Resident Physics Lectures Christensen, Chapter 7 X-Ray Beam Restrictors George David Associate Professor Department of Radiology Medical College of Georgia Restrictors Types Aperture Diaphragms Cones and Cylinders Collimators Aperture Diaphragms Lead sheet with hole in center ◦ simple ◦ large penumbra (shadow) Lead close to focal spot Focal Spot Penumbra Film Cones & Cylinders Focal Spot Focal Spot Film Film With cone angle as above, this functions as an aperture Reduced Penumbra Collimators Advantages ◦ Provides adjustable rectangular field fluoro may also have circular field ◦ light beam indicates x-ray field Configuration ◦ two sets of shutters longitudinal transverse ◦ sets may actually have two pairs of shutters operating together One set; two pairs Collimators Generally include filter & light Filter Mirror Focal Spot Lamp Shutters (only 1 set shown) Collimator Features Numerical field size indicators ◦ Field size is SID dependent 8 10 12 14 17 40” 8 10 12 14 17 Collimator Features Field center indicator (crosshairs) ◦ plastic sheet attached to bottom of collimator two black lines drawn on it ◦ light projects field center on patient Shutters Crosshairs Bottom of Collimator Positive Beam Limitation (PBL or Automatic Collimation) senses film size, orientation, & SID ◦ source - image distance limits x-ray field to size of film No longer required PBL does not insure tube & bucky are aligned! Cassette Collimator Checks X-Ray / Light Field alignment Alignment of x-ray field to bucky Field Size Indicator accuracy PBL accuracy