Physics 4AL Lab 0

advertisement
Jason Fong
702847140
date of experiment: 10-3-00
partner: Rosanne Hong
Physics 4AL Lab 5
Experiment 0
The purpose of this experiment was to calibrate a force transducer and to get acquainted
with the lab equipment. Weights of various masses were hung on the sensor arm, and
measurements were taken for the resulting voltage from the sensor. The voltage and the weight
of each mass were graphed, and a trend line was fitted onto the points of the graph. Then a
regression analysis was done to complete the experiment.
Before beginning the measurements, the force transducer was plugged into channel A of
the PSW box. Then the PSW program was set up for the experiment by establishing an analog
connection to channel A and setting a table to record voltage data from channel A.
Measurements were then taken at two second intervals for seven different masses: 0, 50, 100,
200, 300, 400, and 500 grams. Five readings were taken for each mass in order to obtain more
accurate measurements. The resulting numbers were then transferred to an Excel spreadsheet for
data manipulation. The averages of the voltage readings for each mass were calculated. The
masses in grams were converted to their corresponding weights in Newtons by using the formula
F=ma with the value of gravity g=9.85 m/s^2, and the conversion factor of 1 kg per 1000 g.
Thus the weights in Newtons for each mass was obtained by multiplying the masses in grams by
9.85 and 1/1000. The average voltages were then matched with their corresponding weights, and
the results were plotted on a scatter plot. A trend line was then generated for the scatter plot, and
a regression calculation was performed using the voltages as the x-input and the weights as the yinput. The data, graph, and regression calculation results are included in the attached pages.
The trend line that resulted from the scatter plot gave the expected downward-sloping
graph intersecting the y-axis at 0. However, the values for the graph were roughly double the
values that were expected. Likewise, the values in the data tables were also roughly double the
expected values. Further measurements were taken in an effort to discover the source of this
inaccuracy. However, the same trend line resulted. Thus, the source of the error seems to be
coming from the settings within the PSW box.
Download