1 © 2009 HORIBA, Ltd. All rights reserved. © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Dr. Bernd Bleisteiner Application Scientist Raman Spectroscopy HORIBA Scientific © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Outline Dispersive Raman Instruments Principal setup of dispersive Raman instruments Laser Source – UV-VIS-NIR optics Polarisation control Collection geometries & reduction of Rayleigh scattering – single and triple spectrometers Collection of Raman photons – Macro- & Micro-approach Confocal Raman Microscope and spatial resolution Spectral resolution and spectral coverage Sensitivity in subject to the detector Light flux in subject to dispersion 3 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Principal setup of dispersive Raman instruments A Raman spectrometer requires five elements: 1. 2. 3. 4. 5. Light source Collecting optics Straylight rejection filter Wavelength selector and Detector A dispersive Raman spectrometer uses a monochromator or a spectrograph as wavelength selector 4 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Principal setup of dispersive Raman instruments Filter Collecting lens Detector Wavelength Selector Sample Light source 5 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Principal setup of dispersive Raman instruments Depending on the filtering technique and power of spectral resolution Raman instruments can be divided into two principal groups 1. Single stage instruments which suppress the Rayleigh light by notch or edge filters and 2. Double or triple stage instruments which suppress the Rayleigh light by an intermediate slit Double and triple stage instruments Single stage instruments 6 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments The Light Source Scattering spectroscopy (Raman) experiences much smaller cross section compared to absorption spectroscopy (IR). As a result, Raman scattering intensity is much weaker than that IR absorption intensity. Therefore, a powerful light source is essential. While the first Raman was observed using sun light, all modern Raman spectrometers use lasers exclusively as the light source. The laser light is called the excitation light or incident beam. The optical path from laser to sample is called the Incident Beam Path. 7 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Laser Sources Dispersive Raman instruments can be equipped with lasers sources starting from the UV up to the NIR Most common are laser sources in the VIS region Possible wavelengths are 227, 244, 257, 325, 364, 413, 442, 457, 473, 488, 514, 532, 633, 647, 660, 785, 830, 1064 nm -1 Horizontal lines indicate a relative Raman Shift of 3800 cm 244 325 457 488 514 532 633 785 830 1064 200 400 600 800 1000 1200 1400 1600 1800 Wavelength [nm] 8 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Laser Sources The main reason for switching the laser source is to avoid fluorescence, which is interfering with the Raman spectrum Laser wavelength, λ3 Laser wavelength, λ3 Laser wavelength, λ1 Fluorescence Raman shift, λ1-1+∆ Laser wavelength: λ3 < λ2 < λ1 © 2009 HORIBA, Ltd. All rights reserved. Raman shift, λ3-1+∆ 9 60 000 Green spectrum: 532 nm laser Red spectrum: 633 nm laser Brown spectrum: 785 nm laser 50 000 40 000 Intensity (cnt) 30 000 20 000 10 000 60 000 0 50 000 600 800 Wavelength (nm) 1 000 40 000 Intensity (cnt) Dispersive Raman Instruments Laser Sources 30 000 Fluorescence is wavelength dependent Ordinary Raman is wavelength independent 20 000 10 000 0 1 000 2 000 Raman Shift (cm-1) 3 000 10 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Laser Sources A second important reason for switching the laser source is to do resonance Raman spectroscopy Electronical States Virtual States Polarizability is particularly high if excitation occurs in electronic resonance (Resonance Raman) Excitation light, λ0 Resonance Raman ∆>0 ν Stokes Scattering: λ > λ0 11 © 2009 HORIBA, Ltd. All rights reserved. Resonance Raman of carotinoids in cyano bacteria 17_Cyanobakterium Strang_D3_473nm_H1000_S100_SW_600gr_1x30s 8 000 -20 -10 Y (µm) 7 000 6 000 0 10 Intensity (cnt) Dispersive Raman Instruments Laser Sources – Resonance Raman 5 000 20 4 000 4 µm -20 0 X (µm) 20 3 000 2 000 1 000 0 500 1 000 1 500 2 000 Raman Shift [cm-1] 2 500 3 000 3 500 Laser was attenuated with a density filter OD3. Laser energy was only 16 µW@sample 12 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Laser Sources – VIS optics For Raman instruments equipped with VIS laser sources one optical path is sufficient Confocal hole CCD Microscope Lens Slit Mirror 13 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Laser Sources – UV-VIS optics For Raman instruments equipped with UV and VIS laser sources the optical path has to be adapted to the spectral range For UV instruments a switchable path with (Aluminum) mirrors avoids optical mismatch by chromatic aberrations Special collecting optics (objectives) with a high transmission in the UV are necessary CCD CCD Optional UV optimized CCD Microscope With lenses: Chromatic aberrations in UV Microscope Reflective optics: Achromatic λ1, λ2, λ3 focus at same position © 2009 HORIBA, Ltd. All rights reserved. 14 Dispersive Raman Instruments Laser Sources – VIS-NIR optics For Raman instruments equipped with VIS and NIR laser sources (> 785 nm) the optical path has to be adapted to the spectral range For NIR instruments a switchable path with (Silver) mirrors are necessary Special collecting optics (objectives) with a high transmission in the NIR are necessary An InGaAs detector is a prerequisite for detecting Raman spectra in the NIR CCD CCD InGaAs Microscope Microscope 15 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Polarization control The polarization of the Raman scattered light can contain useful information This property can be measured by using polarized excitation and a polarization analyzer Because the laser light is normally polarized a polarizer in the excitation path is not necessary For polarization experiments the polarization direction of the incident laser path can be controlled by a lambda half wave plate In the Raman path an analyzer set at both perpendicular and parallel to the excitation is necessary 16 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Polarization control Filter Collecting lens Detector Wavelength Selector Sample Analyzer perpendicular or parallel to the excitation plane Polarized Laser source λ/2-plate λ/2-plate to turn the polarization of the polarized laser 17 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Polarization resolved measurements Depending on isotropic or anisotropic samples polarization experiments showing a different result in subject to the relative direction of the excitation polarization and the analyzer Isotropic Samples: Gases and liquids, in which the molecules are randomly oriented Anisotropic Samples: Oriented single crystals, in which the atoms or molecules oriented in a fixed position Attention: Solid amorphous phases or micro crystals are randomly oriented. For the same chemical species, the observed Raman spectra depends on the spatial orientation of the micro sized particles. 18 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Polarization resolved measurements Polarizer Analyzer Isotropic sample Anisotropic Sample same spectra different spectra same spectra different spectra 19 © 2009 HORIBA, Ltd. All rights reserved. Isotropic Sample: Cyclohexane 8000 Normalized spectra 6000 Intensity (a.u.) Dispersive Raman Instruments Polarization resolved measurements 4000 2000 0 800 1000 1200 1400 1600 Wavenumber (cm-1) 20 © 2009 HORIBA, Ltd. All rights reserved. Anisotropic Sample: Acetylic salicylic acid 8000 Normalized spectra 6000 Intensity (a.u.) Dispersive Raman Instruments Polarization resolved measurements 4000 2000 1000 1050 1100 1150 Wavenumber (cm-1) 21 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Raman Scattering Collection Geometry Theoretically, Raman scattering can be observed from any angle, because Raman scattering is generated in all directions Practically and historically, three angles were favored; 0°, 90° and 180°. 0°configuration (forward scattering) was the least favorite because of high Rayleigh scattering and laser interference. 90°configuration (right angle scattering) was an early favorite because Rayleigh scattering was minimized. However, this is viable only for transparent samples such as liquids or solutions Sample Molecule Laser 180° configuration (backscattering) is viable to all forms of samples. However, it posed implementation difficulty - How to guide the laser beam to the sample without blocking the Scattering beam path? 22 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments 180°° Backscattering Collection Geometry Is the most common collection geometry nowadays because almost all Raman system are based on microscope optics The laser is guided to the samples either by a beam splitter or (especially common in single stage instruments) by injection rejection filters (edge-, notch-filters) The filter reflects the laser light and serving to deliver the incident beam to the sample Rejection filter that blocks laser light wavelength is placed in the incident beam path. Sample Molecule The filter is transparent for scattering light because wavelengths are different from the laser light, and passes the Raman signal through. Laser 23 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Notch Filter vs. Edge Filter White light spectrum with a notch filter White light spectrum with an edge filter 150 1500 100 1000 50 500 0 -300 -200 -100 0 100 Wavenumber (cm-1) 200 300 0 -300 -200 -100 0 100 200 300 Wavenumber (cm-1) Zero Raman shift → Excitation laser position • A finite life time • Stokes and Anti-Stokes Raman • A virtually infinite life time • Stokes Raman only There is a cost advantages to the edge filter die to no aging, but Anti-Stokes-Raman is not obtainable 24 © 2009 HORIBA, Ltd. All rights reserved. With double or triple spectrometers the Rayleigh line is filtered by an intermediated slit The cut off gained with this technique is < 5cm-1 in comparison to 100 cm-1 with an edge or notch filter subtractive double monochromator 50000 40000 Notch Filter 20000 10000 3000 Intensity (a.u.) 30000 Intensity (a.u.) Intensity (a.u.) Dispersive Raman Instruments Filter vs. Subtractive Double Monochromator 2000 1000 0 3000 SiGe spectrum 2000 1000 0 -10 -10 -5 -5 0 0 5 5 10 10 15 15 Wavenumber (cm-1) 20 20 25 25 wavenumber (cm-1) 0 0 50 100 150 200 250 300 350 Wavenumber (cm-1) 25 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments 90°° Collection Geometry 90° collection geometry is used in macro Raman experiments The incident beam path of the laser is different from the scattering beam path (= the optical path from sample to the detector) 90° collection geometry is suitable for Raman experiments which need very high excitation power Raman scattering Sample Molecule Macro chamber with interchangeable collection optics covering adjustable collection from 90° to 180° geometries Laser 26 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments 0°° Collection geometry – Transmission Raman 0° collection geometry is used for Raman volume measurements of bulky opaque samples The incident beam path of the laser is in opposite to the scattering beam path Raman scattering Example: Macro -Transmissions Raman for content uniformity control on tablets Sample Molecule Laser 27 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Collection geometry – Various collection angle For micro Raman experiments with various excitation / collection angles customized instruments are available E. g. off-axis excitation for improving very low frequency measurements on solid state samples Sample Molecule Raman scattering Laser 28 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Collection optics Depending on the experiment and the sample different collecting optics are used Therefore micro Raman instruments are equipped with different objectives with different numerical aperture Objective N.A. Working distance [mm] x100 0.90 0.21 x50 0.75 0.38 x10 0.25 10.6 x100 LWD 0.80 3.4 x50 LWD 0.50 10.6 29 © 2009 HORIBA, Ltd. All rights reserved. θ Collection solid angle Large for high N.A. lens Small for low N.A. lens High N.A. lens Sampling volume Small for high N.A. lens Large for low N.A. lens θ Working distance Low N.A. lens Working distance Dispersive Raman Instruments Collection optics Laser spot size Small for high N.A. lens Large for low N.A. lens NA = n · sin (Θ Θ) n: refraction index Θ: aperture angle 30 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Collecting Optics – what objective should be used? A distinction between opaque and transparent samples has to be made For opaque samples, high N.A. lens works better because there is almost no penetration of the laser into the sample. High N.A. lens enables - High laser power density (mW/µm3) → increases sensitivity - Wide collection solid angle → increases sensitivity 100 % 30 000 Example for an opaque sample: Silicon wafer 25 000 70 % 20 000 Intensity (cnt/sec) Y (µm) -20 Silicon x100 – NA = 0.9 – 31.350 C/s x50 – NA = 0.75 - 21.995 C/s x10 – NA = 0.25 - 1.462 C/s 15 000 0 10 000 10 µm 20 5 000 5% 0 X (µm) 0 460 480 500 520 540 Raman Shif t (cm-1) 560 580 600 620 31 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Collecting Optics – what objective should be used? A distinction between opaque and transparent samples has to be made For transparent samples, low N.A. lens works better because of penetration of the laser into the sample. Low N.A. lens enables - Large sampling volume → increases sensitivity x10 3 14 12 Sample: Cyclohexane Instrument: ARAMIS Red: x100LWD, 7,000 cts/s Blue: Macro lens, 14,500 cts/s cyclo_100xLWD cyclo_macro 100 % Intensity (cnt) 10 8 48 % 6 4 2 0 740 760 780 800 820 Raman Shift (cm-1 ) 840 860 880 32 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spatial resolution and confocality The ability to axially discriminate signals using the confocal hole is called confocality. Decreasing the confocal hole diameter increases axial and lateral spatial resolution. Slit Confocal hole Rejection fitter 33 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spatial resolution and confocality Confocal hole (image plane) Wide Hole Objective Sample (object plane) Sampling Volume Narrow Hole 34 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spatial resolution and confocality Confocal z-scan against silicon as a measure for the axial resolution λexc = 633 nm Narrow Hole: Collecting Raman radiation that originates only from within a diffraction limited laser focal volume with a dimension of: Focus waist diameter ~ 1.22 λ / NA Depth of laser focus ~ 4 λ / (NA)2 Laser focus waist diameter Depth of laser focus (d.o.f) 35 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spatial resolution and confocality Integrated Intensity Z-scan of multi layered polymer Film Glue x10 3 Confocal hole diameter 1000 µm 6 4 2 0 0 -20 -40 Glass x10 3 Integrated Intensity 398.0 x10 3 -60 z-Axis (µm) 20 Intensity 15 Film 1730.3 10 5 Glue -80 -100 -120 Confocal hole diameter 100 µm 6 4 2 0 0 -20 -40 -60 z-Axis (µm) -80 -100 -120 0 500 1 000 Raman Shift (cm-1) 1 500 The narrow hole improves axial resolution. 36 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spatial resolution and confocality Spatial resolution especially axial resolution can be improved by (water or oil) immersions objectives (n1 ≈ n2 or ideally n1 = n2) With normal (air) objectives the difference in refraction index (n1 ≠ n2) of air and material results in an optical distortion which reduce the spatial resolution n1 ≠ n2: d.o.f. spreads linearly with z0 n1 ≠ n2: focus is located at deep P, which differs from “mechanical deep” Length or depth of Laser Focus (= d.o.f.) Air Sample Data and figures from: Modeling and Measuring the Effect of Refraction on the Depth Resolution of Confocal Raman microscopy Neil. J. Everall, Applied Spectroscopy, Vol. 54, 6, 2000 © 2009 HORIBA, Ltd. All rights reserved. 37 300 x100 LWD / 100 µm Hole Intensity (cnt) Film Glue Glass The water immersion objective improves axial resolution remarkably in comparison with the x100 LWD air objective 200 100 0 0 -50 -100 -150 -200 Z (µm) -250 -300 -350 x100 water immersion / 100 µm Hole 500 Intensity (cnt) Dispersive Raman Instruments Spatial resolution and confocality 400 300 200 100 0 0 © 2009 HORIBA, Ltd. All rights reserved. -50 -100 -150 -200 Z (µm) -250 -300 -350 38 Exploring lateral spatial resolution -Test on a well structured sample -15 Circles etched by a Gallium ion beam on silicon with different distances of 1400, 1200, 1000, 800, 600, 400, 200 nm. -10 -5 Y (µm) Dispersive Raman Instruments Spatial resolution and confocality Structure were mapped with a mechanical table and x100 try objective with NA = 0.9 0 5 10 1 µm 0 10 X (µm) Sample in courtesy of Dr. Wilhelm, TU Graz © 2009 HORIBA, Ltd. All rights reserved. 39 Dispersive Raman Instruments Spatial resolution and confocality Optical parameters for the lateral spatial resolution Laser spot size 1.22 λ dw = NA Diffraction limited resolution d given by the Rayleigh criterium 0.61 λ 0.61 λ d= = n sin α NA e. g. with 633nm HeNe laser and x100 objective with NA = 0,9 d = 430 nm 40 © 2009 HORIBA, Ltd. All rights reserved. Spatial resolution and confocality Y (µm) -2 -4 -6 -8 -10 -8 0 0 X (µm) -6 4 4 2 -1 Raman Shift [cm ] -4 6 6 -2 8 8 0 10 10 2 12 14 16 12 0.5 µm Intensity [Counts] 30 20 10 0 x [µm] Sample in courtesy of Dr. Wilhelm, TU Graz © 2009 HORIBA, Ltd. All rights reserved. 41 -6.11 -4.17 -1.70 0.97 3.63 40 6.95 10.21 Intensity [Counts] Spatial resolution and confocality 20 0 12 10 8 6 4 2 Raman Shift [cm -1] 0 -2 -4 -6 -8 x [µm] Real distances [nm] Fitted FWHM [nm] Difference 1400 1160 240 1200 1050 150 1000 780 220 800 640 160 600 520 80 400 430 -30 200 270 -70 -10 -8 -6 -4 -2 0 Spatial resolution of object structure which can be resolved is about λ/2 Y (µm) 2 4 6 8 10 12 14 0.5 µm 16 0 X (µm) Sample in courtesy of Dr. Wilhelm, TU Graz © 2009 HORIBA, Ltd. All rights reserved. 42 Dispersive Raman Instruments Spectral resolution and spectral coverage Spectral resolution is a function of 1. dispersion, 2. widths of entrance slit and 3. pixel size of the CCD Dispersion is the relation between refraction of light according to the wavelength of light Dispersion is a function of the 1. focal length of spectrograph the 2. groove density of the grating and 3. the (excitation) wavelength In general, long focal length and high groove density grating offer high spectral resolution. 43 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spectral resolution and spectral coverage Schematic diagram of a Czerny-Turner spectrograph Slit Collimating mirror Raman straylight generated at a certain excitation wavelength Grating with a certain groove density Focusing mirror CCD Detector with a certain pixel size Focal Length 44 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Dispersion in respect of excitation wavelength -1 Horizontal lines indicate a relative Raman Shift of 3800 cm 244 - 269 nm (25 nm) 325 - 371 nm (46 nm) 457 - 553 nm (96 nm) 488 - 599 nm (111 nm) 514 - 639 nm (125 nm) 532 - 667 nm (135 nm) 633 - 833 nm (200 nm) 785 - 1119 nm (334 nm) 830 - 1210 nm (380 nm) 1064 - 1768 nm (704 nm) 400 600 800 1000 1200 1400 1600 1800 Wavelength [nm] Short excitation wavelength Same focal length Same grating CCD Detector CCD Detector 200 Long excitation wavelength 45 © 2009 HORIBA, Ltd. All rights reserved. Length of CCD Chip x10 3 473 nm – 633 nm – 785 nm 22 20 18 Same focal length Same grating Length of CCD Chip 16 Intensity (cnt/sec) Dispersive Raman Instruments Spectral coverage - dependence from excitation wavelength 14 12 10 Length of CCD Chip Relative Raman shift of 3100 cm-1 8 corresponds to 81 nm 6 4 Relative Raman shift of 3100 cm-1 corresponds to 154 nm 2 Relative Raman shift of 3100 cm-1 corresponds to 252 nm 0 500 600 700 800 Wavelength (nm) 900 1 000 46 © 2009 HORIBA, Ltd. All rights reserved. CCD Detector Short focal length Same grating Same excitation wavelength Long focal length CCD Detector Dispersive Raman Instruments Dispersion in respect of the focal length 47 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Dispersion in respect of the focal length Dispersion in cm-1 / pixel 1800 gr/mm Grating LabRAM (F = 300 mm) LabRAM HR (F = 800 mm) Same grating Same excitation wavelength 48 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Dispersion in respect of the grating Grating Equation: sin i + sin i’ = nkλ where n is the groove density, k the diffraction order and λ the wavelength 0th order Polychromatic light 1st order i’ i i 2nd order Grating 49 © 2009 HORIBA, Ltd. All rights reserved. Same focal length Same excitation wavelength © 2009 HORIBA, Ltd. All rights reserved. CCD Detector Low density groove grating CCD Detector Dispersive Raman Instruments Dispersion in respect of the grating High density groove grating 50 Dispersive Raman Instruments Spectral resolution as a function of slit width Slit Slit Slit One of parameters that determines the spectral resolution is the entrance slit width. The narrower the slit, the narrower the FWHM (full width at half maximum), and higher the spectral resolution. When recording a line whose natural width is smaller than the monochromator’s resolution, the measured width will reflect the spectrograph’s resolution. 51 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Spectral resolution as a function of pixel size Detector Intensity Detector Because a CCD detector is made of very small pixels, each pixel serves as an exit slit (pixel size = exit slit width) For the same size CCDs, the CCD with a larger number of smaller pixels produces a larger number of spectral points closer to each other increasing the limiting spectral resolution and the sampling frequency 26 µm pixel vs. 52 µm pixel (simulation) 600 650 Raman Shift (cm 700 -1) 52 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Sensitivity in subject to the detector Comparison of different detectors concerning quantum efficiency between 200 – 1100 nm Front illuminated Back illuminated UV coated Front illuminated open electrode Back illuminated deep depleted 53 © 2009 HORIBA, Ltd. All rights reserved. Comparison of fingerprint region of polyethylene spectra CCD quantum efficiency and scatter efficiency (∼ ∼ ν4) has influence on sensitivity in NIR poly1_633nm BLACK: 785 nm, 30s, 20mW @sample poly1_785nm RED: 633 nm, 5s, 10mW @sample 8 000 7 000 6 000 5 000 Intensity (cnt) Dispersive Raman Instruments Sensitivity in subject to the detector 4 000 3 000 2 000 1 000 0 1 000 1 100 1 200 Raman Shift (cm -1 1 300 1 400 1 500 ) Factor 6 on measurement time, Factor 2 on laser power → means 12 times more input in NIR 54 © 2009 HORIBA, Ltd. All rights reserved. Comparison of C-H region of polyethylene spectra CCD quantum efficiency and scatter efficiency (∼ ∼ ν4) has influence on sensitivity in NIR x10 3 BLACK: 785 nm, 30s, 20mWpoly1_633nm @sample poly1_785nm RED: 633 nm, 5s, 10mW @sample 25 20 Intensity (cnt) Dispersive Raman Instruments Sensitivity in subject to the detector 15 10 5 0 2 600 2 700 2 800 Raman Shift (cm -1 ) 2 900 3 000 3 100 25 times more signal with 633 nm despite 12 times more input in NIR 55 © 2009 HORIBA, Ltd. All rights reserved. Comparison of polyethylene spectra acquired with 633 and 785 nm excitation. Raman signal in the 785 nm spectrum are equal or weaker in comparison to 633 nm spectrum although Factor 6 on measurement time and factor 2 on laser power. Mainly caused due to lower QE of the detector in the NIR region. x10 3 CCD_QE_OP BLACK: 785 nm, 30s, 20mW @sample poly1_633nm RED: 633 nm, 5s, 10mW @samplepoly1_785nm BLUE: QE curve of OE-CCD 20 Intensity (cnt) Dispersive Raman Instruments Sensitivity in subject to the detector 15 10 5 0 650 700 750 800 850 900 wavelength (nm) 950 1 000 1 050 1 100 56 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Light flux in subject to dispersion If dispersion is doubled integrated signal becomes approximately half because every pixel is collecting only half amount of photons 57 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Light flux in subject to dispersion Silicon spectra acquired with 633 nm with 600 and 1200 grating. Ratio is 1 : 2. Dispersion is 2.07 and 0.94 cm-1/pixel @ 654 nm (according to 520,7 cm-1). Ratio is 2.2 : 1 Integrated Raman signal of 520,7 cm-1 band is 265063 and 136982. Ratio is 1.9 : 1 Si_633nm_1200gr Si_633nm_600gr 30 000 Intensity (cnt/sec) 25 000 20 000 15 000 10 000 5 000 0 510 520 Raman Shift (cm-1) 530 Only grating was varied. All other acquisition conditions are unchanged © 2009 HORIBA, Ltd. All rights reserved. 58 Reverse argument: If dispersion is similar or equal by a combination of grating and focal length (e. g. F = 300 mm & 1200 gr/mm and F = 600 mm & 600 gr/mm) the integrated signal becomes approximately the same High density groove grating Long focal length CCD Detector Short focal length CCD Detector Dispersive Raman Instruments Light flux in subject to dispersion Low density groove grating 59 © 2009 HORIBA, Ltd. All rights reserved. Dispersive Raman Instruments Summary – What we have heard about Principal setup of a Raman spectrometer Single and triple stage instruments Laser excitation – switching of excitation wavelength Different optics for VIS, UV-VIS and VIS-NIR Raman instruments Polarization control – isotropic vs. anisotropic samples Collection geometry 0°, 90°, 180°or various angles Different Rayleigh filtering techniques Collection optics with high and low NA Spatial resolution and confocality Spectral resolution and spectral coverage Dispersion in respect of the focal length, grating, wavelength Quantum efficiency of different detectors Wavelength dependence of quantum efficiency of a CCD detector Light flux in respect of dispersion 60 © 2009 HORIBA, Ltd. All rights reserved. Thank you 61 © 2009 HORIBA, Ltd. All rights reserved. © 2009 HORIBA, Ltd. All rights reserved.