SPECIFICATION FOR ADVANCE NEONATAL VENTILATOR (One) The ventilator should be microprocessor controlled designed for neonatal use. With possibility to upgrade with additional features. Continues flow, pressure limited, time cycled ventilator design. Ventilator modes: should have following modes (both pressure and volume) available in the unit. IMV/IPPV CPAP including non-invasive ventilation. SIMV, SIPPV/Assist-control. Volume targeted/guarantee mode of ventilation with ability to deliver and monitor tidal volume as low as 1-2 ml (Range 2 ml to 50 ml) Pressure support mode of ventilation. Apnea back-up ventilation. Should have integrated high resolution LCD screen for real-time display of Scalar (Pressure, Flow and Volume against time) and loop (Pressure-volume, volume-flow and pressure-flow) pulmonary graphics. Digital display of FiO2, peak pressure, means airway pressure, CPAP/PEEP. Expiratory tidal volume, expiratory minute volume, total frequency, spontaneous frequency, lung function monitoring including Compliance, resistance, lung distention coefficient, (C20/C), Lung time constant, Rate volume ratio etc. Should have built-in logbook for recording events like various alarms. Integrated monitoring: integrated volume and pressure monitoring i.e. monitoring of PEEP Pmax, Pmean and VT, VTspont, MV and MVleak. The volume monitoring should have NTPD to BTPS correction. Monitoring of I:E, frequency and Spont. Frequency. Settings range: Trigger Flow/volume, leak adapted PIP 8 to 80 cm H2O PEEP/CPAP 0 to 25 mbar I:E ratio 1:0 to 1:10 Insp. Time 0.1 to 2 Sec Exp. Time 0.2 to 30 Sec Frequency Up to 200 BPM Base Flow (VIVE) 1 to 30 LPM Synchronization Patient synchronization with adjustable Flow trigger Integrated blender for Oxygen 21% to 100% Integrated screen for display of Pressure-Time, Flow-Time and volume-time curves. Integrated monitoring of FiO2. Monitoring of flow: At the Y piece with facility to activate or deactivate it. Audiovisual alarms with advisory on-screen message: MV high/Low, Apnea, tube obstruction, FiO2 high/Low high PIP, low PEEP/CPAP, fail to cycle, gas supply low, power failure, ventilator inoperative, alarm log book. The ventilator should have automatic compensation for leakage and should monitor and display leakages. The ventilator should show trends of important parameters viz. C, R, FiO2, MAP etc. for evaluation of patient improvement. Ventilator should be supplied with Good quality medical air compressor (CE marked) and a heated limb humidified with servo regulation of humidity. Scope of supply. Ventilator on trolley with wheels and brake facility. Integral medical air compressor. Humidifier: Autoclavable humidifier chamber and complete patient circuit (set of 2 patient circuits with each ventilator) Circuit support arm 2 hose sets for conventional neonatal ventilation. Bacterial filters. Flow sensors (5 set) Oxygen cell (Permanent Type) Oxygen connecting hose Air connecting hose Battery back-up (at Least 30 minutes), Instruction manual Training CD/DVD Y sensor if the serving is at ventilator end Others: User manual with trouble shooting guidance, in English with Technical manual with maintenance and first line technical intervention instructions, in English. List with name and address of technical service providers in India. Training and installation at end-user site. Minimum warranty for 5 years (mentioning the parts covered and not covered in detail) Proposal for full service CMC for 5 years following expiry of warranty covering (i) 1 preventing maintenances quarterly, (ii) on-call technical intervention, spare parts and travels. The proposal for CMC should in detail mention the parts covered and parts not covered clearly, accompanied with latest price list. Firm must certify that they have not supplied the equipment at a lower cost to any other MCD or Government hospital (At the time of tender) The firm must certify the availability of the spares and consumables for the 10 years (At the time of tender) and that model quoted is latest and not obsolete, and spares are available for minimum 5 years after warranty. The manufacturing firm must guarantee that the response time of complain will be within one working day and will be rectified in next 48 hrs, if not rectified the standby equipment should be supplied. Complete and exhaustive list of rates of ; priced accessories, consumables and spares should be submitted separately at the time of tender, anything not in this list will be considered to be covered in CMC. o Rates should be quoted in Indian rupees only. o User list with performance report (applicable for both equipment and service provider) from two government hospitals to be submitted, authorized by end user. o The firm must have a service centre in Delhi with 24 hrs backup. o The firm must sign a tripartite agreement with the principal company and the MCD. o The manufacturer will be directly responsible for providing the CMC and the service of the equipment. The department will like to have a live demonstration of the equipment. The ventilator should have following options for upgrade (please give prices for each option) Communication interface with Laptop. PC software for archiving and analysis.