Chemical Analysis of Exhaled Human Breath Using a Terahertz Spectroscopic Approach Alyssa M. Fosnight, Benjamin L. Moran, and Ivan R. Medvedev 1) Department of Physics, Wright State University, 3640 Colonel Glenn Highway, Dayton, Ohio 45435, USA Supplemental Materials In order to precisely quantify the dilution of ethanol, methanol, and acetone present in breath, the preconcentration efficiency of Entech 7100A system was determined in a separate set of experiments. This was done by injecting a Tedlar bag containing 3 L of Nitrogen (measured with a TSI series 4100 thermal flow meter) with 0.75 μL of acetone, methanol, and ethanol. Hamilton Company 7000 series modified MICROLITER syringes with a dispensing volume of 0.5 μL purchased from Fisher Scientific were used to deliver a 0.75 μL of pure liquid samples into a Tedlar bag. The reproducibility of 0.75 μL injection into a Tedlar is unlikely to be better than 5 to 10 %. Given the amount of injected sample and amount of nitrogen in the bag, the expected chemical concentrations in the Tedlar bag was determined. Therefore, by taking the ratio of the concentration found by a spectroscopic measurement and the expected concentration, the preconcentration efficiency could be calculated. In separate set of measurements, 0.05 mL of liquid water was added into the Tedlar bag to simulate moist human breath. The preconcentration efficiency was measured as a function of sampling volume. The sampling volume was varied from 100 cc to 500 cc in increasing and decreasing directions to assess the effects of sampling volume and chemical history on the preconcentration efficiency of Entech 7100A. Figure S1 shows the experimentally determined preconcentration efficiency of Entech 7100A for ethanol, methanol, and acetone as a function of sampling volume, sample wetness (labeled ‘Wet’ and ‘Dry’ in Figure S1), and the chemical history of the preconcentrator. The chemical history dependence was studied by sequentially increasing or decreasing (labeled ‘Up’ and ‘Down’ respectively) volumes of breath samples. No significant correlation between the efficiency and sample wetness or chemical history was observed, but there was a noticeable dependence on the sampling volume for methanol and ethanol. The average of all the measurements taken at 500 cc sampling volume was used as a measure of our preconcentration efficiency. The RMS of the preconcentration efficiency distribution at 500 cc was used as a measure of its uncertainty. 1) Author to whom correspondence should be addressed. Electronic mail: ivan.medvedev@wright.edu FIG S1. Preconcentration efficiency of Entech 7100A as a function of molecular species, samples volume, sample wetness, and the preconcentration history of the system. TABLE SI. Spectroscopic line center-frequencies in MHz for acetone, methanol, and ethanol used for chemical detection. Line frequencies are listed in the order they appear in the snippet spectrum shown in Figure 1. Acetone 226832.08 230176.70 235548.35 237751.08 238868.95 239991.07 245353.09 256259.03 258493.67 259618.37 Methanol 211803.34 227094.78 237970.44 239970.26 241267.89 241791.41 244337.99 261704.48 264325.60 265289.66 Ethanol 225660.44 230991.56 232318.69 233208.71 239478.27 242350.02 242524.40 242870.76 244634.12 259777.30 TABLE SII. Spectroscopically determined partial pressures (Psample) in Torr and least squares signal to noise ratios (SNR) for acetone, methanol, and ethanol as a function of time after the beginning of alcohol consumption. Time (min) -107 -80 34 55 90 125 155 190 230 269 300 335 Acetone Psample SNR 16.7 126 11.3 78 7.4 15 13.2 49 13.8 37 12.0 18 13.3 57 13.6 31 16.9 53 17.7 123 19.5 129 13.8 101 Methanol Psample SNR 1.2 14 0.5 6 0.009 0.03 0.9 5 1.4 6 2.0 5 0.8 5 1.2 4 1.8 9 2.7 29 2.4 24 1.5 17 Ethanol Psample SNR 3.4 17 0.7 4 136.2 186 545.7 1378 456.8 842 469.2 490 123.5 363 171.4 262 120.4 255 64.4 305 23.8 107 6.4 32 TABLE SIII. Experimentally determined dilutions for acetone, methanol and ethanol and corresponding breath alcohol content determined by the breathalyzer measurements and Widmark’s formula as a function of time after the beginning of alcohol consumption. Corresponding experimental uncertainties are listed in parenthesis. BAC is measured in grams per 100 mL of blood volume. Time (min) Acetone (ppb) Methanol (ppb) Ethanol (ppm) -107 -80 34 55 90 125 155 190 230 269 300 335 1042 (127) 706 (87) 462 (67) 826 (103) 864 (109) 753 (103) 831 (103) 854 (109) 1060 (132) 1105 (136) 1223 (150) 863 (106) 302 (109) 136 (55) 2 (93) 217 (91) 347 (140) 504 (215) 207 (85) 311 (137) 460 (172) 683 (241) 587 (208) 379 (135) 0.63 (0.12) 0.14 (0.05) 25 (4) 101 (17) 84 (15) 87 (15) 23 (4) 32 (5) 22 (4) 12 (2) 4.4 (0.8) 1.17 (0.21) Breathalyzer (BAC) 0 0 0.033 0.038 0.032 0.023 0.017 0.008 0 0 0 0 (ppm) 0 0 86.2 99.3 83.6 60.1 44.4 20.9 0 0 0 0 Widmark’s Formula (BAC) (ppm) 0 0.0 0 0.0 0.0427 111.4 0.0374 97.7 0.0287 74.9 0.0199 52.0 0.0124 32.4 0.0037 9.5 0 0.0 0 0.0 0 0.0 0 0.0 TABLE SIV. Experimentally determined dilutions and spectroscopic least squares signal to noise ratios (SNR) for acetone, methanol, and ethanol as a function of time after the beginning of alcohol consumption. Corresponding experimental uncertainties are listed in parenthesis. Time /min -107 -80 34 55 90 125 155 190 230 269 300 335 Acetone (ppb) 1042 (127) 706 (87) 462 (67) 826 (103) 864 (109) 753 (103) 831 (103) 854 (109) 1060 (132) 1105 (136) 1223 (150) 863 (106) SNR 126 78 15 49 37 18 57 31 53 123 129 101 Methanol SNR (ppb) 302 (109) 14 136 (55) 6 2 (93) 0.03 217 (91) 5 347 (140) 6 504 (215) 5 207 (85) 5 311 (137) 4 460 (172) 9 683 (241) 29 587 (208) 24 379 (135) 17 Ethanol (ppm) 0.63 (0.12) 0.14 (0.05) 25 (4) 101 (17) 84 (15) 87 (15) 23 (4) 32 (5) 22 (4) 12 (2) 4.4 (0.8) 1.17 (0.21) SNR 17 4 186 1378 842 490 363 262 255 305 107 32