MSc Chapter4 (Final)

advertisement
CHAPTER 4: ANALYTICAL INSTRUMENTATION
4.1 INTRODUCTION
In this section, a review of the analytical instrumentation used during sample
preparation and analysis is presented which includes an overview of the
instruments, principles and techniques used.
4.2 INSTRUMENTS USED DURING SAMPLE PREPARATION
4.2.1 Industrial microwave systems
The first commercial laboratory microwave unit with pressure feedback control in
1989, and the first with temperature feedback control in 1992 has allowed for
more rigorous design and control of microwave sample preparation procedures
[60, 61]
. These developments, coupled with the evolution of the microwave vessel to
point where it is capable of monitoring reaction conditions throughout digestion,
has allowed researchers to systematically study the decomposition mechanism of
a variety of matrices.
Principle of operation
Microwave chemistry is the science of applying microwave irradiation to
chemical reactions. Microwaves act as high frequency electric fields and will
generally heat any material containing mobile electric charges, such as polar
molecules in a solvent or conducting ions in a solid. Polar solvents are heated as
their component molecules are forced to rotate with the field and lose energy
through collisions. Semiconductive and conductive samples heat up when ions or
electrons within them form an electric current and energy is lost due to the
electrical resistance of the material.
Advantages
Conventional wet-sample preparation methods for the decomposition of solid
samples generally involve heating the samples over long periods of time using
either a hot plate, heating mantle, or oven. The process is stopped when the
analyst considers that the decomposition of the sample is sufficiently complete.
35 This type of open-vessel digestion has many limitations including the use of large
volumes of reagents, a real potential for sample contamination by external
materials and the laboratory environment, and the potential exposure of the
analyst and the laboratory to corrosive fumes.
Closed-vessel microwave decomposition uses significantly different technology to
achieve sample decomposition. Decomposition of most solid samples can be
achieved using only 10 mL reagent, and can be completed in a short period of
time. The higher temperatures achieved in the closed system give microwave
digestion a kinetic advantage over hot plate digestion, as described by the
Arrhenius Equation:
(d ln k)/(dT) = (Ea)/(RT2)
(4.1)
Integration of this equation gives:
(k2/ lnk1) = (Ea)/(2.303R)[(1/T1) – (1/T2)]
(4.2)
In this expression k1 and k2 are rate constants for the reaction of interest at T1 and
T2 respectively, Ea is the activation energy, and R is the ideal gas constant. These
equations show that the reaction rate increases exponentially with increasing
temperature. This translates into approximately a 100-fold decrease in the time
required to carry out a digestion at 175 °C when compared to a digestion at 95 °C.
In addition, because the mineral acid converts microwave energy into heat almost
instantaneously, rapid heating of the sample is achieved, further decreasing the
reaction time. Digestions are also more complete because many acids (e.g. nitric
acid) show improved oxidation potential at elevated temperatures.
A typical closed microwave system with build-in pressure and temperature control
can be seen in Figure 4.1 and Figure 4.2 which also show the microwave
consumables used during routine operation.
36 Figure 4.1 CEM Mars 5 microwave systems in operation,
(Photographed by, Impala laboratory management, 2006)
Figure 4.2 CEM Mars microwave 5 systems with high pressure and temperature
consumables, (Photographed by Impala laboratory management, 2006)
37 4.3 INSTRUMENTS USED FOR SAMPLE MEASUREMENT
4.3.1 Inductively coupled plasma optical emission spectrometry (ICP-OES)
ICP-OES instruments are used for a number of different applications in a variety
of industries. In the mining industry, ICP-OES is used in conjunction with x-ray
fluorescence spectroscopy for the analysis of major elements.
Principle of operation
The first investigations of alkali and alkali earth elements with the aid of a
spectroscope were reported by Bunsen and Kirchhoff in 1860. Spectrochemical
analysis was further developed in the 20th century when flame, arc and spark
technology were introduced [66]. An illustration of an ICP-OES as it appears today
is shown in Figure 4.3.
Figure 4.3 An illustration of a SPECTRO GENESIS ICP-OES as
provided by SPECTRO, S.A.
The principle of operation involves a
sample
introduction
system which
supplies sample at a constant rate to the ICP emission source (plasma),
where desolvation, atomization, excitation and light emission occur. During
the excitation process, molecular, atomic and ionic species in various energy
stages are produced. Energy is released in the form of electromagnetic radiation
38 and, as a result, a wavelength is formed, which is characteristic of the emitting
specie. An expression for the absolute intensity (Iqp) of a spontaneous emission
line arriving from an electronic transition from a higher state q to lower state p
can be shown to be: [64,65]
Iqp = d/4π Aqp hvqp Nq
(4.3)
Where d is the depth of source, vqp is the transition frequency, Aqp is the line
transition probability, h is Planck’s constant and Nq is the number of excited level
species. The intensity of an elemental atomic and ion line is used as the analytical
signal in quantitative atomic emission spectroscopy.
Components of the ICP-OES instrument include the peristaltic pump, the
nebuliser, the spray chamber, the rf generator, the torch, the optical systems, the
detectors and the data processing system. The ICP-OES components are displayed
in figure 4.4 and are discussed briefly thereafter.
Figure 4.4 The SPECTRO CIROS ICP-OES schematic diagram as provided by
SPECTRO, S.A.
39 Nebulizer and spray chamber
The nebulizer works in conjunction with the spray chamber. The major function
of the spray chamber is to act as a droplet size discriminator, passing only
droplets below a certain size. The size of droplets passed by the spray
chamber will largely depend upon the geometry of the spray chamber, but
to a lesser extent will depend upon the gas flow rate. A further function of the
spray chamber is to dampen pulses originating from the peristaltic pump.
Torch, rf generator and coil
Plasma RF power primarily affects the plasma temperature i.e. the greater the
power intake, the higher the plasma temperature. The net effect of power on
analyte sensitivity depends on the ratio of analyte signal to background
noise. Most plasmas are operated between 0.8 – 1.1 KW, with the exact
power chosen in accordance with the most crucial elements to be evaluated.
The plasma is sustained by energy from the RF Generator and coil. The
ICP – OES torch is centered within the induction coil, and the center tube
is 1-2 mm below the bottom of the induction coil. The three concentric
quartz tubes of the torch serve to define three separate gas flow paths i.e.
the coolant, auxiliary and carrier gas paths.
Plasma discharge
Radio frequency energy from the induction coil causes charged particles to
accelerate in a circular pathway in the same plane as the coil windings.
These charged particles collide with Argon gas atoms, causing ionization and
thereby forming a plasma. High in the plasma is the radiation zone, where
excitation, ionization, and emission take place. This region (approximately 6000
to 7000 oK) is normally utilised for analytical measurement.
Optics
The spectrometer isolates analytical wavelengths from the emitted light plasma.
The majority of wavelengths lie within the region 160 to 860 nm .
Separation of light into its component wavelengths is normally achieved
40 using a diffraction grating. There are three types of diffraction grating: ruled,
holographic and echelle grating. Many spectrometers are either flushed with
nitrogen or argon or are maintained under vacuum to remove any oxygen.
Detectors
New generation detectors have recently been introduced. These are solid-state
detectors, but are also referred to as charged - transfer devices. There are two
sub-classifications: charge – coupled devices (CCDs) and charge-injection
devices (CIDs). A CID consists of a two dimensional array of detector
elements which, when coupled to an Echelle spectrometer, is capable of
simultaneous line analysis over the range 170 -800 nm.
Readout devices and data processing
The detector produces an electrical signal which is processed by an
electronic circuit before being measured by a read-out device.
In modern spectrometers the computer controls the operating parameters of
the plasma as well as performing the task of sample logging, operation of
the auto-samplers, construction of calibration curves and facilitates the rapid
and efficient handling of data.
4.3.2 Inductively coupled plasma mass spectroscopy (ICP-MS)
Since the commercialization of ICP-MS in 1983
[64]
, it has undoubtedly been the
fastest growing trace element technique for a series of applications. The platinum
industry however, only truly started employing this technique for the analysis of
exploration samples, Final Tailings streams and Mill Circuit Product streams for
trace levels of precious metals about six years ago. And as yet ICP-MS is still not
perceived as an instrument for routine evaluation analysis unlike FAAS and ICPOES.
Principle of operation
The principle of operation for ICP-MS is similar to that of ICP-OES
and
includes a sample introduction system which supplies sample at a constant
41 rate to the ICP emission source (plasma). It is important to differentiate the
function of the plasma in ICP-MS compared to that in an ICP-OES. In ICP-OES
the plasma is used to generate photons of light, by the excitation of electrons of a
ground-state atom to higher energy level. Wavelength specific photons are
emitted, which are characteristic of the element of interest. A schematic setup of
an ICP-MS is shown in Figure 4.5.
Figure 4.5 A Schematic setup of the SPECTRO MASS 2000 ICP-MS provided
by SPECTRO, S.A
In ICP-MS the plasma is used to generate positively charged ions and not photons.
Once the ions are present in the plasma, they are projected via a low vacuum
interface into the mass spectrometer chamber and focused via an ion lens system
onto a quadrupole mass filter, Figure 4.5. The interface region consists of two
metallic cones (usually nickel), called the sampler and skimmer cones which
allow the ions to pass through to the ion optics, Figure 4.6. The ions which reach
the quadropole are separated based on their mass- to- charge ratio prior to the
detector and to filter out all the nonanalyte, interfering and matrix ions. The final
step is to convert the ions into an electrical signal with a dynode detector.
42 Figure 4.6 The interface region and ion optics of an ICP-MS system as provided
by SPECTRO, S.A
4.3.3
Graphite atomic absorption spectroscopy (GFAAS)
The use of furnaces as atomizers for atomic absorption spectroscopy was a major
breakthrough as it allowed the measurement of very low concentrations when
compared to that achievable by flame atomic absorption spectroscopy.
Principle of operation
There are many different high temperature furnace designs. The objective of
GFAAS is to generate free atoms such that atomic absorption can be measured.
The principle of operation involves three stages: [67,69]
• A drying stage during which, the solvent is removed and dried to a salt
deposit.
• An ashing stage which removes organic or inorganic material.
• An atomization stage in which free atoms are formed within a small zone.
The absorption signal produced in the atomization stage is a sharp peak, the height
of which can be related to the amount of analyte element present. The GFAAS
43 can achieve low limits of detection as the sample is completely atomized and
vaporized and the atoms are kept in the atomic reservoir for an extensive period.
4.4 COMPARING GFAAS, ICP-OES and ICP-MS TECHNIQUES
Atomic spectroscopic methods have many advantages which include: linear
dynamic ranges, low detection limits, easy and rapid qualitative analysis,
simultaneous multi-element analysis, good precision and high sensitivity. Specific
criteria are used to choose appropriate atomic spectroscopic methods for different
applications. A series of criteria shall be utilised to highlight the respective
advantages of the GFAAS, ICP-OES and ICP-MS techniques.
Detection limits
Typical detection limit ranges for major atomic spectroscopy techniques are
shown in Table 4.1.
Table 4.1 Detection limit ranges [70-72]
Atomic spectroscopy techniques
Detection limit range (μg/L)
FAAS
1 – 1000
ICP-OES – Radial view
1 – 100
ICP-OES – Axial view
0.1 – 10
Hydride generation FAAS
0.01 – 1
GFAAS
0.01 – 0.1
ICP-MS
0.001 – 0.01
It is clear from the data in Table 4.1. , that ICP-MS offers the lowest detection
limit followed by GFAAS and axial ICP-OES. Radial ICP-OES and FAAS show
similar detection limits, but ICP-OES can atomize refractory and rare earth
elements more effectively due to the higher temperature that can be achieved by a
plasma as when compared to a flame. Hydride generation FAAS offers
exceptional detection limits for mercury (Hg), arsenic (As), bismuth (Bi),
antimony (Sb), selenium (Se) and tellurium (Te). ICP-MS in conjunction with
44 collision / reaction cells or magnetic sector technology can achieve detection
limits as low as parts-per-quadrillion (ppq) for many elements
[63]
. It is important
to emphasize that these detection limits are only achievable in simple matrixes.
Linear dynamic range (LDR)
ICP-MS is considered to be an ultra-trace element technique which has an LDR in
excess of 105. GFAAS has a limited LDR of 102 – 103, but can be used for higher
concentrations when a less sensitive analytical line is used. ICP-OES is used for
trace and major element analysis with a LDR of 105.
Precision
Short and long term precision are a good indication of how stable an instrument
is. Precision is usually expressed in percent relative standard deviation (%RSD).
The short term precision for ICP-MS is between 1 to 3 %RSD and the long term
precision is less than 5 %RSD. The short term precision for ICP-OES is between
0.3 – 2 %RSD and the long term precision depending on the nebulizer should be
not more than 3 %RSD.
Precision for both ICP-OES and ICP-MS may be
improved by using internal standardisation. GFAAS has a short term precision of
between 0.5 – 5 % RSD, but long term precision is more a function of the number
of graphite tube firings than time [63].
Interferences
Although appropriate techniques are chosen for each application by experienced
staff, interferences still need to be addressed during method development. Specific
interference common to various techniques are shown in Table 4.2.
45 Table 4.2 Summary of instrumental interferences [73,74]
Technique
Type of interference
Method of compensation
FAAS
Ionisation
Chemical
Ionisation buffers
Release agents or nitrous oxide-acetylene
flame
Dilution, matrix matching or standard
addition
Physical
GFAAS
Chemical, physical
Standard temperature platform furnace
conditions, matrix modifiers, standard
addition
Molecular absorption
Zeeman or continuum source background
correction
Spectral
ICP-OES
ICP-MS
Zeeman background correction
Spectral
Background correction
Inter elemental corrections
Alternate analytical line
Matrix effects
Internal standardization
Ionization
Spectral
Ionization buffers
Background correction
Inter elemental correction
Alternate analytical lines
Matrix acids
Doubly charged ions
Matrix effects
Ionization
Higher resolution systems to resolve
masses less than 1 amu apart
Internal standard
Matrix matching, sample dilutions,
standard addition, isotope dilutions
Matrix matching, internal standard
Space charged effects
Isobaric effects
Analysis time and sample throughput
Analysis time and sample throughput rates are influenced by the accuracy and
precision required plus the type of instrument employed. FAAS is a simple and
fast technique with a measurement time of less than 30 s per replicate of 3
integrations. ICP-OES and ICP-MS have measurements times of about 3 min per
46 replicate of 3 integrations, while GFAAS is a very tedious technique with a
measurement time of up to 5 min per replicate of 3 integrations.
General
The most commonly employed instrumental techniques in the platinum industry
are that of FAAS, GFAAS, X-ray fluorescence spectroscopy, ICP-OES and their
capabilities and analytical limitations are well known. ICP-OES would appear to
have become the most popular routine technique for inorganic multi-elemental
analysis of dissolved samples, despite requiring significant capital investment and
having running expenses which are much higher than those for FAAS.
4.5
INSTRUMENTATION FOR THE IDENTIFICATION OF
MINERALS.
4.5.1 Scanning electron microscope (SEM)
SEM uses a focused beam of high-energy electrons that generate signals which
reveal information about the sample such as external morphology, chemical
composition, crystalline structure and the orientation of the sample composition.
Principle of operation
Accelerating electrons in a SEM carrying significant kinetic energy generate
radiation signals when the electrons are decelerated in a solid sample as a result of
electron interactions. These signals include secondary electrons (which produce
SEM images), backscattered electrons (BSE), diffracted backscattered electrons
(EBSD) (which are used to determine crystal structures and the orientation of
minerals), photons (characteristic X-rays that are used for elemental analysis and
continuum X-rays), visible light and heat [76].
Many SEM installations have an energy dispersive X-ray detector system (EDS),
which allows for spectral analysis of X-rays generated from the sample directly
under the electron beam. X-ray generation is produced by inelastic collisions of
incident electrons with electrons in discrete atomic orbitals. As the excited
electrons return to lower energy states, they yield X-rays of a fixed wavelength.
47 Characteristic X-rays are produced for each element in a mineral which is
"excited" by the electron beam.
Advantages and disadvantages
Some advantages of this technique include are as follows:
•
Excellent characterization of solid samples.
•
The electron beam can be scanned over a very small area of the sample.
•
SEM in conjunction with an EDS detector allows morpholograpic,
topographic, crystallographic and compositional information to be
obtained rapidly and simultaneously from the same area.
•
SEM analysis is considered to be "non-destructive", meaning there is no
volume loss of the sample, and as such it is possible to analyze the same
materials repeatedly.
Unfortunately, SEM cannot detect very light elements (H, He and Li) or
elements with an atomic number less than 11. Although SEM with an EDS
detector is fast and easy to use, it suffers from poor energy resolution and poor
sensitivity towards elements present in low abundance compared to
wavelength dispersive X-ray detectors (WDS) or electron probe micro
analyzers (EPMA).
4.5.2 Electron probe micro-analyzer (EPMA)
The EPMA is a micro-beam instrument used for the in situ non-destructive
chemical analysis of minute solid inclusions [75].
Principle of operation
An electron microprobe operates under the principle that if a solid material is
bombarded by an accelerated and focused electron beam, the incident electron
beam has sufficient energy to dislodge inner-shell electrons of the constituent
atoms in the sample to be analyzed. Outer-shell electrons fill these inner-shell
vacancies, losing energy by the emission of characteristic X-rays as in X-ray
48 fluorescence. These electrons can be focused to a very fine beam through a set of
electromagnetic lenses between the electron source and the sample to be analyzed.
The X-rays generated are analysed by a crystal spectrometer with wavelength
dispersive X-ray detector. Of most common interest in the analysis of geological
materials are secondary and back-scattered electrons, which are useful for surface
imaging or obtaining an average composition of the material. The electron optical
path and X-ray spectrometer of an electron microprobe are illustrated in Figure
4.7.
Figure 4.7
microprobe
The electron optical path and x-ray spectrometer of an electron
[68]
.
Advantages and disadvantages
Some advantages of this technique are as follows:
•
Although similar to SEM it is equipped with a range of crystal
spectrometers which enable quantitative chemical analysis at high
sensitivity.
49 •
The EPMA is capable of analysing sample areas as small as 1-2 micron
diameter and even minute single phases in a material.
•
Chemical analyses can be obtained in situ, which allows the detection of
small compositional variations within chemically zoned material.
•
It is considered to be a “non-destructive” technique.
EPMA is also unable to detect the very light elements (H, He and Li) and as a
result cannot detect “water” in hydrous minerals. It is also known that some
elements generate x-rays with overlapping peak positions which need to be
separated. The absolute detection limit for most elements is not as good as that
achievable by x-ray fluorescence because of the presence of a continuum
spectrum.
Nevertheless, the ability of an EPMA to obtain quantitative chemical analysis on a
minute volume of sample or mineral grain ensures its continued popularity in the
study of minerals.
50 
Download