Exam 3 Lecture 4 Now in terms of solvents used in HPLC, you have as MP water + organic solvent (methanol,acetonitrile,THF) if you have problems with one solvent, you can try another First choice in HPLC is acetonitrile gives you very good separation Reversed phase chromatography using C18,C8, or phenyl columns Acetonitrile very low UV cutoff? often times in HPLC you’re using a UV detector 190 means you can set the UV detector at 190 and above 190 and below you will have UV absorption and that will affect your analysis anything above 190 you can use it for detection Second choice is methanol has a UV cut off of 205 problem with methanol is its high viscosity will have higher backtracker as compared to acetonitrile but the advantage of methanol is the expense associated with it acetonitrile is more expensive (shorter of acetonitrile so the price has tripled or quadrupled) THF the UV cutoff is 12 that’s higher than the other solvents problem with THF is the oxidative potential has epoxides in the solvent can affect your analysis by oxidizing compounds because of the problems with THF THF can slowly degrade and oxidize by air to give you peroxides So these are solvents often used in HPLC of course the other components buffer to maintain pH at a certain neutral or acidic point TFA to maintain pH ion-pairing potential of TFA So in terms of detectors for HPLC The criteria for ideal detector is the sensitivity to analytes you are analyzing and the selectivity to those compounds Universability because you don’t want to switch your detectors a lot UV is pretty universal for only compounds that contain a chromoform if they don’t contain chromoform it cant be seen using UV should produce a linear response for when you do quantitation you want a standard curve so the standard curve needs to be linear don’t want the peaks broad either So these are the criteria Will compare the UV detector, fluorescence detector, refractive index detector and the mass spect For both UV and fluorescence this applies mostly to the UV a light source flow cell (solvents coming in from here and out) this is the distance it measures this is the amount of material inside this flow cell that can detect it by detector Since you have compounds coming in here originally the MP acetonitrile and water at certain percentage but if you have compounds coming in and you’re going to have different absorptivity in this range that will be recorded by detector So in terms of sensitivity and the usefulness of these detectors which the gradient because often times in HPLC we are using gradient elution So you are looking at UV, refractive index, and then fluorescence, and then masspect UV is the most commonly used detector in HPLC compatible with gradient elution In terms of sensitivity you’re looking at 0.1 nanogram of material in UV Refractive index refractive index change of the eluent coming off the column so then its not compatible with gradient elution because if you change the gradient (the composition of MP) you will change refractive index and the other problem with refractive index - the sensitivity 1000 times less than the UV detector but the advantage is its universal anything that comes into the solvent stream is gonna change the refractive index universal but less sensitive like the thermal conductivity detector in GC This is used for when you have a compound without a UV chromoform (cant be used by UV detector) Fluorescence detector more sensitive as compared to UV detector because it goes down to 1 pictogram of material and then its also compatible with the gradient elution because as long as the solvents you use (methanol, acetonitrile, and THF) they don’t have fluorescence but in terms of the detector can only detect compounds that can fluoresce its more sensitive, but in terms of selectivity only compounds that are fluorescent Mass spect for GC its getting more and more commonly used, for HPLC as well sensitivity is comparable to UV compatible with gradient dilution the advantage of mass spect is that it gives you the molecular weight of the compound and it gives you fragmentation patterns gives you more structural information than the other 3 in his labs they have GC and HPLC MS But the disadvantage of mass spect much more expensive than the other detectors with the other detectors 30-50,000 dollars with all the pumps but just the mass spect alone will be 100,000 dollars but then the advantage of the mol. Wt and fragmentation pattern is very worthwhile in terms of the expense. Next a few examples of using HPLC in quantitive analysis of drug molecules 1. extraction of APAP tablet from the textbook the drug molecule is paracetamol (book was from England) a. APAP do extraction inject into HPLC UV absorption determined by UV b. The solvent front/ peak coming off at 3.363 minutes rather exact c. Standard if you have just pure compounds, you know the concentration, put it into HPLC comes off at exactly the same place but then this 3.36 and 3.34 because you’re doing 2 separate injections, the difference is only in the second decimal point considered to be the same compound in order to confirm that these 2 are the same you can mix the 2 together and do a coinjection if they give you 1 peak you will have 1 compound d. For this example you are using what is called an external standard HPLC is a very precise measurement of peak size, area, with the concentration so you can use a standard APAP as the compound then you have a series of concentrations made and then do a standard curve like this this is the concentration of the APAP and the y axis is the peak area under the curve what you see is within this concentration range you will have a linear relationship e. inject the sample extract of the tablet if the compound comes out with a peak area in this range what you need to do is draw a line and drop down and then calculate the concentration of your sample can calculate how much you have in the tablet using external standard f. the external standard you are using is the same compound you are analyzing same retention time if you have exactly the same concentration you will have the same peak area will have a linear relationship g. often times when you have extraction processes like for tablets in tablets you have many other materials in it can have excipients that can affect recovery so if your recovery is very good and its at 90-95% and also reproducible you can nuse external standard the experiment HPLC is very precise in measuring concentration but if your recovery for the sample preparation to extract your compounds out of the tablet is very low 50-60% the other half cannot be extracted out then you have to consider using the next standard called internal standard h. internal standard mentioned in GC adding a known compound to the mixture before you do the extraction dissolve and break down tablets into little pieces everything into solution add known amount of internal standard very similar compound to the one you’re analyzing but different enough to produce a different peak so whats the criteria that you use in terms of selecting internal standard i. internal standard should closely relate to the structure of the analyte so the compound we are analyzing here is hydrocortisone so whats chosen is betamethasone (steroid) using an standard that is a steroid is well difference is the methyl group and then there is a double bond in the standard but not in hydrocortisone there is a Fluorine that’s also not in hydrocortisone closely related so the behavior of these 2 compounds ill be similar when you do the extraction during the sample preparation 1% or 5% of hydrocortisone and during the extraction using about 5% of this standard for the extraction ii. compound should be stable because the internal standard should not decompose during the analysis during analysis extraction or storage iii. photographically resolved when you do the separation you have 2 peaks when you mix the 2 you have to have the 2 peaks separated from each other you cannot have overlap and of course the resolution between these 2 you want to have greater than 2 resolution so in this case you will have more than baseline separation iv. even though they should be separated they should be close as possible so that means you don’t want it to have one here and one close to the solvent front bigger separation which means 2 compounds are not very structurally related v. having the same weight of material the response should be similar so if you have 1 nanomolar concentration the response of the peak size should be about the same the 2 compounds that you look at over here the 2 peak sizes are very close together so they’re close in the terms of these 2, but are separated in terms of resolution of 2 or higher and then in terms of peak size, if you have the same amount of material you’re going to have similar peak size several chromatograms over here one over here that’s just hydrocortisone so when you have the compound without internal standard will show 1 peak when you have the internal standard so this solution in terms of hydrocortisone is a cream so you’re talking about extract of the cream hydrocortisone cream so you can add the internal standard and it will go through same extraction process and will have 2 peaks show up need to make sure that these 2 compounds are well-separated from anything else that’s being extracted you don’t want the internal standard to overlap with the excipients in the cream the first one here has known amounts of both the drug molecule and also the internal standard which these 3 graphs , and especially with these 2 you can quantitate how much of hydrocortisone you have in the tablet? that’s the calculation you can do shows you what you can use HPLC for in terms of quantitation External standard and internal standard But the other thing we didn’t discuss In GC there is derivatization but in this case we never mentioned derivatization because in this case its not necessary in HPLC because you are looking at compounds that don’t really need to vaporize But in GC when you have very polar compounds, ones that contain OH or amines you need to derivatize them otherwise the peak shape will be rather tailed doesn’t give you good quantitation very slow/sloped (?) peak and very tailed peak not good in terms of integrating compounds for peak area So we have looked at the different chromatographic separations for both GC and HPLC So what if you were asked to analyze compounds in order to select the method of analysis need to look at the compounds you are analyzing 1. polarity in terms of polarity of analytes compounds at this end will have nonpolar compounds will have strongly polar compounds on this end so this is then looking at the compounds you are asked to analyze so for the nonpolar compounds that means they will be more likely to be volatile for the strongly polar compounds less likely to be volatile so then in terms of determining which method you should peak for this analysis most likely you will be using GC when you have compounds thath are very nonpolar so they are more likely to be lipid soluble , more likely to be volatile so you can use GC analysis if you have compounds with this range which compounds would come out first so the more volatile compounds will be coming out first, this end will be early, this end will be late for GC you are going to increase the temperature of the column oven to elute the compounds off to have them go into the MP (helium) so the more volatile the compounds are, the more likely they will go into MP and will come out earlier the ones that are less volatile will spend more time in SP will come out later 2. HPLC discussed in 2 different possibilities normal and reverse phase a. Normal a more polar stationary phase and a less polar mobile phase – so you have compounds that are strongly polar in this end these analytes will have more interactions with SP will be coming out later less polar compounds will come out earlier because they will have less interaction with SP the organic solvent will elute the less polar compounds out earlier b. Reverse more common SP are less polar and the MP phase is aqueous solution is more polar strongly polar compounds will be coming out first compounds that are less polar will have more interactions with the SP like C18,C8 will come out later c. Compounds with charges + and - and want to separate them from opposite charges, or separate from ones with no charge you can have ion-exchange chromatography running in aqueous solution charged molecules will be in water d. This concludes the discussion on the separations