Progressive report 6 Test of Superlogics and verification of cut off wavelengths of long pass filters 1. Test of superlogics and unit of output signal of photodiode I measured output signal of photodiode with superlogics and verify exact unit of output signal from photodiode. I directly measured output signal of photodiode with oscilloscope and then read output of photodiode with superlogics in IDL again. Below two figures show the setup of devices for measurement. <Figure 1. Setup of devices to measure the output signal from photodiode> I made a measurement from 4 different cases; saturation, intermediate light on, low gain and high gain, respectively. To measure the output signal with low and high gain, I switch up and down during the exposure. Below tables show the measurement of saturation case and intermediate light on case. Type of Measurement with oscilloscope Measurement with Superlogics photodiode (Volts) (Volts) Low gain High gain Low gain High gain SN1(Si) +13.1 ~13.6 +13.1 ~13.6 10.0000 10.0000 SN1(Ge) 0.006 ~ 0.016 -0.013 ~ 0.006 0.02000 0.17500 SN2(Si) +13.0 ~ 13.8 +13.0 ~ 13.8 10.0000 10.0000 SN2(Ge) 0.028 ~ 0.153 -0.006 ~ 0.022 0.00400 0.05100 <Table 1. Comparison of saturation level between Oscilloscope and super logics> Type of photodiode Measurement with Superlogics (Volts) Low gain High gain SN1(Si) 0.251000 2.42900 SN1(Ge) -0.00600000 -0.0490000 SN2(Si) 0.286000 2.81100 SN2(Ge) -0.0170000 -0.152000 <Table 2. Measurement of output with superlogics in the intermediate light on> As you can see above tables, the unit of output through the superlogics is Volts. In addition, real saturation level of photodiode is above 13V but superlogics can only read up to 10V. So we should keep in mind our superlogics can’t measure above 10V. Below figures show well our superlogics can read and give exactly same level of input signal only below 10V. I generated 5 different square wave as for input signals (+/- 0.01V, +/-0.1V, +/-1V, +/5V, +/-10V) using a function generator and then read them with superlogics in IDL. Distortion shown on the plot is caused by impedance matching problem of BNC cable. But we can ignore it for simple comparison of input and output signal. <Figure2. Output signals with superlogics. The output is exactly same of input signal created by function generator> I used same way to check the different channels (Ch2, Ch3, and Ch4) of superlogics. Below figures show the measurement using different channels of super logics. I feed the square wave (+/- 100mV) generated by function generator into the ch2, ch3, ch4 of superlogics. It shows all channels are working properly. 4 channels (ch0~ch3) used for measurement will be connected to two photodiodes to get the signals from them. <Figure 3. output signals from different channels, ch2, ch3, ch4> Based on the above results, our all devices such as photodiode, superlogics and designed bud box are working well and the output unit of photodiode through the superlogics is Voltage. The saturation level of photodiode is around13.6V but our superlogics can measure only up to 10V. In addition, high gain setting multiply data by 9.7 for the Si channel of photodiode1 (SN: 60727-1) and 9.8 for the Si channel of photodiode2 (SN: 60727-2), respectively. For high gain setting of Ge of photodiode1 (SN: 60727-1) and photodiode2 (SN: 60727-2) multiply 8.17 and 8.94, respectively. 2. Verification of cut off wavelength of long pass filters Our monochromator is installed 4 long pass filters with different cut-off wavelengths. Below Table shows currently installed long pass filters and their cut-off wavelengths. Position number 1 2 3 4 5 6 Cut-off wavelength (nm) Open 320 590 655 715 Blocked We measured output voltage to verify cut off wavelengths of long pass filters. Below tablae and figure 4 shows cut off wavelength and output signal using each filter. I measured the output signal with superlogics in some specific wavelengths (10 or more specific wavelengths). Due to the small number of data number, the plot looks discrete. filter2 Wave(nm) 295 300 305 310 315 316 317 318 319 320 322 325 327 330 335 340 350 360 365 370 380 Output(V ) 0.387 0.387 0.388 0.43 0.619 0.681 0.751 0.83 0.92 1.015 1.234 1.621 1.895 2.354 3.242 4.279 6.4 8.119 9.019 10 10 filter3 Wave(nm ) 540 550 560 570 580 585 586 587 590 600 Output(V ) 0.055 0.063 0.094 0.368 3.128 8.101 9.556 10 10 10 filter4 Wave(nm ) 615 625 635 645 655 660 663 665 666 667 670 680 Output(V ) 0.043 0.044 0.047 0.059 0.332 1.601 3.848 6.478 8.27 10 10 10 filter5 Wave(nm ) 665 675 685 695 705 707 710 711 712 715 725 Output(V ) 0.05 0.07 0.146 0.569 3.396 4.803 7.811 9.121 10 10 10 (a) filter2 (320nm cut off) (b) filter3 (590nm cut off) (c) filter4 (665nm cut off) (d) filter5 (715nm cut off) <Figure 4. Measurement of output voltage with long pass filters. Dotted vertical lines in each panel represent cut off wavelength, respectively> Fig 5 is same of figure 4. It is created with smaller readout step than figure 4. All plots look smoother than figure 4 but output values are not exactly matched to the wavelengths because it is continuously varying due to the setup of scan rate. (a) filter2 (320nm cut off) (b) filter3 (590nm cut off) (c) filter4 (665nm cut off) (d) filter5 (715nm cut off) <Figure 5. Measurement of output voltage with smaller step than figure 4. Dotted vertical lines in each panel represent cut off wavelength, respectively> As we can see from figure 4 and 5, filters looks working well in designated cut off wavelengths. But I need to measure output with given wavelength exactly. Due to only one RS232 cable, it is hard to measure output in a given wavelength by monochromator. In addition, filter2 (a figure 4 & 5) plot is different above 330nm. And since output is saturated above the cut off wavelengths, we can not see any slope or change above cut off wavelengths. So we need to decrease the power of light source not to be saturated in the long wavelength. I measured the output again using fitler4 with multiple RS232 connectors with appropriate intensity of light source. But measurement using current my IDL code can not exactly match to the specific wavelength as you can see from figure 6. Measurement value with IDL code is different to directly measured one which is measured with fixed wavelength without using scan rate. This difference is caused by continuously varying of monochroamtor’s gratings when I use the scan rate command. <Figure6. Comparison of filter 3 with and without our IDL code to read the output> NEXT Task. I will update my IDL code to read exactly output signal of photodiode in a given wavelength with an appropriate intensity of light and find characteristics of long pass filters installed in monochromator.