Molecular Spectroscopy Chapter 13 An introduction to Ultraviolet/Visible Absorption Spectroscopy In this chapter, absorption by molecules, rather than atoms, is considered. Absorption in the ultraviolet and visible regions occurs due to electronic transitions from the ground state to excited state. Broad band spectra are obtained since molecules have vibrational and rotational energy levels associated with electronic energy levels. The signal is either absorbance or percent transmittance of the analyte solution where: P Po A = log Po/P or P = P0*10-A %T = (P/Po)*100% The absorbance or %T can be related to concentration and are thus the basis of analyte determination. Factors affecting absorbance, linearity of calibration plots, as well as instruments used in the UV-Vis spectroscopy will be detailed in this chapter. Important Terms and Symbols in UV-Vis Spectroscopy The technique of UV-Vis spectroscopy is one of the oldest spectroscopic techniques where terminology and symbols have changed with time. A list of old and new symbols and terms are summarized in the table below: Term and Symbol (recommended) Definition Radiant power (P, Po) Absorbance (A) Energy of radiation Log (Po/P) Transmittance (T) Path length of radiation (b) Absorptivity (a) Molar absorptivity (ε) P/Po -A/bc (c in g/L) A/bc (c in mol/L) Beer’s Law Old Terms and Symbols (avoid use) Radiation intensity (I, Io) Optical Density (D), or Extinction (E) Transmission (T) l or d Extinction coefficient (k) Molar extinction coefficient The relation between absorbance and concentration is best described by Beer’s law. This law can be easily obtained by simple argument as follows: It is absolutely correct to assume that the absorbance of a solution of analyte, when light of a suitable wavelength passes through, is directly proportional to analyte concentration © as well as the path length (b) of the light beam through solution: A α bc The two terms will be equal when multiplying the second term by a constant (called molar absorptivity, ε). This will give the relation: A = εbc This relation is called Beer’s law. Sometimes this relation is referred to as Lambert law or Beer’s-Lambert law, but we will use the most popular name of the relation which is Beer’s law. The Beer’s law equation suggests a linear relation between absorbance and concentration with no intercept. However, there are some factors that affect the linearity of the relation which will be detailed shortly. Applications of Beer’s Law to Mixtures In case of coexistence of two or more absorbing species in solution, all will absorb if incident beam contains suitable wavelengths. The overall absorbance of the solution will be the sum of all absorbances of the different components: Atotal = A1 + A2 + A3 + ….. Atotal = ε1bc1 + ε2bc2 + ε3bc3 + …. However, it becomes too difficult to perform calculations involving absorbance from more than two absorbing species. Limitations to Beer’s Law Three major reasons for limitations of Beer’s law, which tender the relation between absorbance and concentration deviate from linearity, can be identified: 1. Real Limitations a. Beer’s law is good for dilute analyte solutions only. High concentrations (>0.01M) will cause a negative error since as the distance between molecules become smaller the charge distribution will be affected which alter the molecules ability to absorb a specific wavelength. The same phenomenon is also observed for solutions with high electrolyte concentration, even at low analyte concentration. The molar absorptivity is altered due to electrostatic interactions. b. In the derivation of Beer’s law we have introduced a constant (ε). However, ε is dependent on the refractive index and the refractive index is a function of concentration. Therefore, ε will be concentration dependent. However, the refractive index changes very slightly for dilute solutions and thus we can practically assume that ε is constant. c. In rare cases, the molar absorptivity changes widely with concentration, even at dilute solutions. Therefore, Beer’s law is never a linear relation for such compounds, like methylene blue. 2. Chemical Deviations This factor is an important one which largely affects linearity in Beer’s law. It originates when an analyte dissociates, associates, or reacts in the solvent. For example, an acid base indicator when dissolved in water will partially dissociate according to its acid dissociation constant: HIn = H+ + InIt can be easily appreciated that the amount of HIn present in solution is less than that originally dissolved where: CHIn = [HIn] + [In-] Assume an analytical concentration of 2x10-5 M indicator (ka = 1.42x10-5) was used, we may write: 1.42x10-5 = x2/(2x10-5 – x) Solving the quadratic equation gives: X = 1.12x10-5 M which means: [In-] = 1.12x10-5 M [HIn] = 2x10-5 – 1.12x10-5 = 0.88x10-5 M Therefore, the absorbance measured will be the sum of that for HIn and In-. If a 1.00 cm cell was used and the ε for both HIn and In- were 7.12x103 and 9.61x102 Lmol1 cm-1 at 570 nm, respectively, the absorbance of the solution can be calculated: A = AHIn + AIn A = 7.12x103 * 1.00* 0.88x10-5 + 9.61x102 * 1.00 *1.12x10-5 = 0.073 However, if no dissociation takes place we may have: A = AHIn A = 7.12x103 * 1.00 * 2x10-5 = 0.142 If the two results are compared we can calculate the % decrease in anticipated signal as: % decrease in signal = {(0.142 – 0.073)/0.142}x100% = 49% An example of association equilibria is the association of chromate in acidic solution to form the dichromate according to the equation below: 2 CrO42- + 2 H+ = Cr2O72- + H2O The absorbance of the chromate ions will change according to the mentioned equilibrium and will thus be non linearly related to concentration. 3. Instrumental Deviations a. Beer’s law is good for monochromatic light only since ε is wavelength dependent. It is enough to assume a dichromatic beam passing through a sample to appreciate the need for a monochromatic light. Assume that the radiant power of incident radiation is Po and Po’ while transmitted power is P and P’. The absorbance of solution can be written as: A = log (Po + Po’)/(P + P’) P = Po10-εbc, substituting in the above equation: A = log (Po + Po’)/(Po10-εbc Po’10-ε’bc) Assume ε = ε’ = ε A = log (Po + Po’)/(Po + Po’) 10-εbc A = εbc However, since ε’ # ε, since ε is wavelength dependent, then A # εbc A λ Therefore, the linearity between absorbance and concentration breaks down if incident radiation was polychromatic. In most cases with UV-Vis spectroscopy, the effect small changes in wavelengths is insignificant since since ε differs only slightly; especially at the wavelength maximum. b. Stray Radiation Stray radiation resulting from scattering or various reflections in the instrument will reach the detector without passing through the sample. The problem can be severe in cases of high absorbance or when the wavelengths of stray radiation is in such a range where the detector is highly sensitive as well as at wavelengths extremes of an instrument. The absorbance recorded can be represented by the relation: A = log (Po + Ps)/(P + Ps) Where; Ps is the radiant power of stray radiation.