To determine the presence of PBDEs in local fishes using Gas

advertisement

DETERMINATION OF PBDEs IN CLEAR CREEK MINNOWS

Corliss Harris, Megan O’Connell, Carrie Seltzer, and Monica Silver

CHEM 331

Abstract

Polybrominated diphenyl ethers (PBDEs) are common components of flame retardants used on computer and TV housing, furniture, textiles, carpets, and vehicles. While the current level of PBDEs found in the environment is relatively low, bioaccumulation can concentrate PBDE levels in the fat of animals higher on the food chain. In animals studies, PBDE exposure has been shown to have detrimental effects as an endocrine disrupter. Minnows from

Clear Creek were analyzed for the presence of

PBDEs using Gas Chromatography/ Mass

Spectrometry (GC/MS). Extremely low levels of PBDEs were found in the sample tissue.

Introduction

• Goal: To determine the presence of

PBDEs in local fishes using Gas

Chromatography/ Mass Spectrometry.

• This method allows for the determination of 2,2 ’,4,4’-tetra PBDE only.

• Due to the very low concentrations of

PBDEs found in tissues, the Gas

Chromatography/ Mass Spectrometry method is a suitable choice for detection.

Polybrominated Diphenyl Ether

Experimental Method

Adapted from Covaci, A., et. al

Catch fish using minnow traps

Homogenize fish

Run Soxhlet extraction on 1 g homogenized sample

Perform Solid

Phase Extraction

Gas

Chromatography/Mass

Spectrometry

Reference

Covaci, A., J. de Boer, J.J. Ryan, S. Voorspoelf, P. Schepens.

2000. Determination of Polybrominated Diphenyl Ethers and

Polychlorinated Biphenyls in Human Adipose Tissue by Large-

Volume Injection-Narrow-Bore Capillary Gas Chromatography/

Electron Impact Low-Resolution Mass Spectrometry. Analytical

Chemistry

Soxhlet Extraction of Lipids

Used to extract solute from homogenized sample using boiling solvent.

In Soxhlet thimble:

• 1 gram homogenized sample

• 6 grams anhydrous Na

2

SO

4

Solvent:

• 175 mL of 3:1:1 solution of hexane: acetone: dichloromethane

Run time: 2 hours

Solid Phase Extraction

Column wetting: 20 mL hexane

Column one: 6.0 g salicic acid

Column two: 2.0 g acid silica

1.0 g neutral silica

2.0 g deactivated basic alumina

Rinse: 20 mL 1 hexane: 1 dichloromethane rinse

Collected sample evaporated to near dryness using Rotovapor R3000.

Gas Chromatography/ Mass

Spectrometry

• Machine used: HP 5890/5972 GC/MS

• Column: 30m x 0.25 mm x 0.25 m m DB-5

• Run time: 13.20 minutes

• Level one: 4.00 C/min

• Temperature: 504 K

• Level two: 25.00 C/min

• Temperature: 554 K

Standard Solutions for Calibration

Standard solutions prepared with dichloromethane

• 1.0 ppm

• 2.5 ppm

• 5.0 ppm

• 10.0 ppm

1.2 mL of 50 ppm 2,2’,4,4’tetrabromodiphenylether from

Cambridge Isotope Laboratories

Results

2,2’,4,4’-PBDE concentration

• A small peak with an area of 19 appeared the same time (9.313 minutes) as the 10 ppm standard.

• The concentration of PBDE in the tissue could not be determined from the standard curve because the area fell below the peak area for calibration curve.

• The correlation coefficient of the standard calibration curve was 0.989.

• Small amount of PDBEs was found in sample tissue.

Calibration Curve

6000

4000

2000

0

0 y = 416.89x - 254.37

R

2

= 0.9893

2 4 6

PBDE Concentration (ppm)

8 10

Conclusions

• Overall results were inconclusive.

• Although the peak area was outside the calibration curve, the presence of a small peak at the same retention time as the standard provides cause for further investigation and analysis with more time and resources.

• Standard solutions, as prepared, were insufficient for producing a calibration curve that fit the sample concentration.

Acknowledgements

Earlham College Chemistry Department, Joseph Moore

Museum, Dr. Corinne Diebel, Dr. Michael Diebel, Mary

Hagerman, and Kalani Seu.

Download