Stripping and Recharging Nickel Columns A. Set up FPLC and Strip Columns 1. Make sure there is enough filtered ddH2O, 20% EtOH, buffer A and Stripping Buffer and that they have been recently degassed. Stripping Buffer 20mM Sodium Phosphate 500mM NaCl 50mM EDTA pH 7.4 buffer A (loading/binding buffer): 10 mM Tris pH 8.0 5 mM MgCl2 500 mM NaCl 5 mM Imidazole 2. Remove air from inlet tubing 3. PumpWash A & B w/ ddH2O 4. Also wash out tubing w/ 1-2 ml ddH2O (in load and inject positions) 5. you can strip up to 5 columns at once by attaching them together. Make sure there is a wet connection between each one. Wash Column w/ ddH2O to remove ethanol (5 column volumes (CV); in load position) 6. PumpWash A & B with stripping buffer 7. Run 10 CV of stripping buffer/column @1mL/min (Will usually be stripping multiple columns at once that will be daisy-chained together) 8. PumpWash A & B with binding buffer (A) 9. Wash Column w/ 10 CV of binding buffer (A) 10. PumpWash A & B with ddH2O 11. Wash Columns w/ 10 CV of mqH2O 12. Remove columns one at a time starting with one on the bottom to be recharged. B. Recharge Columns and Clean FPLC 1. Recharge the water-washed column by loading 0.5 ml (for HisTrap FF 1 ml) or 2.5 ml (for a 5ml column) of 0.1 M NiSO4, respectively. This is done by hand using sterile 10mL syringe. 2. Reattach columns to the FPLC 3. Wash columns w/ 5 CV of water 4. Move pump intakes to Buffer A; do PumpWash 5. Wash columns w/ 5 CV of buffer A (to adjust pH) 7. Move pump intakes to water; do PumpWash 8. Wash column w/ water – 5 CV 9. Move pump intakes to 20% EtOH; do PumpWash 10. Wash column w/ 20% EtOH for 5 CV; Remove column and cap and parafilm ends; store @ 4o 11. Do final wash PumpWash; Load Wash and Inject Wash (2-5 ml each) 12. Empty waste containers