Treatment Planning Including Respiratory Motion Based on 4D Computed Tomagraphy Data Eike Rietzel1,2, George T.Y. Chen1, Judith A. Adams1, Alvaro H. Hernandez1, Christopher G. Willett1 (1) Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA; (2) Abteilung Biophysik, Gesellschaft für Schwerionenforschung, Darmstadt, Germany Abstract 4D Computed Tomography records different respiratory states of patient anatomy. These data can be included in dose calculations to explicitly account for intrafractional organ motion caused by respiration. Based on the magnitude of motion, appropriate treatment delivery techniques can be selected. Internal target volumes are generated by combining targets at different respiratory phases. The impact of organ motion on dose distributions is analysed by inspection of dose distributions for time-varying patient anatomy; dose coverage throughout the respiratory cycle can also be visualized. Keywords 4D Computed Tomography, 4D treatment planning, dose calculation, respiratory motion, intrafractional motion, non-rigid registration Introduction Intrafractional organ motion can cause significant errors in radiotherapy imaging, treatment planning, and treatment delivery. In standard CT scanning, severe imaging artifacts can be introduced, primarily due to respiratory motion. To reduce or even eliminate such errors, time-resolved CT data can be acquired. 4D Computed Tomography [1,2] provides multiple spatio-temporally coherent volumetric data sets of patient anatomy, each at a different respiratory state. Such data can be included in treatment planning to ensure adequate dose coverage of the target(s) throughout the respiratory cycle. To generate patient specific composite target volumes that account for respiratory motion, contours from different respiratory volumes are combined. Furthermore, the impact of respiratory motion on dose distributions can be analyzed. Respiration induced density variations must be considered during treatment planning especially for charged particle therapy. Particle ranges strongly depend on time dependent density variations along their path. To analyze organ motion including deformation or to add dose distributions from different respiratory phases, non-rigid registration techniques are needed. Material and methods 4D Computed Tomography 4DCT data are acquired on a 4-slice General Electric Lightspeed CT scanner. The scanner is operated in axial cine mode. At each couch position, data are acquired for a duration equal to the patient’s respiratory cycle; during this interval, ~5 complete x-ray tube rotations occur. The couch is then advanced to the next position. During couch movement data acquisition is paused. At each table index, ~25 images are reconstructed, evenly distributed over the acquisition time. Each image represents a different respiratory state. During CT data acquisition, the patient’s abdominal surface is monitored with the Varian RPM-system. In addition to surface motion, the software records correlation signals sent from the CT scanner whenever data acquisition is turned on or off. GE’s Advantage4D software facilitates retrospective sorting of the images into spatio-temporally coherent volumes. Based on the temporal correlation between surface motion and data acquisition each reconstructed image is assigned a specific respiratory phase, as obtained from the surface motion. A respiratory phase can be chosen and the software automatically assembles images at the selected phase for each couch index. Typically, CT studies at 10 different respiratory phases are generated, equally distributed over the respiratory cycle. Non-Rigid Image Registration Non rigid image registraion is key to 4D CT image and dose analysis. The software package vtkCISG [3] provides voxelbased affine and non-rigid image registration algorithms. Affine registration involves optimizing 12 degrees of freedom. Non-rigid image registration describes local deformations using a free-form deformation model based on B-splines [4]. Different similarity measures and control point spacings for non-rigid registration were evaluated. Independently moving anatomical regions of the body must be registered separately. For the case study presented here, internal anatomy was segmented. Registration results are poor if registration of the complete anatomy is attempted simultaneously. Local deformations cannot be modelled accurately e.g. due to the quasi-static spine and ribs adjacent to moving internal organs. Initially full affine registration is performed to obtain global motion parameters, followed by modelling local deformations via non-rigid registration. Registration results are visually inspected for consistency on a patient-by-patient basis with either difference images or color overlays. Dose calculations The impact of respiratory motion on dose distributions is studied by calculating dose distributions for 4DCT volumes based on a nominal treatment plan optimized for a single CT dataset. Standard treatment planning systems do not support calculations for different anatomical data sets with the same treatment plan. In the CMS FOCUS treatment planning system used here, treatment plan parameters were copied and modified directly at the filesystem level using in-house software. For the example presented here, a two field proton plan was optimized. Results and discussion Motion assessment A standard helical CT scan of the abdominal region is shown in figure 1. Typical motion artifacts are clearly visible. The identical coronal cut at three different respiratory states as obtained by 4DCT scanning is shown in figure 2. Motion artifacts are significantly reduced or absent. Liver motion was assessed to be 2.25 cm along the cranio-caudal axis from 4DCT. The dataset closest to inhale shows minor residual artifacts in the lower abdominal region. These artifacts are due to irregular breathing during 4DCT data acquisition. Figure 1: Standard helical CT scan of the abdomen acquired under light respiration. Typical respiratory artifacts are present, e.g. jagged edges of the liver. Target delineation Dose calculations The hepatocellular tumor of the patient shown in figures 1 and 2 was contoured on the 4DCT datasets closest to inhale and exhale. The union of these GTVs was generated to account for respiratory motion. To assess the adequacy of coverage, the composite target volume was overlayed on all respiratory phases obtained by 4DCT scanning. Visual inspection verified that the composite contour encompassed all intermediate respiratory states of the GTV. Dose distributions were calculated for ten 4DCT volumes based on the optimized treatment plan. Figure 3 shows distributions for respiratory phases close to inhale, mid-exhale, and exhale on a coronal cut. Visual inspection showed that the composite target volume is included in the high dose area for all respiratory states. Dose volume histograms were not calculated for all datasets because contours were only drawn on volumes close to inhale and exhale. The impact of density variations on proton ranges is clearly visible. The shape of iso-dose lines changes between respiratory phases. For the most inferior liver position at inhale low iso-dose lines extend beyond the liver. At inhale, densities in the beam paths are decreased compared to the exhale volume. The liver moves inferior and is replaced by less dense lung tissue. Full dose coverage of the target is maintained at exhale due to the density changes applied during treatment planning. Treatment planning A proton treatment plan was optimized for the composite target volume. Additional target expansions were used to account for interfractional organ motion and residual intrafractional motion uncertainties. The plan consisted of a right lateral and a posterior field. Variations in water equivalent pathlength caused by density changes perturb proton ranges. The liverlung interface was within both treatment fields. Adequate proton ranges to the distal edge of the target throughout the respiratory cycle had to be ensured. Therefore maximum densities throughout the respiratory cycle within each beam’s path had to be considered for treatment planning. A composite liver volume based on contours drawn on exhale and inhale volumes was generated. Densities inside the composite liver were set to liver density. Setting the higher liver density instead of lung tissue density guaranteed sufficient penetration to the distal edge of the target throughout the respiratory cycle. Registrations Figure 4 shows transformation vectors obtained by registration of 4DCT volumes closest to exhale and inhale. Only internal Figure 2: Respiratory organ motion imaged with 4D Computed Tomography. Left to right corresponds to inhale, mid-exhale, and exhale respiratory phases. Note the significant reduction of motion artifacts compared to figure 1. Liver motion was 2.25 cm. anatomy that is subject to respiratory motion was considered for registration as shown in figure 4, top. The similarity measure used for registrations was sum of squared differences. Initially, volumes were registered with translations, rotations, and scaling. The following non-rigid registration was performed using a control grid spacing of 10 mm. Manual segmentation of internal anatomy was performed for one of the volumes (exhale). During registration only voxels within the segmented volume contributed to the similarity measure, other non-overlapping voxels of the inhale volume were disregarded. Figure 5 shows coronal difference images between the registered volumes pre and post registration. Differences between the unregistered volumes clearly show motion of internal anatomy, especially at edges of organs. After registration, only small differences remain in the image. These differences are due to residual artifacts in the 4DCT data and/or volumetric averaging effects during transformations. Conclusion 4D Computed Tomography permits inclusion of information on respiratory organ motion in treatment planning. Composite target volumes can be generated based on multiple contours at different respiratory phases. Visual inspection of composite targets overlaid on 4DCT or dose volume histograms can be used to assess adequate dose coverage of the target throughout the respiratory cycle. For charged particle treatment planning, respiration induced range variations can directly be calculated. Non-rigid registration techniques can relate patient anatomy at different respiratory states to each other. Such information facilitates improved analysis of organ motion and deformation. Furthermore, dose distributions can be calculated including motion. In principle each voxel can be tracked throughout the respiratory cycle to add the deposited dose per voxel. References [1] Pan T, Lee T Y, Rietzel E, Chen G T 2003 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT Med Phys 31(2) 333-340 [2] Rietzel E, Chen G T, Doppke K P, Pan T, Choi N C, Willett C G 2003 4D computed tomography for treatment planning Int J Radiat Oncol Biol Phys 57(2) S232-3 [3] Hartkens T, Rueckert D, Schnabel, J A, Hawkes D J, Hill D L G 2002 VTK CISG Registration Toolkit: An open source software package for affine and non-rigid registration of single- and multimodal 3D images BVM2002, Leipzig, Springer-Verlag. [4] Rueckert D, Sonoda L I, Hayes C, Hill D L G, Leach M O, Hawkes D J 1999 Non-rigid registration using free-form deformations: Application to breast MR images IEEE Transactions on Medical Imaging 18(8) 712-721 Figure 3: Impact of organ motion on a proton dose distribution. Respiratory phases correspond from top to bottom to inhale, midexhale, and exhale. GTVs contoured on the inhale and exhale dataset are shown as thin lines. Iso-dose lines displayed are from 99% to 10% of the prescribed dose. Figure 5: Subtraction images pre (top) and post (bottom) affine followed by non-rigid registration between 4DCT datasets closest to inhale and exhale. Figure 4: Transformation between 4DCT volumes closest to inhale and exhale; top image: segmented internal anatomy at exhale; middle image: transformation vectors from inhale to exhale obtained by affine and non-rigid registration; bottom: non-rigid part of transformation vectors only.