Pacemakers and AICD’s - Chapter 10 Objectives: State the indications for temporary and permanent pacing modalities Discuss the components of the various pacing systems Identify normal and abnormal pacemaker function on a rhythm strip Discuss significant hemodynamic consequences of abnormal pacemaker function Discuss the clinical management of patients with pacemakers Describe the care considerations in a patient with an implantable cardio-defibrillator Pacemaker An artificial device that electrically stimulates the myocardium to depolarize, which begins a contraction Temporary or permanent Indications for Pacing – Page 256 Symptomatic bradyarrhythmias Drug Toxicity Moderate to Severe Heart Failure Support cardiac output after cardiac surgery, or high risk cardiac catheterization Components of a Pacing System – Page 256 Pulse generator Pacing Leads Components of Pacing System Page 256-260 Pulse generator o Houses the battery that creates electrical stimulus to maintain a specified rate o Contains controls: electrical output(mA), rate, sensitivity (mV), mode o Power source Pulse Generator Temporary transvenous or epicardial Pulse Generator – Page260 Figure 10-6 Permanent pulse generator Basic Components of the Temporary Pulse Generator – Page 260 Figure 10-5 On-Off control Mode Dial o Fixed rate or asynchronous o Demand or synchronous pacing Basic Components of the Temporary Pulse Generator – Page 260 Figure 10-5 Rate control o Sets number of beats/minute o Usually set at 60-80 bpm to ensure adequate cardiac output mA or Output Dial o Regulates the amount of energy ( in mA) delivered to the distal electrode. o Ranges from 0.1 to 20 mA Sensitivity o Regulates pacemaker’s ability to detect heart’s intrinsic electrical activity o 1 mV is maximum sensitivity (will see a complex that is very small in size) o 20 mV is asynchronous (will not sense any intrinsic electrical activity) New 7/12 NAM Page 1 Components of a Pacing System – Page 257 Figure 10-1, Page 260 Figure 10-6 Pacing catheter (lead or electrode) o Transmits the patient’s rhythm to pulse generator o Transmits electrical stimulus between generator and chamber paced Transcutaneous Leads Page 258-259 Noninvasive procedure where two large pacing pads are placed on the anterior and posterior chest Pacing pads are attached to a pacing cable, which is then connected to a defibrillator/monitor ECG leads are also attached to the patient Initiate when drug therapy has been ineffective in increasing cardiac output Requires a higher current output (mA) May cause chest wall pain and skin burns Effective when meaningful contractile activity is present only Transvenous Pacemakers – Page 260 Figure 10-5 The pacing electrode is advanced through a vein into the RA, RV, or both Inserted into the RA for atrial pacing Inserted into the RV for ventricular pacing May be unipolar or bipolar leads Epicardial Pacing – Page 259 Electrodes are attached to the epicardial surface of the heart Atrial and/or ventricular wires Permanent Pacemaker Modes of Pacemaker Function – Page 256 Asynchronous (Fixed) o Initiate impulses at a set rate regardless of patient’s intrinsic rate o Used when no intrinsic heart rate o Set at 20 mV on sensitivity dial Synchronous (Demand) o Senses intrinsic rhythm and inhibits discharge when intrinsic heart rate adequate – synchronized with the patient’s own heart activity Permanent Pacemaker Identification Codes – Page 261-262 First letter = chamber paced Second letter = chamber sensed Third letter = response mode Fourth letter = programmable functions Fifth letter = antitachycardia functions Chamber Paced Single-chamber: o Sense and pace either the atrium or ventricle Dual-chamber: o Sense and pace both the atrium and the ventricle Response to Sensing Triggered o Ventricular pacing is triggered by sensing of an atrial event Inhibited o Sensing of a spontaneous ventricular depolarization inhibits ventricular pacing New 7/12 NAM Page 2 Two common pacemakers: VVI o Ventricle paced o Ventricle sensed o Pacing inhibited in response to a sensed beat DDD o Both atrium and ventricle paced o Both atrium and ventricle sensed o Ventricle triggered in response to an atrial beat and inhibited in response to a ventricular beat Advantage of Dual Chamber Restores AV synchronous sequence of the heart (atrial kick) Reduces incidence of atrial fibrillation Pacemaker Terminology - Page 262 Ventricular capture o Ventricle has responded to a pacing stimulus o Stimulus artifact (pacing spike) followed by a wide QRS (depolarization of the right ventricle) Pacemaker Terms – Page 263-264 Figure 10-9 Native Beat (Intrinsic beat) o Produced by patient’s own electrical conduction system Fusion Beat o Pacemaker fires at same time that the patient’s normal electrical impulse has activated the ventricles o Fusion beats are normal and appear as a blend of the two beats Pacemaker Terms – Page 263 Automatic Interval o Heart rate at which pacemaker is set o Measured from one pacing spike to the next consecutive pacing spike Escape Interval o Interval between intrinsic beat and next paced beat o Should be same as automatic interval Paced Rhythm – Page 265 Figure 10-12 Pacemaker Rhythm o Occurs when the heart’s rhythm is completely pacemaker induced o No intrinsic patient beats are seen Ventricular Paced Rhythm Characteristics/Criteria: o Rhythm: Atrial – none Ventricular: Regular o Rate: Automatic interval (spike to spike) o P Waves: none PRI: none present o QRS: wide and bizarre as a ventricular originated beat o QT: Normal for rate Pacemaker Malfunctions Failure to pace (fire) o Pacemaker fails to deliver the pacing stimulus o No pacing artifact even though the patient needs pacing o May result in hypotension, bradycardia New 7/12 NAM Page 3 Pacemaker Malfunctions Page 266 Figure 10-14 Failure to capture o Failure of the heart to respond (depolarize) following a pacing stimulus o Reflected as pacing spike that occurs on time, not followed by a QRS complex o May cause bradycardia, hypotension, fatigue Causes o MA output is too low o Lead dislodgement o Catheter tip in infarcted tissue o Fractured lead wire o Electrolyte imbalances o Faulty connections Interventions o Increase mA till consistent capture is is achieved o Place patient on left side o Physician may need to reposition pacing catheter o Correct electrolytes Undersensing – Page 266 Figure 10-15 Occurs when the pulse generator does not sense the patient’s intrinsic beats Reflected on ECG by a pacing spike that occurs earlier than it should after a native or paced beat May cause palpitations, skipped beats, VT/VF Causes: o Pacing catheter out of place or lying in infarcted tissue o Sensitivity set too low o Pacemaker set on asynchronous (fixed rate) mode Interventions o Increase sensitivity o Place patient on left side Pacemaker Malfunctions – Page 267 Oversensing o Inappropriate sensing of extraneous electrical signals leading to triggering or inhibiting stimulus output – may be sensing tall P or T wave by mistake o Causes pauses on ECG o Pacemaker paces at a slower rate than set o No paced beats Clinical Management Patient/Family Teaching Always check the effect of pacer malfunction on patient’s hemodynamic status A 12-lead ECG should be done to assess pacemaker malfunction Wear gloves when handling wires from the temporary or epicardial pacer to reduce incidence of microshock Clinical Management – Temporary Transvenous Pacing Assess that all connections from the lead connections to the bridging cable to the pulse generator are secure Insulate metal tips of lead wires when they are not in use with a finger cot. Keep negative and positive ends separated Maintain asepsis at all insertion sites Keep dressings dry New 7/12 NAM Page 4 Biventricular Pacing CRT: Cardiac Resynchronization Therapy Paces both of the heart’s ventricles Used to coordinate heart’s 4 chambers to pump more effectively Implantable Cardioverter-Defibrillator (ICD) An implanted electronic device used in treatment of life-threatening tachydysrhythmias Generator is placed in subcutaneous tissue, usually in pectoral region ICD System Components Can Leads Programmer Software ICD System Functions Reliably sense cardiac signals in all rhythms (VT, VF, SR, AF, …) Detect VF, VT, reject SR, SVT Shock or Anti-Tachy pace for VF or VT as appropriate Pace as needed for bradyarrhythmias Store diagnostic data o Episode data o System function data ICD System Continuously monitors and analyzes a patient’s rhythm Initiates low-voltage electrical impulse (0.1 – 0.5 joules) when heart rate becomes very fast (more than 160 – 180) Contains sensing electrodes to recognize dysrhythmias Contains defibrillation electrodes that are in contact with the heart Contains pacing leads Indications for ICD Cardiac arrest caused by VF/VT Spontaneous sustained VT not responsive to drug therapy Hemodynamic compromising VT/VF induced during an electrophysiology study Atrial fibrillation (tachydysrhythmia) THERAPIES Bradycardia pacing Antitachycardia pacing Cardioversion Defibrillation THERAPIES Tachyarrhythmia Shock Therapy • • Cardioversion: Synchronous Electrical Shock Defibrillation: Asynchronous Electrical Shock New 7/12 NAM Page 5 Clinical Management Monitor for complications Avoid MRI, TENS, lithotripsy, electrocautery, welders, transformers If defibrillating or cardioverting, place paddles and/or electrodes about two inches away from device Provide patient education Quick Quiz When a VVI pacemaker senses intrinsic ventricular activity, it responds by: 1. Inhibiting its pacing 2. Triggering its pacing 3. Doing nothing Failure to capture is represented on the ECG as: 1. No pacemaker activity 2. Spikes occurring where they shouldn’t 3. A spike without a complex In a synchronous pacemaker, failure to sense is characterized on the ECG by: 1. Lack of a pacemaker spike 2. A pacemaker spike in the presence of native (intrinsic) activity 3. A pacemaker spike without evidence of cardiac stimulation CHECKPOINT !! Analyze the following rhythm strips for appropriate pacemaker function: Page 271 Strip 10-5 Page 274 Strip 10-13 Page 275 Strip 10-16 Page 275 Strip 10-17 New 7/12 NAM Page 6