Separation and Purification of Two Unknown Liquids by Simple Distillation and Two Unknown Solids by Acid-Base Extraction and Recrystallization and Identification by Mass Spectrometry, Infrared Spectroscopy, and Nuclear Magnetic Resonance Audrey Lullo CHE 231L Section 801. Department of Chemistry, DePaul University, 1110 W. Belden Avenue, Chicago, IL 60614. Annym2693@aol.com. 19 March 2014. ABSTRACT: The separation, isolation, purification and identification of two unknown liquids and two unknown solids is performed through simple distillation, liquid-liquid extraction, mass spectrometry, infrared spectroscopy, and both proton and carbon nuclear magnetic resonance respectively. Simple distillation is used to separate and purify a mixture of two miscible liquids. To separate two unknown solids the process of liquid-liquid extraction is used. The separate and pure liquids and solids were determined to be: acetonitrile for the low boiling liquid unknown #8, 1-butanol for the high boiling liquid unknown #8, 3-bromo benzoic acid for the carboxylic acid solid unknown #20, and benzophenone for the neutral solid unknown #20. INTRODUCTION: There are many ways in which to separate, isolate and purify compounds depending on whether or not these compounds are liquids or solids. One way in which to separate and purify two miscible liquids is simple distillation. Simple distillation only works to separate two liquids that are completely miscible with each other and that boil under 150°C. Not only must both liquids boil under a certain temperature, they must also boil at temperatures that are different by at least 25°C in order to achieve complete separation.1 Simple distillation also works to purify a liquid because impurities will boil off at different temperatures than that of the pure sample. When working with an impure sample of acetone, simple distillation would separate the pure acetone from the impurities because the impurities have very low vapor pressures making condensation and vapor nearly impossible. Therefore, the pure acetone will separate from the impurities resulting in a pure sample. The theory of distillation is essentially that, vapor pressure and temperature have a direct relationship. While one increases so does the other. This is exemplified in the Clausius-Clapeyron equation: π = π°ππ₯π [ −π₯π» 1 1 ( − )] π π π° In this equation p° and T° are both known vapor pressure and temperature respectively, ΔH is the heat of vaporization, R is the gas constant, T is the ideal temperature, and p is calculated.2 This equation shows the relationship between vapor pressure and temperature. Because these liquids have been separated and purified, they can be characterized by physical properties such as boiling point and density. Simple distillation is not only useful in lab, but also is widely prevalent in the food industry. For example, distillation is used in the processing of pectin, hop extract, egg lecithin, and decaf coffee beans.3 Simple distillation is also used in the petroleum oil industry to refine crude oil into a usable form for homes, cars, and airplanes (gasoline, kerosene, and diesel fuel). These different types of fuel are categorized and separated into solutions with similar boiling points by the distillation process.4 Therefore, distillation can be used whenever there is a need to separate and purify a mixture of two miscible liquids. Unlike for liquids, two separate techniques must be used to separate and purify an unknown solid. One technique of separation is extraction. For extraction to work, the solid must be dissolved in a solvent that is not miscible with another solvent. Because these solvents are immiscible, two separate layers form and can therefore be separated from one another. The other extraction method, acid extraction, is useful in separating contaminants such as heavy metal from the soil.5 Therefore, extraction is a useful tool applicable to not only lab but also in the larger environment to separate a solid mixture with two immiscible liquids. To purify this solid that was just separated by liquid-liquid extraction, a process called recrystallization is utilized. In order to recrystallize a solid, the solvent it is dissolved in must be able to dissolve the solid while hot but not while cold, and keep impurities dissolved at both temperatures.2 After, the solids are dissolved in the hot solvents and precipitated back out of solution, the solids will be pure because the impurities stay dissolved no matter what the temperature of solvent. Separating and purifying an unknown solid is an important step in determining the identity of the compound because then the pure sample can be characterized by melting point. In order to accurately determine the structure and identify the solid and liquid unknowns, a combination of mass spectrometry (MS), infrared spectroscopy (IR), and carbon nuclear magnetic resonance (13C NMR) and proton nuclear magnetic resonance (1H NMR) was used. Mass spectrometry is used to determine the molecular weight of a compound. Knowing the molecular weight is crucial in determining possible molecular formulas. The accuracy of various formulas is compared to a calculated index of hydrogen deficiency (IHD), which is used to determine if pi bonds or rings are present, and to what degree. The mass spectrometer works by fragmenting a compound into cations by passing it through a beam of electrons in a vacuum. These fragments appear as different sized peaks graphed based on a mass to charge scale. Therefore, the peak farthest down the spectrum is the molecular ion peak, it expresses the molecular weight of the compound and is denoted (M+). The tallest peak on the spectrum is known as the base peak. Combining all of this information, a relatively accurate molecular formula for a compound can be determined, which is an important step in identifying an unknown.6 Mass spectrometry has practical uses outside of lab. For example, it is used to efficiently test for pesticides and toxins in food. Mass spectrometry can help to identify multiple pesticides at once, therefore, making food healthier and safer.7 Another technique used to determine the identity of a compound is infrared spectroscopy (IR). IR is used to determine possible functional groups in a compound.8 It is possible to determine the functional groups because the functional groups are made of covalent bonds that are depicted like a spring with balls attached to it. This model allows the photons and radiation essential to infrared spectroscopy to excite the atoms until they vibrate. Each atom vibrates at a specific frequency and this frequency is plotted on the IR spectrum. Infrared spectroscopy is not only used in lab to determine the functional groups and types of bonds a compound contains, but it is also a promising technique used in the forensic sciences. Forensic scientists use IR to analyze various inks to determine the presence of forgery and also sweatprints, which have become as valuable as the fingerprint.9 Nuclear magnetic resonance is the final technique necessary to determine the actual structure of a compound. It combines the molecular formula determined from the MS and the functional groups determined from the IR and plots unique hydrogens and carbons on the 1H NMR and 13C NMR respectively. These hydrogens and carbons are considered unique based off of the environments they are in (i.e. bonds, relation to electronegative elements, and the number of atoms they are bonded to). NMR takes advantage of the natural spin individual nuclei have.10 The differences between the magnetic spin and relative magnetic moment each nuclei experiences is the peak seen on the NMR spectra.11 NMR is particularly helpful in the process of making and analyzing drugs to determine the compounds attached to certain enzymes and proteins.12 RESULTS: The results from the acid-base extraction and recrystallization of unknown solid #20 are depicted in Table 1. Upon separation, the aqueous phase formed a powdery, white carboxylic acid solid. After recrystallization, this solid was light and featherlike. The organic phase formed an off-white crystalline solid. After recrystallization this neutral organic solid was no longer off-white, but instead, white and powdery. The percent recovery for the crude carboxylic acid solid was 39.31% and the percent recovery for the crude neutral organic solid was 45.42%. After purification, the percent recoveries for the carboxylic acid solid and neutral organic solid were 32.26% and 35.81% respectively. The melting points were then determined for the pure carboxylic acid solid and the pure neutral organic solid in order to characterize unknown # 20. The melting point for the pure carboxylic acid solid was 158.3-160.3°C. The melting point for the pure neutral organic solid was 38.6-42.5°C. Table 1: Results from the Acid-Base Extraction and Purification of Unknown Solid # 20 Unknown Mixture #20 Carboxylic Acid Solid Neutral Organic Solid (RCO2H) (RZ) Physical Composition Powdery, White, Solid Crystallized, Off-White, Solid Mass of Original Mixture 5.0017 (g) Mass of Crude Product (g) 1.9663 2.717 Mass of Purified Product 1.6135 1.7913 (g) Percent Recovery Crude 39.31 45.42 (%) Percent Recovery Purified 32.26 35.81 (%) Pure Melting Point (°C) 158.3-160.3 38.6-42.5 Table 2 outlines the results obtained from the simple distillation of two unknown miscible liquids. After distilling unknown liquid # 8, the two separate liquids were then characterized by boiling point and density. Table 2: Results from the Simple Distillation of Unknown Liquid #8 Unknown Mixture # 8 Low Boiling Liquid High Boiling Liquid Physical Composition Clear liquid, Sharp odor, Low viscosity Mass of Original Mixture 15.6911 (g) Mass of Separated Product 4.0721 11.0071 (g) Percent Recovery (%) 96.10 Boiling Point (°C) 56.8-57.1 80.2-82.5 Density (g/mL) 0.4566 0.7823 Unknown liquid # 8 was clear and had a low viscosity. Upon heating, it released a sharp odor. After the simple distillation the percent recovery of unknown liquid # 8 was 96.10%. The separate liquids were then heated to determine the temperature at which each boiled. This temperature was also the most stable temperature. The boiling point for the low boiling liquid was 56.8-57.1°C. The boiling point for the high boiling liquid was 80.2-82.5°C. Boiling point and density are both physical properties used to characterize liquid unknown #8. The density of the low boiling liquid was 0.4566 g/mL while the density for the high boiling liquid was 0.7823 g/mL. Table 3 outlines the results obtained from mass spectrometry of the low boiling and high boiling liquids of unknown #8 and the neutral and carboxylic acid solids of unknown #20. The peaks in the spectrum were analyzed to determine possible lost fragments and the IHD was calculated. Table 3: Mass Spectral Data for Unknown Liquid #8 and Unknown Solid #20 Unknown Liquids Unknown Solids Unknown code #8 Low #8 High #20 COOH #20 Neutral Boiling Boiling Molecular formula C2H3N C4H10O C7H5O2Br C13H10O4 IHD 2 0 5 9 M+ (m/z) 41 74 200 182 Base Peak (m/z) 41 56 200 105 Lost Fragment N/A M-18 N/A M-77 Other Peak(s) (m/z) 40, 39,14 41, 31, 27 183,155,75,50 77, 51 Lost Fragment M-1, M-2, M-33, M-43, M-17, M-45, M-105, MM-27 M-47 M-125, M-150 131 The mass spectrum for low boiling unknown liquid #8 expressed a molecular ion peak (M+) peak at m/z of 41, which also happened to be the base peak. The spectrum also showed significant peaks at m/z 40, 39, and 14 with lost fragments at M-1, M-2, and M-27 respectively. The molecular formula was determined to be C2H3N with an IHD of 2. The mass spectrum for the high boiling unknown liquid #8 expressed a molecular ion peak (M+) at m/z of 74 and a base peak at m/z of 56, which corresponds to a lost fragment of M-18. The spectrum also showed significant peaks at m/z 41, 31, and 27 with lost fragments of M-33, M-43, and M-47 respectively. The molecular formula was determined to be C4H10O with an IHD of 0. The mass spectrum for the carboxylic acid unknown solid #20 expressed a molecular ion peak (M+) and a base peak at m/z 200. There were other significant peaks at m/z 183, 155, 75, and 50 with corresponding lost fragments at M-18, M-46, M-126, and M-151 respectively. The molecular formula was determined to be C7H5O2Br with an IHD of 5. The mass spectrum for the neutral solid unknown #20 expressed a molecular ion peak (M+) at m/z of 182 and a base peak at m/z of 105, which corresponds to a lost fragment of M-77. There were other significant peaks at m/z 77 and 51, with lost fragments of M-105 and M-131 respectively. The molecular formula was determined to be C13H10O4 with an IHD of 9. Table 4 outlines the results obtained from IR spectroscopy of the low boiling and high boiling liquids of unknown #8 and the neutral and carboxylic acid solids of unknown #20. The unknowns were characterized by the wavenumber, intensity, and shape of the peaks present in the spectra. Table 4: IR Spectral Data for Unknown Liquid #8 and Unknown Solid #20 Unknown Liquids #8 Unknown Solids #20 Low Boiling High Boiling R-Z R-COOH Wavenumber (cm-1) 3600 3000 2250 1450 - Intensity/ Shape Low intensity, dull Medium intensity, sharp High intensity, sharp Medium intensity, broad Wavenumber (cm-1) 3320 - - 2970 1495 1100 Intensity/ Shape High intensity, broad High intensity, sharp Medium intensity, sharp Medium intensity, sharp Wavenumber (cm-1) 3100 Intensity/Shape 1680 High intensity, sharp 1700 1495 Medium intensity, sharp 1550 1300 High intensity, sharp 1495 - 1290 High intensity sharp 1300 Low intensity, sharp Wavenumber (cm-1) 3000 Intensity/ Shape Medium intensity, dull High Intensity, sharp Medium intensity, sharp Medium intensity, sharp High intensity, sharp The low boiling unknown liquid #8 spectra expressed a low intensity, dull absorption band at 3600 cm-1 and a high intensity, sharp absorption band at 2250 cm-1. There were other significant bands at 3000 cm-1 and 1450 cm-1. The high boiling unknown liquid #8 spectra showed a high intensity, broad absorption band at 3320 cm-1 and high intensity, sharp band at 2970 cm-1. There were other significant bands at 1495 cm-1 and 1100 cm-1. The neutral unknown solid # 20 spectra expressed a low intensity, sharp absorption band at 3100 cm-1 and a high intensity, sharp band at 1680 cm-1. There were other notable bands at 1495 cm-1, 1300 cm-1, and 1290 cm-1. The carboxylic acid unknown solid #20 spectra showed a medium intensity, dull absorption band at 3000 cm-1 and a high intensity, sharp band at 1700 cm-1. There were other notable bands at 1550 cm-1, 1495 cm-1, and 1300 cm-1. Table 5 outlines the results obtained from the 1H NMR and 13C NMR for the low boiling and high boiling liquids of unknown #8 and for the carboxylic acid and neutral solid of unknown #20. The spectra were analyzed by chemical shift, multiplicity, and integration values. Table 5: 1H NMR and 13C NMR Spectral Data for Unknown Liquid #8 and Unknown Solid #20 1H NMR Data 13C NMR Data Unknown πΏ (ppm) πΏ(ppm) M I Low Boiling Liquid # 8 2 s 3h 116 High Boiling Liquid # 8 1 t 3h 14 1.5 m 2h 20 2.2 s 2h 34 3.6 t 2h 63 COOH Solid # 20 7.4 t 1h 85 7.7 d 1h 122 8 t 1h 127 8.1 s 1h 130 132 135 166 Neutral Solid # 20 7.5 m 1h 127 7.8 m 2h 130 132 138 196 The 1H NMR spectra for the low boiling unknown liquid #8 expressed one peak (in pppm) of πΏ 2 (s, 3h) and the 13C NMR expressed one peak at 116 ppm. The 1H NMR spectra for the high boiling unknown liquid #8 showed four peaks (in ppm): πΏ 1 (t, 3h), 1.5 (m, 2h), 2.2 (s, 2h) and 3.6 (t, 2h). The 13C NMR spectra for the high boiling liquid also showed four peaks (in ppm): πΏ 14, 20, 34, and 63. The 1H NMR spectra for the carboxylic acid unknown solid #20 expressed four peaks (in ppm): πΏ 7.4 (t, 1h), 7.7 (d, 1h), 8 (t, 1h), and 8.1 (s, 1h). The 13C NMR spectra for the carboxylic acid solid showed seven peaks (in ppm): πΏ 85, 122, 127, 130, 132, 135, and 166. The 1H NMR for the neutral unknown solid #20 expressed two peaks (in ppm): πΏ 7.5 (m, 1h) and 7.8 (m, 2h). The 13C NMR for the neutral solid expressed five peaks (in ppm): πΏ 127, 130, 132, 138, and 196. DISCUSSION/CONCLUSION: The unknown liquid #8 was separated and purified by simple distillation and then identified using a combination of mass spectrometry, infrared spectroscopy, and 1H NMR and 13C NMR techniques. The unknown solid #20 was separated and purified by liquidliquid extraction and recrystallization, and then identified also by using a combination of mass spectrometry, infrared spectroscopy, and 1H NMR and 13C NMR techniques. Unknown liquid #8 was separated into a high boiling (80.2-82.5°C) liquid with a density of 0.7823 g/mL and low boiling (56.8-57.1°C) liquid with a density of 0.4566 g/mL. Both of these liquids were clear and had a low viscosity and sharp odor. The percent recovery was 96.10%. A loss of product most likely occurred during the transfer from vial to round bottom flask in the first step. The mass spectra for the high boiling point revealed a molecular ion peak (M+) at m/z 74, meaning the compound has a molecular mass of 74 g/mol. The base peak was at m/z 56 meaning the lost fragment was most likely H2O. There was also a significant peak at m/z 31, which corresponds to an OCH3 fragment. Therefore, keeping this in mind, the molecular formula was determined to be C4H10O with an IHD of 0, meaning there were no triple bonds, double bonds, or rings. The IR spectra revealed an O-H bond with a broad absorption band at 3320 cm-1, and encouraged the idea that there were no rings, double bonds, or triple bonds due to the lack of bands in those regions. The proton and carbon NMR each revealed four unique sets of hydrogens and carbons respectively. Therefore, the high boiling liquid for unknown #8 was determined to be 1-butanol. This is a clear, water soluble liquid with a boiling point of 117.6°C and a melting point of -89°C°.13 The difference seen in literature values versus experimental values in boiling point are potentially due to human error. The high boiling liquid would jump between 80°C and 82.5°C and would not go any higher. Therefore, this was determined to be the boiling point range. The mass spectra for low boiling liquid unknown #8 expressed a molecular ion peak (M+) at m/z 41. This means that the compound has a molecular mass of 41 g/mol and most likely contains a nitrogen due to the odd molecular ion peak. The molecular ion peak was also the base peak, meaning the total compound was the biggest fragment. There were other peaks at m/z 40 and m/z 39, which most likely represent the loss of hydrogens. The first peak on the spectrum, at m/z 14 represents a nitrogen fragment. From the mass spectra the molecular formula C2H3N was determined. This formula gave an IHD of 2, meaning a triple bond is likely. The IR spectra agreed with this analysis with a high intensity sharp peak at 2250 cm-1. This peak corresponds to the presence of a nitrile group. The proton and carbon NMR expressed one unique hydrogen and one unique carbon respectively. Therefore, the low boiling compound was determined to be acetonitrile. Acetonitrile is a colorless, water soluble liquid that boils at 81-82°C and melts at -48°C.14 The solid unknown #20 was separated into a neutral organic solid and a carboxylic acid solid. In order to get the solid to dissolve before extraction, double the recommended solvent, dichloromethane, was used. This meant that extraction was done twice. Therefore, smaller amounts of the liquid were being extracted at one time making the extraction more precise. The crude carboxylic acid solid had a percent recovery of 39.31% with a loss of product in the drying process and on the sides of the Erlenmyer flask during filtration. After recrystallization, the carboxylic acid solid was powdery and white with a percent recovery of 32.26%. The loss of product this time was due to vacuum filtration. The melting point of the pure carboxylic acid solid was 158.3160.3°C. The mass spectra for the carboxylic acid solid expressed a molecular ion peak at m/z 201, which indicates a nitrogen. However, upon closer inspection, this molecular ion peak is incorrect, the actual peak is at m/z 200 and the compound does not contain a nitrogen. There was a M+1 peak nearly the same height of the M+ peak indicating the presence of Br. The base peak was also m/z 200. It expressed other significant peaks at m/z 183 and 155, which correspond to an OH fragment and a OCH2CH3 fragment. Therefore, based off of the mass spectra the molecular formula C7H5O2Br was determined. This formula corresponds to an IHD of 5, indicating the possibility of four double bonds and a ring. The IR spectra showed a medium intensity, sharp absorption band at 1550 cm-1 indicating the presence of a ring also. The proton and carbon NMR also indicated a ring because the peaks were in between 7 and 8 ppm for the proton spectra and between 120 and 140 ppm for the carbon spectra. Therefore, the carboxylic acid solid was determined to be 3-bromo benzoic acid. This solid is a white to yellow powder with a melting point of 155-158°C.15 The crude neutral solid had a percent recovery of 45.42% with a loss of product due to product on the walls of the Erlenmyer flask during vacuum filtration. The purified neutral solid was crystalline and off-white in appearance with a percent recovery of 35.81%. This loss of product was also due to the vacuum filtration and the transfer of filter paper to watch glass. The mass spectra showed a molecular ion peak at m/z 182 and a base peak at m/z 105. The lost fragment here is most likely a benzene ring with an M77. There was another notable peak at m/z 77, indicating the presence of another benzene ring. Therefore, the molecular formula determined from the MS was C13H10O and had an IHD of 9. The IHD of 9 indicates two rings and seven double bonds. The IR reveals the presence of ring also with a medium intensity, sharp band at 1600 cm-1, and a carbonoxygen double bond with a high intensity, sharp peak at 1680 cm-1. Both of the proton and carbon NMR spectra reveal the presence of rings in the neutral compound because all of the peaks in the proton NMR are between 7-8 ppm and all but one of the peaks in the carbon NMR appear between 128-140 ppm. Therefore, the neutral solid was determined to be benzophenone. Benzophenone is a white solid with a melting point of 47-51°C.16 EXPERIMENTAL: Lab 2: The Separation of Two Liquid Unknowns by Simple Distillation: Unknown # 8 was obtained. The mass of the distillation flask with five boiling stones and the mass of the receiving flask were obtained before the addition of the liquid unknown. A simple distillation apparatus was built, and the distillation flask was covered in glass wool. The distillation flask was heated with a heating mantle and once condensation began to form on the thermometer tip, the temperature was watched closely. The low boiling liquid was collected in the receiving flask and the stabilized vapor temperature was recorded. Once the temperature dropped the distillation flask was removed from the heating mantle and cooled to room temperature. The mass of both the distillation flask and the receiving flask were taken again and a percent recovery was calculated. Both liquids were poured out of the respective flasks and into two different vials labeled with name, date, section number, and contents. To determine the boiling point of the separated liquids two milliliters of the low boiling point liquid was poured into a test tube containing boiling chips. The test tube was placed into an aluminum heating block and clamped into place, ensuring that the bottom of the test tube did not touch the hot plate. A thermometer was clamped into place to prevent it from touching any part of the test tube. The liquid was then heated until the temperature stabilized for at least one minute. This was the boiling point of the low boiling point liquid. These steps were repeated for the high boiling point liquid. One milliliter volumetric flasks were obtained, rinsed with acetone, and allowed to dry. The mass of the empty volumetric flask was obtained. A Pasteur pipette was used to deliver one milliliter of one of the unknown liquids into the volumetric flask, and then both flask and liquid were massed. The mass and volume were used to determine the density of the liquid. These steps were repeated for the other unknown liquid. Lab 3: The Separation of Two Solid Unknowns by Acid-Base Extraction: Unknown solid # 20 was massed (5.0017g ) and transferred into a 125 mL Erlenmyer flask. Dichloromethane (CH2Cl2) was measured (110.4 mL) and poured into the Erlenmyer flask and the mixture was swirled and stirred with a glass stir rod to dissolve the solid. Once the solid completely dissolved, half of this mixture was poured into a separatory funnel and 20 mL of 1.0 M sodium hydroxide (NaOH) was added. Making sure the stopcock was closed, the funnel was then inverted and shaken for ten seconds and then vented. This process was repeated until venting produced no sound. The separatory funnel was then placed in a ring stand and left to settle undisturbed so the two liquids could separate. Once a clear separation was visible, the organic phase was drained into a beaker below the funnel, and the top layer (the aqueous phase) was poured into a separate beaker. The organic phase was then poured back into the separatory funnel, 20 mL of NaOH was added and the mixture was shaken and vented. The organic phase was once again drained into a beaker and the aqueous phase was poured out the top into the beaker with the aqueous phase from the previous extraction. This process was repeated once more for this half of the original mixture, and then three more times for the second half of the original mixture. In order to precipitate out the carboxylate salt, 23 mL of 6.0 M hydrochloric acid (HCl) was added to the aqueous phase and set on ice. To precipitate out the organic phase, magnesium sulfate (MgSO4), a drying agent, was added until it stopped clumping and the beaker appeared to look like a snow globe. The organic phase was then vacuum filtered through a Buchner funnel to separate the drying agent from the product. The filtrate was then poured into a 100 mL round-bottom flask and then placed in a rotary evaporator until only 1-2 mL of solution remained. This solution was then poured into a massed vial and dried. The aqueous phase, still sitting on ice, was then vacuum filtered through a Buchner funnel and the solid product was rinsed with 7 mL of ice-cold water. After running the vacuum for ten minutes, the solid carboxylic acid was scraped with a spatula onto a labeled watch glass and put in the oven to dry for fifteen minutes. The watch glass was then cooled to room temperature and the carboxylic acid was poured into a massed vial. Lab 4: The Purification and Characterization of Two Unknown Solids by Recrystallization and Melting Point: The mass of both crude solids was obtained and a percent recovery was calculated. In order to recrystallize the carboxylic acid 75 mL of 2:1 MeOH: H2O was poured into a 150 mL Erlenmyer flask with boiling stones and then heated on a hot plate. The carboxylic acid solid was poured into a beaker and the hot 2:1 MeOH: H2O was slowly poured over the solid until it started to dissolve. The beaker was then placed on the hot plate and covered with a watch glass. Solvent was added dropwise until the solid dissolved. The solution was cooled to room temperature and then cooled in an ice-water bath along with at least 5 mL of the solvent used to dissolve the solid. The solid was then separated from the solution by vacuum filtration through a Buchner funnel and rinsed with 5 mL of the cold solvent. The solid was placed onto a massed watch glass and set in a 110°C oven for twenty minutes. All steps were repeated for the neutral organic solid except that it was dissolved in only 25 mL of ethanol and it was not placed in the oven to dry. Each purified solid was transferred into a massed and labeled vial and the pure percent recovery was calculated. To determine the melting points for the pure compounds two melting point samples were obtained using a capillary tube and the MelTemp apparatus. Lab 5: Determination of Molecular Weight and Connectivity by Mass Spectrometry, Lab 6: Infrared Spectroscopy and Functional Group Determination, Lab 7: NMR Spectroscopy: The topics of mass spectrometry (MS), infrared spectroscopy (IR), and nuclear magnetic resonance spectroscopy (NMR) were lectured on during the lab time. Worksheets were passed out to practice the techniques discussed in lab. The unknown samples were not prepared. REFERENCES: (1) Wahrenburg, Zach. CHE 231: Mechanistic Organic Chemistry 1 Lab. https://d2l.depaul.edu/d2l/le/content/280024/viewContent/1804729/View (accessed February 11, 2014) (2) Zubrick, J.W. The Organic Chem Lab Survival Manual: A Student’s Guide to Techniques, 9th ed.; John Wiley and Sons, Inc.: NJ, 2014; pp 325-328, 105. (3) Distillation Applications in Food Industry. http://www.niroinc.com/evaporators_crystallizers/food_distillation.asp (accessed February 11, 2014). (4) Smith, J. G. Organic Chemistry, 4th ed.; McGraw-Hill: New York, 2014; pp 135-136. (5) Chemical Extraction. http://www.cpeo.org/techtree/ttdescript/chemex.htm (accessed February 11, 2014). (6) Wahrenburg, Zach. CHE 231: Mechanistic Organic Chemistry 1 Lab. https://d2l.depaul.edu/d2l/le/content/280024/viewContent/1804732/View (accessed March 10, 2014). (7) Marquit, Miranda. Three Interesting Uses for Mass Spectrometry. http://phys.org/news185029443.html (accessed March 19, 2014). (8) Wahrenburg, Zach. CHE 231: Mechanistic Organic Chemistry 1 Lab. https://d2l.depaul.edu/d2l/le/content/280024/viewContent/1804733/View (accessed March 10, 2014). (9) Yarris, Lynn. New Clues from Infrared Forensics. http://www.lbl.gov/ScienceArticles/Archive/ALS-IR-Forensics.html (accessed March 19, 2014) (10) Wahrenburg, Zach. CHE 231: Mechanistic Organic Chemistry 1 Lab. https://d2l.depaul.edu/d2l/le/content/280024/viewContent/1804734/View (accessed March 15, 2014). (11) Nuclear Magnetic Resonance Spectroscopy. http://www2.chemistry.msu.edu/faculty/reusch/VirtTxtJml/Spectrpy/nmr/ nmr1.htm (accessed March 19, 2014). (12) Michigan State University 900 MHz NMR Facility. http://www2.chemistry.msu.edu/facilities/nmr/900MHz/MCSB_NMR_applications. html (accessed March 19, 2014). (13) Chemical Book: 1-butanol. http://www.chemicalbook.com/ChemicalProductProperty_EN_CB9113046.htm (accessed March 19, 2014). (14) Chemical Book: acetonitrile. http://www.chemicalbook.com/ChemicalProductProperty_EN_CB2127174.htm (accessed March 19, 2014). (15) Chemical Book: 3-bromo benzoic acid. http://www.chemicalbook.com/ProductChemicalPropertiesCB0441823_EN.htm (accessed March 19,2014). (16) Chemical Book: benzophenone. http://www.chemicalbook.com/ProductChemicalPropertiesCB5744679_EN.htm (accessed March 19, 2014). APPENDICES: Calculations: 1. In order to calculate the percent recovery of unknown liquid #8 after simple distillation, add the masses of the separated low boiling liquid and high boiling liquid together. Then divide this sum by the original mass of the mixture and multiply the answer by one hundred percent. πππππππ‘ πππππ£πππ¦ (%) = πππ π ππ πππππππππ‘π (π) × 100% πππ π ππ πππ₯π‘π’ππ (π) 11.0071g +4.0721g = 15.0792 g πππππππ‘ πππππ£πππ¦ (%) = 15.0792π × 100% 15.6911π πππππππ‘ π ππππ£πππ¦ = 96.10% 2. In order to calculate the densities of both the low boiling liquid and the high boiling liquid the mass of 1 mL of each component was divided by the volume of each component (1 mL). π·= π π£ π·πππ€ πππππππ ππππ’ππ = 0.4566 π 1 ππΏ π·πππ€ πππππππ ππππ’ππ = 0.4566 π/ππΏ π·βππβ πππππππ ππππ’ππ = 0.7823π 1 ππΏ π·βππβ πππππππ ππππ’ππ = 0.7823 π/ππΏ 3. In order to determine the percent recoveries for both the crude carboxylic acid solid and the crude neutral organic solid the mass of the component was divided by the mass of the mixture. This answer was multiplied by one hundred percent. πΆππ’ππ πππππππ‘ π ππππ£πππ¦ (%) = πππ π ππ πππ’ππ πππππππππ‘ × 100% πππ π ππ πππ₯π‘π’ππ πΆππ’ππ πππππππ‘ π ππππ£πππ¦ (ππππππ₯π¦πππ ππππ π ππππ) = 1.9663 π × 100% 5.0017 π πΆππ’ππ πππππππ‘ π ππππ£πππ¦ (ππππππ₯π¦πππ ππππ π ππππ) = 39.31% πΆππ’ππ πππππππ‘ π ππππ£πππ¦ (πππ’π‘πππ πππππππ π ππππ) = 2.2717 π × 100% 5.0017 π πΆππ’ππ πππππππ‘ π ππππ£πππ¦ (πππ’π‘πππ πππππππ π ππππ) = 45.42 % 4. In order to determine the percent recoveries for the pure solids carboxylic acid and neutral organic the mass of the pure components after recrystallization and purification was divided by the original mass of the mixture and multiplied by one hundred percent ππ’ππ πππππππ‘ π ππππ£πππ¦ = πππ π ππ ππ’ππ ππππππ’ππ (π) × 100% πππ π ππ πππ₯π‘π’ππ (π) ππ’ππ πππππππ‘ π ππππ£πππ¦ (ππππππ₯π¦πππ ππππ π ππππ) = 1.6135π × 100% 5.0017π ππ’ππ πππππππ‘ π ππππ£πππ¦ (ππππππ₯π¦πππ ππππ π ππππ) = 32.26% ππ’ππ πππππππ‘ π ππππ£πππ¦ (πππ’π‘πππ πππππππ π ππππ) = 1.7913π × 100% 5.0017π ππ’ππ πππππππ‘ π ππππ£πππ¦ (πππ’π‘πππ πππππππ π ππππ) = 35. 81% 5. In order to calculate the index of hydrogen deficiency (IHD) the number of carbons was multiplied by two. The number of Nitrogens plus two was added to this number. The number of hydrogens and the number of halogens were subtracted from this total. This final number was divided by two. πΌπ»π· = 2πΆ + 2 − π» − π₯ + π 2 πΌπ»π·πππ€ πππππππ ππππ’ππ = 2(2) + 2 − 3 + 1 2 IHDlow boiling liquid = 2 πΌπ»π· π»ππβ π΅ππππππ πΏπππ’ππ = 2(4) + 2 − 10 2 IHDHigh Boiling Liquid = 0 πΌπ»π·πΆπππ» = 2(7) + 2 − 5 − 1 2 IHDCOOH = 5 πΌπ»π·πππ’π‘πππ = 2(13) + 2 − 10 2 IHDNeutral = 9