Progress_report7

advertisement

Progressive report 7

Characteristics of long pass filters installed on the Monochromator and verification of 2 photodiodes.

1. Characteristic of long pass filters

As I mentioned in my previous progressive report (Progressive report6), I updated my

IDL code to read exactly output signal of photodiode in a given wavelength and measured the output of photodiode. That measured values are also compared to direct measured values to verify IDL code and characteristics of long pass filters. Generated

Light sources are intermediate level to prevent saturation in the long wavelength.

Below table shows the setup for measurement of each filter

Filter number

2

3

4

5

Start wavelength(nm)

300

560

635

685

End wavelength(nm)

340

630

695

750

Intensity of light source

100

50

35

35

Gain selection

High

Low

Low

Low

Below 4 figures show the comparison between measurement with updated IDL code

(solid line) and direct measurement (dotted line). As we can see from all figures, measurement with updated IDL code coincides with direct measurement one. It represents IDL code can read the output signal exactly in a given wavelength.

In addition, we can see the characteristics of each long pass filters as a function of wavelengths. Cut off wavelength of each filters are 320nm (filter 2), 590nm (filter 3),

665nm (filter 4) and 715nm (filter 5), respectively.

<Figure 1, Verification of long pass filters and IDL code>

2. Verification of 2 photodiodes

I simply tested 4 channels (ch0 – ch 3) with 2 photodiodes (Ch0- Si of photodiode1, Ch1-

Ge of photodiode 1, Ch2-Si of photodiode 2, Ch3-Ge of photodiode 2) to verify their working and IDL code and saved raw data as a function of wavelength (300nm – 1800nm with 100nm increase).

I selected filter 1 which is open and grating 1 which has 1200mm/g with blaze=500nm.

You should note that only photodiode 1 can be installed on the monochromator and measure the output light from monochromator. Si of photodiode 1 looks most sensitive at around 1000nm wavelengths. It shows very similar results of company’s QE documents. Ge of photodiode 1 is also working well in the long wavelength (above

1000nm). But their values are too small since I did not select high gain. So if we use different gratings having different blaze wavelength such as blaze =1μm with high gain, we may possible to get better results from Ge.

<Figure 2, Measurement of output as a function of wavelength of photodiode 1. Left is for Si and right is for Ge>

Below figures show the measurement of photodiode 2. But it (channel 2 and 3) is just connected to bud box and slightly open the cab on the room. This is why figures of ch 2 and 3 show constant in the long wavelengths. Variations of plot on the first part of figures are caused by intentionally moved photodiode cap to see and check their working. Finally, you should note that x-axis does not mean real wavelength. Instead of it, we read all values at a same time when we read the photodiode 1.

<Figure 2, Measurement of output as a function of wavelength of photodiode 2. It is not installed on monochroamator and placed on the table instead with slightly open the cab.

Left is for Si and right is for Ge>

Based above test, I verify all channels and 2 photodiodes are working well.

Download