American Association of Critical Care Nurses (AACN) Recommendations Related to Hemodynamic Monitoring Protocols General Guidelines for Arterial and Pulmonary Artery Monitoring Use saline rather than dextrose flush solutions. Cover all stopcocks with sterile occlusive (dead-end) caps and keep in place at all times. Level transducer stopcock to phlebostatic axis On insertion Whenever patient position is changed Prior to all readings With patient supine, up to 45 headrest elevation Zero transducer at level of phlebostatic axis On insertion Whenever values significantly change When cable is disconnected-reconnected Note: Controversial to zero routinely at beginning of each shift, no studies to support and frequent opening of the system increases chances of infection Assess system adequacy by performing a dynamic response test (square wave test) and mounting strip in chart On insertion At beginning of every shift Whenever accuracy is questioned After zeroing Set alarm parameters according to patient status. Change flush bag/pressure tubing/transducer Every 96 hours (arterial line systems) Every 72 hours (PA line systems) Note: Combined systems of arterial and PA lines are not addressed in AACN Protocols for Practice. Change flush bag ONLY when tubing is changed or when solution is depleted. Document waveform strips at the beginning of each shift or when a significant change in condition occurs. Use waveform strips for pressure determination rather than the digital value on the monitor. Run waveform strips simultaneously with ECG, not separately. Do not apply anti-microbial ointment to insertion sites. Nurses who have demonstrated competency may remove catheters with a physician order. Arterial Lines Document pulses, color, temperature, sensory function of limb distal to arterial line insertion site every 4 hours (or more frequently if there is a problem) while the catheter is in place and for several days after removal. Remove/replace catheter after 4 days. PA Catheters Document all waveforms during insertion. Inflate the balloon with 1.5 ml of air or less for no more than 15 seconds. The PA diastolic pressure may be used as an indicator of PCWP except when There is disease causing high pulmonary vascular resistance (pulmonary hypertension) There is tachycardia of 130/min or >. Do not remove patient from the ventilator to obtain pressures. Notify physician immediately if waveform changes (and document with a strip) when routine measures do not correct Spontaneous RV Spontaneous PCWP Development of large v waves during PCWP inflation Remove/replace catheter after 5 days Use universal precautions (gloves) when handling PA catheters. (Continued) Hemodynamic Monitoring – AACN Recommendations 73 Cardiac Output Determination (Intermittent) Dextrose solution is preferred over saline. Ensure computation constant is accurate. Measure cardiac output with the patient backrest at 20 or <. Use room temperature or iced injectate. Use 10 ml injectate unless fluid restriction is essential, then use 5 ml Use 10 ml iced injectate for patients with very low CO or hyperthermia. Administer injectate within 4 seconds or less. Administer injectate at end expiatory phase of ventilatory cycle. Determine the CO by averaging 3 values within the middle (mean) value. Assess the CO curve and delete the CO value if the curve is abnormal. Index the CO to the body surface area (BSA) to obtain the cardiac index and document. Replace CO flush solution and tubing every 72 hours. Do not infuse vasoactive meds through proximal (right atrial) port used to do cardiac outputs. From AACN Protocols for Practice, 1998 and AACN Procedure Manual for critical Care, 2001 Hemodynamic Monitoring – AACN Recommendations 74