Sample Injection Tube (SIT) de-clogging guide 1. Check sheath tank for pressure. 2. Ensure tube is not cracked. 3. Attempt PRIME and check for bubbles. a. Strong bubble stream i. Clog may be gone, try sample again. b. Weak bubble stream i. Repeat PRIME 2x. ii. Run FACS Clean for 5 min, then DI H2O for 5 min. iii. Repeat PRIME, try sample again. c. No bubbles i. Use stylus wire to manually dislodge clog. ii. Gently feed wire through aperture until there is 1 inch protruding from the bottom. iii. Move wire up and down to dislodge clog. iv. Remove wire and PRIME. Cytometer Troubleshooting Guide Observations No events in acquisition display and RUN button is green. No events in acquisition display and RUN button is orange. Possible Causes Recommended Solutions Air in the sheath filter. Purge the air from the sheath filter. Waste tank is full. Empty the waste tank. Sample injection tube (SIT) is clogged. See SIT de-clogging guide. Connection issue with the cytometer Turn off cytometer and computer and restart both. Gating issue Make sure plot is not gated on a non-existent population. Sheath container is not pressurized. Ensure that the sheath container lid and all connectors are securely seated. Inspect the large O-ring and replace if necessary. Ensure that the sample tube is not cracked, if so, replace it. Sample tube is not pressurized. Ensure that the correct tube is being used. Corning 12x75mm 5mL polystyrene tubes. Check the Bal seal (O-ring that seals the tube onto the SIT). If it is worn, and does not make a tight seal, please ask core staff for assistance. Cytometer Troubleshooting Guide Observations Absent or weak fluorescent signal Possible Causes Incorrect optical filter is installed. Laser malfunction High event rate Low event rate Recommended Solutions Make sure the appropriate optical filter is installed for the fluorochrome in question. Check with core staff to run CST to verify instrument performance. Sample is too concentrated. Dilute the sample or use a lower flow rate. Threshold level is set too low. Increase the threshold level to ignore noise and debris. Air bubble in the sheath filter Bleed the bubbles from the sheath filter. Air bubble in flow cell Perform a PRIME with a tube of water. (de-bubble on CantoII). Sample injection tube (SIT) is clogged. See SIT de-clogging guide. Sample is too dilute. Concentrate the sample. If the flow rate setting is not critical to the application, set the flow rate to MED or HI. Sample is not adequately mixed. Vortex the sample to re-suspend the cells. Cytometer Troubleshooting Guide Observations Erratic event rate Distorted scatter parameters Possible Causes Recommended Solutions Sample tube is cracked. Replace the sample tube Bal seal is worn. Replace the Bal seal. Please ask core staff for assistance. Sample injection tube (SIT) is clogged. PRIME with FACS Clean and H20. Sheath filter is dirty. Replace the sheath filter. Please ask core staff for assistance. Cytometer settings are improperly adjusted. Ensure that FSC and SSC are set to linear scale and voltages are optimized. Air bubble in the sheath filter Bleed the bubbles from the sheath filter. Air bubble in flow cell Perform a PRIME with a tube of water. (de-bubble on CantoII). Flow cell is dirty. Run a tube of FACS Clean for 5 minutes, followed by a tube of dH20 for 5 minutes Cytometer Troubleshooting Guide Observations Possible Causes Excessive amount of debris in display. Threshold level is too low. Sheath filter is dirty. High CV Recommended Solutions Increase the threshold level or lower the forward scatter detector. Replace the sheath filter. Ask core staff for assistance. Excessive debris in sample. Filter sample with a 40um nylon mesh. Air bubble in sheath filter or flow cell Purge the air from the sheath filter. Sample flow rate is set too high. Lower the sample flow rate. Excessive debris in sample. Filter sample with a 40um nylon mesh. Cytometer Troubleshooting Guide Observations Possible Causes Recommended Solutions Software Errors Acquisition rate is too high. Master DAQ Overflow Canto: <15,000 evt/sec Lower flow rate or dilute sample. Increase FSC threshold to ignore noise and debris. Bad connection Turn off both cytometer and computer. Wait 30 sec. Turn on cytometer. Turn on computer. Allow windows to boot up completely (including anti-virus) before opening FACS Diva software. PMT Voltage Mismatch Voltages are not consistent for fluorescent parameters in compensation control tubes. Go through each compensation control and find which voltage is not consistent. Reacquire tube with corrected voltage and overwrite. Recalculate compensation. Software Freeze Usually due to excessive amount of data stored in the Browser. Access ‘Task Manager’ by pressing Control+Alt+Delete. Force quit FACS Diva software. Computer Crash Usually due to excessive amount of data stored on the hard drive. If computer is totally unresponsive, hold the power button for 5 seconds. Wait 30 seconds. Turn computer back on. Cannot Connect to Cytometer