The Role of In-Room kV X-Ray Imaging for Patient Setup and Target Localization (TG104) Fang-Fang Yin, PhD Duke University Medical Center ACMP, May 2009 Members of TG 104 Fang-Fang Yin, Co-Chair, Duke University Medical Center John Wong, Co-Chair, John Hopkins University James Balter, University of Michigan Medical Center Stanley Benedict, Virginia Commonwealth University Jean-Pierre Bissonnette, Princess Margaret Hospital Timothy Craig, Princess Margaret Hospital Lei Dong, M.D. Anderson Cancer Center David Jaffray, Princess Margaret Hospital Steve Jiang, Massachusetts General Hospital Siyong Kim, Mayo Clinic, Jacksonville Charlie Ma, Fox Chase Cancer Center Martin Murphy, Virginia Commonwealth University Peter Munro, Varian Medical Systems Timothy Solberg, University of Nebraska Medical Center Q. Jackie Wu, Duke University Medical Center Objectives 1. Understand the current existing kV x-ray systems used in the radiation treatment room, including system configurations, specifications, operation principles, and functionality. 2. Understand the current clinical application methods about how these systems could be used to improve treatment accuracy and their limitations. 3. Understand issues related to effective implementation in the routine clinical procedures. 4. Understand issues related to acceptance testing and quality assurance. Charges of TG 104 a) Review the current existing kV x-ray systems used in the radiation treatment room, including system configurations, specifications, operation principles, and functionality b) Discuss the current clinical application methods about how these systems could be used to improve treatment accuracy and their limitations. c) Discuss issues related to effective implementation in the routine clinical procedures d) Discuss issues related to acceptance testing and quality assurance What is In-room Image-Guidance? Use of imaging method in the treatment room while patient stay at the treatment position • To localize, monitor, and track surrogates which are associated to the patient and are of interest to radiation treatment • To generate a list of choices for decision-making and intervention for positioning and modification • To direct how the treatment couch or radiation beam should be modified Why In-Room Image-Guidance? • To improve the targeting precision and accuracy so that treatment margin from CTV to PTV could be reduced • Challenges: – uncertain about the target location – uncertain about the target shape – uncertain about the target motion – Limitations of tools used for image-guidance Why In-Room Image-Guidance? Accurate but not precise IGRT Precise but not accurate IMRT Precise and accurate IGRT+IMRT Yin et al Sem Rad Onc 2006 History of In-room kV Imaging Early days 50th ~ 60th: • A separate kV x-ray system and Cobalt-60 treatment unit linked through a mobile couch (Karolinska University Hospital); • A kV x-ray source attached to the beam stopper of a Cobalt-60 unit (Holloway 1958) • A customized Cobalt-60 unit (Johns and Cunningham 1959) linear accelerator (Weissbluth,Karzmark et al. 1959) • A Cobalt-60 unit and a kV x-ray tube mounted at 90 degrees from each other on a circular ring (Netherlands Cancer Institute) • A Cobalt-60 unit with an x-ray tube mounted to the collimator at a well defined angle with a graticule for optical and radiographic projections (Shorvon, Robson et al. 1966) History of In-room kV Imaging kV x-ray source Ontario Cancer Institute's X-otron Cobalt-60 unit History of In-room kV Imaging Continuous efforts 80th ~ 90th: • An off-set kV x-ray source to a 10 MV medical accelerator at MGH (Biggs, Goitein et al. 1985) • A cobalt-60 treatment beam and an offset gantry mounted kV beam on the same screen/film system (Shiu, Hogstrom et al. 1987) • An add-on kV source: RADII product by HRL Inc • A test for a low-Z accelerator targets for generating kV or near-kV xrays for imaging (Galbraith 1989; Ostapiak, O'Brien et al. 1998) • A design of a new x-ray target for producing both kV and MV beams (Cho and Munro 2002) History of In-room kV Imaging Modern IGRT 90th ~ • A CT scanner with the medical accelerator in the treatment room (Akanuma, Aoki et al. 1984; Uematsu, Fukui et al. 1996) • A proposed concept to integrate a kVCT/Linac system for the inroom image guided process (Mackie, Holmes et al. 1993) • The dual orthogonal kV imaging systems, the first version of the CyberKnife system (Murphy and Cox 1996) • A pair of wall-mounted x-ray tubes and a novel portable CCDbased imager (Schewe, Lam et al. 1998) • 4 ceiling mounted fluoroscopic imagers and 4 opposing floor mounted kV sources (Shirato, Shimizu et al. 2000) • Ceiling/floor-mounted kV image guidance methods for radiosurgery by BrainLab Inc. in 2001 (Yin, Ryu et al. 2002) History of In-room kV Imaging Modern IGRT 90th ~ • A dual-beam (kV/MV) imaging system mounted to gantry: 37 degrees apart kV and mV (Sephton and Hagekyriakou 1995) • A dual-beam (kV/MV) imaging system mounted to gantry: 45 and then 90 degrees from the MV source with CBCT (Jaffray, Chawla et al. 1995, Jaffray, Drake et al. 1999). – Synergy accelerator by Elekta – On-Board Imager (OBI) by Varian in 2003 – Artiste solution proposed by Siemens in 2007 • A hybrid imaging system: Varian OBI/CBCT system + the BrainLAB floor/ceiling mounted system in 2008 Rail-track Mounted System Varian-GE ExaCT™ system Curtsey of Lei Dong, Ph.D., MD Anderson Cancer Center, TX Rail-Track Mounted System From C Ma Siemens Primatom system Curtsey of Lisa Grimm, Ph.D., Morristown Memorial Hospital, NJ. Ceiling/Floor-Mounted System Novalis system SDD: 3.62 m SID: 2.34 m Pixel: 0.4 mm Matrix: 512x512 Digital Detector kV x-ray tube F-F Yin Med Phy 2002 Ceiling/Floor-Mounted System Cyberknife system X-ray tube X-ray tube Detector Recessed Detector Detectors under the floor Detectors above the floor Curtsey of Accuray, Inc. Gantry-Mounted System Synergy system Gantry-Mounted System On-board Imager (OBI) system Hybrid Image-Guidance System NovalisTx System Duke University Medical Center KV Detector Video/IR Camera OBI KV Detector OBI KV tube MV Detector Recessed ExactTract KV tube Commercially Available Systems for In-Room kV Imaging X-ray Imaging in Proton Treatment Courtesy of Loma Linda University Courtesy of Heidleberg University In-Room kV Image-Guidance Strategy • Off-line – – – – – 2-D planar imaging for surrogate imaging Real-time fluoroscopic imaging for motion Tomographic imaging for volume 4-D tomographic imaging for motion 3-D images for dose calculation • On-line – – – – – 2-D planar imaging for surrogate imaging Real-time fluoroscopic imaging for motion Tomographic imaging for volume 4-D tomographic imaging for motion 3-D images for dose calculation Off-Line In-Room Image-Guidance Patient planning information/ Patient information system Next treatment Patient setup In-room imaging Treatment On-board images Correct position Y Correction? N Reference images On-Line In-room Image-Guidance Patient planning information/ Patient information system Patient setup In-room imaging I Reference images On-board images Correction? Y In-room imaging III In-room imaging II Feedback Correct position N Treatment 2-D/3-D In-room Image-Guidance • On-line systems • Ceiling/floor-mounted system • Gantry-mounted system • On-line techniques • • • • Real time 2-D radiographic imaging Real time 2-D fluoroscopic imaging 2-D to 3-D image fusion Automatic patient/beam positioning 3-D/4-D In-room Image-Guidance • On-line systems • Rail-track-mounted system • Gantry-mounted system • On-line techniques • • • • Real time 4-D radiographic imaging Real time 2-D fluoroscopic imaging 3-D and/or 4-D image fusion Automatic patient/beam positioning Image-Guidance with ExacTrac 6D Robotics Frameless Radiosurgery Adaptive Gating ExacTrac IGRT Workflow Using Snap Verification Initial 6D setup Snap verification for field 2 Snap verification for field 3 Snap verification for field 4 .... Treat field 1 Treat field 2 Treat field 3 Treat field 4 Use Case: Intra-Fraction Imaging Example of dual x-ray imaging Image-Guidance with CT-on-Rails Planning and R&V System Room Reference CT dataset Treatment Planning R&V System Image Storage Reference CT Control Room Imaging console LINAC console Gating Signal Treatment Room Intranet Alignment Protocol Correction? Couch Shifts LINAC Patient couch In-room kV CT/CBCT Images Use Case: IGRT for Liver Tumor CBCTs 2-D Images On-Board Imaging Use Case: IGRT for Brain with CBCT Onboard verification Plan Use Case: 3-D Free-Breath ITV with CBCT CBCT images after correction CBCT images prior to correction Planning CT with target contours Post-treatment CBCT Wang et al Ref J 2007 Overall Localization Accuracy kV OBI vs Laser Pre-tx CBCT vs 2D kV OBI Post-tx OBI vs Pre-tx CBCT Mean Absolute Displacement (cm) AP CC ML AP CC ML AP CC ML 0.32 0.35 0.36 0.26 0.19 0.22 0.08 0.09 0.07 RMS (cm) 0.60 0.39 0.14 123 Treatments of 40 lesions in 33 patients Yoo et al ASTRO 2007 Image-Guided Implementation • Identification of a Suitable Imaging Technology • Design, Implementation and Maintenance – Image Performance and Objectives – Image Acquisition – Analysis Tools • • • • On-line and Off-line Strategies Margins, Accuracy, and Precision Decision-Making and Intervention Quality Assurance Program for Image-Guided Processes • Imaging Dose Considerations • Manpower and Training Image Analysis & Patient Position Tool Daily Bladder as image Bladder in planning CT as contour overlay In-room CT Hardware and software application and verification Bony Structure is off Variable rectal filling observed In-room CT Prostate target is aligned with the CT image Reference CT Acceptance Testing: Imaging System • The primary goal for acceptance testing is to verify the components, the configurations, the functionality, the safety, and the performance of the system relative to the specifications described in the purchasing agreement and/or installation documentation from the vendors • Data generated in the acceptance testing could be used as the baseline for routine QA Acceptance Testing: Imaging System • Room design and shielding consideration • Verification of Imaging System Installation • Safety and Mechanical Configurations • Geometric Calibration • Localization Accuracy • Image Quality Commissioning: Imaging System • Experimentally determine imaging parameters for optimal image quality and localization accuracy for different anatomical sites • Identify potential limitations of the imaging system • Setup and document operation procedures for different localization purposes Quality Assurance Programs • Goals: – To ensure that the imaging systems, including both hardware and software, function safely and reproducibly, and perform as accepted and commissioned • Contents: – Parameters, phantoms and measurement method – Frequency and criteria – Who and documentation Quality Assurance Programs • • • • • Safety and functionality Geometric accuracy Dosimetric information Software and hardware Imaging system with delivery system alignment/coincidence • Image quality • TG 142 sets the frequencies and criteria QA for Gantry-Mounted System 1. MV Localization (0o) of BB; collimator at 0 and 90o 2. Repeat MV localization of BB for gantry angles of 90o, 180o, and 270o 3. Adjustment of BB to treatment isocenter +1mm θg θg u v -1mm -180 θg +180 Reconstruction 4. Measurement of BB location in kV radiographic coordinates (u,v) vs. θg. 5. Analysis of ‘Flex Map’ and storage for future use 6. Use ‘Flex Map’ during routine clinical imaging Image Quality QA for OBI Measurement setup Image for QA analysis CT number check for CBCT QA for OBI/CBCT • Safety and functionality – Door interlock, collision interlock, beam-on sound, beam-on lights, Hand pendant control, and network-flow. – All test items are verified during tube warm-up (< 5 min) • Geometric accuracy – – – – OBI isocenter accuracy Accuracy of performance for 2D2D match and couch shift Mechanical accuracy (arm positioning of KVS and KVD) Isocenter accuracy over gantry rotation • Image quality – OBI (radiography): contrast resolution and spatial resolution – CBCT (tomography): HU reproducibility, contrast resolution, spatial resolution, HU uniformity, spatial linearity, and slice thickness. Calibration for Ceiling/floor-Mounted System (ExacTrac System) Isocenter calibration phantom x-ray calibration phantom Artifacts in kV CBCT • Cupping and streaks due to hardening and scatter (A&B) • • Gas motion streak (C) Rings in reconstructed images due to dead or intermittent pixels (D) • Streak and comets due to lag in the flat panel detector (E) • Distortions (clip external contours and streaks) due to fewer than 180 degrees + fan angle projection angles (F) Crescent Artifact in CBCT Scans An apparent shift of the bow tie profile from projection to projection deriving most likely from minor mechanical instabilities, such as a tilt of the source or a shift of the focal spot Dose/Exposure vs Imaging Modality Murphy et al Med Phys 2007 TG 76 Report Clinical Imaging Dose Measurements A simple and clinical feasible method to estimated the CBCT imaging dose Detectors Detectors Kim et al, Radiat Prot Dosi. 2008 New Development • • • • CBCT with a Mobile CT Dual X-Ray Tubes with Dual Detectors kV and MV Dual-Energy Imaging Digital Tomosynthesis CBCT with a Mobile C-Arm System Challenge: How to correlate imaging coordinate with treatment unit coordinate C-arm system www.Siemens.com In-Line kVision Image-Guidance System Proposed ARtISTE system by Siemens The kV x-ray axis is in parallel and coincident to the treatment beam but at the opposite direction Radiographic Fluoroscopic imaging kV CBCT www.Siemens.com Dual X-Ray Tubes with Dual Detectors Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantrymounted diagnostic x-ray systems with flat-panel detectors Berbeco et al Phys. Med. Biol. 2004 A New Hybrid Image-Guided Radiotherapy System KAMINO et al, IJROBP 2006 Digital Tomosynthesis (DTS) Scan angle Compare to 2-D radiograph: Patient - Reduce overlying structure - 3D information - more dose (20-40 degrees) Compare to 3-D CT or CBCT: - Fast acquisition (5 s ~ 60 s) - Lower radiation exposure - Less collision limitations On-Board H & N DTS Imaging DRR Onboard R-DTS Onboard-DTS R-CT Onboard-CBCT Wu et al IJROBP 2007 On-Board Liver DTS Imaging 10-degree 20-degree 40-degree Fuller et al ASTRO 2007 On-Board Breast DTS Imaging CBCT DTS Coronal Sagittal Oblique Zhang et al AAPM 2007 On-Board 4D-DTS Imaging 30-degree gantry rotation Maurer et al AAPM 2009 kV/MV Dual-Beam CBCT Images kV+MV CBCT Diag CT Yin et al. Med Phys 2005 Zheng et al 2007 MV-CBCT kV-CBCT Summary The introduction of in-room kV imaging provides new opportunities to further improve treatment accuracy and precision. At the same time, it presents new challenges for its efficient and effective implementation. Each in-room kV imaging method has its strengths and limitations. The user is well advised to match the clinical objective with the appropriate technology; or at least to apply the image guidance information to within the bounds of its validity. Summary Implementation of an in-room kV imaging technology requires rigorous characterization and validation of its performance. Quality assurance measures with phantoms are requisite. Expertise must be developed and must be re-established from time to time. One must also be cognizant that in actual clinical practice, inherent uncertainties of the guidance solution exist, as each technique has its own range of uncertainties. Summary There is uncertainty in the strength of the surrogate information as in the case of implanted fiducials; in the integrity of the information with time as in the case of CT guidance; and in the residual error related to the implementation of the correction. In-room kV guidance clearly offers great potential of improving treatment accuracy. The promise of in-room kV guidance can only be realized with a radiation community that applied the technology with discipline. Acknowledgements: Some research works are partially supported by grants from NIH, Varian Medical Systems, and GE Health Care. Thank you for your attention. The basic QA requirements for in-room kV imaging system should include all items listed below except: 0% 9% 86% 0% 6% 1. 2. 3. 4. 5. Geometric accuracy Dosimetric accuracy Output factor Image quality Imaging dose The differences between on-line and off-line corrections are 83% 2. 0% 3. 11% 4. On-line correction takes image(s) for the current treatment session and makes correction for the next treatment session On-line correction takes image(s) for the current treatment session and makes correction for the current treatment session Off-line correction takes image for the current treatment session and makes correction for the current treatment session Off-line correction cannot correct systematic errors 0% 5. Off-line correction can correct both systematic and random 6% 1. The following statements are all correct except 3% 1. 6% 2. 75% 6% 11% 3. 4. 5. The ceiling/floor-mounted kV imaging systems are able to perform fluoroscopic imaging The gantry-mounted kV imaging systems are able to perform 2-D radiographic imaging and cone-beam CT The ceiling/floor-mounted systems can be used to acquire cone-beam CT The gantry-mounted kV imaging systems can be used to acquire digital tomosynthesis images The rail-track-mounted systems can not be used to acquired real-time 2-D images during the radiation delivery