Directions for using sodium lidar (Updated March 2012) -Start up the seed lasers, from the IP 128. 39.135.160 for the 1064nm and 1319nm -Turn on the standby button for the 1064nm seed -Wait for the actual temperature to reach the set temperature for both diodes of the 1319nm seed laser -Turn on diodes (green button) for the 1319nm laser Standby 1064 Diode on/off Temperature toggle Above: Program to turn on instruments, right is the 1064 (top) and 1319 (bottom) -Make sure that the temperature to the sum frequency generator is well adjusted to obtain maximum power (thermistor reading here is 5.2) Above: Power reading and temperature adjustment -The coupling of the two seed lasers into the fiber can also be adjusted. Each individual laser can be looked at separately to ensure that both are overlapped and properly coupled in. -Generally the knob pointing up and the one pointing down in the following figure are the best ones to increase coupling Above: Stage for coupling of seed lasers into the fiber Frequency Generation Using the piezo scan program to lock the laser -Turn on the power to the doppler free sodium cell PMT from the computer -Start the scan piezo labview program(piezoscan5a) -Choose the direction of scanning (left or right, red or blue) -flip the tune switch in the program -once the dip is seen in the screen, turn off the tune switch -turn on piezo feedback and slow tune feedback -turn piezo amplitude down to about .5V (do this after everything else is ready as the lasers may continue to warm up and the lock may be lost) An image of tuning program note: If there is ever concern about the spectrum, the voltage divider can be disconnected for a larger sweep range than 10V so that a larger range can be seen by the viewer. Also, it is good to let the seed lasers warm up for a period of time before dropping the scan voltage, this way the tuning does not get stuck in a second dip that is not the lowest point. -Turn on the AOM from the computer Note: The controls for the shutter can be set to manual or, when collecting data, they must be set to x_gate for the shutter and ttl for the AOM shutters. Shifted Beam Alignment 1 Shifted Beam Alignment 2 AOM Control Unshifted Beam Alignment Shutter Control -Turn on the PDA pumps from the computer (always do this before turning on the yag or the cells can crack) -check PDA power sources beneath the PDA. These should be between 10-12V and should be checked during the time running the laser. -Be sure that the motor is pumping the dye. This should also be checked throughout the duration of running the laser. (Feel pumps for vibrations). Above: Left, pump power sources. Right, PDA pumps -Turn the key to the water cooler in the hall (to the right) -Turn all values upward (parallel to the pipes) in the hall next to the water cooler. Put the note on that the sodium lidar is running. The one in the middle will be turned if the other lider is on. -If the lidar has not been run in awhile, check the gauges in the back of the cooler to make sure that the system is above the fill line. Above: Valves for cooler and the water cooler with key switch at bottom right -When first adjusting the setup after a period of time, look for the bulls eye pattern at the output of the first and second stage of the PDA output and make sure it looks spherical. This can be done for each shifted frequency and the non-shifted frequency. Look for Bulls Eye for each of the unshifted and shifted beams and a piece of paper. Alignment can be done from the alignment mirrors in the AOM control box. Finer adjustments can be made using a power meter. -Turn the power on for the yag. Sometimes the simmer on the system fails, and if this happens, just retry until it works. -Turn on the choppers using the computer in the back. This is from IP 128. 39.135.160 -Turn on BNC delay box from the back computer -hit run on the actual delay box (in the bottom right corner of the box) -turn on PMTs from the computer, and turn on the high voltage from the control boxes (yellow buttons underneath the table) Chopper Control High Voltage Buttons BNC Delay Run Button -Turn on the yag by choosing run_norm from the control box. The choppers and BNC delay must be on by this point in order for the laser to externally trigger. -It may take a bit for the yag to warm up, but the yag q_switch is started from run_norm. If this does not start the yag, go to monitor->Beamlock on. Additionally, if this still does not work turn d_lock on. Above: YAG Laser control box -Go back to the piezo scan program -In the piezo scan program, recheck that the lowest point is still locked and lower the sweep range to about .5V Using The Faraday Filters Note: for more information on how to replace the sodium cells or open the filters see Katrina’s March 2012 trip report. -Turn on the Faraday Filters using the control at: http://128.39.135.107/ov.html -The filters need to warm up for 1.5-2 hours -Set up the faraday filters and diodes to take a spectrum of the faraday filter transmission -Use the test fiber and connect this to the input side of a chopper (make sure that light is coming through the fiber first) -Make sure that the chopper is open so that light will get through (see figure) -Collimate the light from the fiber. One of the 25mm lenses that is attached between the PMTs and the chopper is good for this purpose. (Katrina left the 25mm lens attached to the Faraday Filter on the March trip) -Attach a 50mm lens to front of the diode and attach the diode at the end of the faraday filter -When taking spectra, BOTH diodes must be on, even if only one filter is being used -The piezo scan MUST be OFF in order to run the scan spectrum program Open chopper -Run the program to take a spectra -The piezo scan MUST be OFF in order to run the program: Run the program ScanandSaveSpectrum10June2010.vi -change the name of the save file and select the save file option -Once a spectrum has been obtained, turn on the piezo scan program again and relock the frequency Above: ScanandSaveSpectrum program -Set up the receiver with the faraday filter -Connect the fiber from the telescope of interest to the input side of the faraday filter where the test fiber was previously connected. Replace the diode with the PMT setup. Note: I used the previous PMT setup (25mm, interference filter, 50mm lens) after the faraday filter and left the other 50mm lens with the diode. This may need to be changed to better focus the light on the PMT detector. I also left a 25mm lens at the front of the faraday filter to collimate light coming out of the fiber. -Chopper 1 does not have a stop plate so be careful Above: Left, diode w/50mm lens Right, chopper w/ 25mm lens before filter Chopper Faraday Filter Input for test fiber or receiver telescope Col. Lens Either 30mm & Diode Or 25mm/filter/50mm & PMT Above: Faraday filter setup Taking Data -The conditions for taking data, in addition to clear weather, are that the winds must be lower than 12m/s. -Be sure to uncover the telescope that you will be using -Make sure that the MCA servers for the counter boards are running before starting the data acquisition program (otherwise it will not work). Go to Programs/FastComtecP7882/MCA-3. -Before taking data, it is good to align the system by steering the beams to align with the telescopes. This can be done by using the DOS program on the right computer. -The Labview program Run3FreqBiff5b.vi can be run with the save data option off. -To make this process easier, it is possible to only send out one frequency during this time by turning the shutter to manual and turning the AOM controls off. -Use the DOS program at this point to adjust the motor control boxes to the left of the computer. Remember to turn these on before starting the adjustments. -There is a spreadsheet of relative motor adjustments for beam steering changes between zenith and off zenith directions on the blue cabinet by the door. It can be helpful to use these positions to adjust the mirror positions, just be sure to look at the difference in the motor positions and use this difference to adjust the mirrors. -Once the signal is good, the shutter should be turned to X_gate enable, and the two AOMcontrols should be turned to ttl. -The save data option in the program can be turned on and the group number should be adjusted to suit the data. For instance if data was taken earlier in the day, the group number should be changed. -Be sure to log the data run in the log book and keep track of the time that the laser has been used and the age of the dye. Above: Data acquisition program Above: stepper motor control program Above: List of relative motor positions Above: MCA Server Shut Down -Stop the data acquisition program -Stop the tune piezo program then turn off the Doppler free detector -If you were using the faraday filters, it may be good to take a second spectrum -Turn off lamps to the yag -Turn off choppers and BNC delay -Be sure to turn off the high voltage to the PMTs in addition to turning off the PMTs -Turn off the PDA pumps -Shut down AOM crystals and shutters -Turn of the seed lasers, first by hand, then using the computer -Turn off the Doppler free detector -Turn off the yag laser -Turn off the power supply and water pumps