UFTI : The 0.8-2.5m Fast Track Imager for the UK Infrared Telescope Patrick F. Rochea*, Philip W. Lucasa,b, Craig D. Mackayc, Eli Atad-Ettedguid, Peter R. Hastingsd, Alan Bridgerd, Nick P. Reese, Sandy K. Leggette, Chris J. Davise, Alan R. Holmesa, AT. Handforda a Physics Department., Oxford University,UK; bSchool of Physics, University of Hertfordshire, UK; c Institute of Astronomy, Cambridge University,UK; d UK Astronomy Technology Centre, Edinburgh; e Joint Astronomy Centre, Hilo, Hawaii 96720, USA. ABSTRACT In 1996, it was proposed to build a near-infrared imager for the 3.8-m UK Infrared Telescope in Hawaii, to exploit the 1024 pixel format detectors that were then becoming available. In order to achieve a fast delivery, the instrument was kept simple and existing designs were reused or modified where possible. UFTI was delivered within 2.5 years of the project start. The instrument is based around a 1k Rockwell Hawaii detector and a LSR Astrocam controller and uses the new Mauna Kea optimised J,H,K filter set along with I and Z broad-band filters and several narrow-band line filters. The instrument is cooled by a CTI cry-cooler, while the mechanisms are operated by cold, internal, Bergelahr stepping motors. On UKIRT it can be coupled to a Fabry-Perot etalon for tunable narrow-band imaging at K, or a waveplate for imaging polarimetry through 1-2.5m; the cold analyser is a Barium Borate Wollaston prism. UFTI was designed to take full advantage of the good image quality delivered by UKIRT on conclusion of the upgrades programme, and has a fine scale of 0.09 arcsec/pixel. It is used within the UKIRT observatory environment and was the first instrument integrated into ORAC, the Observatory Reduction and Acquisition Control System. Results obtained during instrument characterisation in the lab and over the last 3 years on UKIRT are presented, along with performance figures. UFTI has now been used on UKIRT for several hundred nights, and aspects of instrument performance are discussed. Keywords: Infrared Imager, Infrared Detector, Camera, Photometry, Polarimetry, Prism 1. INTRODUCTION The 3.8-m UK Infrared Telescope was originally conceived as a low-cost flux collector, but over the last 20 years, it has been substantially upgraded with the aim of delivering images limited only by the atmospheric conditions above Mauna Kea. A coordinated upgrades programme has installed an active primary mirror support system, an active secondary mirror with fast tip/tilt/focus control, thermal conditioning of the telescope environment and improved dome ventilation (Hawarden et al 1998). By 1996, it was clear that the continuing improvements in image quality and the availability of larger format infrared detector arrays than those employed in the IRCAM instruments, required a new near-infrared imager to exploit the scientific capabilities offered by UKIRT. At that time, the UKIRT instrumentation programme included Michelle, a mid-IR imager spectrometer (Glasse et al 1997) and UIST, a 1-5m imager/spectrometer (Ramsay Howatt et al 2000), scheduled for delivery in 1998 and 2001 respectively, but there was considerable concern that UKIRT’s scientific capabilities would fall substantially behind those of other observatories if we had to wait for UIST to be delivered for an instrument with a 1024 pixel format detector. UFTI was conceived as a fast-track near-infrared imager to fill this gap, and maintain UKIRT’s 1-2.5m imaging capability at a world competitive level. The instrument requirements were for a high-sensitivity imaging capability that fully exploited the best UKIRT imaging performance, with an aim to deliver the instrument within 2 years. UFTI was built as a fully integrated facility instrument and was the first of the UKIRT instruments to be incorporated in ORAC, the new Observatory Reduction and Control environment (Bridger et al 2000). *p.roche@physics.ox.ac.uk http://www-astro.physics.ox.ac.uk/~pwl/UFTI.html Tel: 44 1865 273338 Astrophysics, Denys Wilkinson Building, Keble Road, Oxford OX1 3RH. Fax: 44 865 273390 2. INSTRUMENT REQUIREMENTS On completion of the UKIRT upgrades programme, the delivered image quality at 2m under the best conditions (good atmospheric seeing and low wind) was anticipated to be ~0.25 arcsec. One of the prime requirements for UFTI was to exploit this, which naturally led to a fine pixel scale. A scale of 0.09 arcsec/pixel, which would sample even the best images well, was adopted. Because UKIRT currently operates at a f/36 focal ratio, with a focal plane scale of 1.52 arcsec/mm, a coarser scale would lead to a large entrance window, potentially leading to significant problems with condensation in the central regions, where radiative cooling is most effective. Further, a 90 arcsec field fully covers the clear apertures of the UKIRT waveplates and Fabry-Perot etalons. Cold, interchangeable camera lenses that would give a choice of pixel scales were considered, but were not pursued because of the potential impact on the delivery schedule. The instrument had to fit on the UKIRT instrument support unit (ISU) between two other large UKIRT instruments, CGS4 and Michelle, leading to a relatively narrow format, which determined that the UFTI mechanical layout would be relatively tall and thin. The space between the ISU and the front of the instrument was maximized to allow an ambient temperature Fabry-Perot etalon for narrow-band imaging, or a mask for polarimetry to be inserted. This led to the entrance window being located quite close to the telescope focal plane. Figure 1: Left: UFTI optical layout. Right: JHK spot diagrams at the centre, edges and corners of the field. The circles represent the diameter of the Airy ring at 2m. 2.1 Optical Design The optical design follows conventional IR camera principles. UFTI is mounted on the UKIRT ISU, and at zenith, is fed horizontally by a dichroic mirror, which transmits visible light to the guide camera and reflects infrared light to the instrument. The telescope focal plane lies inside the instrument cryostat, with a field baffle to limit the field of view. An off-axis gold-coated spherical collimator mirror produces an image of the secondary mirror at the position of a cold stop. The collimator was manufactured by diamond machining of a T6061 heat-treated Al-alloy blank by Symons mirror technology of Hertfordshire, UK. 10-position filter wheels are located on either side of the cold stop, allowing broad- or narrow-band filters to be selected. The second wheel also contains a Wollaston prism for polarimetric observations and a Hartmann mask for precise focussing. A lens barrel containing the BaF2/LiF camera lenses forms the image on a Rockwell 1024 x 1024 pixel Hawaii detector array. The optical design was developed to be nearly afocal from 0.9 to 2.5m so that a focusing mechanism would not be required, thus simplifying the mechanical layout and reducing complexity. The IR-grade fused silica entrance windows, together with the spherical lenses, which were supplied mounted in a barrel after cryogenic testing, were manufactured by Graseby-Specac, Orpington, UK. Optimization and tolerancing of the optical design was conducted at the UKATC, Edinburgh using Code-V and Zemax. Supporting structures for optical components and surfaces of lens barrels etc. are black anodized, while the inside of the radiation shield and the collimator support are coated with 3M Nextel black paint to reduce stray and scattered light. The final design gives a scale of 0.0908 arcsec per 18.5m pixel, and a total field of view of 92 x 92 arcsec. The pupil diameter at the cold stop is 15mm, allowing 25mm filters to be used. Over- or under-sized cold stops can be installed between the filter wheel modules, and UFTI currently operates with a slightly undersized stop. The image quality is near diffraction limited down to 1m, with the worst quality at the top of the detector where the off-axis angles are largest (Fig. 1). The total chromatic focal shift between 0.8 and 2.5m is 1.0mm, and with the focus set for = 2m does not contribute significantly to the degradation of image quality. 2.2 Mechanical Layout The constraints of the space available for UFTI, together with packaging to minimize flexure, lead to a folded instrument layout. All of the components and mechanisms are mounted on an optical bench made of Al-alloy, and positioned against a rail mounted on the bench. The bench itself is mounted off the cryostat body on four pairs of G10 Tufnol struts, and covered by an Al-alloy radiation shield. The main cryostat body is an aluminium alloy casting which was subjected to high-pressure isostatic heating to reduce bubbles and flaws. A welded stainless steel lid is attached via quick-release clips, with 4 bolts as a back-up in case of pressure loss. In common with other facility instruments, UFTI is cooled by a closed-cycle refrigerator, which mounts onto the cryostat casting via the anti-vibration mount developed at the UKATC and used on other UKIRT instruments. The cryostat casting is mounted onto the telescope mirror cell via an electrically isolated steel support. The overall dimensions of the cryostat are 40cm wide, 71cm high and 80cm deep, with a total weight, including a stiff telescope mounting frame, of ~150Kg. Figure 2: Left: Optical bench top surface showing the location of the steering mirror, filter modules, shutter and camera lens mount and detector box fitted against the locating rail. Right: Side view of the instrument assembly. Scheduled maintenance, e.g. for filter changes, of the instrument should only require the instrument to be warmed up, removed from the telescope and then opened up by removing the 4 bolts holding the cryostat lid and undoing the clips. Removal of the radiation shield then gives access to the detector housing and all of the optical components, with the exception of the collimator mirror. The only mechanisms provided in UFTI are the two filter wheels and a shutter. The filter wheels are based closely on the proven CGS4 filter wheel designs (Mountain et al 1990), but scaled down to fit the available space. They are driven by cold, internal Bergelahr motors, (with bearings replaced and coated with dry lubricant), via Vespel worm gears. Positional information is provided by a single datum micro-switch. They can each accommodate up to ten 25mm diameter filters, but in practice, one position has to be kept clear. The filter wheels are constructed as modules, which can be removed and re-installed without disturbing the rest of the instrument. A shutter is provided to allow short exposures to be made (shorter than the detector read time), or to prevent over-illumination of the detector array by bright stars. It consists of a sector blade located just beyond the second filter wheel and is rotated by a stepper motor. In practice, this is rarely used, but was included because the detector properties were not fully understood when the instrument was designed. The collimator mirror is located at the end of a turned tube protruding from the back of the instrument, with its mounting face at an angle of 4 degrees to the incoming beam. The beam is reflected back and up to a fixed steering mirror, which passes the beam vertically to the cold stop, camera lenses and detector. The cryostat body was machined and the stainless steel lid fabricated in the Oxford Physics department workshop along with most of the internal components, but some parts were manufactured by local companies to minimize delays to the schedule. Figure 3. The completed instrument. Left: with the cryostat lid and radiation shield removed showing the detector housing, camera lens barrel, filter wheel modules and beam steering mirror mounted on the optical bench, from which locating rods for the radiation shield protrude. Right: the instrument mounted on the UKIRT mirror cell, viewed from the rear. 2.3 Cryogenic Performance UFTI is cooled by a CTI 350 2-stage closed-cycle refrigerator, but only the first stage is used. The first stage cooler is connected to the optical bench by copper braids and also directly to the detector enclosure via a relatively weak thermal link to limit the cooling rate. Under normal operation at the telescope, the optical bench and components stabilize at about 70K, while the detector temperature is actively controlled at 80K by a Lakeshore 330 feeding current through a pair of wire-wound resistors mounted on the detector block. A liquid nitrogen pre-cool tank is included underneath the optical bench; this is not used at UKIRT, but was used in the instrumentation laboratory at Oxford, where the combination of shorter runs of helium gas pipes, warmer ambient temperature and 50Hz cold head operation significantly increased the operating temperature. Lakeshore DT 470 silicon diodes provide temperature monitoring of the detector housing, optical components and the cold head. The heat load onto the optical bench is reduced by multilayer insulation covering the interior cryostat walls. We calculated that the middle of the 120mm diameter cryostat window would cool significantly, due to radiative cooling into the cryostat and relatively poor thermal conduction in the fused silica, leading to condensation under conditions of high humidity. UKIRT does not have a dry air or nitrogen supply available, but condensation can be counteracted to some degree by blowing air across the window; under very high humidity conditions, heated air is required. In practice, conditions sufficiently humid to lead to condensation are unlikely to be very good for astronomical observations. For a normal cycle UFTI takes about 30 hours to warm up and 40 hours to cool down. To prevent potential problems due to differential contraction, the shutter and filter wheel mechanisms are not moved until the temperatures are below 100K. 3 DETECTOR AND CONTROL SYSTEM The heart of UFTI is the Rockwell Hawaii 1024 x 1024 pixel HgCdTe detector array, controlled by a modified LSRAstrocam (now PerkinElmer Life Science) 4100 CCD controller. To minimize development time and effort, the detector control electronics were copied directly from the system developed at the Institute of Astronomy, Cambridge for CIRSI, the University of Cambridge J and H band imager (Mackay et al 2000). The detector is mounted on a CIRSI custom printed circuit board, in which the rear of the detector is spring loaded against a copper block, which houses the temperature control diode and heating resistors, to ensure good thermal contact. The detector package is then mounted inside a die cast box, painted Nextel black on the inside, which itself is mounted onto a support structure attached via the rail to the optical bench. The detector bias, clock and data wires are connected to the pcb via a sub-miniature microD connector and emerge from the die cast box as a single bundle. A W.W. Fischer bulkhead connector on the optical bench allows the signals to pass through to 3 hermetically sealed Amphenol connectors on a plate on the cryostat wall to the detector interface box. Here, pre-amplification and signal conditioning electronics and relays to switch the analogue outputs from the four quadrants of the detector are located, (again identical to the CIRSI electronics) before the signals are passed on to a 14-bit 4100 CCD controller. Initial laboratory testing was conducted using the proprietary Astrocam Pixcel software package to control the instrument, but operation at the telescope uses software developed at JAC, communicating directly with the CCD controller. The Capella 4100 controller runs with custom microcode, written at Cambridge, and stored in sequential access memory chips, to provide the readout clock sequences needed for the HgCdTe detector. The detector controller is mounted in an electronics cabinet under the mirror cell while, the interface box mounted on the cryostat passes the analogue signals from the four detector quadrants to the CCD controller data lines, and the clocks and biases to the detector. More details of the detector control system can be found in Mackay et al (2000). The gain of the detector electronics system is set at ~6 electrons per data number. The small pixels on the sky mean that photon rates with UFTI are relatively low and so even broad-band observations need significant exposures to ensure background-limited data (a few seconds at K and up to 500 sec at I). For short wavelength or narrow band observations (e.g. with the F-P), a highgain mode can be selected which has ~3e- per data number. All observations are made in NDSTARE mode, which is a read-reset-read mode and subtracts the fixed-pattern structure. There are two forms of non-destructive (NDSTARE) readout available. Both do a number of resets to counteract latency on the array. The counts measured are fitted by a slope; the output image gives the increase in counts for the requested exposure time, normalised to one coadd if more than one coadd is taken. The two modes differ, however, in the number of times the array is read out (and consequently the overheads associated with each exposure). In the standard mode, ND_STARE, the array is read out only twice; 10_NDSTARE is read out 10 times, which results in a factor of ~2 improvement in read noise down to about 10e -. 10_NDSTARE may therefore be useful when observing with the FP and sometimes with narrow-band filters, where background signal may be low and so background-limited performance difficult to achieve. The additional overheads associated with 10_NDSTARE are substantial (~25 sec). Sub-arrays may be used for observations of compact sources with a corresponding reduction in read-out overhead. The detector in UFTI has proven to be very easy to use with excellent performance. The NDSTARE read noise is measured at the beginning of every UFTI night via a routine ORAC "array tests" observation, and consistently gives 24- 26e- for a double-correlated read; this can be reduced to ~10e- with multiple reads. The dark current is ~1e- per pixel per minute. For data reduction purposes, standard bad-pixel masks are generated by UKIRT staff, and updated periodically. These typically flag about 0.3% of the pixels as ‘bad’. Over periods of weeks, a number of pixels migrate across the thresholds set, and so the optimum bad-pixel mask changes slowly over time. 3.1 Detector Performance The array begins to saturate at about 10,000 DN and goes into hard saturation at ~14,000 DN, indicating a well depth of ~85000 e-. Linearity corrections have been derived and should be applied for accurate photometry (Fig. 4). At 70% full well, the corrections are <4% and more details are available on the UFTI web pages. Residual images of bright stars are present at a level of 0.30.4% at JHK, even with multiple array resets before each image. This image persistence acts like an enhanced dark current from the highly illuminated pixels, so that it continues to accumulate with long exposures. For example, after exposing standard stars to ~4000 DN, even three frames later, with 80 second blanked off exposures, a latent image of the star was visible ~2 counts above the background at the standard star jitter positions. This persistence makes acquisition of truly uniform flat fields problematic unless care Figure 4. Linearity correction of the UFTI detector. is taken. However, the UFTI flatfield is very stable, and it is possible to take sky flats with a clean array at the start of the night and use them to generate a flat field throughout the night (as opposed to creating flats from median filtered jitter sets) for crowded fields. It is good practice to position any bright objects (e.g. standard stars) in a different part of the array from faint targets, and to keep count levels below ~5000 if possible. Observing sequences are constructed so that the detector is blanked off between objects, so that it is not illuminated during target acquisition UFTI DetectorResponse UFTI and CCD Response 1200 0.5 1000 0.4 800 0.3 600 0.2 400 0.1 200 0 400 500 600 700 Wavelength (nm) 800 900 0 400 500 600 700 800 900 Wavelength (nm) Figure 5. Left: the raw UFTI and ST-6 CCD responses measured with a monochromator. Right: the estimated UFTI detector response between 0.5 and 0.9m obtained by adopting the CCD response curve from the Texas Instruments TC241 data sheet. We wished to include filters in the 0.8-1.0m region, and so investigated the instrument response at short wavelengths in the lab at Oxford using a grating monochromator and tungsten and hydrogen lamps. We did not have access to a flux-calibrated system, but were able to use an ST-6 CCD as a comparison. The detector response appeared to fall smoothly with decreasing wavelength from 1 to 0.8m, but then dropped rapidly near 0.78m. However, the device retains low level (<10% of the sensitivity at 0.9m) residual sensitivity down to wavelengths below 0.5m, with the H line faintly detectable in a hydrogen discharge lamp. Comparison with the measurements made with the CCD suggests that there is a very sharp cut-off in the detector response at 786nm, falling rapidly between 786 and 780nm, and then declining slowly towards 500nm. The tail of residual sensitivity may be due to photons being absorbed in the silicon multiplexor. These measurements indicate that filters used with the Hawaii detectors should block light down to at least 0.4m as well as out to 2.7m or beyond. The 0.78m cut-off is conveniently placed, just longwards of a substantial atmospheric absorption band. 4 MECHANISM AND INSTRUMENT CONTROL SYSTEM UFTI is fully integrated into the UKIRT observatory control system, allowing the instrument observing configuration and telescope positioning sequences to be preset using the UKIRT Observing Tool for automatic execution. Similarly, the data collected are pipeline-processed with standard data reduction recipes for on-line assessment under ORAC. The instrument operates under EPICS (Dalesio, et al 1991) control. The 5-phase stepper motors are powered by a Bergelahr 3-axis MD 5-52 drive unit, which is controlled by a custom electronics rack based upon intelligent motor controllers (RS 440-098). Memory chips on the motor controller boards store control routines for motor datum and filter selection operations, and accept commands and return status information to the EPICS system. Acceleration and deceleration rates, and step speeds and directions for the motors are set using the RS programming language on-board the controller cards. The motor control system is based upon that developed for WHIRCAM, an infrared camera developed at Oxford for the 4.2-m William Herschel Telescope on La Palma. Control of the UFTI detector system is based upon that for the UKIRT auto-guider system, which also uses an Astrocam CCD controller. Much of the software was provided by the vendor in source code form and ported to the VxWorks operating system as part of the UKIRT upgrades project. This was then interfaced to UKIRT's EPICS based control system (which, in turn, interfaces to ORAC), with engineering user interfaces developed using EPICS display tools. The physical connections between the controller and the host computer utilise both a transputer link and a 16 bit parallel link for the video data. The VxWorks transputer connection was implemented using an INMOS B016 board, with driver software from the ESO VLT project. The VxWorks parallel link utilised the same type of interface board as used in the PCI based Astrocam systems, but modified for the PMC form factor by the original supplier. 5. FILTERS AND TRANSFORMATIONS UFTI is equipped with a selection of narrow- and broad-band filters. The broad-band J,H,K filters are from the Mauna Kea Observatory consortium batch supplied by OCLI (MKO-NIR), which are designed to match the atmospheric windows for optimal signal-to-noise (Simons & Tokunaga 2002). A K’ filter is also available, but is most useful for high-emissivity conditions, for example when using warm waveplates and masks for polarimetry. Custom I and Z broad-band filters from Barr Associates are also installed. These filters are significantly different from those installed in optical CCD systems, as they need to be blocked out to 2.7m. The I-band filter bandpass is truncated at the blue end by the detector cut-off at 0.78m and the reflective cut-off of the dichroic. The MKO-NIR J-band filter, and to a lesser extent the H-band filter, is significantly different from the old J and H filters that were used in IRCAM prior to 1999. A number of narrow-band filters from the MKO-NIR consortium and elsewhere are also included. Published magnitudes for near-IR standard stars generally have not used the MKO-NIR filters and so are no longer on the UFTI natural system, but a programme to re-measure the UKIRT standards through the MKO filter set is underway. Colour transformations have been derived in two ways: one empirically based on UFTI photometry and the other calculated by convolving the known filter profiles with spectroscopic data for a representative set of red stars. The two determinations agree well. The transformations between the old IRCAM3 system and the new MKO-NIR system at H and K are well behaved and single-valued. However, for the J filter different terms have to be applied depending whether or not the standard star has intrinsic water absorption features. This is due to the fact that the new J filter cuts off shorter than the old filter, specifically to avoid water absorption in the terrestrial atmosphere. For stars with no intrinsic water features the colour transformations are: KMKO = KUKIRT - 0.020 [+/-0.005] (J-K) UKIRT (J-H) MKO = 0.960 [+/-0.010] (J-H) UKIRT (H-K) MKO = 1.205 [+/-0.010] (H-K)UKIRT (J-K) MKO = 1.040 [+/-0.010] (J-K) UKIRT However, for stars with significant water absorption, stars with spectral type M4 through to the L class (but not including the T class with methane absorption): KMKO = KUKIRT - 0.020 [+/-0.005] (J-K) UKIRT (J-H) MKO = 0.870 [+/-0.010] (J-H) UKIRT (H-K) MKO = 1.205 [+/-0.010] (H-K) UKIRT (J-K) MKO = 0.980 [+/-0.010] (J-K) UKIRT The UFTI I- and Z-band filters are at significantly longer wavelengths than most optical CCD I and Z filters (with 50% transmission at 0.78 and 0.92m and 0.85 and 1.05m respectively). The I-band calibration has been compared to Landolt Cousins-I standards (Landolt 1992), and the Z system to the Sloan Sky Survey standards (Krisciunas et al. 1998). The Sloan Z values have been converted from an AB-system to our Vega=0mag system by subtracting 0.572mag from the Krisciunas et al. values. The following transformations were measured : ZUFTI = ZS - 0.34 [+/-0.03] (IC -ZS ) ZUFTI = ZS - 0.21 [+/-0.03] (ZS -JUKIRT ) IUFTI = IC - 0.72 [+/-0.03] (IC -ZS ) IUFTI = IC - 0.27 [+/-0.03] (IC -JUKIRT ) 6. IMAGING POLARIMETRY AND FABRY-PEROT MODES Linear Polarimetry with UFTI is conducted with a warm half-wave plate located in the instrument support structure above the dichroic. The waveplate is located in the deployable IRPOL arm, which is lowered into the beam and rotated to modulate the polarization direction for polarimetric observations. The analyzer, a Wollaston prism constructed out of Barium Borate, is located in the second filter wheel and separates the orthogonal polarization planes to give two images on the detector. This means that polarimetric observations are restricted to the broad-band filters and the narrow-band 2.12m S1 H2 and continuum filters, which are installed in the front filter wheel. The prism is 18mm on a side and is anti-reflection coated between 0.9 and 2.5m; it was manufactured by Bernard Halle, Berlin. We had hoped to use Barium Borate (-BBO), which appears to have extremely good chromatic performance, but that material proved to be unobtainable and -BBO was used instead. The transmission of the prism was measured with an HP spectrophotometer in the Clarendon laboratory at Oxford, and is presented in Figure 6. It has excellent throughput from the UV atmospheric cut-off to 2m, but an absorption band appears at the long wavelength end of the K window between 2.2 and 2.45m. The chromatic performance is also quite good, but the separation of the e- and o-rays is wavelength dependent, varying from ~19.5 to 17 arcsec between the I and K filters. This can also lead to some slight elongation of images under the best seeing conditions. To prevent overlapping images in extended sources, a mask can be inserted in front of the cryostat window, with clear slots separated by ~18 arcsec. The mask is located about 60mm in front of the telescope focal plane, so is slightly out of focus. The rear surface is matt reflective to reduce the thermal emission seen by the detector in the K-band, but it may be better to use a K filter rather than the standard K filter. Installation of the mask is a manual operation, but all other polarimetric functions can be controlled remotely. Polarimetry is a popular option with UFTI, and like all the other near-IR instruments, instrumental polarization is low (<0.25%). Simultaneous measurement of the e- and o-rays provides good immunity to noise that can arise from variations in atmospheric seeing and transmission. Figure 6: The measured transmission curve between 0.2 and 3.2m for the 7mm thick BBO prism. A K-band Fabry-Perot etalon can be inserted between the UFTI cryostat window and the UKIRT ISU for narrow-band (R ~400km/sec) imaging. The etalon has a spacing of 40m, a Finesse of ~25 across the K band, and is operated in the converging f/36 UKIRT beam. The phase shift between the centre and edge of the usable 70 arcsec field of view is about 70-100 km/s, but is much smaller over the central 50 arcsec. Thus, one setting of the FP is sufficient to accurately image an unresolved line across the entire unvignetted field. For the same reason sky OH lines will either be transmitted or rejected across the entire field (i.e. no sharp rings) and will not cause serious flatfielding problems. The separation of adjacent orders corresponds to d/~ 0.027, so narrow band filters with percentage bandpasses of up to about that value will provide good blocking. The F-P can be used in conjunction with IRPOL to perform spectropolarimetry on extended line emission. With UFTI's small pixels, getting backgroundlimited performance from the array can be very difficult with the narrow bandpass of the FP, and the "High-Gain" readout mode is therefore recommended. Both the Fabry-Perot and IRPOL are controlled within ORAC sequences. A number of ORAC data reduction sequences also exist for standard instrument set-ups in these modes. Figure 7: A raw image showing the unvignetted F-P field lying within the 92 arcsec UFTI field. The cluster of open pixels can be seen at the top left of the array. 7. INSTRUMENT INTEGRATION AND COMMISSIONING The instrument was integrated and tested in the laboratory in Oxford from April to August 1998. The first cool-down with the engineering grade array was in April 1998, about 18 months after the initial funding for the project had been approved, but at that time, several components were still being manufactured. The main cause of delay up to that point had been caused by a faulty main cryostat casting which had not been processed properly and turned out to have substantial numbers of bubbles, inclusions and other flaws. Other delays were caused by lack of effort at critical times. Once we were satisfied that the detector was working properly, the science grade detector was installed. By this time, the detector electronics design had been tested with CIRSI at Cambridge and we were confident that it could meet the performance requirements. In addition to the normal problems encountered in instrument tests, such as extraneous pick-up on the data lines, we came across two main problems in the integration phase. The first was a problem with one of the filter modules, where the filter wheel would seize up periodically. This was fixed temporarily by decreasing the step frequency, but was not finally eliminated until it eventually seized at UKIRT early in 1999; the fault turned out to be due to a small burr on a shaft under a bearing. The second problem proved difficult to track down, and appeared at first sight to be a light leak inside the cryostat. Several attempts to eliminate a putative light leak made no difference, and we eventually concluded that instead the cause was a glowing junction on the detector multiplexor illuminated the device; this was confirmed by reinstalling the engineering array. We returned the detector to Rockwell who very quickly and efficiently provided us with a replacement. This immediately solved the problem and allowed us to ship UFTI to UKIRT for commissioning. The instrument had first light on UKIRT at the end of September 1998, and was commissioned and characterized over the following months, with the first science programmes conducted in December 1998. Several minor problems were solved over the first week of commissioning including software problems encountered in transferring from the pc-based Pixcel detector control to the Observatory system, and in tracking down and eliminating noise sources. The overheads on detector read-out were also substantially reduced in this period, and the ORAC data reduction scripts brought into operation. UFTI generated significantly larger format images than UKIRT had had to cope with previously, and this placed strains on the data acquisition and reduction computing resources, which were dealt with by providing higher specification computers and careful programming of reduction algorithms. Over the next 6 months, the instrument performance was characterized. The system gain was determined from ontelescope tests, and found to differ by about 15% depending upon which of the two detector controllers is in use. Optimum exposure times and observing strategies were developed in the face of instrument properties such as persistence and significant readout overheads. High humidity conditions cause condensation on the cryostat window, which would significantly increase the fraction of UFTI time lost. A fan was installed in front of the UFTI window in December 1998 to aid in clearing condensation, and worked well immediately. This generally allows the instrument to be operated when the humidity is below the UKIRT limit of 90%. Determination of optimum focus is difficult unless long integrations are used to average out atmospheric fluctuations, and a Hartmann mask has been installed to improve the accuracy and speed of focus determination. 8. INSTRUMENT PERFORMANCE The overall throughput of the instrument has been calculated from observations of standard stars to be ~26% at H and K and 13% at J; these figures include the atmosphere, telescope, dichroic, instrument and detector. The lower than expected throughput at J is presumably due to significantly lower detector quantum efficiency at 1.2m than at 2m. The sensitivity of UFTI is measured to be K~ 20.3 mag 5 sigma 1hr in a 2 arcsec aperture under good seeing conditions, and assuming that separate sky frames are not required. This is about 20% better than obtained with IRCAM-3 on UKIRT, because the lower sky background obtained with the HgCdTe detector and lower system emissivity with UFTI more than compensate for the higher quantum efficiency of the IRCAM InSb detector. The sensitivity in the J-band is comparable or slightly worse than that of IRCAM, reflecting the lower detector efficiency. The efficiency of operation, or open-shutter fraction, is strongly dependent on the detector integration time. The time required for calibration observations of standard stars is dominated by detector readout, filter selection and telescope motion overheads, and the overall efficiency can be as low as 30%. Fainter targets need much longer integrations, and the efficiency then approaches 90%. In common with other UKIRT instruments, a set of diagnostic dark exposures is made at the start of every night to check instrument performance. These also provide valuable records of variations and trends in properties such as numbers of dead and hot pixels. UFTI has been used for over one third of the scheduled nights on UKIRT in 1999, 2000 and 2001. Over the last 3 years, there have only been a few failures. Towards the end of 2000, low level noise was seen and the array controller was swapped for the spare early in 2001. Some vignetting of the filters was also detected and in March 2001 the instrument filter wheels were refurbished, solving the problem. After a power outage in November 2000, UFTI had cooling problems, which were traced to a faulty coldhead, which was then replaced. The instrument routinely delivers images with FWHM less than 0.5 arcsec and under the best conditions yields wellsampled images <0.3 arcsec 9. CONCLUSIONS UFTI has fulfilled its principle aim of providing a basic near-infrared imager, which has now been used on over 400 nights for many programmes over a wide range of astronomical research projects. After the initial shake-down, it has proved to be sensitive, reliable and robust, and has required relatively few unscheduled maintenance interventions. Reuse or adaptation of existing instrument designs and solutions saved substantial time and costs in the UFTI project, and future instrumentation projects could probably save effort by similar tactics. The UFTI collaboration worked well, but inevitably, there are frustrations and overheads in dealing with a widely-dispersed project. The instrument was delivered on a relatively short timescale (~2 years from the receipt of first funding) by resisting the temptation to include additional capabilities (e.g. a second imaging pixel scale and/or a spectroscopic mode), and by re-using solutions developed for, and proven in, other instruments. The instrument test and integration phase is the most uncertain, as the most difficult problems cannot be predicted with any certainty. In the case of UFTI, a cold test chamber would have allowed us to check out the filter module operation at cryogenic temperatures without needing to cool the whole instrument, and would also have allowed us to test the detector array in a known dark environment. More details of the current performance and ‘features’ in the instrument, as well as recent images produced, can be found on the UKIRT web site: http://www.jach.hawaii.edu/UKIRT. Finally, maintenance of good image quality places additional burdens on the observatory, as the telescope and instrument have to be maintained near optimum conditions as deteriorations are quickly noticed by visiting observers. ACKNOWLEDGEMENTS We are grateful to the UK Particle Physics and Astronomy Research Council, which funded UFTI through the UKIRT instrumentation programme. Interim initial funding for the project was provided by the Oxford University Physics Department. Thanks are due to many people who contributed to the instrument, including Keith Nobbs, Phil Evans, Ray Knott, Martin Beckett, Mike Snook, David Laird, Maren Purves, Malcolm Currie, Tim Chuter, and Dwight Chan. We thank all of them and others not mentioned here. We are especially grateful to Craig Cabelli at Rockwell Science Center for providing a replacement detector quickly. REFERENCES 1. Bridger A. et al 2000. Proc. SPIE Vol. 4009, p. 227-238, Advanced Telescope and Instrumentation Control Software, Hilton Lewis; Ed 2. Dalesio, L. R., Kraimer, M.R., Kozubal, A. J., "EPICS Architecture," in Proceedings of International Conference on Accelerator and Large Experimental Physics Control Systems, C. O. Pac, S. Kurokowa and T. Katoh, Eds. (ICALEPCS, KEK, Tsukuba, Japan, 1991), pp. 278-282. 3. Glasse, A.C, Atad-Ettedgui, E., Harris, JW., 1997. Proc. SPIE Vol. 2871, p. 1197-1203, Optical Telescopes of Today and Tomorrow, Arne L. Ardeberg; Ed 4. Hawarden T.G. et al, 1998 Proc. SPIE Vol. 3352, p. 52-61, Advanced Technology Optical/IR Telescopes VI, Larry M. Stepp; Ed 5. Krisciunas K, Margon, B., Szkody, P., 1998 PASP 110, 1342 6. Landolt, A.U., 1992 AJ 104, 340 7. Mackay C.D. et al 2000. Proc. SPIE Vol. 4008, p. 1317-1324, Optical and IR Telescope Instrumentation and Detectors, Masanori Iye, Alan F. Moorwood; Eds. 8. Mountain, C.M., Robertson, D.J., Lee, T.J., Wade, R. 1990. Proc SPIE Vol. 1235, 25. 9. Ramsay Howatt, S. K., .et al 2000. Proc. SPIE Vol. 4008, p. 1067-1075, Optical and IR Telescope Instrumentation and Detectors, Masanori Iye, Alan F. Moorwood; Eds. 10. Simons, D.A., Tokunaga, A.T., 2002. PASP 792, 169.