Lab 7 Liquid Chromatography-Mass Spec INTRODUCTION The LC-MS id a chromatography instrument that takes a liquid mixture and separates the compound based on mass and different physical properties. This instrument is using a column in which the mobile phase pulls the sample through and it uses a mass spec analyzer. Once the sample is separate in the column, as the sample comes off the column, it is ionized so that the mass spec can further analyze the sample. Thus, it will yield a mass spec spectra that provides additional information about the structure and what structures are in the sample. This instrument also provides a chromatograph that can be used to note retention times and resolution. PURPOSE The purpose of this experiment is to learn the instruments of chromatography. These instruments are the HPLC and LC-MS. We will run standard samples of caffeine made at different concentrations of ppm. With these concentrations, they will serve as our standard samples. Running these samples on the LC-MS and HPLC and analyze them by evaluating the spectra. Noting the retentions times and by creating a standard curve we can analyze the concentration and separations of an unknown sample. LC-MS is Liquid Chromatography–Mass Spectrometry. This instrument utilizes a similar column separation technique employed in HPLC, but with mass analysis in the form of mass spectrometry. After separation, at the end of the column, solvent is removed and the sample is ionized. A mass detector then scans the molecules (by mass) and produces a spectrum identifying the compounds within the sample. PROCEDURE (i) (ii) (iii) (iv) (v) (vi) Power on the instrument and make sure all reading are well Turn on the nitrogen gas tank in the gas room and on bench top. Before the instruments is setup and ran, make sure are samples are prepared and ready to go. Turn on the software for the LC-MS Make sure all parameters and setting are set and LC-MS is working properly. Set all Standards and unknown to a set vial position and run samples. Sample Preparation (i) Weighted 1.0006g of caffeine and dissolve in 1L of deionized water. This will made 1000ppm solution. Most of the caffeine was dissolve there may have been a couple little pieces of caffeine that weren’t dissolved and that tiny fraction was left behind when the solution was decanted. (ii) Make dilutions using the M1V1= M2V2 and the desired volume was 100mL. The desired concentrations were 50ppm, 100ppm, 150ppm, 200ppm, 250ppm, 300ppm. Once the standards were made, they were assigned a vial and the vial was filled with sample and placed in the LC-MS tray. The unknown samples were collected by heating a tea bag of decaffiened and caffeinated. Once the tea bags were fully brewed 1mL of solution was diluted into a 100mL volumetric flask with deionized water. Assigned it to a vial, filled the vial with sample and placed in the LC-MS tray. (iii) Vial Number # A B C D E F G H L Sample Preparations of Standards and Unknown (mL) Of Stock Dilution Flask (mL) Desired Concen. Solution (ppm) 5 (mL) 100 (mL) 50 ppm 10 (mL) 100 (mL) 100 ppm 15(mL) 100 (mL) 150 ppm 20(mL) 100 (mL) 200 ppm 25(mL) 100 (mL) 250 ppm 30 (mL) 100 (mL) 300 ppm 1 (mL) 100 (mL) (x) 1 (mL) 100 (mL) (x) 1 (mL) 100 (mL) (x) Retention Times (min) 1.8480 (min) 1.8457 (min) 1.8575 (min) 1.8602 (min) 1.8606 (min) 1.8540 (min) 1.7056 (min) 1.7579 (min) 1.6862 (min) RESULTS LC-MS Results For Caffeine Standards (A-F) and Unknown Tea Samples (G-H) Vial Number # Concentration (ppm) Area (Counts) A 50 (ppm) 5495900 B 100 (ppm) 2853700 C 150 (ppm) 20375000 D 200 (ppm) 23934000 E 250 (ppm) 31307000 F 300 (ppm) 33924000 G unknown 528110 H unknown 480140 I unknown 515490 When running the samples on the LC-MS the previous results were obtained. In order to analyze our results we had to make a standard curve of the data of Area Vs. Concentration. The standard curve could be further used to find a concentration for the unknown samples given the equation y = mx + b. thus, with this equation the following information can be gather (i) (ii) The concentration of our unknown species Where the unknown sample concentration fall within our standard curve graph. Graph Standard Curve for Caffiene Standards (LC-MS) Concentration Vs. Area Under Peak 40000000 35000000 Area (counts) 30000000 25000000 20000000 15000000 10000000 y = 117476x - 505400 R² = 0.9804 5000000 0 0 50 100 150 200 250 300 350 Concentration (ppm) CALCULATIONS All o f the calculation done to develop (ppm) concentration were done using the M1V1= M2V2. The desired concentrations were chosen by making an educated guess about what the concentration of the unknown (which was the caffeine composition in the tea). A sample calculation is shown below. For Sample A: (M2) Desired concentration: 50 ppm (V2) Desired Volume: 100 mL Volumetric Flask (M2) Known (Stock) Concentration: 1000ppm (1000 𝑝𝑝𝑚)( 𝑥 ) = (50 𝑝𝑝𝑚)(100 𝑚𝐿) X = 5 mL This means that 5 mL of the stock solution needs to be diluted into a 100 mL volumetric flask to achieve a concentration of 50 ppm. These calculations were done for all standard concentration. The other calculation that had to be done was using the equation of the line obtain from the standard curve graph above to determine the concentrations of the three unknown samples. Simply, you plug in the obtained (y) values (Area) that were obtained from the LC-MS spectra and solve for x. The following calculations were made. For (1) Unknown Sample G: (528110) = 117476 (𝑥) − 505400 𝑥 = 8.797626749 ppm The following concentration was determine using the equation of the line obtained for the sample A and the procedural calculation was repeated for the two additional samples. Sample H: (480140) = 117476 (𝑥) − 505400 x = 8.39892880247880 ppm Sample I: (515490) = 117476 (𝑥) − 505400 x = 8.69020055160 ppm CONCLUSION We determined that the concentrations of all three caffeine samples were 8.79 ppm, 8.39 ppm, and 8.69 ppm respectively. One of the samples was suppose to be decaffeinated. Thus, for sample H there wasn’t supposed to be a caffeine peak. However, there could have been a miss up because the first two dilutions that were made the protocol were to have unknown G be the caffeinated tea sample and unknown H be the decaffeinated sample. Thus, When I was making the dilutions I wasn’t sure if I did just that because I got side tracked while I was making the dilutions. Thus, the peak that appeared on the data output for the LC-MS for all unknown were 1.71 (min), 1.757 (min), and 1.68 (min) for unknown G, H, and I respectively. This was very different then what we obtained from the standards which were about 1.85 on average for all standards. Thus, we were consistent with picking our peaks for the unknown for around 1.70 because that was the characteristic peak that appeared for Unknown G, which was suppose to be the caffeine sample and it was somewhat close to 1.85. Also, I think that since the concentrations for the unknowns were so much lower and possibly a little more diluted that the retention time was so much lower. Also, It could be a possibly that the tea bag that was supposed to be decaffeinated had small amounts of caffeine because when doing the HPLC experiment that was done consecutively with the experiment the same data came about and all unknowns were shown to have caffeine. Thus, we could have also made the mistake of grabbing 3 samples that were caffeinated. To improve this experiment it was be best to repeat the experiment being a more cautious and patient.