Managing Pacemaker-Related Complications And Malfunctions In The Emergency Department September 2014 Volume 16, Number 9 Authors Jennifer Martindale, MD Assistant Professor, Department of Emergency Medicine, Kings County Hospital/SUNY Downstate Medical Center, Brooklyn, NY Ian S. deSouza, MD Assistant Professor, Department of Emergency Medicine, Kings County Hospital/SUNY Downstate Medical Center, Brooklyn, NY Abstract The use of implanted pacemaker devices is increasing worldwide, owing to technological advances, new indications, and an aging population. Despite greater experience in implantation and improved device sophistication, patients continue to face complications associated with hardware implantation and device malfunction. This review summarizes current indications for permanent pacing, reviews epidemiologic data relevant to implant complications, and describes a clinical approach to the patient with potential pacing malfunction. The electrocardiographic diagnosis of hyperkalemia and acute myocardial infarction in paced rhythms is also discussed. Potential sources of electromagnetic interference and special considerations pertaining to the cardiac resuscitation of patients with implanted cardiac devices are reviewed. Finally, a basic approach to implanted cardioverter-defibrillator devices (which often accompany pacemaker devices) is presented. Peer Reviewers William J. Brady, MD Professor of Emergency Medicine and Medicine, Chair, Medical Emergency Response Committee, Medical Director, Emergency Management, University of Virginia Medical Center, Charlottesville, VA Joshua M. Kosowsky, MD, FACEP Vice Chair for Clinical Affairs, Brigham and Women’s Hospital, Boston, MA; Assistant Professor, Harvard Medical School, Boston, MA CME Objectives After reading this article, participants should be able to: 1. State the indications for pacemaker placement and cardiac resynchronization therapy. 2. Anticipate the major complications of pacemaker implantation. 3. Develop a differential diagnosis for pacemaker malfunction. 4. Develop an approach to patients who are shocked by their implanted cardiac devices. Prior to beginning this activity, see “Physician CME Information” on the back page. Editor-In-Chief Andy Jagoda, MD, FACEP Professor and Chair, Department of Emergency Medicine, Icahn School of Medicine at Mount Sinai, Medical Director, Mount Sinai Hospital, New York, NY Associate Editor-In-Chief Kaushal Shah, MD, FACEP Associate Professor, Department of Emergency Medicine, Icahn School of Medicine at Mount Sinai, New York, NY Editorial Board William J. Brady, MD Professor of Emergency Medicine and Medicine, Chair, Medical Emergency Response Committee, Medical Director, Emergency Management, University of Virginia Medical Center, Charlottesville, VA Mark Clark, MD Assistant Professor of Emergency Medicine, Program Director, Emergency Medicine Residency, Mount Sinai Saint Luke's, Mount Sinai Roosevelt, New York, NY Peter DeBlieux, MD Professor of Clinical Medicine, Interim Public Hospital Director of Emergency Medicine Services, Louisiana State University Health Science Center, New Orleans, LA Nicholas Genes, MD, PhD Assistant Professor, Department of Emergency Medicine, Icahn School of Medicine at Mount Sinai, New York, NY Michael A. Gibbs, MD, FACEP Professor and Chair, Department of Emergency Medicine, Carolinas Medical Center, University of North Carolina School of Medicine, Chapel Hill, NC Steven A. Godwin, MD, FACEP Professor and Chair, Department of Emergency Medicine, Assistant Dean, Simulation Education, University of Florida COMJacksonville, Jacksonville, FL Gregory L. Henry, MD, FACEP Clinical Professor, Department of Emergency Medicine, University of Michigan Medical School; CEO, Medical Practice Risk Assessment, Inc., Ann Arbor, MI John M. Howell, MD, FACEP Clinical Professor of Emergency Medicine, George Washington University, Washington, DC; Director of Academic Affairs, Best Practices, Inc, Inova Fairfax Hospital, Falls Church, VA Charles V. Pollack, Jr., MA, MD, Scott Silvers, MD, FACEP Chair, Department of Emergency FACEP Medicine, Mayo Clinic, Jacksonville, FL Professor and Chair, Department of Emergency Medicine, Pennsylvania Corey M. Slovis, MD, FACP, FACEP Hospital, Perelman School of Professor and Chair, Department Medicine, University of Pennsylvania, of Emergency Medicine, Vanderbilt Philadelphia, PA University Medical Center, Nashville, Michael S. Radeos, MD, MPH TN Assistant Professor of Emergency Stephen H. Thomas, MD, MPH Medicine, Weill Medical College George Kaiser Family Foundation of Cornell University, New York; Professor & Chair, Department of Research Director, Department of Emergency Medicine, University of Emergency Medicine, New York Oklahoma School of Community Hospital Queens, Flushing, NY Medicine, Tulsa, OK Ali S. Raja, MD, MBA, MPH Ron M. Walls, MD Director of Network Operations and Professor and Chair, Department of Business Development, Department Emergency Medicine, Brigham and of Emergency Medicine, Brigham Women’s Hospital, Harvard Medical and Women’s Hospital; Assistant School, Boston, MA Professor, Harvard Medical School, Boston, MA Scott D. Weingart, MD, FCCM Associate Professor of Emergency Robert L. Rogers, MD, FACEP, Medicine, Director, Division of FAAEM, FACP ED Critical Care, Icahn School of Assistant Professor of Emergency Medicine at Mount Sinai, New Medicine, The University of York, NY Maryland School of Medicine, Baltimore, MD Shkelzen Hoxhaj, MD, MPH, MBA Chief of Emergency Medicine, Baylor College of Medicine, Houston, TX Alfred Sacchetti, MD, FACEP Assistant Clinical Professor, Eric Legome, MD Department of Emergency Medicine, Chief of Emergency Medicine, Thomas Jefferson University, King’s County Hospital; Professor of Philadelphia, PA Clinical Emergency Medicine, SUNY Downstate College of Medicine, Robert Schiller, MD Brooklyn, NY Chair, Department of Family Medicine, Beth Israel Medical Keith A. Marill, MD Center; Senior Faculty, Family Research Faculty, Department of Medicine and Community Health, Emergency Medicine, University of Icahn School of Medicine at Mount Pittsburgh Medical Center, Pittsburgh, Sinai, New York, NY PA Senior Research Editors James Damilini, PharmD, BCPS Clinical Pharmacist, Emergency Room, St. Joseph’s Hospital and Medical Center, Phoenix, AZ Research Editor Michael Guthrie, MD Emergency Medicine Residency, Icahn School of Medicine at Mount Sinai, New York, NY Federica Stella, MD Emergency Medicine Residency, Giovani e Paolo Hospital in Venice, University of Padua, Italy International Editors Peter Cameron, MD Academic Director, The Alfred Emergency and Trauma Centre, Monash University, Melbourne, Australia Giorgio Carbone, MD Chief, Department of Emergency Medicine Ospedale Gradenigo, Torino, Italy Amin Antoine Kazzi, MD, FAAEM Associate Professor and Vice Chair, Department of Emergency Medicine, University of California, Irvine; American University, Beirut, Lebanon Hugo Peralta, MD Chair of Emergency Services, Hospital Italiano, Buenos Aires, Argentina Dhanadol Rojanasarntikul, MD Attending Physician, Emergency Joseph D. Toscano, MD Medicine, King Chulalongkorn Chairman, Department of Emergency Memorial Hospital, Thai Red Cross, Medicine, San Ramon Regional Thailand; Faculty of Medicine, Medical Center, San Ramon, CA Chulalongkorn University, Thailand Suzanne Y.G. Peeters, MD Emergency Medicine Residency Director, Haga Teaching Hospital, The Hague, The Netherlands failure have translated into an increasing number of biventricular pacing systems with integrated ICD technology. While pacemakers have become more complex and sophisticated, patients continue to present to the emergency department (ED) with symptoms related to device malfunctions. In addition, special considerations with regard to implanted pacemakers are relevant to the diagnostic workup of comorbid illnesses. Case Presentations An ill-appearing 74-year-old woman is brought in on a stretcher by EMS personnel who broke down her door to find her slumped on her couch. Her temperature is normal, her blood pressure is 86/40 mm Hg, and her heart rate is 42 beats/min. On the cardiac monitor, you see extremely wide QRS complexes following pacer artifacts and intermittent pauses where pacing artifacts appear alone. A prior ECG shows a paced rhythm and narrower QRS complexes. What are your critical actions? A 38-year-old woman who fainted in the grocery store is wheeled into the last unoccupied room in your ED with a cardiac monitor. She has cardiac sarcoidosis and has a DDD pacemaker that was implanted 6 months ago. She tells you that this is the third time she has fainted in the past 24 hours. She is normotensive and her heart rate is 48 beats/min. The cardiac monitor shows pacing artifacts that are completely dissociated from QRS complexes. What steps will you take to diagnose the underlying cause of this patient’s recurrent syncopal episodes? Epidemiology Implanted pacemakers are increasingly prevalent in the United States.1 Approximately 370,000 pacemakers are placed annually,2 the most common indication for which is sinus node disease.3 Regardless of the indication for pacing, dual-chamber pacemakers have been adopted as the technology of choice. However, since United States Food and Drug Administration (FDA) approval in 2001, CRT devices are being implanted more frequently.4 Most of these are combined with defibrillation technology, and CRT-defibrillator (CRT-D) devices have come to comprise approximately 40% of all pacemakers in the United States.5 Compared with the general population, patients receiving implanted devices are older (mean age of implantation of 75.6 years) and carry a higher burden of age-adjusted comorbid illness.1,6 Introduction Cardiac pacemakers have evolved from singlechamber devices that deliver a fixed pacing rate to multichamber systems with programmable features that preserve and more closely mimic normal cardiac electrophysiology. Preservation of atrioventricular synchrony, physiologic heart rate, and interventricular synchrony are significant improvements that have been made possible with dual-chamber, rate-adaptive, and biventricular pacing systems. For patients, improved quality of life, increased exercise tolerance, and decreased disease progression are among the clinical outcomes driving new indications for pacemaker implantation and cardiac resynchronization therapy (CRT). At the same time, mortality benefits associated with implanted cardioverterdefibrillator devices (ICDs) in patients with heart Critical Appraisal Of The Literature PubMed and the Cochrane Database of Systematic Reviews were searched for English-language articles pertaining to the management of patients with implanted pacemakers and cardiac devices published between January 1, 1990 and February 21, 2014. Search terms included permanent pacemakers and the thematic topics of pacemaker complications, cardiac resynchronization therapy, pacemaker malfunction, and electromagnetic pacemaker interfer- Table 1. American College of Cardiology/American Heart Association Classification Of Recommendations And Level Of Evidence Certainty of Treatment Effect A Recommendations derived from multiple randomized clinical trials or meta-analyses B Data derived from a single randomized clinical trial or nonrandomized studies C Only consensus opinion of experts, cases studies, or standard of care Size of Treatment Effect Class I Conditions for which there is evidence and/or general agreement that the test is useful and effective Class II Conditions for which there is conflicting evidence and/or a divergence of opinion about the usefulness or efficacy of performing the test Class IIa Weight of evidence/opinion is in favor of usefulness or efficacy Class IIb Usefulness or efficacy is less well established by evidence/opinion Class III Conditions for which there is evidence and/or general agreement that the test is not useful or effective and in some cases may be harmful Copyright © 2014 EB Medicine. All rights reserved. 2 www.ebmedicine.net • September 2014 ence. The bibliographies of review articles were hand-searched and the level of evidence graded according to the guidelines from the American Heart Association Task Force on Practice Guidelines. (See Table 1, page 2.) Guidelines published by the American College of Cardiology Foundation/ American Heart Association (ACCF/AHA), the Heart Rhythm Society (HRS), the European Heart Rhythm Association (EHRA), and the European Society of Cardiology (ESC) were also reviewed.7-11 the left ventricular lead depends on cardiac venous anatomy, avoidance of phrenic nerve stimulation, and location of myocardial scar. Pacemaker Modes A 5-letter code has been adopted to identify different types of pacemakers. (See Table 2.) The first letter corresponds to the lead location. The second letter corresponds to the chamber (or chambers) being sensed by the pacemaker. The third letter indicates the response of the pacemaker to a sensed impulse. The fourth letter indicates the pacemaker’s programmability and capacity for rate modulation. A pacemaker capable of rate modulation (R) is also capable of communicating with external telemetry (C) and multiprogrammability (M), in which 3 or more variables can be modified. Pacemakers capable of rate modulation mimic the rate response of a normal sinus mode when physiological changes are detected by sensors for minute ventilation, QT interval, or stroke volume. In 2002, the code was revised to include a fifth letter indicating whether multisite pacing is present in the atrium, the ventricle, or both.16 Emergency clinicians are most likely to encounter pacers that are AAIR, VVIR, DDD, and DDDR.17 Review Of Normal Pacemaker Function Components And Lead Placement The basic components of a pacemaker are the pulse generator and the lead. The pulse generator consists of hardware, programmable software, and a battery with a 5- to 10-year lifespan. The longevity of any battery depends on the proportion of paced and sensed events, stimulation impedance, and programmed output. Lead components include a conductor, electrode, fixation mechanism, terminal connector pin, and insulation. The connector pin connects the lead to the generator and must be well seated in the connector block of the generator for the device to function. Atrial leads are most commonly positioned in the right atrial appendage. Selective site pacing in the right atrial septum may be chosen to suppress atrial fibrillation. Right ventricular leads are traditionally placed in the right ventricular apex; however, pacing from this site has been shown to result in ventricular dyssynchrony and subsequent impairment in left ventricular function.12-15 Right ventricular leads may be placed in the right ventricular outflow tract (RVOT) or interventricular septum to minimize these effects, but evidence to support improved outcomes from selective site pacing in the RVOT is contested. The left ventricular lead of a biventricular pacing system is placed through the coronary sinus into one of its tributaries, in an epicardial location along the posterior or lateral free wall of the left ventricle. The final position of Pacemaker Indications The indications for permanent pacing are continually amended and are categorized by type of underlying conduction disorder or disease process. The ACC/AHA/HRS guidelines for the use of cardiac pacemakers, ICDs, and CRT were most recently published in 2008,8 with a focused revision on the indications for CRT published in 2013.7 In general, class I recommendations apply in cases of symptomatic impaired conduction. With regard to asymptomatic conduction abnormalities, class I recommendations apply in disorders or diseases that are likely to progress. For more information about the classes and levels of evidence for the indications for permanent pacing, see Figure 1, page 4 for the URL and QR code to access the PDF of the full guidelines. Table 2. The North American Society For Pacing And Electrophysiology/British Pacing And Electrophysiology Group Generic Pacemaker Code I II III IV V Chamber(s) paced Chamber(s) sensed Response to sensing Programmability, rate modulation Multisite pacing A = Atrium V = Ventricle D = Dual (A + V) O = None A = Atrium V = Ventricle D = Dual (A + V) O = None T = Triggered I = Inhibited D = Dual (triggered + inhibited) O = None P = Simple programmable M = Multiprogrammable C = Communicating R = Rate modulation O = None A = Atrium V = Ventricle D = Dual (A + V) O = None Reprinted from Emergency Medicine Clinics Of North America. Volume 24, Issue 1. Theodore C. Chan, Taylor Y. Cardall. Electronic Pacemakers. Pages 179-194, Copyright 2006, with permission from Elsevier. September 2014 • www.ebmedicine.net 3 Reprints: www.ebmedicine.net/empissues of the infarcted area. Patients with atrioventricular block and impaired intraventricular conduction after myocardial infarction (MI) have a high likelihood of mortality that most likely relates to extensive myocardial necrosis rather than impaired conduction. Patients with anterior infarction and impaired systolic dysfunction should be considered for CRT. Sinus Node Dysfunction Pacemakers are placed in patients with symptomatic sinus node dysfunction in order to reduce the frequency of syncopal episodes and the risk of stroke and atrial fibrillation. Sinus node dysfunction includes sinus bradycardia, sinus arrest, sinoatrial block, chronotropic incompetence, atrial tachycardia, and atrial fibrillation. The preferred pacemaker mode in patients with sinus node dysfunction is DDDR. Dual-chamber pacing in patients with sick sinus dysfunction has been associated with reduced pacemaker syndrome symptoms, reduced atrial fibrillation, and improved exercise capacity compared with single-chamber pacing.18,19 Pacemaker syndrome manifests as intolerance to ventricular pacing in the absence of atrioventricular synchrony and will be discussed in more detail on page 9. In patients with preserved atrioventricular conduction, AAIR pacing may be a cost-effective choice to minimize unnecessary ventricular pacing. Subsequent atrioventricular block is a concern factored into the choice of device and develops at an annual rate of 1.4% after atrial pacemaker implantation.20 Cardiac Resynchronization Therapy Systolic heart failure is accompanied by prolonged intraventricular conduction in approximately 30% of patients.23 Delayed conduction results in further reduction of systolic function, ventricular remodeling, and functional mitral regurgitation. Ventricular dyssynchrony, defined by a prolonged QRS duration, is associated with increased mortality in patients with heart failure.24 CRT improves ventricular dyssynchrony by pacing from leads in both right and left ventricles. In patients with heart failure, improved electromechanical synchrony has been shown to induce reverse ventricular remodeling, decrease mitral regurgitation, decrease neurohormonal activation, and improve peak oxygen consumption. Results from several landmark studies have led to the established role of resynchronization therapy in treating heart failure. Initial randomized studies were designed to focus on functional outcomes of CRT. They demonstrated improved quality of life, New York Heart Association (NYHA) functional classifications for heart failure, exercise tolerance, and left ventricular ejection fraction.5,25,26 Later studies demonstrated improved morbidity (unplanned hospitalizations) and mortality outcomes in CRT patients with NYHA Class II,27-29 III, or IV30,31 heart failure. However, improved outcomes are not universal; approximately 10% to 40% of patients are nonresponders, depending on how response is defined functionally.32 Research aimed at defining the subgroups most likely to benefit from CRT has been incorporated into the most recent indications for CRT.7 (See Table 3, page 5.) According to a meta-analysis, the morbidity and mortality benefits of CRT are limited to patients with QRS duration ≥ 150 msec.33 Moreover, patients with left bundle branch block (LBBB) morphology have also been Acquired Atrioventricular Block Pacing is recommended for patients with thirddegree block and type II second-degree atrioventricular block to reduce syncope and improve survival. The indication for pacemaker placement in type I second-degree atrioventricular block is more controversial. Guidelines are based on whether the patient is symptomatic and on the exact location of the block as it relates to the likelihood that the disease will progress to complete heart block. (See Figure 1.) Dual-chamber pacing is recommended over ventricular pacing to reduce the incidence of pacemaker syndrome. Chronic Bifascicular Or Trifascicular Block Syncope is common in patients with bifascicular and trifascicular block, but the risk of progression to complete heart block and death varies based on the presence of atrioventricular block. In the absence of atrioventricular block, the indication for pacemaker placement depends on the presence of prolonged His-ventricular (H-V) intervals during electrophysiologic study. Figure 1. Indications For Permanent Pacing After Acute Myocardial Infarction New-onset atrioventricular block in the acute setting of ST-segment elevation myocardial infarction (STEMI) usually resolves spontaneously within days.21 Temporary pacing is recommended in the setting of acute STEMI for patients with symptomatic bradycardia that is unresponsive to medical treatment (class I, level of evidence [LOE] C).22 In the acute phase of inferior STEMI, atrioventricular block is more often mediated by increased vagal tone; in anterior STEMI, the development of atrioventricular or intraventricular block is related to the extent Copyright © 2014 EB Medicine. All rights reserved. Scan the QR code above with your smartphone or tablet, or go to: https://my.americanheart.org/idc/groups/ahaecc-internal/@wcm/@ sop/documents/downloadable/ucm_423809.pdf 4 www.ebmedicine.net • September 2014 shown to have greater benefit than those with nonLBBB conduction abnormalities, including right bundle branch block (RBBB) and nonspecific intraventricular conduction delay.34 rogation can be performed. Patients should be asked for dates of implantation and revision; if and when programming changes were made; and about a history of exposure to any electromagnetic interference, blunt trauma, or environmental extremes. A complete list of medications should be obtained, as some drugs can alter pacing thresholds. Pacemaker-related symptoms should be reviewed with the patient. These include light-headedness, palpitations, syncope, sense of slowed or racing heart, and extracardiac muscle twitching. Neurocardiogenic Syncope And Carotid Sinus Syndrome The role of pacemaker therapy in neurocardiogenic syncope is controversial, as the evidence supporting its use is mixed. Open-label trials have demonstrated a decrease in syncopal events,35-37 while 2 double-blinded placebo-controlled trials (total of 129 patients) have failed to demonstrate any benefit.38,39 The alleged benefit of pacemaker therapy in carotid sinus syncope is based solely upon observational studies40-42 and open-label randomized trials.43,44 In a double-blind, placebo-controlled crossover trial that included 34 patients, there was no statistically significant difference in unexplained falls between those with the pacemaker turned on and those with it turned off.45 However, this study was considered underpowered because of the study’s high attrition rate. Physical Examination In addition to a standard thorough cardiac examination, the implantation site itself should be examined for erythema, bruising, swelling, and tenderness. Signs of swelling in the neck or ipsilateral arm should raise suspicion for venous thrombosis. Moving the patient’s ipsilateral arm or having the patient perform isometric exercise may help to identify intermittent dysfunction. Diagnostic Studies Emergency Department Evaluation Electrocardiogram The 12-lead ECG is important for determining whether depolarization is intrinsic or paced. Bipolar leads produce smaller pacing artifacts that may be imperceptible in some leads and easily recognized in others. The absence of a pacer artifact before P waves or QRS waves indicates that depolarization is intrinsic. Pacing artifacts preceding depolarizations indicate successful pacing and capture. History Even with the availability of advanced device-monitoring capabilities, obtaining the history from a patient with an implanted device remains important. This includes cardiac history as well as noncardiac medical history and, perhaps most importantly, the original indication for pacemaker placement. The type of pacemaker should be identified so that appropriate inter- Table 3. Indications For Cardiac Resynchronization Therapy7 Class Indication LOE I LVEF ≤ 35%; sinus rhythm; LBBB with a QRS duration ≥ 150 msec; and NYHA class II, III, or ambulatory class IV symptoms on goal-GDMT B LVEF ≤ 35%; sinus rhythm; LBBB with a QRS duration 120-149 msec; and NYHA class II, III, or ambulatory class IV symptoms on GDMT B LVEF ≤ 35%; sinus rhythm; a non-LBBB pattern with a QRS duration ≥ 150 msec; and NYHA class III/ambulatory class IV symptoms on GDMT A Atrial fibrillation and LVEF ≤ 35% on GDMT if: (a) patient requires ventricular pacing or otherwise meets CRT criteria, and (b) AV nodal ablation or pharmacologic rate control will allow near-100% ventricular pacing with CRT B LVEF ≤ 35%, undergoing new placement or replacement of a device with anticipated requirement for significant (> 40%) ventricular pacing C LVEF ≤ 30%, ischemic etiology of heart failure, sinus rhythm, LBBB with QRS duration ≥ 150 msec, and NYHA class I symptoms on GDMT C LVEF ≤ 35%, sinus rhythm, a non-LBBB pattern with QRS duration 120-149 msec, and NYHA class III/ambulatory class IV on GDMT B LVEF ≤ 35%, sinus rhythm, a non-LBBB pattern with a QRS duration ≥ 150 msec, and NYHA class II symptoms on GDMT B NYHA class I or II symptoms and non-LBBB pattern with QRS duration < 150 msec B Patients whose comorbidities and/or frailty limit survival and good functional capacity to less than 1 year C IIa IIb III Abbreviations: AV, atrioventricular; CRT, cardiac resynchronization therapy; GDMT, goal-directed medical therapy; LBBB, left bundle branch block; LOE, level of evidence; LVEF, left ventricular ejection fraction; NYHA, New York Heart Association. September 2014 • www.ebmedicine.net 5 Reprints: www.ebmedicine.net/empissues The 12-lead ECG is also essential for evaluating the morphology of paced QRS complexes. Leads positioned in the right ventricular apex produce a LBBB pattern with appropriate discordance in the ST segments and T waves. A change in this morphology may indicate lead dislodgement. Specifically, the presence of a new RBBB pattern may indicate inadvertent placement of the right ventricular lead into the left ventricle or septal perforation. A RBBB pattern may also appear in a minority of correctly positioned right ventricle leads.46 Simultaneous depolarization of the left and right ventricles from a biventricular pacing system typically produces a dominant R wave in lead V1.47 Hyperkalemia Delayed intraventricular conduction causes paced QRS complexes in the hyperkalemic patient to widen. Severe hyperkalemia may result in a sine wave following each pacing artifact. (See Figure 2.) Severe hyperkalemia may cause ECG changes to be mistaken for pacemaker malfunction. The pacing threshold is elevated in hyperkalemia, leading to increased latency, intermittent capture (including Wenckebach phenomenon), or continuous loss of capture,48,49 as well as loss of sensing.50 Latency or failure to capture may be temporarily overcome by programming pacemaker outputs to maximum voltage, but definitive treatment for hyperkalemia should be instituted without delay. from the GUSTO-1 (Global Utilization of Streptokinase and Tissue Plasminogen Activator for Occluded Coronary Arteries) trial population, that concordant ST elevation, concordant ST depression in leads V1 through V3, and > 5 mm of discordance in leads with predominantly negative QRS complexes were insensitive but specific findings for acute MI.51 (See Figure 3.) In a retrospective review of 57 ECGs of ventricular-paced patients diagnosed with acute MI, discordant ST elevation > 5 mm was the most specific finding (99% specificity; 95% confidence interval [CI], 93%-99%).52 However, discordant ST elevation can be > 5 mm in ventricularpaced rhythms in the absence of MI when negative QRS amplitudes in leads V1 through V3 are large.53 As with application of the Sgarbossa rule to LBBB, an elevated ratio of ST-segment elevation to S-wave depth may be more specific for MI than the absolute measurement of 5 mm.54 Acute anterior MI in a patient with biventricular pacing with a dominant R wave in leads V1 and V2 presents with standard ST elevations across the precordium.55 Radiology A basic knowledge of the normal and abnormal radiographic appearances of cardiac conduction devices and their components can be important in diagnosing complications associated with implantation and device integrity. (See Table 4.) Both posterior-anterior and lateral views should Acute Myocardial Infarction The pattern of depolarization and repolarization in ventricular-paced rhythms may obfuscate the electrocardiographic diagnosis of acute myocardial infarction (MI). Sgarbossa found in a retrospective review of 17 ventricular-paced patients selected Figure 3. Electrocardiogram Tracing Of Myocardial Infarction, Paced Rhythm Figure 2. Electrocardiogram Tracing Of Paced Rhythms In The Setting Of Severe Hyperkalemia Arrows point to examples of exaggerated discordance. Table 4. Radiographic Assessment Of The Permanent Pacemaker/Implantable Cardioverter-Defibrillator • • • • • • Arrows point out pacing artifacts. From Jennifer Martindale, David F. M. Brown. Rapid Interpretation of ECGs in Emergency Medicine: A Visual Guide. Reprinted with permission from Lippincott Williams & Wilkins/Wolters Kluwer, 2012. Copyright © 2014 EB Medicine. All rights reserved. 6 Obtain posterior-anterior and lateral views. Locate pulse generator and compare with previous chest x-ray. Identify manufacturer, if needed. Ensure connector pin is seated deep within header. Identify number of leads and lead location. Confirm integrity of each lead. www.ebmedicine.net • September 2014 leads. ICD leads may contain a single high-voltage coil in the right ventricle or may have an additional coil in the superior vena cava. Complications of implantation that can be diagnosed on plain radiographs include pneumothorax, hemopneumothorax, myocardial perforation, and improper seating of a connector pin. Winding of a lead around the pulse generator may be indicative of twiddler syndrome. (See page 8.) be obtained to confirm correct lead position, and radiographs should be compared with any prior studies. In addition, the manufacturer and model on the pulse generator can often be identified from the chest radiograph. Lead location and integrity should be confirmed systematically. On a lateral view, the right atrial lead has a J-shaped appearance as it enters the right atrium and curves upward and anteriorly into the right atrial appendage. The J portion of the lead is medial on the posteror-anterior projection and anterior on the lateral projection. Atrial leads may also be placed in the interatrial septum. Right ventricular leads should point downward, with the tip between the left of the vertebral column and the cardiac apex on a posterior-anterior radiograph. The lateral radiograph is required to confirm the anterior and caudal course of a correctly positioned right ventricular lead. Leads placed in the left ventricle or coronary sinus should course posteriorly on a lateral radiograph. Magnification may be required to appreciate small defects in leads. A common site of lead fracture is between the first rib and clavicle (subclavian crush syndrome).56 Extraneous wires present on chest radiographs may represent abandoned leads. Malfunctioning leads are disconnected from the generator, capped, and left in place to avoid the risks associated with their extraction. The key to differentiating a permanent pacemaker from an ICD is the presence of shock coils (see Figure 4), which appear as thick bands on the Device Interrogation Noninvasive testing allows acquisition of measured data, interrogation of programmed parameters, and analysis of stored arrhythmias. Basic parameters include pacing mode, and lower and upper tracking rates can be shown during device interrogation. Measurements of the battery’s voltage can help to determine whether battery failure might be the cause of pacemaker malfunction. With some devices, the estimated remaining life of the battery can be determined. Measurements of lead status can help to identify lead failures (such as a break in insulation or conductor fracture). A sudden drop in impedance, for example, might indicate an insulation break. The pacemaker is also capable of telemetry. The relative time spent between paced and native activity in each chamber is recorded. Heart rate measurements are stored and a graphical representation of time spent in different heart rate intervals can be displayed as a histogram. The date, time, and duration of arrhythmia episodes and their associated electrograms are also stored by implanted devices. This capability makes implanted pacemakers valuable diagnostic tools for ruling out tachyarrhythmia as the underlying cause for a patient with recent symptoms. (See Table 5.) Figure 4. Radiograph Of Dual-Chamber, Biventricular Pacemaker/Implantable Cardioverter-Defibrillator System Table 5. Methods To Identify Device Manufacturer Abbreviations: LV, left ventricle; RA, right atrium; RV, right ventricle. September 2014 • www.ebmedicine.net 7 Ask patient for pocket card This will specify the type of pacemaker, type of leads, manufacturer, date of implantation, paced rate, as well as contact information for the patient’s primary physician. Place magnet over pulse generator This will dentify magnet-induced pacing rate (specific to each model). Call manufacturer hotlines Medtronic, Inc.: 1-800-328-2518 The Boston Scientific Corporation: 1-800-CARDIAC St. Jude Medical, Inc.: 1-800-722-3774 Obtain chest x-ray Look for manufacturer code on pulse generator. Reprints: www.ebmedicine.net/empissues Lead Dislodgement Implant-Related Complications A relatively more common lead-related issue is lead dislodgment. Dislodgement occurs in 2.4% of implantations, and may result in inadequate sensing or failure to capture.62 This issue usually occurs very early post procedure but may occur as late as 3 months afterward.57,63 Dislodgement is often suspected when undersensing or failure to capture is detected on follow-up telemetry. Subtle cases may be detectable only upon device interrogation demonstrating changes in pacing thresholds, while more pronounced cases can be visualized by chest radiography.63 Rarely, temporary external or transvenous pacing may be needed until repositioning or replacement of the leads can be performed. Lead dislodgement can sometimes result in lead migration or malposition. Migration of a lead through an occult septal defect can result in pacing of the left ventricle rather than the right. Malposition near the phrenic nerve may stimulate the nerve or the diaphragm directly and cause chronic singultus (hiccoughs).63,64 Lastly, lead dislodgement can cause myocardial perforation in 0.3% of cases in the days to weeks following implantation, resulting in pericardial effusion and, potentially, tamponade.62,64 See Table 6 for a list of common complications arising from the surgical implantation procedure itself or problems with device components. Hematoma Postprocedural bleeding and generator pocket hematoma are relatively common complications that may be encountered immediately after implantation. Bleeding may occur from within the soft tissue of the pocket itself or it may originate from the venous insertion site. Hematomas are managed conservatively in all but 1% to 2% of cases, in which expanding, tense, or painful collections require surgical evacuation and hemostasis.57 Patients who are anticoagulated with clopidogrel or heparin are at greater risk of developing a pocket hematoma.58 Infection As with any surgical procedure, device implantation carries a small risk of surgical site infection (0.2% in one retrospective study).59 Examination of the generator pocket itself may reveal signs of inflammation, dehiscence, or erosion. However, even in the absence of such findings, infection of an implanted device should still be considered when there is fever without an identified source. If infection is confirmed, device removal with lead extraction is typically required, followed by antibiotics and reimplantation at a later date.59,60 A subset of cardiac device infections will result in cardiac device infective endocarditis (CDIE). CDIE is a serious infection occurring at rate of 0.4%.59 Clinical findings, vegetations found on echocardiogram, and positive blood and lead tip cultures are criteria used for diagnosis.59,60 Staphylococci are the predominant bacteria involved.61 CDIE is also treated by removal of the device, prolonged intravenous antibiotics, and possible reimplantation on the contralateral side at a later date.59,60 Twiddler Syndrome Lead dislodgement can also be caused by a patient’s own compulsive manipulation of the pacemaker generator, causing the leads to become retracted and coiled around the manipulated pulse generator. This is referred to as twiddler syndrome, and the diagnosis is made by the radiographic appearance of pacing leads coiled about the generator. Woven Dacron® pouches have been used to prevent migration and flipping of the pulse generator, but subpectoral implantation may be required to prevent recurrent device rotation.65,66 Table 6. Implant-Related Complications Complication Diagnosis Method Management Pocket hematoma Examination Conservative (rarely, surgical evacuation) Pocket infection Examination Explantation, antibiotics Endocarditis Echocardiography, blood cultures, TEE Explantation, antibiotics Lead dislodgement Chest radiography, telemetry/ECG Repositioning or replacement Venous thrombosis Sonography/venography Anticoagulation, angioplasty, or surgery Pneumothorax, hemothorax Chest radiography Conservative or aspiration/thoracostomy Tricuspid regurgitation Echocardiography Lead extraction and/or annuloplasty/valve replacement Cardiac perforation Computed tomography, echocardiography Pericardiocentesis, lead withdrawal and repositioning Abbreviations: ECG, electrocardiogram; TEE, transesophageal echocardiogram. Copyright © 2014 EB Medicine. All rights reserved. 8 www.ebmedicine.net • September 2014 Venous Thrombosis Pacemaker Syndrome Subclavian or brachiocephalic venous thrombosis is a relatively more common occurrence after pacemaker implantation, with reported incidental rates of 2% to 22% from several days up to 9 years postoperatively.67,68 The range of reported rates (both in incidence and time) is a product of how the disease is typically discovered. Venous thrombosis is typically an incidental finding during a preoperative venogram in an asymptomatic patient undergoing lead revision. Thrombosis may be more common in patients who have systemic infection.67 Signs are often absent, but there may be typical manifestations of acute deep venous thrombosis and even superior vena cava syndrome, with the latter occurring in 0.2% of procedures.63 Ultrasonography may not be diagnostic, even in symptomatic cases, and a venogram will be required.68 If thrombosis is found, vascular surgery consultation should be sought. Although it is a relatively common finding, venous thromboses associated with cardiac devices are of questionable clinical importance, as pulmonary embolism is a rare complication in these patients.63,68 Treatments range from heparin/warfarin to percutaneous angioplasty or even open surgery.63,68 Pacemaker syndrome is defined as intolerance to ventricular pacing in the absence of atrioventricular synchrony.70 The Mode Selection Trial (MOST) investigators suggest the following criteria: (1) new or worsened dyspnea, orthopnea, rales, elevated jugular venous pressure, and edema with ventriculoatrial conduction during ventricular pacing; or (2) dizziness, weakness, presyncope or syncope, and a > 20 mm Hg reduction of systolic blood pressure when the patient has VVIR pacing compared with atrial pacing or normal sinus rhythm.70-73 It is hypothesized that ventricular pacing leads to ventriculoatrial contraction with resultant increases in atrial pressures.64,70 Baroreceptor overload then alters vagal tone and there is resultant reduced cardiac output. This reduced output is further complicated by the loss of atrial contribution to preload.64,70 Pacemaker syndrome is usually treated by the restoration of some level of atrioventricular synchrony. This may involve replacing a single-chamber device with a dual-chamber device or reprogramming a dual-chamber device from VVI to DDD function. Pacing System Malfunctions Pneumothorax Pacemakers and ICDs are complex medical devices and, while they are generally regarded as safe and reliable, they are subject to electrical and hardware malfunction. (See Table 7, page 10.) Generator malfunctions occur at a mean annual rate of 4.6 per 1000 implanted pacemakers and 20.7 per 1000 implanted ICDs.74,75 A meta-analysis of prospective device registries demonstrated that rates of generator failure have improved over the past 2 decades, and the most common cause of device failure was battery malfunction.76 The majority of lead failures are caused by insulation defects. Typically, insulation breaks are a late complication, occurring at a median time of 7.5 years (+/- 2.5 y) after implantation.77 Lead problems can be detected by measuring lead impedance. Very high impedance results from an open circuit (conductor fracture), while low impedance implies an insulation defect. Clinical symptoms of device malfunction include syncope, palpitations, dizziness, fatigue, and muscle twitching from extracardiac stimulation. Overall, however, pacemaker malfunction is a rare cause of syncope. In one retrospective study, device malfunction was found to be the cause of syncope in only 8 of 162 patients.78 An assessment of pacing system malfunction should include evaluation of the pulse generator, the leads, programmed settings, and algorithms, as well as any patient status (eg, electrolyte disturbance) that may affect the electrode-myocardial interface. From an electrocardiographic standpoint, pacemaker As in other procedures that involve subclavian venous puncture, device implantation may result in pneumothorax from inadvertent pleural injury. In one study, pneumothorax was the second most common complication overall and occurred in 1.5% of patients.62 Hemothorax may occur due to unintentional arterial puncture or laceration, which can be minimized by proper use of fluoroscopic guidance. Postprocedural pneumothorax and hemothorax are often asymptomatic,63 but some cases may require needle aspiration or tube thoracostomy. Not surprisingly, these complications are less likely with axillary vein cannulation techniques. Tricuspid Regurgitation Tricuspid regurgitation may occur after pacemaker lead placement, although the mechanism of this occurrence is not fully understood. Commonly postulated mechanisms include direct damage to valve leaflets or impairment of valve closure due to scar formation or thrombus. Ventricular asynchrony is also a possible mechanism. Many patients are asymptomatic, but some may present with signs of right-sided congestive heart failure and a systolic murmur that increases with inspiration.69 Diagnosis is made by 3-dimensional transthoracic echocardiography. Leads that are adherent to the valve may require extraction. Unfortunately, the lead extraction procedure itself can lead to progression of tricuspid regurgitation.69 September 2014 • www.ebmedicine.net 9 Reprints: www.ebmedicine.net/empissues malfunction can be categorized as failure to capture, failure to pace, and undersensing. (See Figure 5.) defects. Crosstalk refers to a specific type of oversensing, in which the ventricular lead senses an atrial pacing stimulus, thereby inhibiting ventricular output. A programmed ventricular blanking period following atrial pacing helps to prevent crosstalk. Intrinsic signals that can be oversensed include retrograde P waves, T waves, ventricular afterdepolarizations, and skeletal muscle myopotentials. Myopotentials from the pectoralis, diaphragm,84 and rectus abdominus muscles can all cause oversensing.85 External electromagnetic interference can also result in inappropriate oversensing. Blunt trauma to the pulse generator may cause failure to pace due to output failure.86 Lead fracture, dislodgement, or loose connections can also result in output failure and failure to pace. Undersized pacing artifacts may not be visible on the ECG and can give the appearance of failure to pace. Intrinsic events (such as a premature atrial contractions or premature ventricular contractions) that may not be appreciated on a single telemetry lead can give the false impression of failure to pace when, in fact, there is appropriate inhibition of pacing. Failure To Capture Failure to capture is defined as the delivery of a pacing stimulus without subsequent depolarization. (See Figure 5, view A, page 11.) Failure of depolarization may be functional or pathologic. Functional failure to capture occurs if a pacing impulse is delivered when the myocardium is in a refractory state. Pathologic failure to capture may occur in the setting of myocardial disease, electrolyte disturbance, and antiarrhythmic drugs. The same conditions can also result in latency, which can sometimes be mistaken for failure to capture. Occasionally, isoelectric depolarization can give the appearance of failure to capture, in which case multichannel recording can confirm that depolarization is occurring. Lead dislodgement, perforation, fracture, or an insulation defect can increase the pacing threshold and cause intermittent or persistent failure to capture. Increased pacing threshold can occur several weeks after pacemaker placement due to excessive fibrosis at the electrode-myocardial interface. Capture threshold can also be increased in the setting of hypothyroidism79,80 and infiltrative cardiomyopathies.81 Among the medications that are most likely to affect pacing thresholds are class IC antiarrhythmic drugs.82 Hyperkalemia is also an important cause of failure to capture.83 Failure To Sense A signal is sensed when a myocardial depolarization wave passes up the lead wire and is filtered through the pacemaker. Signals that exceed the sensingthreshold voltage are programmed to inhibit pacing. To prevent double-counting of intrinsic cardiac depolarizations and to prevent sensing of evoked responses as intrinsic depolarizations, pacemakers are programmed to have blanking and refractory periods that immediately follow pacing artifacts and intrinsic depolarizations. Long blanking and refractory periods can cause a functional undersensing of events that occur during these periods. (See Figure 5, view C, page 11.) Undersensing can also occur when there is a Failure To Pace Failure to deliver a stimulus to the heart is referred to as failure to pace. The clearest indication of this is absence of pacing artifact when the intrinsic rate is less than the set lower limit of the device. (See Figure 5, view B, page 11.) Oversensing is the most common cause of failure to pace. Oversensing can occur due to partial lead fracture or insulation Table 7. Causes Of Pacemaker Malfunction, By Category Source of Malfunction Failure To Capture Failure To Sense Failure To Pace Functional failure Pacing output delivered when myocardium functionally refractory Intrinsic event delivered during programmed blanking period or refractory period Appropriate inhibition Pseudofailure Isoelectric depolarization; recording artifacts Not applicable Small bipolar pacing artifacts; ventricular avoidance algorithms Pulse generator Circuit failure; battery depletion Battery depletion Blunt trauma; battery depletion Lead Lead fracture; lead insulation break; lead dislodgement Lead fracture; lead insulation break; lead dislodgement Lead fracture; lead insulation break Electrode-myocardial interface Electrolyte abnormality (hyperkalemia); drug effect (class IC antiarrhythmic drugs); myocardial infarction; defibrillation/cardioversion Intracardiac signals Not applicable Change in intrinsic complex (new bundle branch block) Retrograde P waves; afterdepolarizations; T waves; crosstalk; far-field sensing Extracardiac signals Electromagnetic interference Electromagnetic interference Electromagnetic interference Copyright © 2014 EB Medicine. All rights reserved. 10 www.ebmedicine.net • September 2014 Nonmedical Sources Of Interference change in the morphology or vector of the depolarization pattern. Premature ventricular contractions, bundle branch block, and certain forms of ventricular tachycardia may not exceed the sensing thresholds applied to the previous intrinsic depolarization pattern; however, undersensing is most commonly the result of a break in lead insulation. Undersensing that occurs soon after implantation can be caused by lead dislodgement or perforation. Class IC antiarrhythmic drugs and hyperkalemia may also alter the sensing threshold.50 Finally, undersensing can occur as the pacemaker battery becomes depleted. Common household appliances (such as televisions or microwave ovens) do not interfere with pacemaker function. Potential nonmedical sources of electromagnetic interference include cell phones, security gates, and electronic article surveillance devices. Pacemaker device manufacturers have developed filters that account for common frequencies employed by cell phones. In a study of 679 patients, interaction between cell phones and pacemakers was found in only 0.3% of patients, and there were no clinical manifestations.89 The consequences of the interaction were mostly activation of asynchronous pacing and, rarely, oversensing or inappropriate tracking. Interactions occurred at higher rates with unipolar lead pacemakers and only during the ringing phase when the phone was positioned within 10 cm of the pacemaker. Cell phones have not been shown to interact with ICDs.90 Despite the low incidence of interaction with cardiac devices, the common-sense recommendation for patients should be to position the phone at the ear contralateral to the device. Patients with cardiac devices may trigger alarms at airport security gates when they are screened, but no alteration of function should be expected. In a study of 388 patients screened with handheld metal detectors, there were no observed pacing or sensing abnormalities or reprogramming.91 Interactions with article surveillance devices in retail stores are rare but may occur during prolonged exposure, resulting mostly in asynchronous pacing.92 It should be recommended that patients avoid lingering near these security gates. Additionally, digital music players can cause some interference but do not directly affect the device.93 However, portable headphones often include a powerful magnetic substance that can produce electromagnetic interference and interactions when placed within 3 cm of the cardiac device.94 Electromagnetic Interference For simplification, recent reviews have separated electromagnetic interference sources into nonmedical sources and medical sources.87,88 Figure 5. Electrocardiogram Appearance of Pacemaker Failure Failure to Capture Failure to Pace Failure to Sense Electrocardiographic appearance of different types of pacemaker malfunction. (A) The appearance of appropriately timed pacing artifacts that are not immediately followed by expected P or QRS waves may indicate that a pacemaker's failure to capture is the cause of a patient's symptoms. In this case of atrioventricular pacing, atrial pacing artifacts are followed by inverted P waves, whereas ventricular pacing artifacts are not immediately followed by any QRS waves. The asterisk identifies where QRS waves would be expected to appear in the first 2 beats. The first beat in this strip is a native atrial impulse (appropriately sensed) followed by ventricular pacing artifact that fails to capture. The following beats are atrial-paced with capture and ventricle-paced with failure to capture. (B) Failure of pacing artifacts to appear when expected may indicate that a pacemaker is malfunctioning by failing to pace. This is a ventricular-paced rhythm. The absence of a fifth pacing artifact represents failure to pace, commonly caused by oversensing. (C) The appearance of a pacing artifact at an inappropriate location on the electrocariogram may indicate that the pacemaker is failing to sense native conduction. The asterisk indicates a pacing artifact appearing too soon after a native QRS wave. Medical Sources Of Interference Medical sources of electromagnetic interference include electrocautery (most common), magnetic resonance imaging (MRI), ventricular assist devices, extracorporeal shock wave lithotripsy, therapeutic ionizing radiation, and external cardioversion.87 Physical movement of the cardiac device has not been reported during MRI,88,95 but MRI may affect pacemaker function through 3 mechanisms: (1) the static magnetic field of the MRI will typically close the reed switch and lead to asynchronous pacing; (2) the gradient magnetic fields of the MRI may cause alteration of function or dysrhythmia through direct current effects on pacemaker output circuits;95,96 and (3) the radiofrequency energy of the MRI may be captured by pacemaker leads, which act as antennae to concentrate the energy to myocardium.95,96 In one study, 555 MRI studies were performed From Jennifer Martindale, David F. M. Brown. Rapid Interpretation of ECGs in Emergency Medicine: A Visual Guide. Reprinted with permission from Lippincott Williams & Wilkins/Wolters Kluwer, 2012. September 2014 • www.ebmedicine.net 11 Reprints: www.ebmedicine.net/empissues on 438 patients with cardiac devices.97 Only 3 devices went into a temporary back-up mode, and there were no significant changes in function, leading the authors to conclude that MRI can be done safely in monitored patients with selected devices.97 The AHA and the American College of Radiology (ACR) offer guidelines that state that the decision to perform an MRI on patients with cardiac devices should be patient-specific, with assessment of risks and benefits, and coordination by both cardiology and radiology services.95,98 Therapeutic radiation can produce electromagnetic interference that may directly damage the device circuitry. Manufacturers recommend against radiation therapy if the device is within the treatment field, and there are recommendations for maximum doses. External direct cardioversion or defibrillation may also cause device malfunction. Pacing and sensing function may be affected as a result of associated increases in the myocardial stimulation thresholds, some of which may be permanent.99 Activation of the backup function/rate or mode switch may also occur.99 In patients with bipolar systems, anteriorposterior cardioversion with pads more than 8 cm away from the device will rarely cause adverse effects.100 Contemporary extracorporeal shock wave lithotripsy can be performed safely in patients with cardiac devices, as the rate of interaction is estimated to be less than 1%.101 dysrhythmia.103 After a pacemaker-mediated tachycardia event is successfully treated, the device will need to undergo adjustments to its atrial sensing thresholds. Most modern devices include programming to prevent pacemaker-mediated tachycardia. With such programming, the device may disregard atrial stimuli when atrial rates exceed its own upper limit, or it can mode-switch to discontinue atrial tracking altogether. (See Figure 6, views B and C.) In patients who have older devices with unipolar leads, abnormal, rapid ventricular pacing may occur through a different mechanism. Unipolar systems are more susceptible than bipolar devices to sensing of myopotentials from neighboring muscle activity, and these myopotentials can trigger a rapidly paced rhythm.104 In patients who are ventricular paced, sinus tachycardia (whether physiologic or pathophysiologic) will result in rapid ventricular pacing up to the preset rate limit. Some pacemakers have rate-responsive sensors that can detect changes in physiological parameters (ie, minute ventilation, QT interval, and stroke volume) and permit higher heart rates when dictated by the environment.71 It must be noted that sensory-induced tachycardia will appear identical to pacemaker-mediated tachycardia on an ECG; differentiation between them may be possible only through device interrogation. Management Approach To The Pacemaker Patient With Tachycardia Figure 6. Pacemaker-Mediated Tachycardia Pacemaker-mediated tachycardia (also known as pacemaker re-entrant tachycardia or endless loop tachycardia) is observed exclusively in patients with dual-chamber devices. Pacemaker-mediated tachycardia is similar to other re-entrant dysrhythmias except that the pacemaker itself forms part of the re-entry circuit. An intrinsic premature complex is sensed by the atrial lead of the pacemaker, which responds by generating a ventricular impulse. This ventricular impulse is then conducted retrograde through the atrioventricular node to the atria. The impulse, now in the atria, is sensed by the atrial lead to complete the loop and again trigger the pacemaker to generate a ventricular impulse. The ECG will show a regular wide-QRS tachycardia with a pacing artifact prior to each complex, and retrograde P waves may be seen. (See Figure 6, view A.) This reentrant tachycardia will not exceed the pacemaker’s programmed upper rate limit; however, tachycardia may be significant enough to cause symptoms necessitating emergent treatment. A trial of adenosine may or may not be effective in treating this tachycardia.102 On the other hand, placing a magnet over the pacemaker will definitively terminate this re-entrant Copyright © 2014 EB Medicine. All rights reserved. A Retrograde Ps B Lack of atrial capture Lack of atrial capture C Atrial capture and termination of PMT (A) First 3 asterisks indicate premature ventricular contractions with retrograde P that triggers onset of PMT. (B) Arrow shows lack of atrial capture; a failed attempt at termination by device. (C) Arrow indicates successful atrial capture and termination of PMT. Abbreviation: PMT, pacemaker-mediated tachycardia. Anton Verrijcken, Wim Huybrechts, and Rik Willems. PacemakerMediated Tachycardia With Varying Cycling Length: What is the Mechanism? Europace. 2009, Volume 11, Issue 10, page 14001402. By permission of Oxford University Press. 12 www.ebmedicine.net • September 2014 Magnets turn off the antitachycardia functions of an ICD without affecting its ability to perform backup bradycardia pacing. As such, magnets can be used to terminate inappropriately delivered device shocks. Some ICDs are programmed to play alert tones to indicate high lead impedance, pacing impedance, battery depletion, and delivery of more than 3 shocks. These tones can be replayed to the clinician by applying a magnet over the ICD.106 The Magnet Most pacemakers and ICDs have magnetic reed switches. In pacemakers, placement of a magnet turns off the sensing function, effectively making the pacemaker fire asynchronously at a specified rate. Removal of the magnet causes the pacemaker to switch back to its programmed mode. The magnet response is specific to the manufacturer, model, and programmed response of the device. Each pacemaker has a fixed rate programmed to reflect different levels of battery status. For example, in a Medtronic pacemaker (Medtronic, Inc., Minneapolis, MN), a rate of 85 beats/min reflects beginning-of-life status and 65 beats/min indicates that elective replacement is indicated. At the battery’s end of life, the response of a pacemaker to the magnet is unpredictable.105 If there is no electrocardiographic response to a magnet, there are several possibilities to consider. (See Table 8.) The battery may be at the end of its life or completely depleted, the magnet may need to be repositioned, or the pacemaker may be distanced by adipose tissue. More than 1 magnet may be needed to activate the reed switch of a pulse generator placed in an obese patient. Rarely, a device may also be programmed to ignore magnet application (magnet “OFF” mode). There are several clinical applications for placing a magnet on a pacemaker, one of which is to terminate pacemaker-mediated tachycardia, as the ventriculoatrial limb of the re-entrant circuit is broken when atrial sensing is disabled by a magnet. (See Table 9.) Inappropriate sensing (whether from electromagnetic interference or another cause) will be inhibited by placement of a magnet. Absent pacing artifacts on an ECG can distinguish output failure or lead disruption from oversensing. Some devices are programmed to initiate ECG storage (electrogram mode) under the influence of a magnet. By applying a magnet, the patient can trigger the pacemaker to store ECG data during episodes of palpitations or light-headedness. Advanced Cardiovascular Life Support In Patients With Pacemakers Or Implantable Cardioverter-Defibrillators With few exceptions, Advanced Cardiovascular Life Support (ACLS) procedures should be performed in patients with pacemakers or ICDs in the same manner as in those without the devices. Shocks delivered by an implanted device during a cardiac arrest do not pose a risk to ACLS providers or other equipment.107 External defibrillation is ordinarily not necessary in a patient with a device that has antitachycardia function; however, it should be noted that a dysrhythmia may persist after a normally functioning pacemaker or ICD has completed its programmed sequence of therapy. To conserve battery life, typical sequences are limited to 5 successive shocks. In some cases, the implanted device Table 9. Clinical Applications Of Magnet On Implanted Cardiac Device Permanent Pacemaker • Break pacemaker-mediated tachycardia • Prevent pacing inhibition from electromagnetic interference • Demonstrate ability to pace • Electrocardiograph storage (electrogram mode) Implantable Cardioverter-Defibrillator • Cessation of inappropriate shocks • Inhibit antitachycardia therapy during procedures with electromagnetic interference Table 8. Expected Response To Magnet Placement Over Pacemaker/Implantable CardioverterDefibrillator And Clinical Implications Of Different Magnet Responses Implantable Cardioverter-Defibrillator Permanent Pacemaker Tachyarrhythmia therapy Suspended Not applicable Effect on pacing Not applicable • Asynchronous pacing (DOO, AOO, VOO) l l l Upon magnet removal Tachyarrhythmia therapy restored At 85 beats/min*: Functioning battery at beginning of life At 65 beats/min*: Elective replacement interval No pacing response: Depleted battery; battery at end of life; pacemaker is programmed to ignore magnet for safety (electrogram mode); magnetic field doesn’t reach device (deep placement or obese patient); poorly positioned magnet Sensing function restored *Rates specific for Medtronic pacemaker (Medtronic, Inc., Minneapolis, MN, USA). For an explanation of pacemaker codes, see Table 2, page 3. September 2014 • www.ebmedicine.net 13 Reprints: www.ebmedicine.net/empissues Clinical Pathway For Emergency Department Management Of Multiple Clinical Pathway For Suspected Pacemaker Malfunction Shocks Obtain device history Obtain 12-lead ECG Pacemaker malfunction apparent on ECG? YES YES NO Patient asymptomatic during ECG but reports recent symptomatic episodes Electrolyte abnormality? NO Correct electrolyte abnormality (Class II) Change in ST segments or positive cardiac markers? NO Interrogate device YES • Antithrombotic therapy • Revascularization (Class II) • Check battery status • Check lead status • Check event record (Class II) Replace battery, if necessary If abnormal lead impedance, obtain chest x-ray (Class II) Abbreviation: ECG, electrocardiogram. Class Of Evidence Definitions Each action in the clinical pathways section of Emergency Medicine Practice receives a score based on the following definitions. Class I Class II • Always acceptable, safe • Safe, acceptable • Definitely useful • Probably useful • Proven in both efficacy and effectiveness Level of Evidence: Level of Evidence: • Generally higher levels of evidence • One or more large prospective studies • Nonrandomized or retrospective studies: are present (with rare exceptions) historic, cohort, or case control studies • High-quality meta-analyses • Less robust randomized controlled trials • Study results consistently positive and • Results consistently positive compelling Class III • May be acceptable • Possibly useful • Considered optional or alternative treatments Level of Evidence: • Generally lower or intermediate levels of evidence • Case series, animal studies, consensus panels • Occasionally positive results Indeterminate • Continuing area of research • No recommendations until further research Level of Evidence: • Evidence not available • Higher studies in progress • Results inconsistent, contradictory • Results not compelling This clinical pathway is intended to supplement, rather than substitute for, professional judgment and may be changed depending upon a patient’s individual needs. Failure to comply with this pathway does not represent a breach of the standard of care. Copyright © 2014 EB Medicine. 1-800-249-5770. No part of this publication may be reproduced in any format without written consent of EB Medicine. Copyright © 2014 EB Medicine. All rights reserved. 14 www.ebmedicine.net • September 2014 Risk Management Pitfalls For Implantable Devices 1. “The ventricular-paced rhythm makes the ECG uninterpretable for diagnosing MI. I’ll call cardiology if the troponin comes back positive.” While concordant ST-segment changes and exaggerated discordance are insensitive findings for diagnosing acute MI in paced rhythms, they are diagnostically useful, when present. 6. “I didn’t think to compare the chest x-ray with a previous one or consider lead dislodgement as a complication that could occur so far out from implantation.” Lead migration more commonly occurs shortly after implantation, but it can be a late complication. Comparison with previous x-rays is helpful in detecting macro lead dislodgements. 2. “Pacing artifacts appear stranded among widened and regular QRS complexes at a ventricular rate of 42. There must be a problem with the pacemaker.” Hyperkalemia can cause QRS widening, bradycardia, and failure to capture. Obtain a serum potassium level quickly and consider immediate administration of calcium. Always consider hyperkalemia before attributing failure to capture to a hardware problem or programming error. 7. “The 12-lead ECG appeared normal. I didn’t think the patient needed device interrogation.” Pacemaker interrogation is critical in evaluating the possibility of pacemaker dysfunction that might not be apparent on a 12-lead ECG. 8. “The patient experienced syncope and has a pacemaker, so she is a high-risk cardiac patient and needs to be admitted to telemetry.” Pacemaker interrogation can be performed in the ED to rule out dysrhythmia and pacing malfunction as underlying causes of palpitations, syncope, and light-headedness. Device interrogation may help to avoid unnecessary hospital admission in some of these patients. 3. “I thought electrical cardioversion was contraindicated in patients with permanent pacemakers.” Electric direct-current cardioversion is safe as long as the electrodes are placed in anteriorposterior orientation at least 8 cm away from the pacemaker device. Device interrogation for proper functioning should be performed after cardioversion. 9. “I didn’t think that hiccoughs could be related to the patient’s pacemaker.” Direct stimulation of the diaphragm or the phrenic nerve occurs with lead dislodgement or with high output from a left ventricular lead placed in the coronary sinus. 4. “Pacemaker-mediated tachycardia was on my differential, but I was afraid to place a magnet over the pacemaker.” A pacemaker magnet only disables the sensing function, not the pacing function, and it can be effective in terminating pacemaker-mediated tachycardia as well as for differentiating among various types of potential pacemaker malfunction. 10. “Ventricular tachycardia clearly wasn’t the cause of his symptoms. If he had ventricular tachycardia, his CRT-D device would have shocked him.” Ventricular tachycardia may occur at rates below the programmed tachycardia detection rate (TDR). Above the TDR, ICD therapy is administered. A low TDR can be programmed, with the risk of administering therapy inappropriately to supraventricular tachyarrhythmias. Antitachycardia pacing and algorithms to improve the specificity of VT detection have reduced the incidence of inappropriate ICD therapy for supraventricular arrhythmias. 5. “I thought all newer pacemakers were MRIcompatible.” While advances in pacemaker design have led to the development of pacing leads and pulse generators that are safe for the magnetic environment, safe MRI requires careful screening, established protocols, and physician supervision. September 2014 • www.ebmedicine.net 15 Reprints: www.ebmedicine.net/empissues of 3 or more appropriate shocks within a 24-hour period is termed electrical storm), the patient should be evaluated for possible electrolyte abnormalities (potassium and magnesium), myocardial ischemia, proarrhythmic drug effect, or other causes of QT prolongation. In cases of electrical storm, amiodarone, intravenous beta-adrenergic receptor antagonists, and propofol are suggested drug therapies.113-115 Refractory cases of electrical storm may require general anesthesia, left ventricular assist devices, and emergent radiofrequency ablation. The diagnosis of MI may be difficult in these patients, as transient postshock ST-segment deviation is common and myocardial injury can be caused by the shocks themselves.107,109,116,117 may fail to deliver appropriate therapy, as in situations where rhythm sensing or shock delivery is compromised. For example, an ipsilateral tension pneumothorax may raise the defibrillation threshold, leading to ineffective shocks.64 Antiarrhythmic agents that block fast inward sodium channels (ie, lidocaine) may also increase the threshold, but agents that prolong repolarization (ie, sotalol) may decrease the threshold.108 Keeping all of these in mind, if external electrical therapy is required, transcutaneous pads should be placed anterior-posteriorly if possible, and as far from the pacemaker as possible to avoid damage to the device.64,100,109 The rescuer may also consider magnet placement to avoid unintentional device-mediated therapies during resuscitation.109 After successful resuscitation, the device should be interrogated to ensure proper status of its parameters.109 If central venous access is required, the femoral vein is the safest approach in a patient with a pacemaker.64,110 Pacemaker wires in the subclavian vein are often thrombosed, and may preclude normal ipsilateral subclavian vein cannulation. In addition, contact between a procedural guidewire and a lead can trigger an inappropriate response.64 Disposition The majority of patients presenting with implantrelated complications or pacemaker malfunction will require consultation with a cardiologist and, potentially, hospital admission. Pacemaker interrogation can be performed in the ED to rule out arrhythmia and pacemaker malfunction, and to avoid unnecessary hospital admission in some of these patients. Approach To The Pacemaker/Implantable Cardioverter-Defibrillator Patient Who Receives A Shock Controversies And Cutting Edge Many contemporary pacemakers include antitachycardia functions via combinations of overdrive pacing, cardioversion, and/or defibrillation programmed to occur at designated detection zones. Newer-generation devices employ algorithms to distinguish ventricular from supraventricular tachycardia in order to reduce the risk of inappropriate shocks. These devices use rhythm characteristics such as onset, rate stability, electrogram morphology, and atrioventricular dissociation to differentiate ventricular from supraventricular tachycardia.109,111 In one study, a dual-chamber algorithm reduced the rate of inappropriate detection by 50%.112 The occurrence of a single shock in a patient without preceding symptoms does not require immediate ED consultation with an electrophysiologist; referral within the week is generally appropriate.111 However, in the presence of symptoms such as dyspnea, chest pain, or weakness, underlying etiologies such as myocardial ischemia, decompensated heart failure, or electrolyte imbalance should be considered. The patient who receives multiple shocks should receive immediate attention and cardiac monitoring in the ED until the device can be interrogated. Multiple inappropriate shocks may occur in response to supraventricular tachycardia and may require reprogramming by an electrophysiologist or representative from the device manufacturer. (See Figure 6, page 12.) If defibrillation therapy is found to be appropriate (delivery Copyright © 2014 EB Medicine. All rights reserved. Remote Monitoring Some devices allow for remote monitoring of data, including heart rate variability, tachyarrhythmia frequency and duration, intrathoracic impedance, and percentage of time that the rhythm is paced. Integration of these data points can be clinically useful in predicting heart failure hospitalizations.118-120 Future research is required to define how remote monitoring might lead to improved clinical outcomes. Leadless Pacing As described above, many pacemaker complications relate either to transvenous leads or the subcutaneous pulse generator. Leadless pacing systems are currently under development and early experience has demonstrated safety and efficacy in single-chamber ventricular pacing.121 Device Deactivation Patients with terminal illness may request device deactivation along with do not resuscitate (DNR) status. The HRS has published a consensus statement that declares deactivation of an implanted device as ethical and legally permissible.122 Emergency clinicians should engage patients in conversation about advance directives as they relate to their cardiac device. 16 www.ebmedicine.net • September 2014 1. Summary Emergency clinicians must be familiar with the indications for pacemaker placement and aware of the complications associated with implantation. A systematic approach to evaluating the pacemaker is key to expeditious diagnosis of device malfunction. This includes careful history taking, a thorough physical examination of the patient, and accurate interpretation of the 12-lead ECG and chest radiographs. Device interrogation is a critical step in troubleshooting with regard to both pacemaker and antitachycardia functions. 2. 3. 4. Case Conclusions Before reflexively calling the electrophysiologist to interrogate the pacemaker of the 78-year-old woman with widened QRS complexes and bradycardia, you paused to consider a life-threatening cause of both QRS widening and failure to capture. You sent for a potassium level on a blood gas sample and administered calcium gluconate before finding out the lab result. You watched the QRS complexes on the monitor narrow to their previous width as the heart rate increased. After another dose of calcium gluconate, you noticed that all pacing artifacts were captured. Lab results returned, showing a serum potassium level of 8.7 mEq/L, with prerenal azotemia. You consulted the nephrologist and admitted the patient to the ICU. You obtained a chest x-ray on the 38-year-old woman with sarcoidosis and compared it to the x-ray obtained shortly after implantation of her pacemaker. You noticed that both atrial and ventricular pacing leads were retracted into the superior vena cava and wound around the vertical axis of the pulse generator. The patient denied manipulating the device, but the x-ray was consistent with twiddler syndrome. The patient was admitted to the cardiac ICU and later taken to the operating room for lead revision and reinforcement of the pulse generator with a Dacron® pouch. 5.* 6. 7.* 8. 9. References Evidence-based medicine requires a critical appraisal of the literature based upon study methodology and number of subjects. Not all references are equally robust. The findings of a large, prospective, random­ized, and blinded trial should carry more weight than a case report. To help the reader judge the strength of each reference, pertinent information about the study will be included in bold type following the ref­ erence, where available. In addition, the most informative references cited in this paper, as determined by the authors, will be noted by an asterisk (*) next to the number of the reference. September 2014 • www.ebmedicine.net 10. 11. 12. 17 Greenspon AJ, Patel JD, Lau E, et al. Trends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures. J Am Coll Cardiol. 2012;60(16):1540-1545. (Retrospective observational; 2.9 million patients) Go AS, Mozaffarian D, Roger VL, et al. Heart disease and stroke statistics--2014 update: a report from the American Heart Association. Circulation. 2014;129(3):e28-e292. (Executive summary) Mond HG, Proclemer A. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009--a World Society of Arrhythmia project. Pacing Clin Electrophysiol. 2011;34(8):1013-1027. (Survey; 1.02 million implantations) van Veldhuisen DJ, Maass AH, Priori SG, et al. Implementation of device therapy (cardiac resynchronization therapy and implantable cardioverter defibrillator) for patients with heart failure in Europe: changes from 2004 to 2008. Eur J Heart Fail. 2009;11(12):1143-1151. (Retrospective observational) Young JB, Abraham WT, Smith AL, et al. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial. JAMA. 2003;289(20):2685-2694. (Prospective randomized controlled; 369 patients) Uslan DZ, Tleyjeh IM, Baddour LM, et al. Temporal trends in permanent pacemaker implantation: a population-based study. Am Heart J. 2008;155(5):896-903. (Retrospective observational; 1291 patients) Epstein AE, DiMarco JP, Ellenbogen KA, et al. 2012 ACCF/ AHA/HRS focused update incorporated into the ACCF/ AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2013;61(3):e6-e75. (Guideline) Epstein AE, DiMarco JP, Ellenbogen KA, et al. ACC/AHA/ HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/ AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. J Am Coll Cardiol. 2008;51(21):e1-e62. (Guideline) European Society of Cardiology, European Heart Rhythm Association, Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on Cardiac Pacing and Resynchronization Therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace. 2013;15(8):1070-1118. (Guideline) Gillis AM, Russo AM, Ellenbogen KA, et al. HRS/ACCF expert consensus statement on pacemaker device and mode selection. Developed in partnership between the Heart Rhythm Society (HRS) and the American College of Cardiology Foundation (ACCF) and in collaboration with the Society of Thoracic Surgeons. Heart Rhythm. 2012;9(8):1344-1365. (Guideline) European Heart Rhythm Association, European Society of Cardiology, Heart Rhythm Society, et al. 2012 EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management. Europace. 2012;14(9):1236-1286. (Guideline) Andersen HR, Nielsen JC, Thomsen PE, et al. Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome. Lancet. Reprints: www.ebmedicine.net/empissues 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 1997;350(9086):1210-1216. (Prospective observational; 225 patients) Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003;107(23):2932-2937. (Retrospective observational trial; 1339 patients) Sweeney MO, Bank AJ, Nsah E, et al. Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease. N Engl J Med. 2007;357(10):1000-1008. (Prospective randomized; 1065 patients) Thambo JB, Bordachar P, Garrigue S, et al. Detrimental ventricular remodeling in patients with congenital complete heart block and chronic right ventricular apical pacing. Circulation. 2004;110(25):3766-3772. (Prospective cohort; 53 patients) Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptiverate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25(2):260-264. (Guideline) Chan TC, Cardall TY. Electronic pacemakers. Emerg Med Clin North Am. 2006;24(1):179-194. (Review) Nielsen JC, Thomsen PE, Hojberg S, et al. A comparison of single-lead atrial pacing with dual-chamber pacing in sick sinus syndrome. Eur Heart J. 2011;32(6):686-696. (Prospective randomized; 1415 patients) Dretzke J, Toff WD, Lip GY, et al. Dual chamber versus single chamber ventricular pacemakers for sick sinus syndrome and atrioventricular block. Cochrane Database Syst Rev. 2004(2):CD003710. (Systematic review) Brandt J, Anderson H, Fahraeus T, et al. Natural history of sinus node disease treated with atrial pacing in 213 patients: implications for selection of stimulation mode. J Am Coll Cardiol. 1992;20(3):633-639. (Prospective observational; 213 patients) Gang UJ, Hvelplund A, Pedersen S, et al. High-degree atrioventricular block complicating ST-segment elevation myocardial infarction in the era of primary percutaneous coronary intervention. Europace. 2012;14(11):1639-1645. (Retrospective observational; 2073 patients) American College of Emergency Physicians, Society for Cardiovascular Angiography and Interventions, O’Gara PT, et al. 2013 ACCF/AHA guideline for the management of ST-elevation myocardial infarction: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2013;61(4):e78-140. (Guideline) Farwell D, Patel NR, Hall A, et al. How many people with heart failure are appropriate for biventricular resynchronization? Eur Heart J. 2000;21(15):1246-1250. (Retrospective observational; 721 patients) Shamim W, Francis DP, Yousufuddin M, et al. Intraventricular conduction delay: a prognostic marker in chronic heart failure. Int J Cardiol. 1999;70(2):171-178. (Retrospective observational; 241 patients) Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002;346(24):1845-1853. (Prospective randomized controlled; 453 patients) Higgins SL, Hummel JD, Niazi IK, et al. Cardiac resynchronization therapy for the treatment of heart failure in patients with intraventricular conduction delay and malignant ventricular tachyarrhythmias. J Am Coll Cardiol. 2003;42(8):14541459. (Prospective randomized controlled; 490 patients) Linde C, Abraham WT, Gold MR, et al. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricu- Copyright © 2014 EB Medicine. All rights reserved. 28. 29. 30. 31. 32. 33. 34. 35. 36. 37. 38. 39. 40. 41. 42. 18 lar dysfunction and previous heart failure symptoms. J Am Coll Cardiol. 2008;52(23):1834-1843. (Randomized controlled; 610 patients) Tang AS, Wells GA, Talajic M, et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N Engl J Med. 2010;363(25):2385-2395. (Prospective randomized; 1798 patients) Moss AJ, Hall WJ, Cannom DS, et al. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med. 2009;361(14):1329-1338. (Prospective randomized controlled; 1820 patients) Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004;350(21):2140-2150. (Prospective randomized; 1520 patients) Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005;352(15):1539-1549. (Prospective randomized; 813 patients) Birnie DH, Tang AS. The problem of non-response to cardiac resynchronization therapy. Curr Opin Cardiol. 2006;21(1):2026. (Review) Stavrakis S, Lazzara R, Thadani U. The benefit of cardiac resynchronization therapy and QRS duration: a meta-analysis. J Cardiovasc Electrophysiol. 2012;23(2):163-168. (Meta-analysis; 6501 patients) Sipahi I, Chou JC, Hyden M, et al. Effect of QRS morphology on clinical event reduction with cardiac resynchronization therapy: meta-analysis of randomized controlled trials. Am Heart J. 2012;163(2):260-267. (Meta-analysis; 5356 patients) Connolly SJ, Sheldon R, Roberts RS, et al. The North American Vasovagal Pacemaker Study (VPS). A randomized trial of permanent cardiac pacing for the prevention of vasovagal syncope. J Am Coll Cardiol. 1999;33(1):16-20. (Prospective randomized; 54 patients) Sutton R, Brignole M, Menozzi C, et al. Dual-chamber pacing in the treatment of neurally mediated tilt-positive cardioinhibitory syncope : pacemaker versus no therapy: a multicenter randomized study. The Vasovagal Syncope International Study (VASIS) Investigators. Circulation. 2000;102(3):294-299. (Prospective randomized; 42 patients) Ammirati F, Colivicchi F, Santini M, et al. Permanent cardiac pacing versus medical treatment for the prevention of recurrent vasovagal syncope: a multicenter, randomized, controlled trial. Circulation. 2001;104(1):52-57. (Prospective randomized; 93 patients) Raviele A, Giada F, Menozzi C, et al. A randomized, doubleblind, placebo-controlled study of permanent cardiac pacing for the treatment of recurrent tilt-induced vasovagal syncope. The vasovagal syncope and pacing trial (SYNPACE). Eur Heart J. 2004;25(19):1741-1748. (Prospective randomized controlled; 29 patients) Connolly SJ, Sheldon R, Thorpe KE, et al. Pacemaker therapy for prevention of syncope in patients with recurrent severe vasovagal syncope: Second Vasovagal Pacemaker Study (VPS II): a randomized trial. JAMA. 2003;289(17):2224-2229. (Prospective randomized; 100 patients) Huang SK, Ezri MD, Hauser RG, et al. Carotid sinus hypersensitivity in patients with unexplained syncope: clinical, electrophysiologic, and long-term follow-up observations. Am Heart J. 1988;116(4):989-996. (Prospective observational; 76 patients) Sugrue DD, Gersh BJ, Holmes DR, Jr., et al. Symptomatic “isolated” carotid sinus hypersensitivity: natural history and results of treatment with anticholinergic drugs or pacemaker. J Am Coll Cardiol. 1986;7(1):158-162. (Prospective observational; 56 patients) Morley CA, Perrins EJ, Grant P, et al. Carotid sinus syncope treated by pacing. Analysis of persistent symptoms www.ebmedicine.net • September 2014 43. 44. 45. 46. 47. 48. 49. 50. 51. 52. 53. 54. 55. 56. 57. 58. 59. and role of atrioventricular sequential pacing. Br Heart J. 1982;47(5):411-418. (Case series; 70 patients) Brignole M, Menozzi C, Lolli G, et al. Validation of a method for choice of pacing mode in carotid sinus syndrome with or without sinus bradycardia. Pacing Clin Electrophysiol. 1991;14(2 Pt 1):196-203. (Prospective observational; 60 patients) Brignole M, Menozzi C, Lolli G, et al. Natural and unnatural history of patients with severe carotid sinus hypersensitivity: a preliminary study. Pacing Clin Electrophysiol. 1988;11(11 Pt 2):1628-1635. (Prospective randomized; 35 patients) Parry SW, Steen N, Bexton RS, et al. Pacing in elderly recurrent fallers with carotid sinus hypersensitivity: a randomised, double-blind, placebo controlled crossover trial. Heart. 2009;95(5):405-409. (Randomized controlled crossover trial; 34 patients) Okmen E, Erdinler I, Oguz E, et al. An electrocardiographic algorithm for determining the location of pacemaker electrode in patients with right bundle branch block configuration during permanent ventricular pacing. Angiology. 2006;57(5):623-630. (Retrospective observational; 25 patients) Barold SS, Herweg B, Giudici M. Electrocardiographic follow-up of biventricular pacemakers. Ann Noninvasive Electrocardiol. 2005;10(2):231-255. (Review) Kahloon MU, Aslam AK, Aslam AF, et al. Hyperkalemia induced failure of atrial and ventricular pacemaker capture. Int J Cardiol. 2005;105(2):224-226. (Case report) Ortega-Carnicer J, Benezet J, Benezet-Mazuecos J. Hyperkalaemia causing loss of atrial capture and extremely wide QRS complex during DDD pacing. Resuscitation. 2004;62(1):119-120. (Case report) Schiraldi F, Guiotto G, Paladino F. Hyperkalemia induced failure of pacemaker capture and sensing. Resuscitation. 2008;79(1):161-164. (Case report) Sgarbossa EB, Pinski SL, Gates KB, et al. Early electrocardiographic diagnosis of acute myocardial infarction in the presence of ventricular paced rhythm. GUSTO-I investigators. Am J Cardiol. 1996;77(5):423-424. (Retrospective observational; 17 patients) Maloy KR, Bhat R, Davis J, et al. Sgarbossa Criteria are highly specific for acute myocardial infarction with pacemakers. West J Emerg Med. 2010;11(4):354-357. (Retrospective cohort; 57 patients) Madias JE. The nonspecificity of ST-segment elevation ≥ 5.0 mm in V1-V3 in the diagnosis of acute myocardial infarction in the presence of ventricular paced rhythm. J Electrocardiol. 2004;37(2):135-139. (Case report) Smith SW, Dodd KW, Henry TD, et al. Diagnosis of STelevation myocardial infarction in the presence of left bundle branch block with the ST-elevation to S-wave ratio in a modified Sgarbossa rule. Ann Emerg Med. 2012;60(6):766-776. (Retrospective cohort; 162 patients) Karumbaiah K, Omar B. ST-elevation myocardial infarction in the presence of biventricular paced rhythm. J Emerg Med. 2013;45(2):e35-e40. (Case report) Altun A, Erdogan O. Pacemaker lead failure suggestive of crush injury. Cardiol Rev. 2003;11(5):256. (Case report) Kiviniemi MS, Pirnes MA, Eranen HJ, et al. Complications related to permanent pacemaker therapy. Pacing Clin Electrophysiol. 1999;22(5):711-720. (Retrospective observational; 446 patients) Kutinsky IB, Jarandilla R, Jewett M, et al. Risk of hematoma complications after device implant in the clopidogrel era. Circ Arrhythm Electrophysiol. 2010;3(4):312-318. (Prospective observational; 935 patients) Osmonov D, Ozcan KS, Erdinler I, et al. Cardiac devicerelated endocarditis: 31-years’ experience. J Cardiol. 2013;61(2):175-180. (Retrospective observational; 5287 patients) September 2014 • www.ebmedicine.net 60. Rodriguez Y, Garisto J, Carrillo RG. Management of cardiac device-related infections: a review of protocol-driven care. Int J Cardiol. 2013;166(1):55-60. (Retrospective observational; 350 patients) 61. Athan E, Chu VH, Tattevin P, et al. Clinical characteristics and outcome of infective endocarditis involving implantable cardiac devices. JAMA. 2012;307(16):1727-1735. (Prospective cohort; 2760 patients) 62. Ellenbogen KA, Hellkamp AS, Wilkoff BL, et al. Complications arising after implantation of DDD pacemakers: the MOST experience. Am J Cardiol. 2003;92(6):740-741. (Retrospective observational; 2010 patients) 63.* Pavia S, Wilkoff B. The management of surgical complications of pacemaker and implantable cardioverter-defibrillators. Curr Opin Cardiol. 2001;16(1):66-71. (Review) 64.* McMullan J, Valento M, Attari M, et al. Care of the pacemaker/implantable cardioverter defibrillator patient in the ED. Am J Emerg Med. 2007;25(7):812-822. (Review) 65. Higgins SL, Suh BD, Stein JB, et al. Recurrent twiddler’s syndrome in a nonthoracotomy ICD system despite a Dacron pouch. Pacing Clin Electrophysiol. 1998;21(1 Pt 1):130-133. (Case report) 66. Furman S. Defibrillator twiddler’s syndrome. Ann Thorac Surg. 1995;59(2):544-546. (Comment) 67. Bracke F, Meijer A, Van Gelder B. Venous occlusion of the access vein in patients referred for lead extraction: influence of patient and lead characteristics. Pacing Clin Electrophysiol. 2003;26(8):1649-1652. (Retrospective observational; 89 patients) 68. Mandal S, Pande A, Mandal D, et al. Permanent pacemakerrelated upper extremity deep vein thrombosis: a series of 20 cases. Pacing Clin Electrophysiol. 2012;35(10):1194-1198. (Case series; 20 patients) 69. Al-Bawardy R, Krishnaswamy A, Bhargava M, et al. Tricuspid regurgitation in patients with pacemakers and implantable cardiac defibrillators: a comprehensive review. Clin Cardiol. 2013;36(5):249-254. (Review) 70. Link MS, Hellkamp AS, Estes NA 3rd et al. High incidence of pacemaker syndrome in patients with sinus node dysfunction treated with ventricular-based pacing in the Mode Selection Trial (MOST). J Am Coll Cardiol. 2004;43(11):20662071. (Prospective cohort; 182 patients) 71.* Trohman RG, Kim MH, Pinski SL. Cardiac pacing: the state of the art. Lancet. 2004;364(9446):1701-1719. (Review) 72. Mollazadeh R, Mohimi L, Zeighami M, et al. Hemodynamic effect of atrioventricular and interventricular dyssynchrony in patients with biventricular pacing: implications for the pacemaker syndrome. J Cardiovasc Dis Res. 2012;3(3):200-203. (Cross-sectional; 40 patients) 73.* Lamas GA, Lee KL, Sweeney MO, et al. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med. 2002;346(24):1854-1862. (Randomized controlled trial; 2010 patients) 74. Maisel WH, Moynahan M, Zuckerman BD, et al. Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports. JAMA. 2006;295(16):19011906. (Retrospective observational) 75. Tracy CM, Epstein AE, Darbar D, et al. 2012 ACCF/AHA/ HRS focused update of the 2008 guidelines for devicebased therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. [corrected]. Circulation. 2012;126(14):1784-1800. (Guidelines) 76. Maisel WH. Pacemaker and ICD generator reliability: metaanalysis of device registries. JAMA. 2006;295(16):1929-1934. (Meta-analysis; 3365 patients) 77. Hauser RG, Hayes DL, Kallinen LM, et al. Clinical experience with pacemaker pulse generators and transvenous leads: an 8-year prospective multicenter study. Heart Rhythm. 19 Reprints: www.ebmedicine.net/empissues 2007;4(2):154-160. (Prospective observational; 2652 pulse generators, 615 leads) 78. Ofman P, Rahilly-Tierney C, Djousse L, et al. Pacing system malfunction is a rare cause of hospital admission for syncope in patients with a permanent pacemaker. Pacing Clin Electrophysiol. 2013;36(1):109-112. (Retrospective observational; 162 patients) 79. Emilsson K, Oddsson H, Allared M, et al. An unusual cause of high threshold values at pacemaker implantation. Pacing Clin Electrophysiol. 1997;20(2 Pt 1):366-367. (Case report) 80. Patton KK, Levy M, Viswanathan M. Atrial lead dysfunction: an unusual feature of hypothyroidism. Pacing Clin Electrophysiol. 2008;31(12):1650-1652. (Case report) 81. Takasugi N, Kubota T, Kawamura I, et al. Sudden reversible pacemaker failure in a patient with cardiac sarcoidosis: an unfortunate case of ventricular septal pacing. Europace. 2012;14(7):1061-1062. (Case report) 82. Montefoschi N, Boccadamo R. Propafenone induced acute variation of chronic atrial pacing threshold: a case report. Pacing Clin Electrophysiol. 1990;13(4):480-483. (Case report) 83. Barold SS, Herweg B. The effect of hyperkalaemia on cardiac rhythm devices. Europace. 2014. (Review) 84. Santos KR, Adragao P, Cavaco D, et al. Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices. Europace. 2008;10(12):1381-1386. (Retrospective observational; 37 patients) 85. Gross JN, Platt S, Ritacco R, et al. The clinical relevance of electromyopotential oversensing in current unipolar devices. Pacing Clin Electrophysiol. 1992;15(11 Pt 2):2023-2027. (Prospective observational; 34 patients) 86. Brown KR, Carter W, Jr., Lombardi GE. Blunt traumainduced pacemaker failure. Ann Emerg Med. 1991;20(8):905907. (Case report) 87.* Misiri J, Kusumoto F, Goldschlager N. Electromagnetic interference and implanted cardiac devices: the nonmedical environment (part I). Clin Cardiol. 2012;35(5):276-280. (Review) 88. Misiri J, Kusumoto F, Goldschlager N. Electromagnetic interference and implanted cardiac devices: the medical environment (part II). Clin Cardiol. 2012;35(6):321-328. (Review) 89. Tandogan I, Temizhan A, Yetkin E, et al. The effects of mobile phones on pacemaker function. Int J Cardiol. 2005;103(1):51- 58. (Cross-sectional; 679 patients) 90. Tandogan I, Ozin B, Bozbas H, et al. Effects of mobile telephones on the function of implantable cardioverter defibrillators. Ann Noninvasive Electrocardiol. 2005;10(4):409-413. (Cross-sectional; 43 patients) 91. Jilek C, Tzeis S, Vrazic H, et al. Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis. Ann Intern Med. 2011;155(9):587-592. (Cross-sectional; 388 patients) 92. Gimbel JR, Cox JW Jr. Electronic article surveillance systems and interactions with implantable cardiac devices: risk of adverse interactions in public and commercial spaces. Mayo Clin Proc. 2007;82(3):318-322. (Case series; 2 patients) 93. Webster G, Jordao L, Martuscello M, et al. Digital music players cause interference with interrogation telemetry for pacemakers and implantable cardioverter-defibrillators without affecting device function. Heart Rhythm. 2008;5(4):545- 550. (Cross-sectional; 51 patients) 94. Wolber T, Ryf S, Binggeli C, et al. Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators. Heart Rhythm. 2007;4(1):1-4. (Cross-sectional; 70 patients) 95. Levine GN, Gomes AS, Arai AE, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Copyright © 2014 EB Medicine. All rights reserved. Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation. 2007;116(24):2878-2891. (Scientific statement) 96. Hayes DL, Holmes DR Jr, Gray JE. Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol. 1987;10(4):782-786. (Cross-sectional; 8 patients) 97. Nazarian S, Hansford R, Roguin A, et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011;155(7):415-424. (Prospective nonrandomized trial; 438 patients, 555 MRI studies) 98. Kanal E, Borgstede JP, Barkovich AJ, et al. ACR Blue Ribbon Panel response to the AJR commentary by Shellock and Crues on the ACR white paper on MR safety. AJR Am J Roentgenol. 2003;180(1):31-35. (Comment) 99. Das G, Staffanson DB. Selective dysfunction of ventricular electrode-endocardial junction following DC cardioversion in a patient with a dual chamber pacemaker. Pacing Clin Electrophysiol. 1997;20(2 Pt 1):364-365. (Case report) 100. Waller C, Callies F, Langenfeld H. Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes Is external cardioversion contraindicated in patients with permanent pacing systems? Europace. 2004;6(2):165-168. (Case report; 3 cases) 101. Platonov MA, Gillis AM, Kavanagh KM. Pacemakers, implantable cardioverter/defibrillators, and extracorporeal shockwave lithotripsy: evidence-based guidelines for the modern era. J Endourol. 2008;22(2):243-247. (Guideline) 102. Fishberger SB, Mehta D, Rossi AF, et al. Variable effects of adenosine on retrograde conduction in patients with atrioventricular nodal reentry tachycardia. Pacing Clin Electrophysiol. 1998;21(6):1254-1257. (Prospective nonrandomized trial; 13 patients) 103. Sarko JA, Tiffany BR. Cardiac pacemakers: evaluation and management of malfunctions. Am J Emerg Med. 2000;18(4):435-440. (Review) 104. Lewalter T, Schimpf R, Kulik D, et al. Comparison of spontaneous atrial fibrillation electrogram potentials with the P-wave electrogram amplitude in dual chamber pacing with unipolar atrial sensing. Europace. 2000;2(2):136-140. (Prospective observational; 42 patients) 105. Jacob S, Panaich SS, Maheshwari R, et al. Clinical applications of magnets on cardiac rhythm management devices. Europace. 2011;13(9):1222-1230. (Review) 106. Becker R, Ruf-Richter J, Senges-Becker JC, et al. Patient alert in implantable cardioverter defibrillators: toy or tool? J Am Coll Cardiol. 2004;44(1):95-98. (Prospective cohort; 240 patients) 107. Stevenson WG, Chaitman BR, Ellenbogen KA, et al. Clinical assessment and management of patients with implanted cardioverter-defibrillators presenting to nonelectrophysiologists. Circulation. 2004;110(25):3866-3869. (Review) 108. Glikson M, Friedman PA. The implantable cardioverter defibrillator. Lancet. 2001;357(9262):1107-1117. (Review) 109. Pinski SL. Emergencies related to implantable cardioverterdefibrillators. Crit Care Med. 2000;28(10 Suppl):N174-N180. (Review) 110. McPherson CA, Manthous C. Permanent pacemakers and implantable defibrillators: considerations for intensivists. Am J Respir Crit Care Med. 2004;170(9):933-940. (Review) 111.* Gehi AK, Mehta D, Gomes JA. Evaluation and management of patients after implantable cardioverter-defibrillator shock. JAMA. 2006;296(23):2839-2847. (Review) 112. Friedman PA, McClelland RL, Bamlet WR, et al. Dualchamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study. Circulation. 2006;113(25):2871- 20 www.ebmedicine.net • September 2014 2879. (Randomized controlled trial; 400 patients) 113. Credner SC, Klingenheben T, Mauss O, et al. Electrical storm in patients with transvenous implantable cardioverter-defibrillators: incidence, management and prognostic implications. J Am Coll Cardiol. 1998;32(7):1909-1915. (Prospective cohort; 136 patients) 114. Nademanee K, Taylor R, Bailey WE, et al. Treating electrical storm: sympathetic blockade versus advanced cardiac life support-guided therapy. Circulation. 2000;102(7):742-747. (Prospective nonrandomized trial; 49 patients) 115. Mulpuru SK, Patel DV, Wilbur SL, et al. Electrical storm and termination with propofol therapy: a case report. Int J Cardiol. 2008;128(1):e6-e8. (Case report) 116. Sham’a RA, Nery P, Sadek M, et al. Myocardial injury secondary to ICD shocks: insights from patients with lead fracture. Pacing Clin Electrophysiol. 2014;37(2):237-241. (Prospective observational; 22 patients) 117. Gurevitz O, Lipchenca I, Yaacoby E, et al. ST-segment deviation following implantable cardioverter defibrillator shocks: incidence, timing, and clinical significance. Pacing Clin Electrophysiol. 2002;25(10):1429-1432. (Prospective observational; 28 patients) 118. Cowie MR, Sarkar S, Koehler J, et al. Development and validation of an integrated diagnostic algorithm derived from parameters monitored in implantable devices for identifying patients at risk for heart failure hospitalization in an ambulatory setting. Eur Heart J. 2013;34(31):2472-2480. (Retrospective observational; 2231 patients) 119. Sack S, Wende CM, Nagele H, et al. Potential value of automated daily screening of cardiac resynchronization therapy defibrillator diagnostics for prediction of major cardiovascular events: results from Home-CARE (Home Monitoring in Cardiac Resynchronization Therapy) study. Eur J Heart Fail. 2011;13(9):1019-1027. (Prospective cohort; 377 patients) 120. Whellan DJ, Ousdigian KT, Al-Khatib SM, et al. Combined heart failure device diagnostics identify patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF (Program to Access and Review Trending Information and Evaluate Correlation to Symptoms in Patients With Heart Failure) study. J Am Coll Cardiol. 2010;55(17):1803- 1810. (Retrospective observational trial; 694 patients) 121. Reddy VY, Knops RE, Sperzel J, et al. Permanent leadless cardiac Pacing: results of the LEADLESS trial. Circulation. 2014. (Prospective nonrandomized trial; 33 patients) 122. Lampert R, Hayes DL, Annas GJ, et al. HRS expert consensus statement on the management of cardiovascular implantable electronic devices (CIEDs) in patients nearing end of life or requesting withdrawal of therapy. Heart Rhythm. 2010;7(7):1008-1026. (Consensus statement) CME Questions Take This Test Online! Current subscribers receive CME credit absolutely free by completing the following test. Each issue includes 4 AMA PRA Category 1 CreditsTM, 4 ACEP Category I credits, 4 AAFP Prescribed Take This Test Online! credits, and 4 AOA Category 2A or 2B credits. Monthly online testing is now available for current and archived issues. To receive your free CME credits for this issue, scan the QR code below with your smartphone or visit www.ebmedicine.net/E0914. 1. Cardiac resynchronization therapy is indicated for patients with heart failure and: a. Recurrent ischemic events b. Ventricular dysrhythmias c. A left bundle branch block d. Atrioventricular nodal conduction delay 2. Electrocardiographic findings indicative of severe hyperkalemia in a patient with a paced rhythm include: a. QRS widening b. Failure to capture c. Sine wave QRS morphology d. All of the above 3. Acute myocardial infarction in a paced rhythm: a. Is always electrocardiographically silent b. Can be detected by elevated ST segments in a pattern of exaggerated discordance c. Is the most common cause of failure to sense d. Can be detected by applying a magnet 4. The manufacturer of a pacemaker can be identified by: a. The amplitude of pacing artifacts on the 12- lead ECG b. The size of the pulse generator c. Placing a magnet over the pulse generator d. None of the above September 2014 • www.ebmedicine.net 21 Reprints: www.ebmedicine.net/empissues 5. Management of an infected device usually involves: a. Intravenous antibiotics and device removal b. Needle aspiration of the infected pocket and intravenous antibiotics c. A prolonged course of oral antibiotics d. Incision and drainage of the pocket followed by wound care Get 4 Hours Stroke CME Credit From The July and August 2013 Issues Of EM Practice Guidelines Update 6. What is the most common cause of failure to pace? a. Lead fracture b. Oversensing c. Lead dislodgement d. Battery depletion The July and August 2013 issues of our online journal supplement, EM Practice Guidelines Update, are devoted to stroke topics, and each offers 2 hours of Stroke CME credit. All subscribers to Emergency Medicine Practice automatically receive free subscriptions to EM Practice Guidelines Update; to access the articles, simply log in to your www.ebmedicine.net account (or give us a call and we’ll help you get your account set up). The July 2013 issue reviews the 2009 guideline on transient ischemic attack (TIA) and the revised American Heart Association/American Stroke Association (AHA/ASA) “tissue-based” diagnosis of TIA. Jonathan Edlow, MD, of Harvard Medical School, offers a guest editorial on the evolution of effective emergency care options for TIA patients, and Editor-in-Chief Sigrid Hahn, MD reviews and comments on portions of the guideline relevant to emergency clinicians. Read the issue online at: www.ebmedicine.net/TIA. 7. MRI in patients with pacemakers: a. Is uniformly contraindicated b. Can be performed without complications as long as the pacemaker was manufactured after 2002 c. Can be performed, as long as a magnet is placed over the implanted device d. Can be performed in some cases, but only if specific guidelines are met 8. What is the mechanism of pacemaker-mediated tachycardia? a. A re-entrant circuit b. Increased automaticity at the site of the ventricular electrode tip c. Cardiac remodeling d. Delayed afterdepolarizations 9. Placing a magnet on a pacemaker effectively: a. Switches the pacemaker into an asynchronous mode b. Turns pacing off c. Turns sensing on d. Switches the pacemaker into a demand pacing mode The August 2013 issue reviews 2 different guidelines published in 2013 on acute ischemic stroke and the use of intravenous t-PA (tissue plasminogen activator) from: (1) the American College of Emergency Physicians jointly with the American Academy of Neurology, and (2) the AHA/ASA. Christopher Hopkins, MD of the University of Florida College of Medicine-Jacksonville, the guest editor, provides an assessment of these controversial new guidelines, which have been 8 years in development. Read this issue online at: www.ebmedicine.net/Stroke. 10. External defibrillation in a patient with an implanted cardiac device: a. Is contraindicated b. Is not necessary if the device has anti- tachycardia function (ICD) c. Can be performed if pads are placed a safe distance away from the device d. Can be performed, but only at 50 joules Copyright © 2014 EB Medicine. All rights reserved. 22 www.ebmedicine.net • September 2014 Order Your LLSA Study Guide Today! Receive full reprints of the original articles and gain all the knowledge you need, plus earn free CME. The 2015 LLSA Study Guide is available for $199. EB Medicine's 2015 LLSA Study Guide is designed exclusively to give you the freedom to prepare for the ABEM exam your way, helping you gain practical knowledge of the material. For each of the 13 chosen readings, you receive print and digital tools that include the materials you need to ace the exam. The print and digital tools include full reprints of the articles, a summary and in-depth discussion of each article, sample questions and answers, a critique of the article, and links to previous EB Medicine articles covering those topics. Both print and digital formats will be included with your order! Plus, you earn 35 AMA/ACEP Category 1 CME credits at no extra charge. You will receive: • The traditional spiral-bound book, which lays flat for study and is easily portable. • The digital version, which takes portability to a whole new level, allowing you to read and study from any computer, tablet, or smartphone. TABLE OF CONTENTS: 1. Evaluation and initial treatment of supraventricular tachycardia 2. Prehospital epinephrine use and survival among patients with out-ofhospital cardiac arrest Go to 3. Questioning the use of epinephrine to treat cardiac arrest www.ebmedicine.net/LLSA 4. Therapeutic hypothermia: a state-of-the-art emergency medicine or call us 1-800-249-5770 perspective to get your copy. 5. Accidental hypothermia 6. Cardiovascular disasters in pregnancy 7. Coronary CT angiography versus standard evaluation in acute chest pain 8. Infective endocarditis 9. Lean thinking in emergency departments: a critical review 10. Relief of pain and anxiety in pediatric patients in emergency medical systems 11. Clinical practice guideline for emergency department ketamine dissociative sedation: 2011 update 12. Electrocardiogram findings in emergency department patients with syncope 13. New oral anticoagulants in the ED setting: a review OUR MONEY-BACK GUARANTEE: If you are unsatisfied for any reason, simply call to receive a full refund immediately. No questions asked. INFORMATION ON THE LLSA EXAMINATION: The American Board of Emergency Medicine Continuous Certification (ConCert) program requires the reading of up to 20 designated articles each year. In order to maintain board certification, emergency physicians must complete the annual Lifelong Learning Self-Assessment (LLSA) test on the ABEM website based upon the contents of these articles. A new LLSA test and associated reading list are developed each year. You have up to three years to take each test. The 2015 LLSA exam will be available from April 2015 until April 2018. September 2014 • www.ebmedicine.net 23 Reprints: www.ebmedicine.net/empissues Physician CME Information Date of Original Release: September 1, 2014. Date of most recent review: August 10, 2014. Termination date: September 1, 2017. Accreditation: EB Medicine is accredited by the Accreditation Council for Continuing Medical Education (ACCME) to provide continuing medical education for physicians. This activity has been planned and implemented in accordance with the Essential Areas and Policies of the ACCME. Credit Designation: EB Medicine designates this enduring material for a maximum of 4 AMA PRA Category 1 Credits™. Physicians should claim only the credit commensurate with the extent of their participation in the activity. ACEP Accreditation: Emergency Medicine Practice is approved by the American College of Emergency Physicians for 48 hours of ACEP Category I credit per annual subscription. AAFP Accreditation: This Medical Journal activity, Emergency Medicine Practice, has been reviewed and is acceptable for up to 48 Prescribed credits by the American Academy of Family Physicians per year. AAFP accreditation begins July 31, 2013. Term of approval is for one year from this date. Each issue is approved for 4 Prescribed credits. Credit may be claimed for one year from the date of each issue. Physicians should claim only the credit commensurate with the extent of their participation in the activity. AOA Accreditation: Emergency Medicine Practice is eligible for up to 48 American Osteopathic Association Category 2A or 2B credit hours per year. Needs Assessment: The need for this educational activity was determined by a survey of medical staff, including the editorial board of this publication; review of morbidity and mortality data from the CDC, AHA, NCHS, and ACEP; and evaluation of prior activities for emergency physicians. Target Audience: This enduring material is designed for emergency medicine physicians, physician assistants, nurse practitioners, and residents. Renew your subscription to Emergency Medicine Practice now for $279 (a savings of $50 off our regular subscription rate) You'll also get a free binder to hold all your 2014 issues! Use CME dollars to renew your subscription and get organized at the same time. Call us at 1-800-249-5770 and use promotion code RGEAJ or go online to www.ebmedicine.net/RGEAJ to take advantage of this limited-time offer. Goals: Upon completion of this article, you should be able to: (1) demonstrate medical decision-making based on the strongest clinical evidence; (2) cost-effectively diagnose and treat the most critical ED presentations; and (3) describe the most common medicolegal pitfalls for each topic covered. Discussion of Investigational Information: As part of the newsletter, faculty may be presenting investigational information about pharmaceutical products that is outside Food and Drug Administration-approved labeling. Information presented as part of this activity is intended solely as continuing medical education and is not intended to promote off-label use of any pharmaceutical product. Faculty Disclosure: It is the policy of EB Medicine to ensure objectivity, balance, independence, transparency, and scientific rigor in all CME-sponsored educational activities. All faculty participating in the planning or implementation of a sponsored activity are expected to disclose to the audience any relevant financial relationships and to assist in resolving any conflict of interest that may arise from the relationship. In compliance with all ACCME Essentials, Standards, and Guidelines, all faculty for this CME activity were asked to complete a full disclosure statement. The information received is as follows: Dr. Martindale, Dr. deSouza, Dr. Brady, Dr. Kosowsky, Dr. Jagoda, Dr. Damilini, Dr. Toscano and their related parties report no significant financial interest or other relationship with the manufacturer(s) of any commercial product(s) discussed in this educational presentation. Commercial Support: This issue of Emergency Medicine Practice did not receive any commercial support. Method of Participation: •Print Semester Program: Paid subscribers who read all CME articles during each Emergency Medicine Practice 6-month testing period, complete the CME Answer and Evaluation Form distributed with the June and December issues, and return it according to the published instructions are eligible for up to 4 hours of CME credit for each issue. •Online Single-Issue Program: Current, paid subscribers who read this Emergency Medicine Practice CME article and complete the test and evaluation at www. ebmedicine.net/CME are eligible for up to 4 hours of Category 1 credit toward the AMA Physician’s Recognition Award (PRA). Hints will be provided for each missed question, and participants must score 100% to receive credit. Hardware/Software Requirements: You will need a Macintosh or PC to access the online archived articles and CME testing. Additional Policies: For additional policies, including our statement of conflict of interest, source of funding, statement of informed consent, and statement of human and animal rights, visit http://www.ebmedicine.net/policies. CEO & Publisher: Stephanie Williford Director of Editorial: Dorothy Whisenhunt Content Editors: Erica Carver, Jenique Meekins Editorial Projects Manager: Kay LeGree Director of Member Services: Liz Alvarez Office Manager: Kiana Collier Member Services Representative: Paige Banks Director of Marketing: Robin Williford Marketing Coordinator: Katherine Johnson Subscription Information: Direct all questions to: EB Medicine Phone: 1-800-249-5770 or 1-678-366-7933 Fax: 1-770-500-1316 5550 Triangle Parkway, Suite 150 Norcross, GA 30092 E-mail: ebm@ebmedicine.net Website: www.ebmedicine.net To write a letter to the editor, please email: jagodamd@ebmedicine.net 12 monthly evidence-based print issues; 48 AMA PRA Category 1 CreditsTM, 48 ACEP Category 1 credits, 48 AAFP Prescribed credits, and 48 AOA Category 2A or 2B CME credits; and full online access to searchable archives and additional CME: $329 Individual issues, including 4 CME credits: $30 (Call 1-800-249-5770 or go to http://www.ebmedicine.net/EMP issues to order) Emergency Medicine Practice (ISSN Print: 1524-1971, ISSN Online: 1559-3908, ACID-FREE) is published monthly (12 times per year) by EB Medicine (5550 Triangle Parkway, Suite 150, Norcross, GA 30092). Opinions expressed are not necessarily those of this publication. Mention of products or services does not constitute endorsement. This publication is intended as a general guide and is intended to supplement, rather than substitute, professional judgment. It covers a highly technical and complex subject and should not be used for making specific medical decisions. The materials contained herein are not intended to establish policy, procedure, or standard of care. Emergency Medicine Practice is a trademark of EB Medicine. Copyright © 2014 EB Medicine. All rights reserved. No part of this publication may be reproduced in any format without written consent of EB Medicine. This publication is intended for the use of the individual subscriber only and may not be copied in whole or part or redistributed in any way without the publisher’s prior written permission — including reproduction for educational purposes or for internal distribution within a hospital, library, group practice, or other entity. Copyright © 2014 EB Medicine. All rights reserved. 24 www.ebmedicine.net • September 2014