ADVIA® 360 Hematology System PEP to Go Answers for life. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 1 2014-07-22 HILS# 00001401 Effective date 04/09/2015 Global Product Education Table of Contents 1. System Overview 2. Daily Setup 3. Sample Processing 4. Technologies 5. Calibration 6. Quality Control 7. Maintenance © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 2 2014-07-22 Global Product Education System Overview Front, Top, and Right Side External Panel Components: 2 1. LCD Capacitance Touch-screen 2. On-board Printer (Prints to 58mm wide thermal paper) 5 1 3. Right Side Door Lock 4. Sample Rotor Station 5. Status light – Green (Ready), Red (Busy), Orange (Standby) 6. Barcode scanner (Scans sample and control barcodes and product and IFU reagent barcodes) 6 4 3 Note: The 2D barcode reader/scanner that ships with the system can be connected via one of the USB (A) ports. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. System Overview © Siemens rights reserved. Page AG 3 2013. All 2014-07-22 Global Product Education System Overview Back Panel and Reagent Connectors: 1. Reagent Connectors 2. Waste Connectors 3. Power Switch 4. Power Connector 3 5. USB (A) Port 1 and 2 6. LAN/Ethernet Port 4 7. USB-B Port 5 6 7 1 2 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 4 2013. All 2014-07-22 Global Product Education System Overview Reagent Connectors and Waste Outlets: 1. Diluent 2. Lyse 3. Cleaner 4. Waste Outlet Connector 1 1 5. Waste Outlet Connector 2 2 3 4 5 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 5 2013. All 2014-07-22 Global Product Education System Overview Left Side External Components: 1. USB (A) Port 1 2. USB (A) Port 2 Note: The 2D barcode reader/scanner that ships with the system can be connected via one of these USB (A) ports. Note: You may connect an external USB keyboard and/or a USB mouse to the system, but these do not ship with the system. 1 2 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 6 2013. All 2014-07-22 Global Product Education System Overview Right Side Internal Components: 1. Needle Chamber / Drain Chamber 2. Mixing Chamber 5 3. RBC Chamber 4. X-Y Sampling Unit with Piercing Needle 4 7 6 5. Sampler Head 6. Reagent Sensors 7. Vacuum Receiver 8 8. Sample micro-dilutor 9. Lyse Dilutor 10. Dilutors 9 10 1 2 3 12 11 11. Valves 12. WBC Aperture Assembly Chamber and HGB Head © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 7 2013. All 2014-07-22 Global Product Education System Overview Home Screen Components: 1. Measure – New Sample, Re-run, Print, Discard 2. Database – Details, Print, Filter, Statistics, Manage 3. Maintenance – Cleaning, Calibration, Diagnostics, Reagent Status 1 2 4. Exit – Shutdown, Preparing for Shipment, Logout Admin, User Management 5. Settings – General, Communication, Measurement, Printer, Date and Time 6 3 6. Quality Control – New Lot, Delete, Print, Export, Monthly View 5 4 Note: The Header at the top of the screen displays warning messages, the title of the Menu or function and the current time © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 8 2013. All 2014-07-22 Global Product Education Daily Setup Power On/Start Up the System: Turn on all external and peripheral devices prior to powering on the system Press the Main power switch located on the back of the system above the DC power input connector During startup, the ADVIA 360 screen appears, followed by the Home screen The system automatically runs a priming cycle if the fluidic system requires additional liquid and the fluid sensors are turned on Caution: Wait 5 minutes before initiating any measuring processes after turning on the power to allow the instrument to warm up. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 9 2013. All 2014-07-22 Global Product Education Daily Setup Check/Empty Waste: Check the contents of the waste tank If the level is above 80% full, empty the tank After disposing of the waste contents, navigate to: 1. Home 2. Maintenance 3. Reagent status 4. Empty Waste (located within the Waste box) Selecting Empty waste resets the waste counter to 0% The waste level % calculation is based on the setting of the tank capacity Note: If your system empties directly into a drain, the Waste tracking setting will be set to OFF, and you will not need to track the waste level. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 10 2013. All 2014-07-22 Global Product Education Daily Setup Check/Replace Reagents: To check reagents navigate to: 1. Home 2. Maintenance 3. Reagent status Check if the there is sufficient reagent for the expected number of daily measurements. The reagent level % calculation is based on the setting of the tank capacity To replace reagents navigate to: 1. Home 2. Maintenance 3. Reagent status 4. Replace 5. Reagent type 6. Scan the reagent barcode 7. Accept If you are replacing the lyse reagent, insert the reagent hardware key into the key slot on the back of the system. Note: Run a Priming cycle and a Blank cycle after the replacement of any reagents. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Daily Setup © Siemens rights reserved. Page AG 11 2013. All 2014-07-22 Global Product Education Daily Setup Reagents: 1. ADVIA 360 Lyse 3P Diff 1L: Color code – Yellow Creates hemolysate for 3-part WBC differential, WBC count and HGB determination Reagent key is attached to the bottle 2. ADVIA 360/560 Dil 20L: Color code – Green Dilutes whole blood samples Rinses the fluidic system between measuring cycles 3. ADVIA 360 Cleaner 1L: Color code – Dark Blue Cleans the tubing, counting chambers, and apertures 1 2 3 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 12 2013. All 2014-07-22 Global Product Education Daily Setup More Reagents: 4. ADVIA 360/560 Hypoclean CC 100mL: Color code – Black Used for the Hard Cleaning procedure to remove protein and lipids from sample tubes, chambers, and apertures 1mL should be dispensed into a 5mL sample tube and run before a Hard cleaning is started 4 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 13 2013. All 2014-07-22 Global Product Education Daily Setup Reagent Lock System/Hardware (HW) Key Each ADVIA 360 Lyse 3P Diff reagent comes with a hardware key containing 900 measurement credits. When the number of measurement credits reaches zero (0), the system can no longer perform a measurement, and Error 6003 displays: Hardware key not found, or is not suitable for this device! Please connect a proper hardware key! Each measurement cycle decreases the number of measurement credits by one, except for initialization, standby, and wakeup cycles which do not affect the credits To restore measurement credits, obtain a new reagent hardware key, remove the depleted key, and insert it into the HW Key reagent slot To display the number of measurement credits: Navigate to Home > Maintenance > Reagent status > Measurement credits © Siemens rights reserved. Page AG 14 2013. All 2014-07-22 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Global Product Education Daily Setup Perform a Blank Measurement: To ensures the cleanliness of the system Prior to running patients, controls, or calibrations on the system At the beginning of each shift and before sample analysis If reagents are replaced and/or maintenance is performed If the system has not been used for more than 3 hours Navigate to: 1. Home > Measure > New Sample 2. If a Blank measurement was not performed before running a sample, a message appears: “There is no valid blank measurement”. Select OK to continue with the Blank measurement 3. Otherwise, select Blank After one minute, the system displays the blank values. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 15 2013. All 2014-07-22 Global Product Education Daily Setup Perform a Blank Measurement Continued: If the Blank values are within the acceptable ranges, select Accept If one or more parameters are out of range, re-run the blank by selecting Re-blank If Blank values are consistently high for any of the parameters, run a Cleaning by selecting Maintenance > Cleaning > Cleaning Repeat Blank once Cleaning is complete If Blank measurement continues to be out of range after performing a Cleaning, perform a Hard Cleaning on the system: 1. Put 1mL of ADVIA Hypoclean CC reagent in a 5mL sample tube 2. Place sample tube in the sample rotor 3. Select Maintenance > Cleaning > Hard Cleaning > OK 4. When complete, repeat Blank measurement and select Accept if Blank is in range 5. If the Blank measurement is still out of range call the Customer Care Center – Technical Solutions for assistance © Siemens rights reserved. Page AG 16 2013. All 2014-07-22 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Global Product Education Daily Setup Check for Air Bubbles: Inspect the rear reagent liquid connectors and syringes (except waste line) If you see bubbles, prime the reagent and perform a blank run To prime the reagent, navigate to: 1. Home 2. Maintenance 3. Reagent status 4. Prime (located within each individual reagent box) or Prime all (located at the bottom of the screen) © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 17 2013. All 2014-07-22 Global Product Education Daily Setup Perform a Daily Shutdown / Cleaning: Initiate a Shutdown of the system from the Home screen by selecting Exit > Shutdown Turn off the power via the power switch when the system displays a warning and emits a sustained tone During a shutdown, the system: 1. First, saves data and settings, and closes the databases 2. Then, automatically cleans and flushes debris from the system to prevent clogs and contamination 3. Lastly, moves mechanical parts (sample needle, syringe pumps, and valves) into their home positions Warning: Do not turn off the instrument by flipping the power switch on the back panel. Doing so may result in erroneous operating, loss of data, or a prolonged start-up cycle during later use. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 18 2013. All 2014-07-22 Global Product Education Sample Processing Run a Patient Sample (Barcoded and NonBarcoded) 1. Mix sample gently by inverting 8 – 11 times 2. Navigate to Home > Measure > New Sample 3. Select a sample type (Human, Male, Female, Profile 1 – 7) 4. Identify the sample (Sample ID) by scanning the sample barcode or manually entering Sample ID 5. Insert the sample tube into the Sample rotor 6. Select Run 7. During this time the status light turns red, the sample rotor rotates the sample into the system for aspiration (100µL). After a few seconds, the sample rotor rotates back out to its original position 8. When sampling is complete, the status light turns green indicating that the system is ready for the next sample Important: If QC was not performed, perform QC prior to running a patient sample. Note: If a Blank was not performed before running a sample, a message appears: “There is no valid blank measurement. Press OK to measure a blank first”. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Sample Processing © Siemens rights reserved. Page AG 19 2013. All 2014-07-22 Global Product Education Sample Processing Interpret Patient Results: Results display in the Measure screen (Home > Measure) Results are divided into three categories: 1. Parameters 2. Normal Ranges 3. Tech Details The ID header displays: PID – Patient ID, date and time of sample run SID – Sample ID and sample Type Warning flags when appropriate Results screen – Parameters Values within normal range are white Values outside of the normal range are red and flag as L or H Histograms display graphical information about volume distribution of WBC, RBC, PLT Discriminators are indicated by vertical lines that separate different cell populations © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 20 2013. All 2014-07-22 Global Product Education Sample Processing Review Patient Results: Results screen – Normal Ranges Normal ranges display in a graphical format on the right of the screen Each value is indicated by a downwardpointing white arrowhead within the normal range graphics Low and high limits of the normal range display on either side of the range line © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 21 2013. All 2014-07-22 Global Product Education Sample Processing Review Patient Results: Results screen – Tech Details Probe voltage: shows electrode voltages for WBC and RBC counts WBC lyse volume: default value = 0.9 Diagnostic flags: shows evaluations made by the system based on results Warning flags display for the measurement © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 22 2013. All 2014-07-22 Global Product Education Sample Processing Review Patient Results: Flags Out of Normal Range Flags • Parameters are highlighted in red and display an L flag if under the normal range, and an H flag if above the normal range Measurement Condition Flags • If the system cannot interpret measurements, condition flags are attached to the result and display in the header of the Measure screen with a Warning, and in the Tech Details screen with a brief explanation. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 23 2013. All 2014-07-22 Global Product Education Sample Processing Review Patient Results: Flags Blank Cycle Flagging • The system flags parameter values if a high Blank has been accepted • Associated value is displayed but flagged with an asterisk (*) • Associated value is dashed out (---) and flagged by an E flag Note: Refer to the tables titled Flagging on WBC and HGB channels, Flagging on RBC/PLT channel, and Warning Flags of Differential Parameters in the Sample Measurement and Interpretations of Results section of the Operator’s Guide, for more information regarding flag symbols, their meaning, and explanation and recommended actions. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 24 2013. All 2014-07-22 Global Product Education Technologies Measurement Methods: Photometric Light Absorbance Method Measures the WBC dilution to determine the hemoglobin (HGB) concentration of the erythrocytes (RBC) Volumetric Impedance Method Measures the cellular concentrations and volume distributions of RBCs and PLTs from RBC dilution A lytic reagent is used to determine the distribution of leukocytes (WBC), and their three sub-populations (LYM, MID, GRA) 3-Part Differential Analysis Method Counts and simultaneously sizedifferentiates the white blood cells by using a WBC lytic process © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 25 2013. All 2014-07-22 Technologies Global Product Education Technologies Measurement Sequence: 1. The sampling needle moves into the sample vial and aspirates 100µL of EDTA anticoagulated whole blood sample 2. 25µL of this sample is separated in the sampler head while the remaining 75µL is disposed of through the sampling needle into a dedicated washing chamber 6. The WBC cell counting is performed, followed by an HGB measurement 7. Then an RBC and PLT counting is performed 8. The system drains and cleans the measuring chambers and related tubing, and prepares for the next blood sample. Apertures are cleaned with high-voltage burning pulses 3. The 25µL whole blood sample is transported to the mixing chamber through the sampling needle with an addition of 4mL of diluent to make a 1:160 primary dilution 4. 25µL of the diluted sample is aspirated into the sampler head and then mixed with another 4mL of diluent which is dispensed into the RBC chamber forming a 1:25,000 overall RBC dilution 5. The remaining mix dilution is moved to the WBC chamber where 0.9mL of lysing reagent is added. Bubbles are used to mix the lyse and the mix dilution. This creates a dilution of 1:196 which is suitable for photometric measurement and WBC counting Note: Measuring times: WBC count (adaptive) 2 – 6 seconds HGB measurement 2 seconds RBC/PLT count 8 seconds © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. © Siemens rights reserved. Page AG 26 2013. All 2014-07-22 Global Product Education Calibration Prepare and Perform a Calibration: 1. Prior to performing a Calibration, run a Selftest and a Blank measurement: To perform a Self-test navigate to: 1. Home 2. Maintenance 3. Diagnostics 4. Self Test 5. Start 2. Run a Blank measurement (Refer to the section titled Daily Startup in this documentation for instructions) 3. To perform a calibration navigate to: 1. Home 2. Maintenance 3. Calibration 4. Calibration with measurement © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Calibration Page 27 2014-07-22 Maintenance Global Product Education Calibration Prepare and Perform a Calibration: 4. If using a new lot of calibrator, scan the barcode on the assay sheet to upload the lot number, expiration date, and target values for the calibrator 5. Select the number of runs desired: 3 or 7 and select Accept 6. Mix the calibrator material, insert the tube into the sample adaptor, and select Run 1. The system rotates the sample for aspiration and begins measuring (Status light turns red) 2. The system rotates the sample to the original position once aspiration is complete 7. If the results look good, and there are no warning flags, run Calibration with Measurement again until you reach the required number of runs: 3 or 7 8. If a warning flag is displayed (C or c, clogging) select Discard to delete that particular calibration run, then run Calibration with measurement again Note: Mix the calibrator sample thoroughly in between runs © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 28 2014-07-22 Maintenance Global Product Education Calibration Review and Interpret Calibration Results: After performing all the runs, the results display showing statistical evaluation of calibration runs Each calibration factor must be within 1.00 +/0.20 limits If a calibration is outside this range, it is highlighted in red The left part of the screen displays: Target values, Mean Values of Runs, CV%, and the newly calculated calibration factors If the CV% values and the new factors are satisfactory, select Evaluate > Accept If the results are not acceptable, select Cancel. Repeat the calibration process using a new vial of calibration material Page 29 2014-07-22 © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Maintenance Global Product Education Calibration View Calibration History and Database To view the history of calibration events by factors, navigate to: 1. Home 2. Maintenance 3. Calibration 4. History The Calibration History screen shows the calibration factors that were generated by each calibration © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 30 2014-07-22 Maintenance Global Product Education Calibration View Calibration History and Database To view the calibration database, navigate to: 1. Home 2. Maintenance 3. Calibration 4. Database The calibration database screen shows the details of each calibration run © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 31 2014-07-22 Maintenance Global Product Education Quality Control Enter a New Lot of QC Material: 1. Navigate to Home > Quality control > New Lot 2. Scan the QR code located on the Assay sheet shipped with your control materials, to load the Lot information, expiration date, and target and range values 3. Allow time for the prompt “Do you want to install controls…?” to appear 4. Select Yes © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Quality Control Page 32 2014-07-22 Maintenance Global Product Education Quality Control View QC Reference Values: 1. Select Quality control 2. Select > next to the expiration date for that control 3. Select References © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Quality Control Page 33 2014-07-22 Maintenance Global Product Education Quality Control Run a Quality Control Sample 1. To measure QC (all levels), navigate to: 1. Quality control 2. Select the check box in front of the Lot number and level of QC you wish to run 3. Select > next to the expiration date for that control 4. Select Measure 5. Mix the control material, insert the tube into the sample adaptor, and select Run 6. The system rotates the sample for aspiration and begins the QC measurement (Status light turns orange and then red) 7. The system rotates the sample to the original position once aspiration is complete 8. The QC results display and the system beeps when the measurement is complete 9. Select Run to rerun QC, Discard to discard the QC results, and Print to print the results Page 34 2014-07-22 Important: Verify that a valid Blank has been performed and accepted prior to running QC. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Maintenance Global Product Education Quality Control Review and Interpret QC Results Results display in the QC screen (Home > Quality control) Results are divided into three categories: 1. Parameters 2. Normal Ranges 3. Tech Details The Lot number, date and time of run, sample ID and sample type display in the header Results screen – Parameters Values within normal range are white Values outside of the normal range are red and flag as L or H Histograms display graphical information about volume distribution of WBC, RBC, PLT Discriminators are indicated by vertical lines that separate different cell populations © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 35 2014-07-22 Maintenance Global Product Education Quality Control Review and Interpret QC Results Results screen – Normal Ranges Normal ranges display in a graphical format on the right of the screen Each value is indicated by a downwardpointing white arrowhead within the normal range graphics Low and high limits of the normal range display on either side of the range line © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 36 2014-07-22 Maintenance Global Product Education Quality Control Review and Interpret QC Results Results screen – Tech Details Probe voltage: shows electrode voltages for WBC and RBC counts WBC lyse volume: default value = 0.9 Diagnostic flags: shows evaluations made by the system based on results Warning flags display for the measurement © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 37 2014-07-22 Maintenance Global Product Education Quality Control To Display QC Results in Database (Table) View: 1. Select Quality control 2. Select the check box to the left of the Lot file that you wish to view 3. Select > next to the expiration date for that control 4. Select Database at the bottom of the screen 5. Navigate through all the parameters by selecting the left/right arrows at the bottom of the screen © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 38 2014-07-22 Maintenance Global Product Education Quality Control View, Print, or Export Control Lot Data Files 1. Navigate to Home > Quality Control 2. Select the check box to the left of the Lot file that you wish to view 3. Select Monthly View at the bottom of the screen 4. Navigate through the months (mm/yyyy) using the arrows at the top of the screen 5. Select Print or Export for the selected file(s) Important: Selecting Export writes the selected records as an ASCII file (.txt) to the USB storage device if a USB device is present in the USB port, otherwise the system requests that the USB device be inserted. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 39 2014-07-22 Maintenance Global Product Education Quality Control View a Levey-Jennings Chat 1. Navigate to Home > Quality Control 2. Select the check box to the left of the Lot file that you wish to view 3. Select > to the right of the expiration date so that the Levey-Jennings chart displays 4. To navigate through the chart to view the results for the different parameters, select the up and down arrows towards the bottom of the screen 5. To display results for a specific measurement point, select the Left/Right arrows on either side of the View option, to move the white triangle pointer to the specific measurement you wish to view 6. Select View 7. Select Normal ranges and Tech Details to view those results for that particular measurement Important: Any QC that has been measured via the Measure screen will not be included in the statistical data. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 40 2014-07-22 Maintenance Global Product Education Maintenance Daily Maintenance: Perform a Daily Shutdown / Cleaning: 1. To shut down the system select Exit > Shutdown • The software initiates a Cleaning sequence 2. When the system displays a warning message and emits a sustained tone, it is safe to turn off the power 3. Power off the system by pressing the Power switch Perform a Daily Startup: 1. Power on all external devices (Printer) if installed 2. Power on the system by pressing the Power switch (Wait 5 minutes before initiating any measuring processes to allow system to warm up) 3. Follow Startup as per laboratory policy © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 41 2014-07-22 Maintenance Global Product Education Maintenance Weekly Maintenance: Perform a Hard Cleaning Put 1 mL of the ADVIA Hypoclean solution into a sample tube Run the Hard Cleaning function by navigating to: 1. Home 2. Maintenance 3. Cleaning 4. Hard Cleaning Repeat blank and select Accept if in range Check Tubing System Open the side door and check for any leaks Warning: The sharp tip of the piercing needle can cause injury. Always use rubber gloves and a cotton swab to clean the wash head. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 42 2014-07-22 Global Product Education Maintenance Weekly Maintenance: Clean the Needle Wash Head 1. Run the shut-down procedure and power down the system 2. Open the side door 3. Use a moistened cotton swab with deionized water to remove salt and blood residuals from the bottom of the wash head 4. Power on the system by pressing the Power switch 5. Follow Startup as per laboratory policy Warning: The sharp tip of the piercing needle can cause injury. Always use rubber gloves and a cotton swab to clean the wash head. © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 43 2014-07-22 Global Product Education Note: This document is for supplemental use only, and not meant to be used in place of primary technical materials. © 2015 Siemens Healthcare Diagnostics Inc. All rights reserved. This guide and the software described within, are copyrighted. No part of this may be copied, reproduced, translated, or reduced to any electronic medium or machine-readable form without the prior written consent of Siemens Healthcare Diagnostics. ADVIA is a trademark of Siemens Healthcare Diagnostics Inc. or its affiliates. Other trademarks used herein are the property of the respective owners. Global Siemens Headquarters Siemens AG Wittelsbacherplatz 2 80333 Muenchen Germany Global Siemens Healthcare Headquarters Siemens AG Healthcare Sector Henkestrasse 127 91052 Erlangen Germany Global Division Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591-5005 USA www.siemens.com/diagnostics © Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. Page 44 2014-07-22 Global Product Education