latter effect results from the selective attenuation of the higher -Electro-mechonical king characteristics layers frequency components of the pulse. The shift of i towards rical matching -Beam distributions lower frequencies is more marked for a wide than for a narrow Testing bandwidth pulse. From (1), it is easily seen that this would be sitivity manifested as reduced lateral resolution. In Section IV, we am profile ctrical impedance examine the frequency shift problem in which an empirical ndwidth and pulse ape formula has been devised for selecting the optimum bandwidth. 3) Transducer Materials: Having established the critical f an ultrasound transducer designed for pulse-echo transducer parameters, the piezoelectric material, the backing, imaging. or the quarter-wave matching materials must be selected. In Section II, we present a discussion of piezoelectric materials. Over the past 25 years, ferroelectric ceramics have gained aldepth of penetration, the speed of image most universal acceptance as the active piezoelectric compoired resolution, and the physical restrictions nents in transducers. The success of the ceramic materials is and shape. largely due to their extremely high piezoelectric and electroCompromises: Once these objectives have mechanical coupling coefficients. Unfortunately, some of this variety of pulse-echo and other less conven- benefit is lost because of the large acoustic mismatch between and the biological material. Special matching chniques are available. After an appropriate the ceramic M-Mode n selected, it is necessary to make compro- arrangements such as quarter-wave layers or passive electrical of the transducer -to achieve adequate circuits are used to improve the bandwidth and sensitivity of ign and beam properties. Fig. 3 illustrates some thesecorresponding transducers, butsignal complicate their Display the A-mode to repeated input pulses in separate construction. Recently, a 2D image, for a fixed transducer position the of low-acoustic impedance doxes that confront the transducer designer. column piezoelectric polymer polyvinyli– Motion of an object point along the transducer axis (z) is revealed by a bright trace moving up and down across the image – Used to image motion of the heart valves, in conjunction with the ECG Application of A-Mode B-Mode Display time depth Applications: ophthalmology (eye length, tumors), localization of brain midline, liver cirrhosis, myocardium infarction Frequencies: 2-5 MHz for abdominal, cardiac, brain (lower for brain); 5-20 MHz for ophthalmology, pediatrics, peripheral blood vessels Used in ophthalmology to determine the relative distances between different regions of the eye and can be used to detect corneal detachment – High frequecy is used to produce very high axial resolution – Attenuation due to high frequency is not a problem as the desired imaging depth is small Move the transducer in x-direction while its beam is aimed down the z-axis, firing a new pulse after each movement Received signal in each x is displayed in a column Unlike M-mode, different columns corresponding to different lateral position (x) Directly obtain reflectivity distribution of a slice! Application of B-Mode Transducer Array Can be used to study both stationary and moving structures High frame rate is needed to study motion Directly obtain reflectivity distribution of a slice – No tomographic measurement and reconstruction is necessary! 40-MHz annular array transducers for dynamic focusing Array types a) b) c) d) e) With a single crystal, manual or mechanical steering of the beam is needed to produce a two-dimensional image Practical systems today use an array of small piezoelectric crystals – Allow electronic steering of the beam to optimize the lateral resolution Linear Sequential (switched) ~1 cm × 10-15 cm, up to 512 elements Curvilinear similar to (a), wider field of view Linear Phased up to 128 elements, small footprint → cardiac imaging 1.5D Array 3-9 elements in elevation allow for focusing 2D Phased Focusing, steering in both dimensions 5-element array pattern Prototype transducer Ketterling et al, IEEE Trans UFFC 2005 Annular array transducer improves focusing in depth E11.5 Mouse Embryo Fixed-Focus Array-focus 3-D Imaging By mechanically or manually scanning a phased array transducer in a direction perpendicular to the place of each B-mode scan By electronically steering the beams to image different slices Focal Zone Ventricle Segmentation Aristizábal et al, Ultrasound Med Biol, 2006 B-mode Scanner Types Phased Arrays B-mode scanners use multiple transducers Phased array: – Much smaller transducer elements than in linear array – Use electronic steering/focusing to vary transmit and receive beam directions Delays for Steering Beam Steering (Transmit) Extra distance that T0 travels than T1: Δd = d sinθ For the wave from T1 to arrive at a point at the same time as T0, T1 should be delayed by Δt = Δd/c = d sinθ/c If T0 fires at t=0, Ti fires at ti = iΔt = id sinθ/c Beam Focusing (Transmit) Delays for Focusing Receive Beamforming Receive Dynamic Focusing T0 fires in direction θ, and all Ti’s receive after a certain delay, so that they are all receiving signal from the same point at a particular time Delays for Dynamic Focusing First consider a stationary scatterer at (x,z) Time for a wave to travel from T0 to the scatterer and then to Ti is ti = {(x2+z2)1/2 + [(id-x)2+z2]1/2}/c Practicalities of dynamic focusing Steer and focus the transmit beam in direction θ Focus the receive beam dynamically along that direction Increment steering direction to θ + Δθ Repeat for the new direction / image line Time difference between arrival time at T0 and at Ti Δti = t0 - ti Desired time delay is a function of t: Steering and Focusing: Summary Doppler Ultrasound: Reminder Beam steering and focusing are achieved simply by applying time delays on transmit and receive The time delays are computed using simple geometrical considerations, and assuming a single speed of sound These assumptions may not be correct, and may lead to artifacts Doppler Equation: fd = 2fo.v.cosθ/c Transducer fo fo+fd θ Blood flow – fo is the frequency transmitted – v is the velocity of the moving blood – c is the sound speed in the medium (blood, ~1600 m/s) Doppler Ultrasound Instrumentation CW Doppler (2 transducers) Pulse Mode Doppler Measurement Doppler Imaging via Time Correlation Performing correlation of two signals detected at two different times Deducing the time shift (correspondingly distance traveled) that yields maximum correlation Determine the velocity Duplex Imaging Doppler Data Processing Combines real-time B-scan with US Doppler flowmetry B-Scan: linear or sector Doppler: C.W. or pulsed (fc = 2-40 MHz) Duplex Mode: – Interlaced B-scan and color encoded Doppler images ⇒ limits acquisition rate to 2 kHz (freezing of B-scan image possible) – Variation of depth window (delay) allows 2D mapping (4-18 pulses per volume) (Aristizabal, Ultrasound in Medicine & Biology 1998) Duplex: Imaging + Doppler Blood Velocity Use only one transducer – Transmits short pulses and receives backscattered signals a number of times Can measure Doppler shifts at a specific depth Time Color Doppler of a Mouse Embryo Color Dopper Imaging Example Homework Clinical Applications Ultrasound is considered safe; instrument is less expensive and imaging is fast Clinical applications – Obstetrics and gynecology » Widely used for fetus monitoring – Breast imaging – Musculoskeletal structure – Cardiac diseases Contrast agents Reading: – Prince and Links, Medical Imaging Signals and Systems, Chapters 10 & 11 Problems: – – – – – – Work through example 11.3 in text (not to be handed in) P11.2 P11.3 P11.6 P11.9 P11.14