SOLAS HALOCARBON INSTRUMENTATION AND CALIBRATION Measurement of the important halogenated trace gases relevant to ozone depletion and climate change are measured using the MADS GC-MS (Modified Adsorption Desorption System) instrument. At the heart of the MADS is the Peltier cooler which maintains a temperature of 40C, even with a substantial heat load, using two simple three-stage Peltier coolers. The microtrap can independently be heated resistively to any temperature from 40C to about +210C in about four seconds. The use of a microtrap permits the desired analytes from 3-litre air samples to be pre-concentrated and effectively separated from more-abundant gases that would otherwise interfere with chromatographic separation or ion generation in the GC-MS source, such as N2, O2, Ar, H2O, CO2, CH4, Kr and Xe. Importantly, the capillary column dimensions of the microtrap, without the need for internal packings mean that the flow rate through the trap is unimpeded, blank and artifact formation are negligible and also means that the resulting injections to the Agilent 5973 GC-MS are sharp and reproducible. By trapping and eluting analytes at low temperatures, the range of compounds that can be measured include a number of important volatile compounds. The MADS system uses a highprecision integrating mass flow controller (MFC) for accurate measurement of sample volumes. In addition, significant advances have been made in the software to control and acquire data from the MADS and the GC-MS itself, so that the entire system has programmability, versatility and ease of operation comparable to that of the Advanced Global Atmospheric Gases Experiment (AGAGE) GC-MD instruments. Each ambient air measurement is bracketed by the measurement of fixed loop injection of halocarbons from an oven containing permeation tubes of halocarbon standards. For compounds not contained in permeation tubes, a secondary method of calibration utilises calibration standards, which are generated at the University of Bristol from high concentration cylinders diluted using a specially developed system. The field standards (contained in pressurized 35L Essex Cryogenic stainless steel canisters) are inter-compared, before and after use, by reference to gold standards maintained at the University of Bristol and Scripps Institution of Oceanography (SIO). Gold standards have a lineage, which is linked to primary calibration standards maintained at SIO. Figure 1 The MADS measures 3-liter samples, at a frequency of 60 minutes per measurement. A list of the currently calibrated halocarbon gas species it measures is given in Table 1. The microtrap consists of a 75 cm long length of ultimetal tubing (PoraPLOT Q), 0.53mm I.D. Cooling of the coiled microtrap (Figure 2) is achieved by holding it in an anodized block between two triple stage thermoelectric coolers (Peltier coolers). Figure 2 Species C2H5I CHBrCl2 CH2ClI 2-C3H7I 1-C3H7I CHBr2Cl CHBr3 CH2I2 Table 1 The flow scheme of the MADS (Figure 3) permits direct concentration and focusing of analytes prior to GC-MS analysis. The “strip chart” shows (Figure 4) how key operating parameters are changed over the course of the 60-minute measurement cycle. A sample is taken from any of 6 pressurized inlets flows (V1) through a counter-purged Nafion drier before passing through the microtrap (V4) to a filter and MFC (set to 100 scc/min) where the sample volume is determined before exhausting to vacuum. After trapping, the microtrap is post-flushed with He from an electronic pressure controller to remove residual N2, O2, Ar and CH4. The microtrap is then heated to 210C and the sample is back-flushed into the GC/MS. Separations are performed on a PoraBOND-Q capillary column, which is ramped from 40C to 220°C during the main analysis to allow less-volatile analytes to elute sharply. Figure 3 Figure 4 In routine operation of the MADS, there are substantial investments in optimization and data collection and processing. The Agilent GC-MS control, tuning and acquisition software has been completely replaced by custom software under the Linux operating system which runs both the MADS “front end” and the GC-MS in selective ion mode (SIM). This software builds upon the base “gccontrol” developed for the AGAGE program, but now includes the mass/charge ratio as a variable, as well as the many control and diagnostic parameters of the MADS. Blanks and instrument linearity’s are measured routinely. This software is continually being updated in response to the needs of the station operators and the data processors. An important advance in the MADS is its ability to check its linearity by injecting a wide range of sample loops (from permeation tubes) or standard gas volumes (from pressurized cylinder). Such measurements were used throughout the development process to optimize trapping materials and operating parameters, and they are now part of routine diagnostics which confirm that the MADS is linear over wide ranges of sample volume, concentration, and composition. Such linearity and compositionindependence are critical to accurate calibration, especially when propagating synthetic primary standards or when measuring samples spanning wide concentration ranges. By using quantifier and qualifier ions for each measured species, the MADS also offers improved peak identification and immunity from interference by co-eluting species. As noted above, using custom AGAGE software has had many benefits, including complete source-code control over all instrument operation software, integration and data processing algorithms, and the ability to improve the software interactively. Its success is borne out by the fact that portions of the software have been adopted by other leading laboratories engaged in atmospheric and oceanic trace gas measurements, including NOAA/GMD, CSIRO and the European SOGE program. Chromatograms are acquired and displayed in real time, and are stored with a 4:1 compression ratio. The software allows the operator to quickly review and batchintegrate chromatograms, and produce time series and diagnostic plots of integration results to assess instrumental performance.