JLab-TN-02-025 Preliminary Results from Cavity and Process Particulate Collection By: John Mammosser and Andy T. Wu Particulates from the cavity production High Pressure Rinse (HPR) filter and Jl009 cavity after vertical test were collected and analyzed to determine the source of recent poor cavity RF performances. Commonalities in particles found from the cavity, trapped in the HPR filter media and the filter media material, suggest that many of the particulates found in JL009 could have come from the DI water and HPR system due to maintenance and replacement of this filter prior to processing of JL009, however it is not known which of these particulates would cause field emission, or if they were the material causing the field emission in JL009. A method for collecting and analysis of particulates collected for this study, the results from the particulate analysis and several suggestions to improve procedures are reported in this document. Background In preparation for building the SL21 cavity string, cavities must first be qualified for string assembly by demonstrating that they can operate above design specification at which they will be operated in the accelerator. The SL21 module cavities are required to operate at a Q-value of 6.5e9 at gradients above 12.5MV/m. Recent cavity qualification tests were failing due to low Q-values caused by field emission loading at gradients lower then these design values. The cause of the field emission is believed to be surface contamination that enters in during processing and assembly stages for these cavities, prior to the vertical RF tests. The cavity processing procedures were recently modified to address two identified critical areas in which particulates could come in and to improve chances of their removal. The two areas identified were ineffective high pressure rinsing (not enough time) and test flange cleaning. The test flanges were cleaned in a different way and were not high pressure rinsed like the cavity. The decision was to extend the amount of high pressure rinsing time to four hours in two-hour intervals, the second interval having all test flanges attached with the exception of the bottom flange (input coupler) to allow the wand a path for entering the cavity. After several cavities were processed in this way and RF tested in the VTA, it became evident that a field emission was not reduced by the procedure changes in all tests. Also the cavity performance seemed to vary even more with few good tests and the rest dominated by field emission, most likely caused by particulate contamination on cavity interior surfaces. To investigate the root cause of this contamination, particulate samples were taken from the high-pressure rinse filter, which was removed for inspection and replaced, and JL009 cavity after the vertical RF tests and after the filter replacement. Particulate Collection Method In-order to collect particulate samples and be able to identify them on the Scanning Electron Microscope (SEM), double sided carbon tapes (manufactured by SPI) were used for collecting particulates on both the filter and cavity surfaces. These tapes are designed for clean collection and analysis of small size particulates. Particulate Collection from HPR Filter On 6/18 the HPR filter was removed for inspection and a new filter was put in its place. The filter is a 20-inch long borosilicate glass fibers held together with a polypropylene binder. The filter thickness is ½ inch and is sealed at both ends and designed for filtering liquids to 0.2um. The following procedure was used to remove the filter and collect the particulates. A clean nylon bag was cut to length and sealed at the bottom in the main chemroom and placed in the pass thru. In the cleanroom inside the HPR cabinet he filter housing was unscrewed and the filter was removed and immediately placed into the nylon bag. The unsealed end was folded back and taped across the entire opening to the bag itself. The filter was transported to the offline cleanroom where the outside of the bag was wiped with a cleanroom wipe with methanol. The laminar flow bench was then wiped down with acetone and then methanol to clean work surfaces. The bag was then cut open with scissors below the taped end and the filter was removed from the bag. JLab-TN-02-025 Next the scissors were cleaned and outer plastic guard was removed and the filter cut open down the middle. Carbon double sided tape was then used to pick samples from the filter using blotting at areas of visible particulates Inspection of the filter showed the internal surfaces where the water enters was a brownish color and after a few layers in the filter media was white and clean. The outside of the filter was inspected and no visible particles were found. But at the base of the housing there was also a brownish stain. JL009 History and Performance JL009 was assembled with the modified procedure mentioned above and RF tested in the VTA on 6/23, see Figure 1. JL009 Vertical Test Results. The processing and assembly of this cavity encountered no problems and was typical for this type of cavity. The results from this test showed early onset of field emission at 5.8MV/m and a low Q-value at low gradients of 8E9. The performance of this cavity test was typical of a performance dominated by heavy field emission and was identified as a good candidate for collecting particulates from the interior surfaces to gain knowledge on what could be causing the failure. JL009b Vertical Test 6/23/02 10 10 10 9 1 2 3 4 5 6 7 8 Eacc (MV/m) Figure 1. JL009 Vertical Test Results. Particulate Collection from the Cavity The following procedure was used to collect particulates from the cavity interior surfaces: First the cavity was moved to the cleanroom on the test stand, cavity in the vertical position and was back filled with filtered nitrogen to 1 ATM so it could be opened. Several clean nylon bags were cut to length and sealed at one end in the chemroom. The bottom flange hardware was removed and then the test flange was separated from the cavity flange. JLab-TN-02-025 Next the surface between the vacuum port opening and the input probe port was blotted with the carbon tape and the tape was immediately placed into the nylon bag and closed. The cavity interior beam tube, FPC end was also blotted with tape and added to the nylon bag. Finally the field port test flange was removed in the same way (cavity top) and the flange and the HOM beam tube end were blotted with the tape. The nylon bag opening was folded back and taped across the entire opening to the bag itself. Cleaning and preparing samples for the SEM The following procedure was used to prepare the SEM chamber: Prior to starting the analysis, the SEM chamber was vented and cleaned by wiping the interior surfaces with a cleanroom wipe and methanol. Next the stage was removed and brought to the laminar flow bench where it was cleaned by wiping with methanol. Next a new package of silicon wafers was opened and a single wafer was removed and wiped off with methanol. Next the nylon bag was opened, by cutting with a clean scissors and the carbon tape was pealed from the nylon bag and the backside protection strip was removed. Finally the tape was adhered to the silicon wafer. All samples from the cavity were mounted to this wafer and the wafer was placed into the stage and clipped into place for a good electrical contact. Next the stage was carried to the chamber and placed onto the stage holder and the door was closed. The chamber was then evacuated and identification and analysis was ready to start. Particulate Identification The two sample tapes were collected from the HPR filter and these showed many very large size particles with at least one side having a length of greater then 100m and many smaller sizes less then 100m in maximum length. The type particle could also be categorized as a solid element with some additional trace elements or consisting of a composite of many different elements forming the particle. The picture John04 shows a typical smaller size solid element that consists of the following makeup: Element Si S Fe % 97.56 0.81 1.63 Picture John04 -A small solid element mostly Silicon, collected from the HPR filter. JLab-TN-02-025 The picture John02 shows a large particulate that is a composite and has the following makeup: Element S Mo Fe Cr % 63.1 34.12 2.18 0.58 These pictures show clearly the filter media glass fibers with their polypropylene binder and a clear texture to the particles that were removed from the filter. The small solid element John04, has a smooth texture and the large composite John02, a course texture. Both pictures show a rectangular box locating the position on the particle where the electron beam was focused for analysis. For smaller size particles a spot marked by a Picture John02 – A large composite particle collected from the HPR filter. cross hairs was used to reduce background from entering into the collection process. The identification of the particle makeup was performed with the SEM X-ray analysis (EDAX) system. This system has a lower energy cutoff of about 0.5keV so it cannot detect elements with atomic numbers below fluorine and this is why carbon tape makes for an excellent collection media. The carbon tape background was analyzed to determine if any energy peaks were identified and is provided at the end of this document with the picture locating the electron beam spot. For large composite particles, the particle makeup can be dependent on where on the sample the beam is focused and no attempt was made to fully characterize each particle. The picture John03 shows a large composite particle that has a very different texture that that of John02. At the rectangle, smaller coarser particles are evident and the makeup consists of 6 different elements as follows: Element S % 38.09 JLab-TN-02-025 F Fe Si Mg Cr 34.22 12.78 6.76 4.24 3.92 Picture John03 – A large composite particulate collected from the HPR filter. The summary of ten particles analyzed shows that eleven different elements were found and that some of these elements are consistent with the type of suspended matter that can be found in high purity water (complex colloids). For example, silica is a common colloid that can contain heavy metal and organic ions that collect and form complex particulates and are suspended in the water and pass through most ion exchange beds but should be filter out with ultra filtration stages1, see Figure 2. Results from analysis of particles trapped in the HPR filter. Particulates Collected From The HPR Filter 200 150 100 50 0 S Al Fe Si Mo Ti F Cr Ni Mg Cl Ca P Cu Au Na K Figure 2. Results from analysis of particles trapped in the HPR filter. JLab-TN-02-025 Additionally, the filter media was analyzed to compare to the elements found in the cavity, see figure 3. Analysis of the HPR filter media. The filter media is important because it is the final filtration and the bulk volume of high-pressure water must interface with the media. Filter Media Make-up 500 400 300 200 100 0 Si Ca Al Na K Mg Fe Figure 3. Analysis of the HPR filter media. Observations Many very large particles were present on the input side of the filter media, which should have been filtered out, in the polishing loops of the DI water plant. This system has two polishing loops each with final 0.1micron ultra filters. Filter looked clean outside and no visible particles were identified. Filter interior was brown as well as the bottom of the filter-holder, this should be analyzed to find out what it is. Discussion: These particles could have come into the filter during modification to the DI water system, installation of the filter or from the HPR components itself. More effort to collect particulates from the outside of the filter could have revealed smaller size particulates that might have traveled to the cavity. Can these large particles breakup and travel through the filter media and end up in the cavity. Does the filter media glass fibers field emit subjected to high electric fields. During the analysis several of the glass fibers were charged by the electron bean in the SEM and stood straight up on the carbon tape while scanning the surface. Tests should be carried out to see if the filter media causes field emission. JLab-TN-02-025 Results From Cavity Particulates Identification Twenty different particles collected from cavity JL009 surfaces (after vertical test), were analyzed to determine their elemental makeup and characteristic sizes. The particles varied in size from 1m to 75 m in maximum length and elemental makeup with 16 different elements identified. These particles were typically smaller in size then the particles found in the HPR filter (see Figure 4. Particle Sizes from both JL009 and HPR Filter.). Particulate Sizes (um) 250 200 150 100 50 0 F2 F3 C2 F4 C10 C3 C5 C9 C11C20 C4 C8 C6 C18C12C17C19 C7 Particle Number (F-filter,C-cavity) Figure 4. Particle Sizes from both JL009 and HPR Filter Analysis of the particles showed that most were a composite of at least two elements. Only four of the twenty particles were solid elements such as Picture JL009-7, which was pure copper and 1 by 2m in size. The sizes of the smallest of the particles were from 1 to 8m in maximum length and most were metals such as gold, copper and aluminum. Also it is important to note that all but three of the elements found in the cavity were present in the filter and filter media. These three elements were copper, gold and phosphorus. Picture JL009-7 Solid Copper Particle. JLab-TN-02-025 The Picture JL009-8, shows a typical small composite particle found on the in the cavity JL009 along with its energy dispersive X-ray data that clearly shows seven different elements. This collection of elements has metals such as iron and titanium as well as the typical dominant element silicon. Although many particles were present on these samples not all were analyzed but the majority of the particles, especially the larger sizes, contained silicon as seen in Figure 5. Particulates collected from JL009 cavity surfaces. Particulates Collected from JL009 Cavity Surfaces 600 500 400 300 200 100 0 Si S Fe Cl Al Ca Cu Au Ti Mg F Na K Cr P Ni Mo Elements Figure 5. Particulates collected from JL009 cavity surfaces. The cavity particulates also varied in texture as did the ones collected from the filter and is shown in Picture JL009-9 were six different particulates are in the field of view. JLab-TN-02-025 Picture JL009-9. In the picture JL009-9, there are particles present that have elements representing most the elements found and large composites (upper left) as well as smaller solids (lower center). Observations Discussion Silicon seems to be the most dominant large element found in JL009 but the small metals such as aluminum and copper are also present. These samples only represent a small sample of the particulates from JL009 and it is not known if these are a good representation of the interior cell surfaces. All but three of the elements in the cavity were found in the water system. The particulate collection method described in this paper was simple and effective in collecting particle of sizes ranging from sub-micron to few tenths of micron from SRF cavity surfaces. Many particulates were collected from the cavity and test flanges. These particulates need to be reduced through improved assemble and process procedures. More attention to system operation and maintenance could reduce the number of particulates entering these components. It would be important to clearly identify where process improvements need to take place by analyzing critical process performance such as the HPR system, DI water quality, vacuum system performance and test-stand contamination. We would like to thank Ralph Afanador and Danny Forehand for their contributions with collecting samples. Appendix 1 Metzer, Theodor H. “High-Purity Water Preparation for The Semiconductor, Pharmaceutical, and Power Industries”, Tall Oaks Publishing, Inc., pg 9-10, 1997. JLab-TN-02-025 Background analysis of the carbon tape.