Sample QC guidelines for ChIP v1.4

advertisement
ChIP Sample QC Guidelines
We expect that your samples will be supplied in the correct containers and at the
required concentration and quality (requirements are stated below). Timely
completion of your project depends on you following these guidelines.
To prevent delays to projects that are batch-processed using our automated
systems, in the event that a sample does not meet our QC criteria, we will still
proceed with the library preparation. Your project manager will contact you to
inform you of the problems with sample QC. It will then be at your discretion
whether you sequence the library. If you choose not to sequence the library and
instead provide us with a replacement sample(s), drop that sample or cancel your
project you will still be charged for library prep and sample QC. Additional costs for
QC and library preparation will be levied for each replacement sample(s). All new
samples will be added to the end of the queue for the next batch of samples to be
processed.
For further details on packing and submitting your samples, please see the following
document ‘Guidelines for Preparing and Submitting Samples v1.4’.
Sample QC
Please check the following guidelines to ensure that your samples meet our QC
criteria. If you are unsure please contact your project manager prior to sending the
samples. If your sample(s) do not meet our QC criteria we will still proceed to make a
library. A sample of poor quality is likely to produce a poor, or biased, library.
ChIP
We require 10ng of material. Quantification should be done by Qubit or picogreen as
nanodrop overestimates the amount of material present. Samples should be in a
volume of no more than 20μl of ultrapure water, only 15μl will be used for the
library preparation. Input samples may need dilution, however, please do not dilute
ChIP samples as this may make them undetectable during the in-house QC.
Please note, due to the very low amounts of DNA provided by our customers, ChIP
libraries have a higher failure rate than other library types. This is largely as a direct
result of the low input amounts but is also due to the fact that we rarely have
1
Version 1.4
March 2015
enough material to troubleshoot when needed.
If it is not possible for you to generate 10ng of material in a single ChIP experiment
and you do not wish to pool experiments then we recommend that you perform
your experiment in duplicate, reserving one replicate so that it can be supplied at a
later date if needed.
Samples should be run on a High Sensitivity bioanalyzer chip or High Sensitivity
Screen Tape (TapeStation) to test the fragmentation. For the library prep, we require
that there is material in the 50-350bp range. We recommend that a control real-time
PCR is run to confirm that the expected enrichment is present prior to submitting the
sample for sequencing. Although there is no defined limit for the amount of
enrichment that can be detected by sequencing, in some cases, 2-3 fold enrichment
by qPCR has not been detected adequately by sequencing.
Summary Table
2
Type of
Sample
Amount
required
ChIP
10ng
Version 1.4
March 2015
Concentration Volume
>0.33ng/μl
20 μl
Bioanalyzer
High
Sensitivity
Chip
Download