Medical Equipment MASH 203 Ultrasound Equipment (US) The Physics of Ultrasound The Principles of Ultrasound Diagnostic Equipment Dr. Mohammed Fayez Al Rez 1 Ultrasound Equipment - Unit 3 The Physics of Ultrasound The term Ultrasound refers to acoustical waves above the range of human hearing (frequencies higher than 20 kHz). Ultrasound waves are vibrations or disturbances consisting of alternating zones of compression and rarefaction in physical medium such as gas, liquid, or solid matter. The following will be studied: - Frequency - Velocity - Generation - Reflection - Beam pattern - Resolution Dr. Mohammed Fayez Al Rez 2 Ultrasound Equipment - Unit 3 Ultrasound Frequency Humans can hear only a limited rang of frequencies (Audible spectrum). The frequency range of audible sound is approximately 20 Hz to 20 kHz. Frequencies lower than 20 Hz are Infrasound and frequencies higher than 20 kHz are called Ultrasound. Diagnostic ultrasound equipments used ultrasound frequencies in the range of approximately 1 MHz to 15 MHz. Dr. Mohammed Fayez Al Rez 3 Ultrasound Equipment - Unit 3 Velocity of Ultrasound Waves Velocity of ultrasound waves to each material of human body Material Velocity m/sec Air 330 Blood 1570 Brain 1540 Fat 1450 Soft tissue 1540 Kidney 1560 Liver 1550 Muscle 1580 Skull bone 4080 Water 1480 The average velocity in human body is about 1530 m/sec, this average velocity is significant in the study of equipment principles. Dr. Mohammed Fayez Al Rez 4 Ultrasound Equipment - Unit 3 Generation of Ultrasound Waves Phenomena of Piezo-Electric Effect If we apply voltage in pulses the transducer will expand and contract. By applying this voltage in very quick pulses we can vibrate the transducer and produce the ultrasound wave. If any external forces is applied to the transducer a voltage is generated. Dr. Mohammed Fayez Al Rez 5 Ultrasound Equipment - Unit 3 Voltage-Ultrasound Conversion In the practice, high voltage pulses of about 100 V are applied on the transducer. The transducer converts the voltage pulses into the ultrasound waves, which may be transmitted into the human body. Dr. Mohammed Fayez Al Rez 6 Ultrasound Equipment - Unit 3 Ultrasound-Voltage Conversion When the ultrasound waves reach an object they are reflected back to the transducer. The reflection echoes are then converted by the transducer into voltage to be used as reception signals. Dr. Mohammed Fayez Al Rez 7 Ultrasound Equipment - Unit 3 Voltage-Ultrasound Conversion Dr. Mohammed Fayez Al Rez 8 Ultrasound Equipment - Unit 3 Ultrasound Reflection Ultrasound waves transmitted in the human body. At the border of different structures some of the transmitted waves are reflected back according to the acoustic impedance different between two structures. Acoustic impedance: Z=ρ*C o Z: Acoustic impedance o ρ: Density o C: Velocity Dr. Mohammed Fayez Al Rez 9 Ultrasound Equipment - Unit 3 Acoustic Impedance Acoustic impedance of each material in the human body Material Acoustic impedance Air 0.00043 Blood 1.62 Brain 1.6 Fat 1.38 Soft tissue 1.6 Kidney 1.62 Liver 1.66 Muscle 1.7 Skull bone 7.8 Water 1.52 The average impedance in human body is about 1.6. Acoustic impedance of each material is different from each other and that is why US waves are reflected from the different structures. Dr. Mohammed Fayez Al Rez 10 Ultrasound Equipment - Unit 3 Ultrasound Beam Pattern As it shown above, if there is no acoustic lens on the transducer surface, the ultrasound beam is transmitted in a straight line in the near field then begins divert from a certain point in the far field as shown above. This is called an unfocused beam. If there is an acoustic lens on the transducer surface, the ultrasound beam will be focused and convert as a certain point. This is called the focal point. Generally, the focused beam can give a clearer image. Dr. Mohammed Fayez Al Rez 11 Ultrasound Equipment - Unit 3 Resolution There are two major parameters, which influence the diagnostic image quality these are axial and lateral resolution. Resolution can be defined as the ability of the equipment to distinguish two separate objects. Dr. Mohammed Fayez Al Rez 12 Ultrasound Equipment - Unit 3 Axial Resolution Axial resolutions describe resolution in the axial direction. Axial resolution is determined by the US wave length and frequency. If the US beam uses a short wave length as is shown above we can distinguish two separate objects which are based closed together in line in axial direction. This is because wave length of the US beam is shorter than the distance between the two objects. Dr. Mohammed Fayez Al Rez 13 Ultrasound Equipment - Unit 3 Axial Resolution If the US beam uses a longer wave length as shown below, we cannot distinguish two separate objects, this is because the wave length is longer than the distance between the two objects. Because a higher frequency means a shorter wave length, we can say generally that: the higher the US frequency, the higher the axial resolution. Dr. Mohammed Fayez Al Rez 14 Ultrasound Equipment - Unit 3 Lateral Resolution Lateral resolution describes resolution in the lateral direction. If the US beam is wide, as shown in the top figure, the displayed object size may be bigger than the actual object size, so lateral resolution is poor. If the US beam is narrow the displayed object size will be a closer to the actual size, so lateral resolution is better. Generally, the narrow the US beam, the higher the lateral resolution. Dr. Mohammed Fayez Al Rez 15 Ultrasound Equipment - Unit 3 Relationship between Frequency, Resolution and Penetration The table below shows the relationship between frequency, resolution and penetration. In general, the higher the US frequency the higher the resolution and the lower the penetration. Conversely, the lower the U.S frequency the lower the resolution but the higher the penetration. To determine a wide range of diagnostic requirements we need a wide range of US frequencies. That is why there is a wide selection of probe frequency for studies of different organs. Dr. Mohammed Fayez Al Rez 16 Ultrasound Equipment - Unit 3 The Principles of Ultrasound Diagnostic Equipment Ultrasound Equipment US diagnostic equipment uses a pulse reflection system Dr. Mohammed Fayez Al Rez 17 Ultrasound Equipment - Unit 3 The Principles of Ultrasound Diagnostic Equipment Pulse generator: A pulse circuit generates a high voltage pulses, then the T-delay line circuit applies appropriate delay time for each transmitted pulse for electronic focusing of the US beam. When the US probe touches the patient skin surface, US waves are transmitted to the patient. The delayed high voltage pulses are applied to the piezo-electric transducer elements. They produce the US waves, which will entire the patient, and are reflected back by the intern structures. Dr. Mohammed Fayez Al Rez 18 Ultrasound Equipment - Unit 3 The Principles of Ultrasound Diagnostic Equipment The reflected echo signals are received by the same probe. Then they are input into the R-delay line circuit to compensate for the transmission delay factor of the pulses and mix the echo signals. Then the receiver circuit amplifies the mixed echo signal from the R-delay line. Dr. Mohammed Fayez Al Rez 19 Ultrasound Equipment - Unit 3 The Principles of Ultrasound Diagnostic Equipment In new equipment the echo signals are produced using a Digital Scan Converter D.S.C. So that the image data is thought enter the frame menu and read sequentially out in a format suitable for display on TV-monitor. The result is real time imaging. Dr. Mohammed Fayez Al Rez 20 Ultrasound Equipment - Unit 3 Absorption The absorption of MHz US waves in each material Material Absorption at 1 MHz dB/cm Air 12 Blood 0.2 Brain 0.9 Fat 0.6 Soft tissue 0.8 Kidney 1 Liver 0.9 Muscle 2.3 Skull bone 13 Water 0.002 Absorption is measured in decibels per centimeters (dB/cm). The absorption in air and skull bone is very high. If absorption is very high, the US waves cannot pass through. As a general rule US diagnosis is not restructures containing gas or structures hidden behind bon such as lung, stomach and brain. Dr. Mohammed Fayez Al Rez 21 Ultrasound Equipment - Unit 3 Relationship between Depth and Time The average velocity of US in the human body is about 1530 m/sec. At this velocity an US wave will take about 6.5 µsec to travel from the probe transducer to a target 1 cm below the skin as shown. An additional 6.5 µsec is required for the US wave to return to the probe transducer. As a result the total time taken for the US wave to be transmitted and received to and from a target, which is 1 cm from the transducer, is about 13 µsec as shown. 1sec :1530 m X :1cm X Dr. Mohammed Fayez Al Rez 1 6.5 sec 1530 100 cm 22 Ultrasound Equipment - Unit 3 Relationship between Depth and Time The US transmission timing and reception period are determined by rate pulse. The rate pulse allows a reception period of up to 260 µsec. This 260 µsec translate to distance of about 20 cm depth in the human body. Generally, the maximum scanning depth is limited to this 20 cm. Dr. Mohammed Fayez Al Rez 23 Ultrasound Equipment - Unit 3 Display Modes for US Imaging The most common display modes for US imaging are: A-Mode. B-Mode. M-Mode. Dr. Mohammed Fayez Al Rez 24 Ultrasound Equipment - Unit 3 Principles of A-Mode A-Mode display gives information about the distance between tissue boundaries e.g. to measure organ thickness. If we observe an iconoscopes display, let the vertical direction be the time or depth axis and let the horizontal direction be the amplitude of the received signal wave. This display method is called Amplitude Mode or just A-Mode. Dr. Mohammed Fayez Al Rez 25 Ultrasound Equipment - Unit 3 Principles of B-Mode The B of B-Mode stands for brightness. Shifting the position of US transmission and reception results in scanning. Echo signals received and then go to intensity modulation so that across sectional image can be displayed on CRT. Using this display method we can see internal structure in a patient. Dr. Mohammed Fayez Al Rez 26 Ultrasound Equipment - Unit 3 Principles of M-Mode The M of M-Mode stands for motion. This display mode is useful for studying moving targets such as inertial mitral valve. The transducer position should remain fixed when using this mode, while the display mode used customarily screened in a movement code scrolling. Scrolling allows movement of the valve to be displayed as an image. Dr. Mohammed Fayez Al Rez 27 Ultrasound Equipment - Unit 3 Principles of M-Mode Dr. Mohammed Fayez Al Rez 28 Ultrasound Equipment - Unit 3 A, B and M-Mode Dr. Mohammed Fayez Al Rez 29 Ultrasound Equipment - Unit 3 Principles of Electronic Focus In actual practice a block of transmission elements are driven for transmission each US wave. That the transducer elements are not driven at the same time, the first US pulse will be driven from transducers 1 and n then transducers 2 and n-1 and so on. Dr. Mohammed Fayez Al Rez 30 Ultrasound Equipment - Unit 3 Principles of Electronic Focus This sequential driving causes the US beam to convert at a focal point and that is called Electronic Focusing. The distance of the focal point from the probe is determined by the delay time between transmission pulses. Dr. Mohammed Fayez Al Rez 31 Ultrasound Equipment - Unit 3 T & R - Delay Line For transmission a pulses input to the transmission delay line. The transmission delay line generates a delay time between pulses to achieve a desired focusing distance. For reception the echo signals are input to the reception delay line, where they are a phase corrected and mixed together. Dr. Mohammed Fayez Al Rez 32 Ultrasound Equipment - Unit 3 B-Mode Scanning Method The two most common scanning modes are: Linear scanning Sector scanning Dr. Mohammed Fayez Al Rez 33 Ultrasound Equipment - Unit 3 Principles of Linear Scanning In practice many tiny transducers called transducer elements arrayed in straight line. The US scanning beam is produced by simultaneously driving a block of several transducer elements. Dr. Mohammed Fayez Al Rez 34 Ultrasound Equipment - Unit 3 Principles of Linear Scanning The block is new to across the links of the elements array by electronic switching. This effect moving the scanning beam across the links of the transducer. One scanning line of the US image is produced from the information gathered between each switching interval of the US beam. Dr. Mohammed Fayez Al Rez 35 Ultrasound Equipment - Unit 3 Linear Scanning On the left hand side is the probe and the shape of the beam. On the right hand side is the display format of linear scanning. The advantages of linear scanning are: o A large field of view at shallow regions of the body, and o Easier identification of skin orientation. Dr. Mohammed Fayez Al Rez 36 Ultrasound Equipment - Unit 3 Principles of Sector Scanning For sector scanning all the transducer elements are used for transmission and reception. Dr. Mohammed Fayez Al Rez 37 Ultrasound Equipment - Unit 3 Principles of Sector Scanning The angle of the US beam is changed by the delay time between transmission pulses as shown. Dr. Mohammed Fayez Al Rez 38 Ultrasound Equipment - Unit 3 Sector Scanning On the left hand side is the probe and the shape of the beam. On the right hand side is the display format of sector scanning. The advantages of sector scanning are: o A large field of view at deep region of the body, and o A small skin contact area. Dr. Mohammed Fayez Al Rez 39 Ultrasound Equipment - Unit 3 Sector Scanning Therefore the sector probe can transmit US waves to a very small acoustic window such as between the ribs to exam the heart as shown. Dr. Mohammed Fayez Al Rez 40 Ultrasound Equipment - Unit 3 Sensitivity Time Control (STC) The transmitted US waves attenuated gradually through absorption as they penetrate the human body. This means that echoes reflected from near the skin surface will be stronger than echoes reflected from deep regions of the human body. The STC or sensitivity time control can be used to compensate for this difference in echo signals strength from shallow to deep regions as shown. As echo strength decreases over distance, gain of the signal for different depth can be increased independently, so that the echoes all have the same strength. Dr. Mohammed Fayez Al Rez 41 Ultrasound Equipment - Unit 3 Sensitivity Time Control (STC) Dr. Mohammed Fayez Al Rez 42 Ultrasound Equipment - Unit 3 Types of Ultrasound Probes Linear Array: o Frequency range: 5 MHz to 13 MHz o Length: 30 to 60 mm o Applications: Orthopedics, Breast, Musculoskeletal Convex array: o Frequency range: 3.5 MHz to 9 MHz o Length: 20 to 60 mm o Applications: Abdomen, GYN, OB, Urology Micro-convex: o Application: Pediatrics Dr. Mohammed Fayez Al Rez 43 Ultrasound Equipment - Unit 3 Types of Ultrasound Probes Phased array: o Frequency range: 2 MHz to 8 MHz o Angle: 60 to 90 degree o Applications: cardiology (heart) Multi-Plane TEE-Transducer: o Frequency range: 4.0 to 8.0 MHz o Image plane rotation: 0 and 180 deg Dr. Mohammed Fayez Al Rez 44 Ultrasound Equipment - Unit 3 Types of Ultrasound Probes Fingertip / Intraoperative: o Frequency range: 5 MHz to 10 MHz Endocavity o Frequency range: 5 MHz to 10 MHz o Application: Endovaginal, Endorectal Endoscopic / Mini-probes o Frequency range: 5 MHz to 20 MHz Dr. Mohammed Fayez Al Rez 45 Ultrasound Equipment - Unit 3