UV-Vis/Fluorescence Lab Introduction The UV-vis is used to probe the range of the electromagnetic spectrum from 200nm to 800nm. The energy is absorbed by the bonding electrons of a molecule. Qualitative data can be obtained but this technique is mainly quantitative. This experiment will determine the levels of salicylic acid in aged aspirin samples. Fluorescence is used in the detection of analyte molecules that are found at low concentrations. Not all molecules are capable of fluorescing so the systems are naturally selective. This lab uses the standards quinine and sulfuric acid to make a calibration curve. Procedure Day 1: Investigate the absorbance properties of SA by preparing a stock solution of 10mL SA in hexane 1000ppm. Determine an appropriate wavelength monitor. Create a calibration curve for the range you expect. Prove the calibration curve by creating an unknown. Day 2: Create a new calibration curve while the aspirin sample is dissolving. Filter and run the aspirin to make sure it falls in the range of the calibration curve. UV-Vis Thermo Evolution 600 UV-Vis Turn on Instrument (Power switch on the left side near the electrical cord) Open the “VisionPro” program on the adjacent computer The instrument will perform a startup check to see that everything is in working order Select the Scan icon on the tool bar (names of icons that the cursor is over will appear in the bottom left corner of the program) Four viewing windows should be on screen, in the top center one make the following adjustments: Wavelength: 190nm-400nm Scanning for: Absorbance Lamp: Tungsten Set the Baseline Correction for 100%T Prepare Blanks: Wash a quartz cuvette with hexane multiple times, fill 3/4th full, and dry with Kim wipes Place one blank in each sample holder After setting the Baseline Correction the program will have had a request for blanks, and a simple approval of the request will start the baseline scan, if denied, simply press baseline/zero icon In the same window from step 5s adjustments; set the results dropdown menu to peak pick Prepare sample/standard the same way as step seven and place in the front holder and leave blank in the back one Press the Run icon (appears to be a person running) Run each sample/standard three times without opening the instrument Repeat steps 10-12 for remaining sample/standards Click on sample name on left and right click. Click plot sample. This puts all samples on one graph. When finished: Close out program, turn off UV-Vis, turn off computer monitor [LEAVE COMPUTER ITSELF ON] Fluorescence: Shimadzu: Turn on the instrument (bottom right) and computer Open the PX-150x program Check lower right corner to see if the indication says the instrument is on. Go to Configure > go down to the PC Configuration > make sure the First Portal is selected for the instrument Configure menu > go to the Instrument and make sure that On is selected Set the instrument to Quantitative Analysis by going to Acquire Mode > selecting Spectrum With the 1000 ppm stock solution, make a 100 ppm stock solution Make 2 ppm, 3 ppm, 4 ppm, 5 ppm standards of quinine in 0.1M H2SO4 Fill a cuvette with some 2ppm solution, run the instrument under these settings: Spectrum Type: Excitation EM Wavelength: 400nm Ex Wavelength Range: 35o nm to wavelength that contains peak Sensitivity: Low Scan Speed: Fast Recording Range: -10.00 to an amount that shows entire peak EX/EM Slit Width: 10/10 Response Time: 0.02 Repeat Scan/ Auto File: No setting needed/ disable Then click Search λ and click Search. Ensure that the default range for the EX (Excitation) and EM (Emission) are 350 - 600 nm and 240 - 650nm respectively. Record optimal wavelengths found for EX and EM. Spectrum Type: Emission EX Wavelength: The ideal wavelength found in the optimal wavelength search EM Wavelength Range: Make sure that the optimal EM wavelength is included in the range Sensitivity: Low Scan Speed: Fast Recording Range: -10.00 (low); 500.00 (high) EX/EM Slit Width: 10/10 Response Time: 0.02 Repeat Scan/ Auto File: No setting needed/disable Run a scan of each concentration and record the wavelength and intensity of the largest peak. Repeat 3 times for each Make a solution of tonic water by diluting 5mL of tonic water with 95mL of H2SO4 Run a scan of this solution and record wavelength and intensity of largest peak. Repeat 3 times Calculations (UV-vis) The desired concentrations were made from a stock solution of 1000ppm and the dilutions were transferred and prepared in a 25mL volumetric flask. The dilutions were made using a M1V1 = M2V2 equation. x = 0.25 mL The calculations of the unknown data were done using the slope of the line equation: y = 0.0181x + 0.2323. Absorbance of Unknowns Samples at 236nm Trial 1 0.555 Trial 2 0.546 Trial 3 0.530 Average: 0.543 The average absorbance of the unknown data was 0.543. If you use that number in the equation of the line that was generated by linear regression analysis, as the y-value, you will obtain a fitted value for the concentration of the sample. x = 17.1657ppm The standard curve showed that there was a high percent correlation given that the rsquared value was 0.978. This shows that there was a 97.8% correlation based on the linear regression analysis. The equation that we obtained from the data showed to be very accurate and precise. Absorbance Values for Standard Concentrations (ppm) Of Salicylic Acid at 240nm Absorbance Concentration 10 0.379 10 0.37 10 0.372 20 0.587 20 0.587 20 0.583 30 0.77 30 0.771 30 0.758 40 1.025 40 1.02 40 1.004 50 1.107 50 1.113 50 1.102 60 1.427 60 1.441 60 1.416 80 1.602 80 1.598 80 1.588 Conclusion We were able to find the unknown concentration to be 17.165ppm. The standard curve generated used all three samples because we thought that this would give us the best results. The equation that was generated was adequate because the linear regression/best fit line showed a high correlation, being 98.7%. This means that there was a strong relationship. Results (Fluorescence) Standard Concentrations of Quinine at 250nm Concentrations (ppm) Absorbance 0.1 2.477 1 2.491 1 2.4889 5 3.006 5 3.006 10 3.006 10 3.006 The results showed the equation of the line to be y = 0.058x = 2.5172. The regression analysis showed that it was not very strong. However, it was good enough to be used for determination of quinine in an unknown sample. Absorbance of Quinine if Unknown sample at 508nm Trial 1 3.006 Trial 2 3.006 Trial 3 3.006 The results showed that the absorbance of quinine in the unknown sample was 3.006, which was similar to was obtained in the higher concentrations. Calculations The desired concentrations were calculated similar to how the standard concentrations were determined in the UV-Vis experiment. The concentrations were prepared and made by using the M1V1=M2V2 equation. x = 2.5mL The desired standard concentrations were all made using this calculation. The unknown concentration was found using the equation of the line generated by the calibration curve. x = 8.4258ppm The unknown concentration was found to be 8.4258ppm for the unknown sample of quinine. Conclusion We found the unknown concentration of quinine to be 8.4258ppm. This does not seem to be as accurate, when compared to the UV-vis, because the experiment showed to give consistent absorbance values from all of the standards after 5ppm. We were not able to determine if this result was accurate.