UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Table of Contents
1.
Scope
2.
Objective
3.
Definitions
4.
Responsibility
5.
Procedure
6.
Precautions
7.
References
8.
Appendices
9.
Revision History
Page
6
34
35
35
35
2
2
2
5
Valid for:
2 years from approval
Original
Authorised Copy for:
Reading Copy
Written by:
Reviewed by:
Approved by:
Signature/Date:
Signature/Date:
Signature/Date:
Page 1 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
1.
Scope
This Standard Operating Procedure (SOP) applies to the staff and students using the
™
Medtronic Minimed CGMS
®
System Gold in the Pharmacy Practice Resource Unit
(PPRU) at the Pharmacy Department, University of Malta.
2.
Objective
To describe the procedure for the set up, operation, maintenance and troubleshooting of the Medtronic Minimed CGMS
®
System Gold
™
.
3.
Definitions
3.1.
ACT (Activate) Button: To confirm chosen parameters during monitor set up and to turn off the alarms.
3.2.
Belt Clip: Used to attach the monitor to a belt or a piece of clothing.
3.3.
Cable: Links the sensor to the monitor via 2 types of connectors i.e. the sensor and the monitor connectors. The sensor connector allows the cable to be securely connected to the sensor whilst the monitor connector allows the cable to be securely connected to the monitor.
3.4.
Cable Tool: Used to securely attach the cable to the monitor.
3.5.
Calibration: A procedure to ensure that the monitor is able to convert electronic signals received from the sensor into glucose values. This is achieved by entering periodic finger stick glucose readings into the monitor.
3.6.
Com-Station: A device that allows the transfer of data from the monitor to a computer. Infrared ports are the communication link with the monitor and an
RS-232 cable is the communication link with the computer. The following are the operating components of the Com-Station:
3.6.1.
AC Adapter Port: The location in which the power supply connection is connected to the Com-Station.
3.6.2.
Communications Ports A and B: Through a computer cable, provides a data link between the Com-Station and the computer (Com-Port A) or a memory glucose meter (Com-Port B)
Page 2 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
3.6.3.
Device Selector Switch: Selects the data download source from either a continuous glucose monitor (the wave symbol) or a memory glucose meter (B symbol).
3.6.4.
Monitor Cradle: A depression in the Com-Station where the monitor is placed to download data from its memory. It contains various infrared ports which act as the communication link between the monitor and the computer.
3.6.5.
On/Off Switch: To switch the Com-Station On or Off.
3.7.
Down Arrow: To choose between “ No
” and “
Yes
” commands and to find the next lowest preset value.
3.8.
Events Menu: Enables user to record a number of events such as meals and snacks, insulin dosages, exercise and illnesses or any other activity that may affect blood glucose levels.
3.9.
Exporting Files: A function that allows data to be imported into other software applications such as Microsoft™ Excel for further analysis.
3.10.
Finger Stick Glucose Measurement: A glucose reading obtained through the use of another separate point-of-care testing glucose device and is used for monitor calibration.
3.11.
Home Screen: Displays the current time.
3.12.
Initialisation Period: The process of equilibrating the electronic signal generated by the sensor after it has been inserted in the subcutaneous tissue.
During this period, the sensor becomes wet with the interstitial fluid and the glucose oxidase reaction starts to become stabilised.
3.13.
ISIG: The acronym for Interstitial Signal i.e. the electrical current, measured in nanoamperes (nA), that is sent to the monitor via the sensor.
3.14.
Medtronic Minimed CGMS
®
System Gold
™
: A continuous glucose monitoring device used to continuously monitor glucose levels in the interstitial fluid within a range of 40-400mg/dL (or 2.2-22mmol/L). The enzyme glucose oxidase converts the glucose at the surface of the sensor into electrical signals which are sent through the cable to the monitor. The monitor will sample the received signals once every 10 seconds and records an average signal every 5 minutes. The stored signals are then converted into glucose concentrations once these are downloaded in a computer.
Page 3 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
3.15.
Memory: Stores up to 14 days of data which can then be transferred to a computer for further analysis.
3.16.
Monitor: Displays all the functions that are available for the device and continuously stores electronic signals that are received from the sensor. All the stored data is downloaded in a computer, via the Com-Station, for conversion to glucose readings.
3.17.
On/Off Button: To switch the monitor On or Off. Switching the monitor Off will not allow new glucose data to be recorded into Memory.
3.18.
Operating Menus: Four different menus that can be accessed from the Home
Screen i.e. [ METER BG ], [ EVENT ], [ SIGNALS ] and [ SETUP ].
3.19.
Patient Entry and Edit Screen: Allows the user to create new patient files, to edit existing ones, to open prior version files, to change an individual patient’s glucose target range and to change an individual patient’s time range blocks.
3.20.
Reports Menu: Consists of 5 report types that can be displayed and these are:
3.20.1.
Sensor Daily Details Report: Provides the user with up to 14 days of individual data plots.
3.20.2.
Sensor Data Report: A computerised logbook that presents all of the data stored in the monitor memory for a patient download.
3.20.3.
Sensor Modal Day Report: Represents all the glucose data over a 24hour period were each day is represented as a separate plot line in a different colour.
3.20.4.
Sensor Modal Time Periods Report: Provides an analysis of each of the Time Range Blocks.
3.20.5.
Sensor Summary Report: Includes a tabular summary of the statistical data from the CGMS
®
and the glucose data that was entered into the monitor.
3.21.
SEL (Select) Button: To scroll through each of the displays and menus available.
3.22.
Sen-serter: Used to help insert the sensor through the skin and into the subcutaneous tissue.
Page 4 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
3.23.
Sensor: A small device that is placed in the subcutaneous tissue which produces a signal that is proportional to the concentration of glucose in the interstitial fluid. It is to be stored in a refrigerated environment.
3.24.
Sensor Insertion Site: A site that contains adequate subcutaneous tissue for sensor insertion. The lower abdomen and the upper, outer quadrant of the buttocks are the sites most frequently used.
3.25.
Set Up Menu: Consists of 7 options which are used to set the monitor according to the user’s preferences i.e. [ CLEAR ], [ PATIENT ], [ INIT ], [ SET
TIME ], [ UNIT ], [ LOCK ] and [ EXIT ]. Refer to SOP/PD/214_01/A1 for further details on each option.
3.26.
Shower-Pak: A waterproof, clear-plastic pouch used to keep the monitor dry during a shower or in any other situation that may cause the monitor to get wet. However, it is not intended to protect the monitor from submersion in water.
3.27.
Solutions™ Software:
A computer programme designed to extract glucose data from the Medtronic Minimed CGMS
®
System Gold
™
.
3.28.
Test Plug: A device used to provide diagnostic information that can help identify if a sensor, cable or monitor is/are operating in the correct manner.
Refer to SOP/PD/214_01/A2 for further details on how to use the test plug.
3.29.
Up Button: To select between “ No
” and “
Yes
” commands and to find the next highest preset value.
3.30.
User Preferences Menu: A menu in the Solutions™ Software computer programme consisting of 3 sub-menus i.e. [ General ], [ Reports ] and [ Clinic
Info ] that are used to set the default patient information for all the patient’s reports.
4.
Responsibility
4.1.
The members of the Department of Pharmacy (staff and students) are responsible for following this SOP.
4.2.
The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed.
Page 5 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.
Procedure
5.1.
Set Up
5.1.1.
Monitor Set Up
5.1.1.1.
Insert 2 new AAA alkaline batteries into the battery compartment.
5.1.1.2.
Press the On/Off button to switch the monitor On.
5.1.1.3.
Wait until the monitor performs its routine diagnostic test and the Home Screen is displayed.
5.1.1.4.
Press the SEL button 4 times. [ SETUP ] will be displayed.
5.1.1.5.
Press the ACT button.
5.1.1.6.
Repeatedly press the SEL button to scroll through the 7 different options available in the Set Up menu (Refer to
SOP/PD/214_01/A1).
5.1.1.7.
Press the ACT button whilst [ EXIT ] is displayed to exit the
Set Up menu and return to the Home Screen.
5.1.2.
Sensor Insertion
5.1.2.1.
Wash hands well with soap and water.
5.1.2.2.
Wear a suitable pair of disposable gloves.
5.1.2.3.
Take out a sensor package from the refrigerator.
5.1.2.4.
Allow the sensor package to reach room temperature.
5.1.2.5.
Choose an appropriate sensor insertion site (see 3.24)
5.1.2.6.
Clean sensor insertion site with alcohol.
5.1.2.7.
Allow the sensor insertion site to dry completely before inserting a sensor.
5.1.2.8.
Remove the sensor from its packaging.
5.1.2.9.
Place the sensor in the Sen-serter.
5.1.2.10.
Hold the Sen-serter upright, secure the white tape of the sensor and remove the clear tape using an anticlockwise motion.
5.1.2.11.
Place fingers on the back of the white tape and push the carrier down until it clicks into place.
5.1.2.12.
Remove the needle guard from the introducer needle. o
5.1.2.13.
Rest the Sen-serter legs against the skin at a 45 angle.
5.1.2.14.
Press the white button on the Sen-serter.
5.1.2.15.
Check that the sensor has been inserted properly through the skin.
5.1.2.16.
Hold the sensor in place and gently slide the Sen-serter away from the sensor.
Page 6 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.1.2.17.
Remove the white paper from the adhesive pad and press it against the skin.
5.1.2.18.
Hold the sensor with two fingers on its base and gently remove the introducer needle at a 45 o
angle.
5.1.2.19.
Dispose the introducer needle into a sharps container.
5.1.3.
Connecting the Cable
5.1.3.1.
Place the monitor connector end of the cable above the monitor connection port until slight contact is made.
5.1.3.2.
Align the arrow on the monitor connector with the arrow on the monitor connection port.
5.1.3.3.
Place the groove on the side of the cable tool on the corresponding groove on the monitor connector.
5.1.3.4.
Apply pressure to advance the monitor connector into the monitor connection port.
5.1.3.5.
Check that both arrows on the monitor connector and on the monitor connection port are still properly aligned.
5.1.3.6.
Grasp the sensor base between the thumb and forefinger.
5.1.3.7.
Orient the sensor connector end of the cable to be in close proximity to the connection of the sensor.
5.1.3.8.
Plug the sensor connector end of the cable to the sensor.
5.1.3.9.
Press both parts together until they snap in place.
5.1.3.10.
Cover the insertion site with a transparent sterile dressing.
5.1.3.11.
View the [ ISIG ] screen (see section 5.2.2) for 1-2 minutes to ensure that the sensor current is adequate and stable.
5.1.4.
Sensor Initialisation
5.1.4.1.
Repeatedly press the SEL button, while in Home Screen, until [ SETUP ] is displayed.
5.1.4.2.
Press the ACT button.
5.1.4.3.
Repeatedly press the SEL button until [ INIT ] is displayed.
5.1.4.4.
Press the ACT button. Monitor will beep and [ NO ] will flash.
5.1.4.5.
Press the Up or Down arrow button to select [ YES ].
5.1.4.6.
Press the ACT button to begin the initialisation process.
5.1.4.7.
Wait for one hour for the initialisation process to be complete. The monitor will continuously countdown the time remaining.
5.1.4.8.
Wait for the monitor to beep to indicate that the initialisation process is complete.
Page 7 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.1.5.
Sensor Calibration
5.1.5.1.
Perform the first sensor calibration when sensor initialisation is complete.
5.1.5.2.
Perform a finger stick glucose measurement.
5.1.5.3.
Press the SEL button once while in Home Screen. [ METER
BG ] is displayed.
5.1.5.4.
Press the ACT button and wait for [ --] and [ ENTER BG ] to be displayed.
5.1.5.5.
Use the Up and Down arrow buttons to enter the glucose measurement reading in increments of 1mg/dL (or
0.1mmol/L).
5.1.5.6.
Press the ACT button to confirm.
5.1.5.7.
If an alarm is heard and [ CAL ERR ] is displayed within approximately 1-2 minutes, the entered glucose reading is outside the sensitivity check expected ranges (Refer to section 5.4 for possible solutions).
5.1.5.8.
Repeat sensor calibration procedure at least 4 times daily. A combination of fasting, preprandial and postprandial values are ideal.
5.1.6.
Using the Monitor Belt Clip
5.1.6.1.
Slide the triangular-shaped ridge on the back of the belt clip into its corresponding groove on the back of the monitor.
5.1.6.2.
Keep pushing the belt clip until it snaps into place.
5.1.6.3.
If belt clip is to be removed, lift the grooved edge of the belt clip with thumb and gently slide the clip out of its groove.
5.1.7.
Installing the Solutions™ Software
5.1.7.1.
Insert the Medtronic MiniMed CGMS
®
Solutions
™
Software
CD into the CD drive.
5.1.7.2.
Double click on the My Computer icon on the Windows desktop.
5.1.7.3.
Right click on the CD drive.
5.1.7.4.
Click on Open .
5.1.7.5.
Double click on the SETUP.EXE file.
5.1.7.6.
Click the Next button.
5.1.7.7.
Read the Software License Agreement.
5.1.7.8.
Click the Yes button.
5.1.7.9.
Click the Next button to accept the default destination location for installation.
Page 8 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.1.7.10.
Click the Next button to accept the default destination folder to save the patients download files. If another file is desired, click the Browse button to select another location.
5.1.7.11.
Click the Next button. The Select Program Folder window will now be displayed.
5.1.7.12.
Click the Next button to accept the MiniMed Solutions
CGMS
®
Sensor software.
5.1.7.13.
Click the Next button when the Start Copying Files window is displayed.
5.1.7.14.
Wait until the setup status reaches 100%.
5.1.7.15.
Click the Finish button when the Setup Complete window is displayed. If a reboot request message appears, accept the request in order to finish the installation process.
5.1.7.16.
The MiniMed Solutions CGMS
®
Sensor icon will now be displayed on the Windows desktop.
5.1.8.
Setting Up the MiniMed Solutions CGMS
®
Sensor Programme
5.1.8.1.
Double click the MiniMed Solutions CGMS
®
Sensor icon on the Windows desktop.
5.1.8.2.
Check that the device selector switch of the Com-Station is in the wave symbol position.
5.1.8.3.
Select [ User Preferences ] from the File Menu.
5.1.8.4.
From the [ General ] Menu, click on the arrow of the drop down menu under Com-Port to select the serial port from which the Com-Station is connected to the computer.
5.1.8.5.
Click the [ Check Com-Station ] button. If correct Com-
Station port was selected, [ Com-Station Ready ] will be displayed.
5.1.8.6.
Click the Ok button.
5.1.8.7.
If the wrong Com-Port was selected, [ No Com-Station
Detected ] will be displayed. Click the Ok button and recheck the connections.
5.1.8.8.
Select between a 12-hour or a 24-hour time display option.
5.1.8.9.
Select between mg/dL or mmol/L from the [ Report Glucose
Units ] section.
5.1.8.10.
Click the Up and Down arrow buttons to change the default glucose target for the Upper and Lower Limits from the
[ Default Glucose Target ] section.
5.1.8.11.
From the [ Reports ] Menu, click the checkbox of the Report that needs to be printed.
5.1.8.12.
Click on the time range block that needs to be changed.
5.1.8.13.
Click the Up and Down arrows to adjust its time.
Page 9 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.1.8.14.
Click outside of the time range block to save the changes made.
5.1.8.15.
Repeat steps 5.1.8.12 to 5.1.8.14 to change the time setting of other time range blocks.
5.1.8.16.
From the [ Clinic Info ] Menu, input the desired clinic name, address and phone number.
5.1.8.17.
Press the Ok button to save all the changes made to the User
Preferences Menu. The Patient Entry and Edit screen will now be displayed.
5.1.9.
Connecting the Com-Station to the Computer
5.1.9.1.
Connect the computer cable into an available serial Com-
Port connector on the computer.
5.1.9.2.
Connect the other end of the computer cable to Com-Port A of the Com-Station.
5.1.9.3.
Connect the AC adapter into a suitable power source.
5.1.9.4.
Connect the other end of the AC Adapter cable to the Com-
Station.
5.2.
Operation
5.2.1.
Recording Events in Monitor
5.2.1.1.
Press the SEL button twice while in Home Screen. [ EVENT ] will be displayed.
5.2.1.2.
Press the ACT button.
5.2.1.3.
Wait until the word [ EVENT ] or the most recently entered event, if a previous event had been entered, is flashing.
5.2.1.4.
Press the ACT button.
5.2.1.5.
Press the Up and Down buttons to select the desired event i.e. [ FOOD ], [ MEDS ], [ EXERCISE ] or [ OTHER ].
5.2.1.6.
Press the ACT button to accept the chosen event.
5.2.2.
Reading Current and Voltage Sensor Signals in Monitor
5.2.2.1.
Press the SEL button 3 times while in Home Screen.
[ SIGNALS ] will be displayed.
5.2.2.2.
Press the ACT button to display the current, in nanoamperes, in the upper right-hand corner of the screen, together with
[ + ISIG ].
Page 10 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.2.2.3.
Press the SEL button whilst [ ISIG ] is displayed to display the voltage, in volts, in the upper right-hand corner of the screen.
5.2.2.4.
Press the SEL button to return to [ ISIG ] display.
5.2.2.5.
Press the ACT button to return to the Home Screen.
5.2.3.
Entering and Editing Patient Information and Downloading Data in Computer
5.2.3.1.
Double click the MiniMed Solutions CGMS
®
Sensor icon on the Windows desktop.
5.2.3.2.
Select [ New Patient ] from the File menu.
5.2.3.3.
Enter the patient’s first and last name.
5.2.3.4.
Enter the patient’s ID number.
5.2.3.5.
Enter the name of the physician.
5.2.3.6.
Use the Up and Down arrow buttons if the Upper and Lower
Limits of the [ Glucose Target Range ] need to be changed.
5.2.3.7.
Click on any of the Time Range Blocks if the previously set default times need to be changed.
5.2.3.8.
Press the On/Off switch of the Com-Station to switch it on.
5.2.3.9.
Check that the monitor is switched on and that it is displaying the Home Screen.
5.2.3.10.
Place the monitor face up into the Com-Station cradle.
5.2.3.11.
Check that the monitor fits well in the cradle to ensure that the infrared communications ports are aligned.
5.2.3.12.
Select [ Download Glucose Monitor ] from the [ Download
Data ] Menu.
5.2.3.13.
Click the Download button to initiate the download.
5.2.3.14.
Wait until the status bar moves through its entire range.
5.2.3.15.
Click the Ok button when [ Download Complete ] is displayed.
5.2.3.16.
Select [ Save As ] from the File Menu.
5.2.3.17.
Press the Save button to save the patient file with its default name (i.e. last name, first initial, patient ID and date).
5.2.3.18.
Select [ Open Existing Patient ] from the File Menu to open and/or edit an existing patient file.
5.2.3.19.
Select the patient file that needs to be opened.
5.2.3.20.
Click the Open button.
5.2.3.21.
The Patient Entry and Edit screen will be displayed containing information about the patient.
5.2.3.22.
Select [ Rename Patient File ] from the File Menu if the first name, second name or patient ID needs to be changed.
5.2.3.23.
Click the Ok button.
Page 11 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.2.3.24.
Click the Save button when the Save As window is displayed.
5.2.4.
Displaying and Printing Report Results
5.2.4.1.
Open an existing patient file.
5.2.4.2.
Click on the Reports Menu and choose the desired report to be viewed.
5.2.4.3.
Wait until the report is displayed for that selected patient.
5.2.4.4.
Select [ Print ] from the File Menu if report is to be printed.
5.2.4.5.
Click the Print button on the Sensor List Screen.
5.2.4.6.
Click the Ok button.
5.2.5.
Exporting Files
5.2.5.1.
Close any reports that may be open.
5.2.5.2.
Select [ Export File ] from the File Menu. The Save As window will be displayed.
5.2.5.3.
Check that the correct patient file is being exported.
5.2.5.4.
Click the Save button.
5.2.5.5.
Open the desired programme you want to use to view the data.
5.2.5.6.
Select [ Open ] from this programme’s File Menu.
5.2.5.7.
Locate the .fst file. The default location is: C:\ Program
Files\ MiniMed Inc\ MiniMedSolutions CGMS Sensor\
PatientData.
5.2.5.8.
Click on the .fst file to be opened.
5.2.5.9.
Click the Open button. The Text Import Wizard window will be displayed.
5.2.5.10.
Check that [ Delimited ] is selected.
5.2.5.11.
Click the Next button.
5.2.5.12.
Select the [ Other ] text box and type in [ | ].
5.2.5.13.
Click the Next button.
5.2.5.14.
Click the Finish button.
Page 12 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.3.
Maintenance
5.3.1.
Monitor Maintenance
5.3.1.1.
Replace batteries if [ LOW BATT ] or [ NO POWER ] is displayed and accompanying alarm is heard.
5.3.1.2.
Clean the outside surface using a moist cloth with a mild cleaning agent.
5.3.2.
Sen-serter Maintenance
5.3.2.1.
Clean using a moist cloth with soapy water.
5.3.2.2.
Disinfect using a moist cloth with 10% bleach or 70% alcohol solution.
5.3.2.3.
Allow to dry before use or storage.
5.3.2.4.
Store in the released position to maintain optimum product performance.
5.3.3.
Cable Maintenance
5.3.3.1.
Clean the outside surface using a moist cloth with a mild cleaning agent
5.3.3.2.
Disinfect using a moist cloth with 70% alcohol solution.
Page 13 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
[
[
[
5.4.
Troubleshooting
Alarm Message
[ DISCONN-] displayed and alarm beeps every 15 seconds until
Cause/s
Monitor is not detecting signal from acknowledged.
Alarm will repeat every 5 minutes as long as monitor continues not to detect a signal sensor
Sensor may be disconnected from monitor
Sensor may have pulled out of insertion site
CAL ERR
ISIG HI
] is > 90% full displayed and alarm beeps every 15 seconds until acknowledged
MEM FULL seconds until seconds until
] displayed and alarm beeps every 15 acknowledged.
Alarm will repeat every 6 hours as long as memory
] displayed and alarm beeps every 15 acknowledged.
Alarm will repeat every 3
The entered finger stick blood glucose reading is either higher or lower than expected
Memory used to store glucose data is > 90% full
Monitor detected sensor current [ ISIG ] that is significantly higher than expected
Solution/s
Check sensor-to-cable and cable-to-monitor connections
Read the [ ISIG ] for 15 seconds
If [ ISIG ] value is > 5 and remains fairly constant, wait
15 minutes and recalibrate monitor before returning to normal operation
If [ ISIG ] value is < 5 or highly variable, perform Test
Plug Procedure (Refer to
SOP/PD/214_01/A2)
Press SEL and then ACT to switch alarm off
Check [ METER BG ] screen to ensure that blood glucose reading was inputted correctly
Check monitor and cable function by performing Test
Plug Procedure
(Refer to
SOP/PD/214_01/A2)
Replace sensor if above steps do not resolve error
Press SEL and then ACT to switch alarm off
Download glucose data to a computer within 24 hours after first alarm is heard
Clear all glucose data
(Refer to
SOP/PD/214_01/A1)
Press SEL than ACT to switch alarm off
Take a finger stick glucose measurement.
Page 14 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval minutes as long as sensor current remains higher than expected
If reading > 400mg/dl (or >
22mmol/l), treat patient accordingly and recalibrate sensor once patient glucose levels become < 400mg/dl
Replace sensor if reading was < 400mg/dl, sensor connections were all properly attached and no moisture was found in the connections
Install new batteries and recalibrate sensor
[ LOW BATT ] displayed and alarm beeps once.
Alarm will repeat every hour until batteries are replaced or loose all power
Batteries have only about
72 hours of remaining power
[ NO POWER ] displayed Batteries are only expected to last less than one hour
Diagnostic test detected [ E-OX ] displayed and alarm beeps every 10 seconds until data is downloaded to pc an error in the memory programme
Immediately install new batteries and recalibrate sensor
Immediately download all data from monitor while this is switched on and then replace monitor
Page 15 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.
Flow Charts
5.5.1.
Set Up – Monitor Set Up
Start
Insert 2 new AAA alkaline batteries into battery compartment
Press On/Off button to switch monitor On
Wait until monitor performs its routine diagnostic test and Home
Screen is displayed
Press SEL button 4 times.
[ SETUP ] will be displayed
Press the ACT button
Repeatedly press SEL button to scroll through the 7 different options available in Set Up menu
(Refer to SOP/PD/214_01/A1)
Need to exit Set Up menu
Yes
Press ACT button whilst [ EXIT ] is displayed in Set Up menu
No
Page 16 of 38
End
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.2.
Set Up – Sensor Insertion
Start
Wash hands well with soap and water
Wear suitable pair of disposable gloves
Take out sensor package
from refrigerator
Allow sensor package to reach room temperature
Choose an appropriate sensor insertion site
Clean sensor insertion
site with alcohol
Allow site to completely dry
before inserting sensor
Remove sensor from its packaging
Place sensor in Sen-serter
Hold Sen-serter upright, secure white tape of sensor and remove the clear tape using an anticlockwise motion
Place fingers on back of white
tape and push carrier down until it clicks into place
Remove needle guard from introducer needle
Rest Sen-serter legs against the skin at 45 o
angle
1
Page 17 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
1
Press white button on Sen-serter
Check that sensor has been inserted properly through the skin
Hold sensor in place and gently slide
Sen-serter away from sensor
Remove white paper from adhesive pad and press it against skin
Hold sensor with 2 fingers on its base and gently remove introducer needle at a 45 o
angle
Dispose introducer needle into sharps container
End
Page 18 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.3.
Set Up – Connecting the Cable
Start
Place monitor connector end of cable above monitor connection port until slight contact is made
Align arrow on monitor connector with arrow on monitor connection port
Place the groove on side of cable tool on corresponding groove on monitor connector
Apply pressure to advance monitor connector into monitor connection port
Check that both arrows on monitor connector and on monitor connection port are still properly aligned
Grasp sensor base between
thumb and forefinger
Orient sensor connector end of cable to be in close proximity
to sensor connection
Plug sensor connector end of
cable to sensor
Press both parts together until they snap
Cover insertion site with transparent sterile dressing
View [ ISIG ] screen (Refer to 5.5.10) for 1-2 minutes to ensure that sensor current is adequate and stable
End
Page 19 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.4.
Set up – Sensor Initialisation
Start
Repeatedly press SEL button, while in Home Screen, until
[ SETUP ] is displayed
Press ACT button
Repeatedly press SEL button until [ INIT ] is displayed
Press ACT button
Monitor will beep and
[ NO ] will flash
Press Up or Down arrow button to select [ YES ]
Press ACT button to begin initialisation process
Wait for 1 hour for the initialisation process to be complete
Wait for monitor to beep to indicate that initialisation process is complete
End
Page 20 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.5.
Set Up – Sensor Calibration
Start
Sensor initialisation complete
No
Yes
Perform first sensor calibration
Perform a finger stick
glucose measurement
Press SE L button once
while in Home Screen
[ METER BG ] is displayed
Press ACT button and wait for [ --] and [ ENTER BG ] to be displayed
Use Up and Down arrow buttons to enter glucose measurement reading in increments of 1mg/dl or 0.1mmol/l
Press ACT button to confirm
Alarm heard and
[ CAL ERR ] displayed within
1-2 minutes
Yes
Entered glucose value is outside the sensitivity check expected ranges
Refer to section 5.4 for further details
No
End
Wait for initialisation to be complete
Repeat sensor calibration procedure at least 4 times daily incorporating a combination of fasting, pre-prandial and post-prandial values
Page 21 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.6.
Set Up – Using the Monitor Belt Clip
Start
Slide triangular-shaped ridge on back of belt clip into its corresponding groove on back of monitor
Keep pushing belt clip until
it snaps into place
Need to remove belt clip
Yes
Lift grooved edge of belt clip with thumb and gently slide clip out of its groove
No
End
Page 22 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.7.
Set Up – Installing the Solutions™ Software
Start
Insert CGMS
®
Solutions
™
Software CD into CD drive
Double click on My Computer icon on Windows desktop
Right click on the CD drive
Click on Open
Double click on SETUP.EXE file
Read instructions carefully and when ready click Next button
Read Software License Agreement
Click Yes button
Click Next button to accept default destination location for installation
Yes
Use default destination folder to save patient files
Click Next button
The Select Program Folder window will now be displayed
Click Next button to accept MiniMed Solutions
CGMS
®
Sensor software
1
No
Page 23 of 38
Click on Browse to select another location
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
1
Click Next button when Start Copying
Files window is displayed
Wait until setup status reaches 100%
Yes
Reboot request message displayed
Accept request to finish installation process
No
Click Finish button when Setup
Complete window is displayed
MiniMed Solutions CGMS
®
Sensor icon will now be displayed on Windows desktop
End
Page 24 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
SYSTEM GOLD
™
® Valid for:
2 years from approval
Start
5.5.8.
Set Up – Setting up the MiniMed Solutions CGMS Sensor
Programme
Double click MiniMed Solutions
CGMS
®
Sensor icon on desktop
Yes
Check that device selector switch of Com-Station is in
the wave symbol position
Select [ User Preferences ] from File Menu
Enter General
Tab Menu
No
Yes
Enter Reports
Tab Menu
No
2
Click [ General ] Tab
Click on arrow of drop down menu under Com-Port to select serial port from which Com-Station is connected to computer
Click Check Com-Station button
No
[ Com-Station
Ready ] displayed
Yes
Click Ok button
Select between a 12-hr or 24-hr time display option
Select between mg/dL or mmol/L from [ Report Glucose Units ] section
1
[ No Com Station ] will be displayed to indicate that wrong
Com-Port was selected
Click Ok button and recheck connections
Yes
Click [ Reports ] Tab
Click the check box of the
Report type to be printed
Need to change time range block
Click on Time Range Block
that needs to be changed
Click Up and Down arrows to adjust time
Click outside of time range block
to save changes made
Change more time range blocks
Yes
No
No
2
2
Page 25 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
2
Need to change
Upper and/or
Lower limit/s of default glucose target
Yes
Use the Up and Down arrow
buttons respectively
Yes
1
Enter Clinic
Tab Menu
Click [ Clinic Info ] Tab
Input desired clinic name, address and phone number
Press Ok button to save all changes made to User Preferences Menu
End
No
No
Page 26 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
5.5.9.
Set Up - Connecting the Com-Station to the Computer
Start
Connect computer cable into an available serial Com-Port
connector on computer
Connect other end of computer cable to Com-Port A of Com-Station
Connect AC adapter into a suitable power source
Connect other end of AC Adapter cable to Com-Station
End
Page 27 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
SYSTEM GOLD
™
® Valid for:
2 years from approval
Yes
Start
5.5.10.
Operation – Recording Events and Reading Current and Voltage
Sensor Signals in Monitor
1
No
No
Need to record
an event
Press SEL button twice while in Home Screen
Yes
Need to read current and voltage sensor signals
Press SEL button 3 times while in Home Screen
[ EVENT ] will be displayed
Press ACT button
Yes
Previous event had already been entered
Wait until the most recently entered event is flashing
Press ACT button
Press Up and Down buttons to select the desired event i.e. [ FOOD ],
[ MEDS ], [ EXERCISE ] and [ OTHER ]
No
Wait until the word
[ EVENT ] is flashing
[ SIGNALS ] will be displayed
Press ACT button to display current, in nA, together with [ +ISIG ]
Press SEL button whilst [ ISIG ] is displayed to display voltage
Press SEL button to return
to [ ISIG ] display
Press ACT button to return to Home Screen
End
1
Page 28 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
SYSTEM GOLD
™
® Valid for:
2 years from approval
5.5.11.
Operation - Entering and Editing Patient Information and
Downloading Data in Computer
Start
Double click MiniMed Solutions
CGMS
®
Sensor icon on desktop
Select [ New Patient ] from File Menu
Enter patient’s first and last name
Enter patient’s ID number
Yes
Enter name of physician
Need to change
Upper and/or
Lower Limits of glucose target range
No
Yes
Set default times need to be changed
Click on the desired Time Range
Block/s to be changed accordingly
No
Press On/Off switch of
Com-Station to switch it on
Check that monitor is switched on and that it is displaying Home Screen
Use Up and Down arrow buttons to change accordingly Place monitor face up into
the Com-Station cradle
Check that monitor fits well in cradle to ensure that infrared communications ports are aligned
Select [ Download Glucose Monitor ] from Download Data Menu
Click Download button to
initiate download
Wait until status bar moves
through its entire range
1
Page 29 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
1
Click Ok button when
[ Download Complete ] is displayed
Select [ Save As ] from File Menu
Press Save button to save patient file with its default name
Yes
Need to open and/or edit an existing patient file
Select patient file that
needs to be opened
Click Open button
Patient Entry and Edit screen will be displayed
Need to change 1 st name, 2 nd
name or patient ID
Yes
Select [ Rename Patient File ] from File Menu
No
Edit the information in the
[ Rename Patient File ] window
Press Ok button
Click Save button when [ Save As] window is displayed
End
No
Page 30 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
SYSTEM GOLD
™
® Valid for:
2 years from approval
Yes
5.5.12.
Operation – Displaying and Printing Report Results and
Start
Exporting Files
Need to display a report for a patient
Open an existing patient file
No
Yes
Need to print data
Select [ Print ] from File Menu
No
Click on Reports Menu and choose desired report to be viewed
Click Print icon on
Sensor List Screen
Click Ok button
Wait until report is displayed for the selected patient
No
Yes
Need to export file
Close any reports that may be open
Select [ Export File ] from File Menu
Save As window will be displayed
Check that the correct file is being exported
Click Save button
Open desired programme you want to use to view data
Select Open from this
programme’s File Menu
Locate the .fst file
2
1
Page 31 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
1
2
The default location is: C:\Program
Files\ MiniMed Inc\
MiniMedSolutions CGMS Sensor\
PatientData
Click on .fst file
Click Open button
Text Import Wizard window
will be displayed
Check that [ Delimited ] is selected
Click Next button
Select [ Other ] text box and type in [|]
Click Next button
Click Finish button
End
Page 32 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
Ref. No.
SOP/PD/214_01
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
5.5.13.
Maintenance
Start
[ LOW BATT ] or
[ NO POWER ] displayed and accompanying alarm heard
Yes
Replace batteries
Monitor needs cleaning
Yes
Clean using a moist cloth with
a mild cleaning agent
No
No
Yes
Yes
Sen-serter needs cleaning
Clean using moist cloth with soapy water
Sen-serter needs to be disinfected
Disinfect using moist cloth with 10% bleach or 70% alcohol solution
No
No
Allow to dry before use or storage
Store in released position to maintain optimum product performance
Yes
Valid for:
2 years from approval
Cable needs cleaning
Clean using moist cloth with
a mild cleaning agent
Yes
Cable needs to be disinfected
Disinfect using moist cloth with
70% alcohol solution
End
No
No
Page 33 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
6.
Precautions
6.1.
Sensor / Insertion Site Precautions
6.1.1.
Do not use a sensor if the temperature-limit indicator on its storage box is black in colour.
6.1.2.
Do not use sensor if its package has been already opened or damaged.
6.1.3.
Avoid touching any sensor surfaces that will eventually come in contact with the inside of the body.
6.1.4.
Check that the sensor is securely placed in the Sen-Serter prior to insertion since this may result in pain and minor injury.
6.1.5.
Do not point a loaded Sen-serter towards any body part where insertion is not desired.
6.1.6.
Do not twist, bend or lift the Sen-serter while removing it from the sensor.
6.1.7.
Select a new insertion site that is at least 5cm from the previous insertion site each time a new sensor is to be inserted.
6.1.8.
Do not insert sensors in locations that are constrained by clothing or are subjected to rigorous movements.
6.1.9.
Avoid sensor insertion in the area of the waistline, around navel and in scarred or atrophied tissue
6.1.10.
For users who inject insulin, instruct patient or carer to administer the injections at least 8 cm away from the sensor insertion site.
6.1.11.
For users who wear an insulin pump, ensure that the sensor insertion site is at least 5 cm away from the insulin infusion site.
6.1.12.
Do not rotate the introducer needle while it is being removed.
6.1.13.
Periodically check the insertion location for any redness, bleeding, pain, tenderness and swelling and to ensure that the sensor remained in place.
6.1.14.
Advise the patient to check the sensor regularly especially before going to bed and after waking up.
6.1.15.
Infection and/or site irritation may result from improper insertion of the sensor at the insertion site.
6.1.16.
Enter sensor calibration values immediately within 5 minutes of obtaining a fingerstick glucose value.
6.1.17.
Recalibrate the sensor whenever the sensor is disconnected and reconnected again.
6.2.
Monitor precautions
6.2.1.
Do not use the monitor in close proximity of strong electromagnetic sources such as medical imaging equipment, television and radio transmitters.
Page 34 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
6.2.2.
Check that both the sensor and the monitor are securely fastened to the patient before starting any form of exercise.
6.2.3.
Advise the patient to place the monitor in a Shower-Pak™ prior to having a shower or engaging in other activities that may wet the monitor.
6.2.4.
Keep the monitor in its original case while it is in use to protect it against bumps and scratches of everyday use.
6.2.5.
Whilst changing batteries, do not leave the monitor for more than 5 minutes without batteries to avoid losing glucose data and programme information.
6.2.6.
Ensure that the finger stick glucose meter readings inputted into the monitor are accurate since inaccurate entries may reduce the accuracy of the calculated glucose readings.
6.2.7.
Avoid taking finger stick blood glucose measurements when the glucose levels may be unstable such as immediately following a meal, insulin delivery or exercise since this may result in readings which are outside the sensitivity check expected ranges.
7.
References
Medtronic MiniMed. CGMS
®
System Gold
™
MMT-7102W user guide. Northridge
USA: Medtronic MiniMed; 2003.
8.
Appendices
SOP/PD/214_01/A1 – Set Up Menu Table
SOP/PD/214_01/A2 – Test Plug Procedure Flow Chart
9.
Revision History
Version Number
01
Amendments/ Reasons for change
Initial Release
Page 35 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
SOP/PD/214_01/A1 – Set Up Menu Table
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
Display Symbol
CLEAR
PATIENT
INIT
SET TIME
To clear the monitor from all glucose results history and calibration data once all information from monitoring period has been downloaded to a computer
To enter a patient’s unique identification number
Function Set-Up
Press ACT when [ CLEAR ] is displayed
Wait until the words [ CONFIRM ] and
[ NO ] are displayed
Use Up and Down arrows to select between [ YES ] and [ NO ].
Press ACT to clear memory if [ YES ] is chosen
Press ACT to cancel the clear memory operation if [ NO ] is chosen
Press ACT when [ PATIENT ] is displayed
Wait until [ 000-0000 ] is displayed
Use Up and Down arrows to select a number between 0-9 for the first blinking digit
Press ACT to confirm and advance to the next digit
Repeat last two steps for all 7 digits
See section 5.1.4 To start sensor initialisation period
To programme the current time and date into the monitor
Press ACT when [ SET TIME ] is displayed
Press ACT
Wait for [ HOUR ] to be displayed
Use Up and Down arrows to select hour between 1 and 12
Press ACT to confirm hour
Wait for [ MINUTE ] to be displayed
Use Up and Down arrows to select minute between 00 and 59
Press ACT to confirm
Wait for [ YEAR ] to be displayed
Use Up and Down arrows to select last two digits of current year
Press ACT to confirm year
Wait for [ MONTH ] to be displayed
Use Up and Down arrows to select digit for month between 1 and 12
Page 36 of 38
UNIT
LOCK
EXIT
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
To select between mg/dL or mmol/L as the unit for measuring blood glucose
To prevent the set parameters from being altered by patient
Press ACT to confirm month
Wait for [ DAY ] to be displayed
Use Up and Down arrows to select digit for current day between 1 and 31
Press ACT to confirm day
Press ACT when [ UNIT ] is displayed
Use Up or Down arrow to select between the two units
Press ACT to confirm chosen unit
Press ACT when [ LOCK ] is displayed
Use Up or Down arrow to select between [ YES ] and [ NO ]
Press ACT to confirm if [ YES ] is chosen
Press ACT again
Wait for [ 0 ] to be displayed
Press the following sequence of buttons to lock the set up menu:
UP ARROW
SEL
DOWN
ARROW
ACT
Wait for [ 1,2,3 ] and [ INIT ] to be displayed
User is now given the option to start the initialisation process
To unlock set up menu:
Press ACT when
[ UNLOCK ] is displayed
Press ACT again when
[ YES ] is displayed
Wait for [ 0 ] to be displayed
Press the same button sequence used to lock set up menu
Wait for [ 1,2,3 ] and
[ CLEAR ] to be displayed
Press ACT when [ EXIT ] is displayed To exit the Set Up menu and return to Home
Screen
Page 37 of 38
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
STANDARD OPERATING PROCEDURE
MEDTRONIC MINIMED CGMS
®
SYSTEM GOLD
™
SOP/PD/214_01/A2 - Test Plug Procedure Flow Chart
Ref. No.
SOP/PD/214_01
Valid for:
2 years from approval
Start
Disconnect cable from sensor
Connect test plug to cable
Read ISIG value
No
ISIG reading between 24-29nA
Yes
Sensor no longer useful / has been pulled out of insertion site or sensor connector may have gotten wet
Remove old sensor and insert new one
Initialise and recalibrate new sensor
Cable or monitor may be faulty
Disconnect cable from monitor
Yes
Replace cable with test plug
Read ISIG value
ISIG reading between 24-29nA
Cable may be malfunctioned
Disconnect test plug from monitor
Plug in a new cable to monitor
Remove old sensor and
insert a new one
No
Monitor may be malfunctioned
Call supplier
Connect new cable to sensor
Initialise and calibrate new sensor
End
Page 38 of 38